Validation Set: 515 characterized samples (171 celiac disease, 20 IgA deficient celiac disease, 34 dermatitis herpetiformis, 290 control samples).
—
Indications for Use
The Aptiva Celiac Disease IgG Reagent is an immunoassay utilizing particle-based multi-analyte technology for the semi-quantitative determination of anti-tissue transglutaminase IgG autoantibodies and anti-deamidated gliadin peptide IgG autoantibodies in human serum. The presence of these antibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of celiac disease and dermatitis herpetiformis, particularly in patients with selective IgA deficiency. The Aptiva Celiac Disease IgG Reagent is intended for use with the Inova Diagnostics Aptiva System.
Device Story
Aptiva Celiac Disease IgG Reagent is a semi-quantitative immunoassay for use on the Inova Diagnostics Aptiva System. It utilizes paramagnetic particles coated with recombinant tissue transglutaminase (tTG) and synthetic deamidated gliadin peptide (DGP), plus a control particle coated with goat anti-human IgG. The system dilutes patient serum, incubates with particles, performs wash cycles, and adds a phycoerythrin (PE)-labeled anti-human IgG conjugate. The system captures multiple images to identify particles and measure median fluorescent intensity (MFI). MFI is proportional to bound IgG concentration. Software calculates Fluorescent Light Units (FLU) using lot-specific master curves and instrument-specific working curves. Results are reported as positive or negative based on a 5 FLU cut-off. The device is intended for clinical laboratory use to aid in diagnosing celiac disease and dermatitis herpetiformis, particularly in IgA-deficient patients.
Clinical Evidence
Clinical validation used 515 samples: 171 celiac disease (CD), 20 IgA-deficient CD, 34 dermatitis herpetiformis (DH), and 290 controls (autoimmune/infectious diseases). For CD diagnosis, Aptiva DGP IgG showed 82.2% sensitivity and 97.9% specificity; Aptiva tTG IgG showed 60.7% sensitivity and 100% specificity. For DH diagnosis, DGP IgG showed 70.6% sensitivity and 97.9% specificity; tTG IgG showed 26.5% sensitivity and 100% specificity. Method comparison with predicates showed TPA of 86.7% (DGP) and 86.8% (tTG).
Technological Characteristics
Particle-based multi-analyte technology (PMAT) using paramagnetic particles. Reagents include tTG/DGP-coated particles, PE-labeled anti-human IgG tracer, and buffers. Energy source: Aptiva System instrumentation. Connectivity: RFID-labeled reagent cartridges for master curve data. Software: Automated image-based particle counting and fluorescence quantification. Sterilization: Not specified.
Indications for Use
Indicated for semi-quantitative detection of anti-tTG IgG and anti-DGP IgG in human serum to aid in diagnosis of celiac disease and dermatitis herpetiformis, specifically in patients with selective IgA deficiency.
Regulatory Classification
Identification
A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).
Predicate Devices
QUANTA Flash DGP IgG (k113863)
QUANTA Flash h-tTG IgG (k101644)
Submission Summary (Full Text)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K200230
B Applicant
Inova Diagnostics, Inc.
C Proprietary and Established Names
Aptiva Celiac Disease IgG Reagent
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| MVM | Class II | 21 CFR 866.5660 - Multiple Autoantibodies Immunological Test System | IM - Immunology |
| MST | Class II | 21 CFR 866.5750 - Radioallergosorbent (RAST) immunological test system | IM - Immunology |
## II Submission/Device Overview:
A Purpose for Submission:
New device
B Measurand:
Anti-deaminated gliadin peptide (DGP) IgG antibodies
Anti-human tissue transglutaminase (h-tTG) IgG antibodies
C Type of Test:
Semi-quantitative particle-based multi-analyte technology (PMAT)
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below
B Indication(s) for Use:
The Aptiva Celiac Disease IgG Reagent is an immunoassay utilizing particle-based multi-analyte technology for the semi-quantitative determination of anti-tissue transglutaminase IgG autoantibodies and anti-deamidated gliadin peptide IgG autoantibodies in human serum. The presence of these antibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of celiac disease and dermatitis herpetiformis, particularly in patients with selective IgA deficiency.
The Aptiva Celiac Disease IgG Reagent is intended for use with the Inova Diagnostics Aptiva System.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
D Special Instrument Requirements:
Inova Diagnostics Aptiva System
IV Device/System Characteristics:
A Device Description:
The Aptiva Celiac Disease IgG Reagent is an immunoassay utilizing particle-based multi-analyte technology. The kit contains one Aptiva Celiac Disease IgG Reagent Cartridge with the following reagents for 250 determinations:
- tTG, DGP, and Control paramagnetic particles, preserved
- Assay buffer – containing protein stabilizers and preservatives
- PE Tracer IgG – PE labeled anti-human IgG antibody, containing buffer, protein stabilizers and preservative
- Rehydration Buffer – containing protein stabilizers and preservatives
The Aptiva Celiac Disease IgG Calibrators and Aptiva Celiac Disease IgG Controls are required but sold separately:
- The Aptiva Celiac Disease IgG Calibrators include three Calibrators: Calibrator 1, Calibrator 2, and Calibrator 3. Each Calibrator is provided with two barcode labeled tubes each containing 0.3 mL pre-diluted and ready to use reagent. Calibrators contain human antibodies to DGP and tTG in stabilizers and preservatives.
- The Aptiva Celiac Disease IgG Controls include two controls: Control 1 and Control 2. Each Control is provided with two barcode labeled tubes each containing 0.5 mL, ready to use reagent. Controls contain human antibodies to DGP and tTG in stabilizers and preservatives.
K200230 - Page 2 of 16
{2}
K200230 - Page 3 of 16
## B Principle of Operation:
Aptiva Celiac Disease IgG Reagent kit contains two different populations of particles; one particle coated with recombinant tissue transglutaminase antigen and one particle coated with a synthetic deamidated gliadin peptide and an additional third particle coated with goat anti-human IgG antibody as control verification.
The Aptiva System dilutes the patient sample 1:23 then combines an aliquot of diluted patient sample, and reagent into a cuvette. The mixture is incubated at 37°C. After a wash cycle, conjugated anti-human IgG antibody is added to the particles and this mixture is incubated at 37°C. Excess conjugate is removed in another wash cycle, and the particles are re-suspended in system fluid.
The system generates multiple images in order to identify and count the two unique analyte particles, as well as determine the amount of conjugate on each particle. A third particle, coated with goat anti-human IgG antibodies, is present in the reagent as a control to flag low concentrations of IgG in the sample as an assay verification step. The median fluorescent intensity (MFI) for each analyte is proportional to the concentration of conjugate bound to human IgG, which is proportional to the concentration of IgG antibodies bound to the corresponding particle region. The system uses the MFI from at least 50 particles of each region. The identity of the particles is determined by the unique signature of the particles. Each analyte in the Aptiva Celiac Disease IgG Reagent is assigned a predefined lot specific Master Curve. The analyte specific Master Curve is stored on the reagent cartridge RFID label (Radio-frequency identification). Based on results obtained by running calibrators (supplied separately), the system creates instrument specific Working Curves. Working curves are used by the software to calculate Fluorescent Light Units (FLU) for each analyte from the MFI values obtained for each sample.
Based on the defined cut-off for each analyte, the test results are reported for each sample as "positive" or "negative" with test value in FLU for each assay, i.e., DGP IgG and tTG IgG.
## V Substantial Equivalence Information:
### A Predicate Device Name(s):
- QUANTA Flash DGP IgA, QUANTA Flash DGP IgG, QUANTA Flash DGP IgA Calibrators, QUANTA Flash DGP IgG Calibrators
- QUANTA Flash h-tTG, QUANTA Flash h-tTG IgG Calibrators, QUANTA Flash h-tTG IgG Controls
### B Predicate 510(k) Number(s):
K113863
K101644
{3}
C Comparison with Predicate(s):
Comparison with DGP IgG:
| Device & Predicate Device(s): | K200230 Device | K113863 Predicate |
| --- | --- | --- |
| Device Trade Name | Aptiva Celiac Disease IgG Reagent (DGP IgG) | QUANTA Flash DGP IgG |
| General Device Characteristic Similarities | | |
| Assay Methodology | Solid phase (heterogenous) immunoassay | Same |
| Sample Matrix | Serum | Same |
| Solid Phase | Paramagnetic particles | Same |
| General Device Characteristic Differences | | |
| Intended Use / Indications for Use | The Aptiva Celiac Disease IgG Reagent is an immunoassay utilizing particle-based multi-analyte technology for the semi-quantitative determination of anti-tissue transglutaminase IgG autoantibodies and anti-deamidated gliadin peptide IgG autoantibodies in human serum. The presence of these antibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of celiac disease and dermatitis herpetiformis, particularly in patients with selective IgA deficiency. The Aptiva Celiac Disease IgG Reagent is intended for use with the Inova Diagnostics Aptiva System. | The QUANTA Flash DGP IgG is a chemiluminescent immunoassay for the semi-quantitative detection of IgG antibodies to synthetic, deamidated gliadin peptides in human serum. The presence of IgG deamidated gliadin peptide antibodies can be used in conjunction with clinical findings and other laboratory tests to aid in the diagnosis of celiac disease in both IgA sufficient and IgA deficient subjects, as well as dermatitis herpetiformis. |
| Assay Technology | Fluorescent immunoassay | Chemiluminescent immunoassay |
| Antigen | DGP and recombinant tTG | DGP |
| Detection Antibody (Conjugate) | Phycoerythrin conjugated polyclonal goat anti-human IgG antibody | Isoluminol conjugated monoclonal anti-human IgG antibody |
| Units | Fluorescent Light Units (FLU) | Chemiluminescent Units (CU) |
| Cut-off | 5 FLU | 20 CU |
| Analytical Measuring Range | DGP IgG: 0.56–250.00 FLU | DGP IgG: 2.8–1936.7 CU |
K200230 - Page 4 of 16
{4}
Comparison with tTG IgG:
| Device & Predicate Device(s): | K200230
Device | K101644
Predicate |
| --- | --- | --- |
| Device Trade Name | Aptiva Celiac Disease IgG Reagent (tTG IgG) | QUANTA Flash h-tTG IgG |
| General Device Characteristic Similarities | | |
| Assay Methodology | Solid phase (heterogenous) immunoassay | Same |
| Antigen | Recombinant tissue transglutaminase | Same |
| Sample Matrix | Serum | Same |
| Solid Phase | Paramagnetic particles | Same |
| General Device Characteristic Differences | | |
| Intended Use/Indications for Use | The Aptiva Celiac Disease IgG Reagent is an immunoassay utilizing particle-based multi-analyte technology for the semi-quantitative determination of anti-tissue transglutaminase IgG autoantibodies and anti-deamidated gliadin peptide IgG autoantibodies in human serum. The presence of these antibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of celiac disease and dermatitis herpetiformis, particularly in patients with selective IgA deficiency.
The Aptiva Celiac Disease IgG Reagent is intended for use with the Inova Diagnostics Aptiva System. | The QUANTA Flash™ h-tTG IgG is a chemiluminescent immunoassay (CIA) for the semi-quantitative detection of IgG anti-human tissue transglutaminase (h-tTG) antibodies in human serum. The presence of IgG anti-h-tTG antibodies, in conjunction with clinical findings and other laboratory tests, can aid in the diagnosis of the gluten sensitive enteropathy celiac disease, particularly in celiac patients with selective IgA deficiency. |
| Assay Technology | Fluorescent immunoassay | Chemiluminescent immunoassay |
K200230 - Page 5 of 16
{5}
| Detection Antibody (Conjugate) | Phycoerythrin conjugated polyclonal goat anti-human IgG antibody | Isoluminol conjugated monoclonal anti-human IgG antibody |
| --- | --- | --- |
| Units | Fluorescent Light Units (FLU) | Chemiluminescent Units (CU) |
| Cut-off | 5 FLU | 20 CU |
| Analytical Measuring Range | tTG IgG: 0.82–250.00 FLU | tTG IgG: 3.75–2560.0 CU |
| Controls | 2 Controls (tTG IgG) with lot specific values assigned:
Control 1: 8.18 FLU
Control 2: 16.39 FLU | 2 Controls (h-tTG IgG) with lot specific values assigned:
Negative Control: 11.4 CU (6.8–16.0)
Positive Control: 50.4 CU (30.2–70.6) |
| Calibration | Lot Specific Curve + 3 Calibrators (sold separately) | Lot Specific Curve + 2 Calibrators (sold separately) |
VI Standards/Guidance Documents Referenced:
- EP05-A3, Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline, Third Edition
- EP06-A, Evaluation of Linearity of Quantitative Measurement, Approved Guideline, Second Edition
- EP07-A2, Interference Testing in Clinical Chemistry, Approved Guideline, Second Edition
- EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline, Second Edition
- EP28-A3c, Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline, Third Addition
VII Performance Characteristics (if/when applicable):
A Analytical Performance: All results presented below met the manufacturer's pre-determined acceptance criteria
1. Precision/Reproducibility:
Within-Laboratory Precision:
The precision of the Aptiva Celiac Disease IgG reagent was evaluated in accordance with CLSI EP05-A3 using eight samples for DGP IgG and tTG IgG. Samples containing various concentrations of antibodies were run in duplicate, twice a day, for 20 days (total of 80 replicates per sample). The standard deviation (SD) and %CV of the within-run (repeatability), between-run, between-day and total within-laboratory precision were calculated for each sample and the results are summarized in the following tables:
K200230 - Page 6 of 16
{6}
| Aptiva DGP IgG Precision | | Repeatability | | Between-Run | | Between-Day | | Within-Laboratory | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | Mean (FLU) | SD (FLU) | CV | SD (FLU) | CV | SD (FLU) | CV | SD (FLU) | CV |
| 1 | 1.73 | 0.10 | 6.1% | 0.07 | 4.0% | 0.09 | 5.1% | 0.15 | 8.9% |
| 2 | 4.29 | 0.17 | 3.9% | 0.07 | 1.6% | 0.19 | 4.5% | 0.27 | 6.2% |
| 3 | 4.47 | 0.17 | 3.7% | 0.19 | 4.3% | 0.14 | 3.1% | 0.29 | 6.5% |
| 4 | 5.42 | 0.23 | 4.3% | 0.30 | 5.5% | 0.22 | 4.0% | 0.44 | 8.1% |
| 5 | 16.17 | 0.38 | 2.3% | 0.48 | 2.9% | 0.81 | 5.0% | 1.01 | 6.3% |
| 6 | 31.27 | 0.81 | 2.6% | 1.00 | 3.2% | 0.90 | 2.9% | 1.57 | 5.0% |
| 7 | 131.92 | 3.25 | 2.5% | 6.32 | 4.8% | 0.94 | 0.7% | 7.16 | 5.4% |
| 8 | 216.46 | 9.76 | 4.5% | 7.11 | 3.3% | 12.23 | 5.6% | 17.19 | 7.9% |
| Aptiva tTG IgG Precision | | Repeatability | | Between-Run | | Between-Day | | Within-Laboratory | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | Mean (FLU) | SD (FLU) | CV | SD (FLU) | CV | SD (FLU) | CV | SD (FLU) | CV |
| 1 | 1.88 | 0.13 | 7.1% | 0.00 | 7.1% | 0.14 | 3.0% | 0.14 | 7.7% |
| 2 | 2.43 | 0.15 | 6.2% | 0.04 | 1.7% | 0.12 | 5.1% | 0.20 | 8.2% |
| 3 | 4.53 | 0.23 | 5.1% | 0.05 | 1.2% | 0.18 | 3.9% | 0.30 | 6.5% |
| 4 | 5.29 | 0.27 | 5.1% | 0.00 | 0.0% | 0.15 | 2.8% | 0.31 | 5.8% |
| 5 | 11.49 | 0.42 | 3.7% | 0.12 | 1.0% | 0.44 | 3.8% | 0.62 | 5.4% |
| 6 | 46.27 | 1.40 | 3.0% | 1.04 | 2.3% | 1.02 | 2.2% | 2.02 | 4.4% |
| 7 | 117.59 | 6.73 | 5.7% | 2.14 | 1.8% | 2.51 | 2.1% | 7.49 | 6.4% |
| 8 | 210.59 | 14.77 | 7.0% | 0.00 | 0.0% | 9.17 | 4.4% | 17.38 | 8.3% |
# Lot to Lot Imprecision:
To evaluate lot-to-lot variability, six serum samples for DGP IgG and seven serum samples for tTG IgG were tested according to CLSI EP05-A3 using three different lots. Samples were run in replicates of five, once a day, for five days, to generate 25 data points per sample for each lot (75 data points total for each sample). The results are summarized in the tables below:
K200230 - Page 7 of 16
{7}
K200230 - Page 8 of 16
| Aptiva DGP IgG | | Repeatability | | Between-Day | | Within-Lot | | Between-Lot | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | Mean (FLU) | SD (FLU) | CV | SD (FLU) | CV | SD (FLU) | CV | SD (FLU) | CV | SD (FLU) | CV |
| 1 | 2.95 | 0.13 | 4.4% | 0.30 | 10.3% | 0.33 | 11.2% | 0.00 | 0.0% | 0.33 | 11.2% |
| 2 | 5.79 | 0.25 | 4.3% | 0.45 | 7.7% | 0.51 | 8.9% | 0.40 | 7.0% | 0.65 | 11.3% |
| 3 | 8.88 | 0.41 | 4.6% | 0.80 | 9.0% | 0.89 | 10.0% | 0.26 | 2.9% | 0.93 | 10.5% |
| 4 | 24.09 | 0.80 | 3.3% | 1.36 | 5.7% | 1.58 | 6.6% | 0.00 | 0.0% | 1.58 | 6.6% |
| 5 | 93.99 | 2.60 | 2.8% | 5.68 | 6.0% | 6.25 | 6.7% | 5.13 | 5.5% | 8.09 | 8.6% |
| 6 | 122.73 | 5.36 | 4.4% | 5.23 | 4.3% | 7.49 | 6.1% | 10.70 | 8.7% | 13.06 | 10.6% |
| Aptiva tTG IgG | | Repeatability | | Between-Day | | Within-Lot | | Between-Lot | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | Mean (FLU) | SD (FLU) | CV | SD (FLU) | CV | SD (FLU) | CV | SD (FLU) | CV | SD (FLU) | CV |
| 1 | 1.87 | 0.12 | 6.2% | 0.06 | 3.0% | 0.13 | 6.9% | 0.03 | 1.8% | 0.13 | 7.1% |
| 2 | 5.17 | 0.21 | 4.1% | 0.15 | 2.9% | 0.26 | 5.0% | 0.38 | 7.4% | 0.46 | 8.9% |
| 3 | 5.68 | 0.22 | 3.8% | 0.18 | 3.2% | 0.28 | 5.0% | 0.33 | 5.9% | 0.44 | 7.7% |
| 4 | 11.01 | 0.40 | 3.6% | 0.61 | 5.5% | 0.73 | 6.6% | 0.41 | 3.7% | 0.83 | 7.6% |
| 5 | 25.83 | 1.12 | 4.3% | 0.99 | 3.8% | 1.49 | 5.8% | 1.42 | 5.5% | 2.05 | 8.0% |
| 6 | 73.07 | 3.51 | 4.8% | 6.18 | 8.5% | 7.11 | 9.7% | 0.00 | 0.0% | 7.11 | 9.7% |
| 7 | 170.49 | 11.27 | 6.6% | 6.87 | 4.0% | 13.20 | 7.7% | 4.66 | 2.7% | 14.00 | 8.2% |
Site-to-Site Reproducibility:
Seven serum samples for DGP IgG and tTG IgG were tested according to CLSI EP05-A3 at three different sites using the same lot of Aptiva Celiac Disease IgG reagent. Samples were run in replicates of five, once a day, for five days, to generate 25 data points per sample at each site (75 data points per sample for all sites combined). The results are summarized in the following tables:
| Aptiva DGP IgG | | Repeatability | | Between-Day | | Within-Site | | Between-Site | | Reproducibility | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | Mean (FLU) | SD (FLU) | CV | SD (FLU) | CV | SD (FLU) | CV | SD (FLU) | CV | SD (FLU) | CV |
| 1 | 2.26 | 0.25 | 11.2% | 0.00 | 0.0% | 0.25 | 11.2% | 0.14 | 6.2% | 0.29 | 12.8% |
| 2 | 5.41 | 0.27 | 5.0% | 0.32 | 5.9% | 0.42 | 7.7% | 0.27 | 5.1% | 0.50 | 9.2% |
| 3 | 6.42 | 0.35 | 5.4% | 0.38 | 5.9% | 0.51 | 8.0% | 0.24 | 3.8% | 0.57 | 8.9% |
| 4 | 12.40 | 0.57 | 4.6% | 0.86 | 7.0% | 1.04 | 8.4% | 0.00 | 0.0% | 1.04 | 8.4% |
| 5 | 44.17 | 1.40 | 3.2% | 2.34 | 5.3% | 2.73 | 6.2% | 2.16 | 4.9% | 3.48 | 7.9% |
| 6 | 119.06 | 6.15 | 5.2% | 5.38 | 4.5% | 8.17 | 6.9% | 6.72 | 5.6% | 10.58 | 8.9% |
| 7 | 160.22 | 10.82 | 6.8% | 4.54 | 2.8% | 11.74 | 7.3% | 9.78 | 6.1% | 15.28 | 9.5% |
{8}
| Aptiva tTG IgG | | Repeatability | | Between-Day | | Within-Site | | Between-Site | | Reproducibility | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | Mean (FLU) | SD (FLU) | CV | SD (FLU) | CV | SD (FLU) | CV | SD (FLU) | CV | SD (FLU) | CV |
| 1 | 2.33 | 0.12 | 5.0% | 0.11 | 4.5% | 0.16 | 6.7% | 0.14 | 6.2% | 0.21 | 9.1% |
| 2 | 5.33 | 0.25 | 4.5% | 0.19 | 3.4% | 0.31 | 5.7% | 0.32 | 5.8% | 0.45 | 8.1% |
| 3 | 6.03 | 0.25 | 4.1% | 0.18 | 3.0% | 0.30 | 5.0% | 0.34 | 5.6% | 0.45 | 7.5% |
| 4 | 12.12 | 0.38 | 3.1% | 0.23 | 1.9% | 0.44 | 3.7% | 0.45 | 3.7% | 0.63 | 5.2% |
| 5 | 30.43 | 1.35 | 4.4% | 0.55 | 1.8% | 1.46 | 4.8% | 1.12 | 3.7% | 1.84 | 6.0% |
| 6 | 96.08 | 5.94 | 6.2% | 5.19 | 5.4% | 7.88 | 8.2% | 4.31 | 4.5% | 8.98 | 9.4% |
| 7 | 192.87 | 13.07 | 6.8% | 4.58 | 2.4% | 13.85 | 7.2% | 8.86 | 4.6% | 16.44 | 8.5% |
# 2. Linearity:
The linearity across the analytical measuring range (AMR) of each analyte was evaluated in accordance with CLSI EP6-A. The study was done by using four human serum samples with various anti-DGP and anti-tTG antibody concentrations which were combined with another human serum sample containing low levels of antibodies in $10\%$ increments (from $0\%$ to $90\%$ of low sample) to obtain values that cover the entire AMR. Each dilution was tested in duplicate. Percentage recovery of obtained mean results was calculated compared to the expected results. Results were analyzed according to the guideline performing regression analysis and identifying the best fitting polynomial.
The linear regression analysis was performed using the samples falling within the master curve and the results of samples within AMR are summarized in the following tables:
| DGP IgG Sample | Test Range (FLU) | Slope (95% CI) | Y-Intercept (95% CI) | R² | % Recovery |
| --- | --- | --- | --- | --- | --- |
| 1 | 54.85–274.25 | 1.04 (1.00–0.09) | -8.73 (-16.69– -0.77) | 0.99 | 93.4–103.4% |
| 2 | 7.13–71.33 | 1.04 (1.02–1.06) | -2.29 (-3.27– -1.31) | 1.00 | 80.0–101.5% |
| 3 | 4.82–48.20 | 1.04 (1.00–1.08) | -0.16 (-1.26–0.94) | 0.99 | 86.7–106.4% |
| 4 | 0.52–5.19 | 1.02 (0.99–1.05) | -0.15 (-0.24– -0.05) | 1.00 | 86.5–100.3% |
| Combined | 0.52–274.25 | 1.00 (0.99–1.01) | -0.50 (-1.49–0.50) | 1.00 | 80.0–106.4% |
For the anti-DGP IgG assay, the data support linearity for the claimed AMR from 0.56 FLU to 250.00 FLU.
K200230 - Page 9 of 16
{9}
| tTG IgG Sample | Test Range (FLU) | Slope (95% CI) | Y-Intercept (95% CI) | R² | % Recovery |
| --- | --- | --- | --- | --- | --- |
| 1 | 32.78–327.80 | 0.93 (0.88–0.99) | 13.20 (1.59–24.82) | 0.99 | 93.7–113.2% |
| 2 | 7.48–74.77 | 1.00 (0.96–1.05) | 1.40 (-0.72–3.53) | 0.99 | 100–113.2% |
| 3 | 0.99–9.91 | 1.03 (1.00–1.06) | -0.22 (-0.42– -0.02) | 1.00 | 80.2–102.4% |
| Combined | 0.99–327.80 | 0.98 (0.96–1.00) | 2.32 (-0.08–4.73) | 0.99 | 80.2–113.2% |
For the anti-tTG IgG assay, the data support linearity for the claimed AMR from 0.82 FLU to 250.00 FLU
## Auto-rerun:
To validate the auto-rerun function with 1:10 dilutions, two positive specimens with anti-DGP IgG (599.54 and 1085.45 FLU) and two positive specimens with anti-tTG IgG (1266.11 and 1705.55 FLU) concentrations well above the assay measuring range were run with the auto-rerun function enabled on the Aptiva instrument. The same set of samples were manually diluted 1:10, tested and used as reference values. Comparing the values obtained by the auto-rerun to the manual dilution, the recovery values ranged from 91.4% to 94.1% for DGP IgG, and 89.4% to 99.9% for tTG IgG.
## Hook Effect:
The possibility of hook effect for Aptiva Celiac Disease IgG Reagent on the Aptiva System was evaluated with testing three anti-DGP IgG high positive samples (378.94, 379.85 and 723.10 FLU) and three high positive samples with anti-tTG IgG (264.21, 405.53 and 465.52 FLU). No antigen excess hook effect was observed up to 723.10 FLU for DGP IgG and up to 465.52 FLU for tTG IgG.
## 3. Analytical Specificity/Interference:
An interference study was performed according to CLSI EP07-A2 using a set of three human serum specimens for DGP IgG: one positive (12.64 FLU), one near the cut-off (5.31 FLU) and one negative (1.14 FLU), and three serum specimens for tTG IgG: one positive (46.65 FLU); one near the cut-off (5.22 FLU); and one negative (2.29 FLU). Each sample was tested in triplicate with the spiked interfering substances in the table below. The percent (%) recovery for the sample spiked with the potential interfering substance was calculated by comparing its result to that of the corresponding control sample without the interfering substance. For both assays, no significant interference (defined as % recovery between 90–110%) was observed in: 1 mg/mL for bilirubin, 2 mg/mL for hemoglobin, 1000 mg/dL for triglycerides, 332.5 mg/dL for cholesterol, and 250 IU/mL for rheumatoid factor (IgM). In the DGP assay, human IgG at a concentration of 35 mg/mL showed % recovery of 114.1% in the sample around the cut-off.
K200230 - Page 10 of 16
{10}
4. Assay Reportable Range:
DGP IgG assay: 0.56 FLU–250.00 FLU
tTG IgG assay: 0.82 FLU–250.00 FLU
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Traceability:
There is no international reference material for anti-DGP or anti-tTG IgG antibodies that allows for the standardization of anti-DGP or anti-tTG IgG antibodies detection assays. Calibrator and control values are directly traceable to in-house standards that are used to create the master curves for the Aptiva Celiac Disease IgG Reagent.
Stability:
Kit stability (unopened): An accelerated stability study the Aptiva Celiac Disease IgG Reagent was performed using three lots stored at 37°C±3°C for 5 weeks where one week is equal to 6 months at 5±3°C. Seven samples were run in duplicate for DGP IgG, and 12 samples run in duplicate for tTG IgG. The accelerated stability study supports a claim for a 24-month shelf-life.
Real time: Real-time stability study was performed at 3–6-month intervals on the Aptiva Celiac Disease IgG reagents stored at 2–8°C. At each timepoint, a low positive sample, a moderately high sample and a high positive patient sample (DGP IgG: 5.83, 12.58 and 47.14 FLU; and tTG IgG: 8.09, 15.03 and 51.23 FLU) were tested at time zero, 7 months, 12 months, 20 months, 24 months, and 25 months. The two lots support a 24-month stability claim.
On-board (In-use) stability: An on-board stability study was performed for the reagent cartridge. One lot of the reagent cartridge was tested using a set of human serum samples: seven DGP IgG; 12 tTG IgG. In addition, 14 samples of IgG Control bead were tested. The specimens were tested periodically for 31 days at each timepoint with a total of 12 timepoints for the DGP IgG, tTG IgG reagent cartridges and IgG Control beads. At day 15, the reagent cartridge was recalibrated. The results support an on-board stability claim of 28 days.
Sample stability: Six samples ranging from 4.1 FLU to 103.7 FLU for DGP IgG and seven samples ranging from 2.7 FLU to 110.4 FLU for tTG IgG were tested. All samples were tested in duplicate for up to 21 days while stored at 2–8°C, up to 48 hours while stored at room temperature (RT), and after repeated freeze/thaw cycles up to six cycles. Results were compared to those obtained on control samples (time zero / zero cycles). The results support sample stability for 48 hours at RT, 14 days at 2–8°C, and up to five freeze/thaw cycles when samples are stored at or below -20°C.
K200230 - Page 11 of 16
{11}
K200230 - Page 12 of 16
## 6. Detection Limit:
The Limit of Blank (LoB) was determined by assaying eight blank samples in five replicates per sample over three days with two reagent lots. One hundred twenty data points were generated for each lot. LoB was calculated as the 95th percentile using the non-parametric method for each lot as the dataset showed non-normal distribution. The highest LoB was determined to be 0.02 FLU for both DGP IgG and tTG IgG.
The Limit of Detection (LoD) was determined by using four samples with low level of anti-DGP IgG and anti-tTG IgG antibodies. Each sample was tested in five replicates over three days with two reagent lots. LoD was calculated as the LoB+1.652 x SD of the replicates for the samples. The LoD was determined to be 0.15 FLU for DGP IgG and 0.13 FLU for tTG IgG.
The Limit of Quantitation (LoQ) was determined by assaying four low level samples of anti-DGP IgG and anti-tTG IgG with two reagent lots. Each sample was run in replicates of five, twice per day, for three days, with a total of 120 data points generated on each assay, on each reagent lot. The LoQ was defined to be the lowest concentration level that meets the within-laboratory imprecision of < 20% for each lot. The LoQ for each assay was determined as the highest LoQ across the two lots and set as the lower limit of the AMR. The claimed LoQ is 0.56 FLU for the DGP IgG assay and 0.82 FLU for the tTG IgG assay.
## 7. Assay Cut-Off:
The Aptiva Celiac Disease IgG Reagent cut-off was established by testing 11 celiac disease samples and 192 presumably negative autoimmune disease samples [30 infectious disease, 20 systemic sclerosis, 13 systemic lupus erythematosus, 15 Crohn's disease, 29 autoimmune thyroid disease, 30 primary biliary cholangitis, 40 rheumatoid arthritis, and 15 ulcerative colitis samples]. The assay cut-off was established as 5.00 FLU (109 MFI for DGP IgG and 175 MFI for tTG IgG) to ensure optimal differentiation between DGP IgG and tTG IgG positive and negative samples.
For both DGP IgG and tTG IgG, an interpretation of "positive" is for samples ≥ 5.00 FLU and an interpretation of "negative" is for samples with FLU < 5.00 FLU.
## B Comparison Studies:
### 1. Method Comparison with Predicate Device:
All samples (n=515) from the clinical validation study (described in Section C.1. below) were tested on the Aptiva Celiac Disease IgG Reagent and on their predicates, the QUANTA Flash DGP IgG and tTG IgG assays. Among them, 218 samples have anti-DGP IgG levels within the AMR for DGP IgG of the Aptiva Celiac Disease IgG Reagent and the QUANTA Flash DGP IgG, and 265 samples have anti-tTG IgG levels within the AMR for tTG IgG of the Aptiva Celiac Disease IgG Reagent and the QUANTA Flash tTG IgG. The positive percent agreement (PPA), negative percent agreement (NPA) and total percent agreement (TPA) were calculated by comparing the Aptiva Celiac Disease IgG Reagent with the predicate devices on samples within the AMR. The results are summarized in the following tables for each analyte.
{12}
| DGP IgG Method Comparison (N=218) | QUANTA Flash DGP IgG | | | |
| --- | --- | --- | --- | --- |
| | | Negative | Positive | Total |
| Aptiva DGP IgG | Negative | 48 | 4 | 52 |
| | Positive | 25* | 141 | 166 |
| | Total | 73 | 145 | 218 |
| NPA: 65.8% (48/73) (95% CI: 54.3–75.6%) PPA: 97.2% (141/145) (95% CI: 93.1–98.9%) TPA: 86.7% [(48+141)/218] (95% CI: 81.5–90.6%) | | | | |
* 18 out of 25 samples tested as negative by the predicate Quanta Flash DGP IgG, but positive by the Aptiva DGP IgG were celiac disease samples; seven (7) of the discordant samples (28%) were within +25% of the Aptiva DGP assay cut-off.
| tTG IgG Method Comparison (N=265) | QUANTA Flash tTG IgG | | | |
| --- | --- | --- | --- | --- |
| | | Negative | Positive | Total |
| Aptiva tTG IgG | Negative | 139 | 8 | 147 |
| | Positive | 27** | 91 | 118 |
| | Total | 166 | 99 | 265 |
| NPA: 83.7% (139/166) (95% CI: 77.4–88.6%) PPA: 91.9% (91/99) (95% CI: 84.9–95.8%) TPA: 86.8% [(139+91)/265] (95% CI: 82.2–90.3%) | | | | |
**Among 27 samples tested as negative by the predicate Quanta Flash tTG IgG, but positive by the Aptiva tTG IgG Assay, 24 were celiac disease samples (six of 27 were IgA deficient celiac disease samples) and three (3) were dermatitis herpetiformis samples); eleven (11) of the discordant samples (41%) were within +25% of the Aptiva tTg assay cut-off.
2. Matrix Comparison:
Not Applicable
C Clinical Studies:
1. Clinical Sensitivity and Clinical Specificity:
A cohort of 515 characterized samples (none of which were used to establish the reference range), were used to validate the clinical performance of the Aptiva Celiac Disease IgG Reagent. This validation set included 171 samples from celiac disease (CD) patients, 20 samples from patients with IgA deficient celiac disease, 34 samples from dermatitis herpetiformis (DH) patients, and 290 samples from patients with various types of
K200230 - Page 13 of 16
{13}
autoimmune and infectious diseases. The distribution of the cohort and the DGP and tTG positivity rate are shown in the Table below:
| Patient Group | N | DGP IgG N Positive | DGP IgG % Positive | tTG IgG N Positive | tTG IgG % Positive |
| --- | --- | --- | --- | --- | --- |
| Celiac Disease | 171 | 142 | 83.0% | 100 | 58.5% |
| IgA Deficient Celiac Disease | 20 | 15 | 75.0% | 16 | 80.0% |
| Dermatitis Herpetiformis | 34 | 24 | 70.6% | 9 | 26.5% |
| Total | 225 | - | - | - | - |
| | | | | | |
| Rheumatoid Arthritis | 69 | 4 | 5.8% | 0 | 0.0% |
| Ulcerative Colitis | 31 | 0 | 0.0% | 0 | 0.0% |
| Crohn's Disease | 31 | 0 | 0.0% | 0 | 0.0% |
| Hepatitis C Virus | 28 | 0 | 0.0% | 0 | 0.0% |
| Hepatitis B Virus | 25 | 0 | 0.0% | 0 | 0.0% |
| Syphilis | 21 | 1 | 4.8% | 0 | 0.0% |
| Sjögren’s Syndrome | 20 | 0 | 0.0% | 0 | 0.0% |
| Systemic Sclerosis | 19 | 0 | 0.0% | 0 | 0.0% |
| Autoimmune Gastritis | 15 | 0 | 0.0% | 0 | 0.0% |
| Human Immunodeficiency Virus | 13 | 0 | 0.0% | 0 | 0.0% |
| Systemic Lupus Erythematosus | 12 | 0 | 0.0% | 0 | 0.0% |
| Epstein-Barr Virus | 6 | 1 | 16.7% | 0 | 0.0% |
| Total Controls | 290 | 6 | 2.1% | 0 | 0.0% |
Clinical sensitivity and specificity for the Aptiva DGP IgG and Aptiva tTG IgG in diagnosis of celiac disease (CD) were analyzed and results are shown in the following tables:
| | Diagnosis | | | |
| --- | --- | --- | --- | --- |
| | | CD* | Non-CD† | Total |
| Aptiva DGP IgG | Positive | 157 | 6 | 163 |
| | Negative | 34 | 284 | 318 |
| | Total | 191* | 290 | 481 |
| Sensitivity: 82.2% (157/191) (95% CI:76.2–87.0%)
Specificity: 97.9% (284/290) (95% CI: 95.6–99.0%) | | | | |
* The 191 CD patient samples in the study above include 20 samples from CD patients with selective IgA deficiency: 15 out of 20 tested positive on DGP IgG for a sensitivity of 75% (53.1–88.1%).
† Non-CD (290 samples) does not include the dermatitis herpetiformis (DH) samples
K200230 - Page 14 of 16
{14}
| | Diagnosis | | | |
| --- | --- | --- | --- | --- |
| | | CD* | Non-CD† | Total |
| Aptiva tTG IgG | Positive | 116 | 0 | 116 |
| | Negative | 75 | 290 | 365 |
| | Total | 191** | 290 | 481 |
| Sensitivity: 60.7% (116/191) 95% CI: 53.7–67.4% Specificity: 100% (290/290) 95% CI: 98.7–100% | | | | |
** The 191 CD patient samples in the study above include 20 samples from CD patients with selective IgA deficiency: 16 out of 20 tested positive on tTG IgG for a sensitivity of $80.0\%$ (58.4-91.9%).
† Non-CD (290 samples) does not include the DH samples
Clinical sensitivity and specificity for the Aptiva DGP IgG and Aptiva tTG IgG in diagnosis of dermatitis herpetiformis (DH) were analyzed and results are shown in the following tables:
| | Diagnosis | | | |
| --- | --- | --- | --- | --- |
| | | DH | Non-DH | Total |
| Aptiva DGP IgG | Positive | 24 | 6 | 30 |
| | Negative | 10 | 284 | 294 |
| | Total | 34 | 290 | 324 |
| Sensitivity:70.6% (24/34) 95% CI: 53.8–83.2% Specificity: 97.9% (284/290) 95% CI: 95.6–99.0% | | | | |
| | Diagnosis | | | |
| --- | --- | --- | --- | --- |
| | | DH | Non-DH | Total |
| Aptiva tTG IgG | Positive | 9 | 0 | 9 |
| | Negative | 25 | 290 | 315 |
| | Total | 34 | 290 | 324 |
| Sensitivity: 26.5% (9/34) 95% CI: 14.6–43.1% Specificity: 100% (290/290) 95% CI: 98.7–100% | | | | |
D Clinical Cut-Off:
Same as assay cut-off
K200230 - Page 15 of 16
{15}
K200230 - Page 16 of 16
E Expected Values/Reference Range:
A panel of 120 apparently healthy blood donors (64 females/56 males, ages 17 to 57 years, with an average and median age of 32 years) were tested on the Aptiva Celiac Disease IgG Reagent.
For DGP IgG, with a cut-off of 5.00 FLU, the cohort ranged from 0.03 to 8.06 FLU; three samples were positive.
For tTG IgG, with a cut-off of 5.00 FLU, the cohort ranged from 0.05 to 3.76 FLU; no samples were positive.
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.