Mirus Ligating Clip, Mirus Ligating Clip Applier

K200166 · M/s. Meril Endo Surgery Private Limited. · FZP · Nov 23, 2020 · General, Plastic Surgery

Device Facts

Record IDK200166
Device NameMirus Ligating Clip, Mirus Ligating Clip Applier
ApplicantM/s. Meril Endo Surgery Private Limited.
Product CodeFZP · General, Plastic Surgery
Decision DateNov 23, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4300
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Mirus™ Ligating clip is intended for use in open general surgery procedures on tubular structures or vessels wherever a metal ligating clip is indicated. The tissue being ligated should be consistent with size of the clip. Mirus™ Ligating Clip applier is intended to apply ligating clips on tubular structures or vessels wherever a metal ligating clip is indicated.

Device Story

Mirus Ligating Clip is a sterile, single-use implantable device; used in open general surgery to ligate tubular structures or vessels. Device consists of titanium clips housed in plastic cartridges and a multi-use stainless steel applier. Surgeon manually operates the applier to place and deform the clip around the target vessel/tissue. Output is a secured, ligated structure. Benefit is effective hemostasis or vessel closure during surgical procedures.

Clinical Evidence

Bench testing only. Performance evaluated via dimension compliance, shape, radial/axial pull-off force, leakage, applier performance, and comprehensive biocompatibility testing (cytotoxicity, sensitization, intracutaneous reactivity, acute/sub-acute systemic toxicity, intramuscular implantation, mutagenicity, hemolysis, and pyrogenicity).

Technological Characteristics

Materials: Implant-grade titanium (ISO 5832-2 / ASTM F 67) for clips; stainless steel for applier. Principle: Mechanical ligation via clip deformation. Form factor: Various clip sizes in plastic cartridges; manual surgical applier. Sterilization: Not specified for clips (implied sterile), applier is non-sterile. Connectivity: None.

Indications for Use

Indicated for patients undergoing open general surgery requiring ligation of tubular structures or vessels. Contraindicated if tissue size is inconsistent with clip size.

Regulatory Classification

Identification

An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION". November 23, 2020 M/s. Meril Endo Surgery Private Limited Umesh Sharma General Manager - Ouality Assurance/Regulatory Affairs Third Floor , E1-E3, Meril Park, Survey No. 135/2/B & 174/2, Muktanand Marg Chala, Vapi, 396191 In Re: K200166 Trade/Device Name: Mirus Ligating Clip, Mirus Ligating Clip Applier Regulation Number: 21 CFR 878.4300 Regulation Name: Ligating clip Regulatory Class: Class II Product Code: FZP Dated: October 14, 2020 Received: October 20, 2020 Dear Umesh Sharma: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {2}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Cindy Chowdhury, Ph.D., M.B.A. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K200166 Device Name Mirus™ Ligating Clip Indications for Use (Describe) The Mirus™ Ligating clip is intended for use in open general surgery procedures on tubular structures or vessels wherever a metal ligating clip is indicated. The tissue being ligated should be consistent with size of the clip. Mirus™ Ligating Clip applier is intended to apply ligating clips on tubular structures or vessels wherever a metal ligating clip is indicated. Type of Use (Select one or both, as applicable) | <span style="font-size: 10pt;"> <span style="font-family: Symbol;">V</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | |------------------------------------------------------------------------------------------------------------------------------------| | <span style="font-size: 10pt;"> <span style="font-family: Symbol;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | X Prescription Use (Part 21 CFR 801 Subpart D) _ Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/3/Picture/1 description: The image contains the logo for Meril Endo Surgery. The logo features the word "Meril" in a clean, sans-serif font, with a small yellow dot above the "i". Below the logo, the words "Endo Surgery" are written in a smaller, gray font. The background of the logo is a light blue color. ## 510(k) Summary #### Submitter l. M/s. Meril Endo Surgery Private Limited Third Floor, E1 – E3, Meril Park, Survey No. 135/2/B & 174/2, Muktanand Marg, Chala, Vapi – 396191 Gujarat, India. Tel. No: +91-260-3052100 Extension: 3811, Fax: +91-260-3052125 Web site: www.merillife.com E-mail: umesh.sharma@merillife.com Contact Person: Umesh Sharma Date Prepared: November 10tt, 2020 #### II. Device | Trade Name | Mirus™ Ligating Clip & Clip Applier | | |---------------------|-------------------------------------|-----------------------------------------------| | Panel | General and Plastic Surgery | | | Common/ Usual Name | Ligating clip | Clip Applier | | Classification Name | Clip Implantable | Manual surgical instrument for<br>general use | | Regulation Number | 21 CFR 878.4300 | 21 CER 878.4800 | | Classification | Class II | Class I | | Product Code | FZP | GAG | ## III. Predicate Device K834267: Ligaclip™ Titanium Ligating Clip, Ethicon, Inc.(Primary Predicate Device) K982313: Horizon™ Titanium Clips, Weck Closure Systems {5}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for Meril Endo Surgery. The word "Meril" is in a large, sans-serif font, with the "i" dotted with an orange circle. The words "Endo Surgery" are in a smaller, sans-serif font below the word "Meril". The background is a light blue color. ### IV. Device Description Mirus™ Ligating clip are sterile, single use, implantable device made of implant grade titanium (ISO 5832-2 / ASTM F 67). The clips are available in various sizes with 6 clips packed in a single plastic cartridge. Mirus™ Ligating clips are supplied with multiuse non sterile stainless steel Mirus™ Ligating Clip applier, allows the end user to ligate a wide range of vessels and tissue structures using the ligating clips #### V. Intended Use The Mirus™ Ligating clip is intended for use in open general surgery procedures on tubular structures or vessels wherever a metal ligating clip is indicated. The tissue being ligated should be consistent with size of the clip. Mirus™ Ligating Clip applier is intended to apply ligating clips on tubular structures or vessels wherever a metal ligating clip is indicated. ### VI. Substantial Equivalence The device design, material of construction, performance, packaging and intended uses are similar to the predicate device. Substantial equivalence is conducted based on the following parameters: - 1. Intended use - 2. Device Size - 3. Principle of Operation - 4. Single use - 5. Sterilisation method - 6. Packaging & Labelling - 7. Performance {6}------------------------------------------------ Image /page/5/Picture/1 description: The image contains a logo for a company called "Meril Endo Surgery". The word "Meril" is in a large, sans-serif font and is white. Above the "i" in Meril are two small yellow dots. Below the logo is the text "Endo Surgery" in a smaller, sans-serif font and is gray. ### VII. Performance Data Mirus™ Ligating Clip was subjected to the performance testing. The safety and effectiveness of Mirus™ Ligating Clip has been evaluated for the following performance and safety requirements. - 1. Dimension compliance - 2. Shape - 3. Radial pull off force - 4. Axial Pull off force - 5. Leakage - 6. Applier Performance - 7. Biocompatibility - a. In Vitro Cytotoxicity Study - b. Skin Sensitization Study - Intracutaneous Reactivity Test c. - d. Acute Systemic Toxicity Study - e. Sub Acute Systemic Toxicity - f. Intramuscular Implantation Test - Bacterial Reverse Mutation Test ರು. - h. In Vitro Hemolysis Test - Material Mediated Pyrogen Test i. #### Conclusion VIII. Mirus™ Ligating Clips are composed of implant grade Titanium. Also Mirus™ Ligating Clips have the same intended use and the same basic technology as that of the predicate devices. Meril's device is substantially equivalent to the predicate in term of material, features and design and it does not pose any new issues concerning safety and effectiveness.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...