# 3MP Monochrome Digital Mammography LCD Monitor MS-S300 (K200161)

_Jvckenwood Corporation · PGY · Feb 21, 2020 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K200161

## Device Facts

- **Applicant:** Jvckenwood Corporation
- **Product Code:** [PGY](/productcode/PGY.md)
- **Decision Date:** Feb 21, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

MS-S300 is intended to be used in displaying medical images for diagnosis by trained medical practitioners or certified personnel. It's intended to be used in digital mammography PACS, digital breast tomosynthesis and modalities including FFDM.

## Device Story

MS-S300 is a 21.3-inch monochrome LCD monitor for medical image display; inputs digital video signals (DVI-D, DisplayPort) from PACS/imaging modalities; transforms signals into grayscale images compliant with DICOM GSDF; features built-in front sensor, ambient light sensor, and human presence sensor for luminance stabilization and power management; utilizes digital uniformity correction to minimize luminance unevenness; operated by clinicians in clinical environments; output used for diagnostic interpretation of mammography and tomosynthesis images; benefits include high-luminance, stable, and uniform image quality for accurate diagnosis.

## Clinical Evidence

Bench testing only. Performance validated against AAPM-TG18 guidelines, ISO13406-2, and VESA 2001 standards. Tests included MTF measurement, pixel defect analysis, luminance response/stability, DICOM GSDF conformance, angular luminance dependency, uniformity, chromaticity, spatial noise, reflection coefficients, and veiling glare.

## Technological Characteristics

21.3-inch TFT Monochrome LCD (IPS) panel; LED backlighting; 3MP (2048x1536/1536x2048) resolution; 10.3-bit grayscale (1276 gradations); built-in front, ambient light, and human presence sensors; digital uniformity correction; DVI-D and DisplayPort connectivity; DICOM GSDF calibration; AAPM-TG18 compliant.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- 21.3 inch (54cm) Monochrome LCD Monitor MS35i2 ([K133686](/device/K133686.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{1}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.

February 21, 2020

JVCKENWOOD Corporation % Hideki Tengeiji Senior Manager 3-12, Moriya-cho, Kanagawa-ku Yokohama, Kanagawa 221-0022 JAPAN

Re: K200161

Trade/Device Name: 3MP Monochrome Digital Mammography LCD Monitor MS-S300 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: January 16, 2020 Received: January 22, 2020

Dear Hideki Tengeiji:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{2}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

### Indications for Use

510(k) Number (if known)

K200161

Device Name

3MP Monochrome Digital Mammography LCD Monitor MS-S300

Indications for Use (Describe)

MS-S300 is intended to be used in displaying medical images for diagnosis by trained medical practitioners or certified personnel. It's intended to be used in digital mammography PACS, digital breast tomosynthesis and modalities including FFDM.

| Type of Use (Select one or both, as applicable)                                                                                                                                                                         |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">❌</span> Prescription Use (Part 21 CFR 801 Subpart D)     </span> <span> <span style="font-size:16px">❍</span> Over-The-Counter Use (21 CFR 801 Subpart C)     </span> </div> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

 Department of Health and Human Services  

 Food and Drug Administration  

 Office of Chief Information Officer  

 Paperwork Reduction Act (PRA) Staff  

 PRAStaff@fda.hhs.gov

*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*

{4}------------------------------------------------

## JVCKENWOOD

# 510(k) SUMMARY K200161

| Submitted Information: | JVCKENWOOD Corporation<br>3-12, Moriya-cho, Kanagawa-ku,<br>Yokohama-shi, Kanagawa, 221-0022 Japan                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |  |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Contact Person:        | Hideki Tengeiji, Senior Manager<br>Email: tengeiji.hideki@jvckenwood.com<br>Tel: +81-45-450-2715<br>Fax: +81-45-450-1926                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |  |
| Date Prepared:         | January 16, 2020                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |  |
| Device Name:           | 3MP Monochrome Digital Mammography LCD Monitor MS-S300                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |  |
| Common Name:           | MS-S300 (MS-S300xxxxx)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |  |
| Classification Name:   | Class II<br>(Part 892 Radiology Devices<br>Sec. 892.2050 Picture Archiving and Communication System)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |  |
| Predicate Device:      | 21.3 inch (54cm) Monochrome LCD Monitor MS35i2<br>(MS35i2/ K133686)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |  |
| Device Description:    | 21.3 inch Monochrome Digital Mammography LCD Monitor<br>2048 x 1536 (landscape), 1536 x 2048 (portrait)<br>■ High-luminance monochrome LCD panel, which has wide view<br>angle, is used for this product. It is designed for medical image<br>display.<br>■ Luminance stabilization function composed with luminance<br>sensor and luminance control circuit always observes the<br>luminance and makes it stable.<br>■ Images are faithfully displayed along grayscale characteristics<br>(DICOM GSDF) based on the calibrated data stored to the lookup<br>table of the monitor.<br>■ It minimizes luminance unevenness by Uniformity Correction<br>Function to achieve the uniformity of luminance on the whole<br>screen.<br>■ Quantitative evaluation and visual evaluation are done before<br>the shipment. Quality control along the QC guideline is conducted. |  |

{5}------------------------------------------------

## JVCKENWOOD

- MS-S300 is intended to be used in displaying and viewing medical Intended Use: images for diagnosis by trained medical practitioners or certified personnel. lt's intended to be used in digital mammography PACS, digital breast tomosynthesis and modalities including FFDM.
- MS-S300 shares the same technical characteristics, application, and Substantial Equivalence: intended use as our predicate device MS35i2/ K133686.

MS-S300 is substantially equivalent to the predicate device with Conclusion: respect to technical characteristics, application and intended use. The specification of the primary component employed by the proposed device is the same as one of the predicate device, and other differences have been independently validated. Any differences between the devices do not affect safety or effectiveness.

> JVCKENWOOD Corporation Healthcare Business Division 3-12, Moriya-cho, Kanagawa-ku, Yokohama-shi, Kanagawa, 221-0022 Japan

{6}------------------------------------------------

### Substantial Equivalence Comparison

This comparison table shows the differences on the technical characteristics between the proposed device and the predicate device.

### <MS-S300>

|                                      | Predicate device<br>LCD Monitor MS35i2<br>(ML21035)                                                                                                                                                                                                                                                                                      | LCD Monitor MS-S300                                                                                                                                                                                                                                                                                                                      | Explanation of<br>Differences                                                |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| 510(k) Number                        | K133686                                                                                                                                                                                                                                                                                                                                  | —                                                                                                                                                                                                                                                                                                                                        | —                                                                            |
| Indication for Use                   | MS35i2 is intended to be used in<br>displaying and viewing medical images for<br>diagnosis by trained medical practitioners<br>or certified personnel. It is to be used in<br>digital mammography PACS and<br>modalities including FFDM.                                                                                                 | MS-S300 is intended to be used in<br>displaying and viewing medical images for<br>diagnosis by trained medical practitioners<br>or certified personnel. It is intended to be<br>used in digital mammography PACS,<br>digital breast tomosynthesis and modalities<br>including FFDM.                                                      | —                                                                            |
| Response Time<br>(typical)           | 40ms (On/Off)                                                                                                                                                                                                                                                                                                                            | 28ms (On/Off)                                                                                                                                                                                                                                                                                                                            | Response Time of MS-S300 is better<br>than the predicate's one.              |
| Resolution or<br>Matrix Size         | 3MP(1536 x 2048)                                                                                                                                                                                                                                                                                                                         | 3MP(1536 x 2048)                                                                                                                                                                                                                                                                                                                         | —                                                                            |
| Screen Technology                    | TFT Monochrome LCD Panel (IPS)                                                                                                                                                                                                                                                                                                           | TFT Monochrome LCD Panel (IPS)                                                                                                                                                                                                                                                                                                           | —                                                                            |
| Backlighting                         | LED                                                                                                                                                                                                                                                                                                                                      | LED                                                                                                                                                                                                                                                                                                                                      | —                                                                            |
| Maximum Luminance                    | Typ. 1700cd/m2                                                                                                                                                                                                                                                                                                                           | Typ. 2000cd/m2                                                                                                                                                                                                                                                                                                                           | The maximum luminance of MS-S300<br>is much better than the predicate's one. |
| DICOM<br>Calibrated Luminance        | 500cd/m2                                                                                                                                                                                                                                                                                                                                 | 500cd/m2                                                                                                                                                                                                                                                                                                                                 | —                                                                            |
| Viewing Angle                        | CR>10<br>Horizontal: Typ.176<br>Vertical: Typ.176                                                                                                                                                                                                                                                                                        | CR>10<br>Horizontal: Typ.178<br>Vertical: Typ.178                                                                                                                                                                                                                                                                                        | Viewing angle of MS-S300 is wider<br>than the predicate's one.               |
| Display Area                         | Horizontal: 324.86mm<br>Vertical: 433.15mm                                                                                                                                                                                                                                                                                               | Horizontal: 324.86mm<br>Vertical: 433.15mm                                                                                                                                                                                                                                                                                               | —                                                                            |
| Aspect Ratio                         | 3:4                                                                                                                                                                                                                                                                                                                                      | 3:4                                                                                                                                                                                                                                                                                                                                      | —                                                                            |
| Pixel Pitch                          | Horizontal: 0.2115mm<br>Vertical: 0.2115mm                                                                                                                                                                                                                                                                                               | Horizontal: 0.2115mm<br>Vertical: 0.2115mm                                                                                                                                                                                                                                                                                               | —                                                                            |
| Contrast Ratio                       | Typ. 1400:1                                                                                                                                                                                                                                                                                                                              | Typ. 1500:1                                                                                                                                                                                                                                                                                                                              | Contrast ratio of MS-S300 is wider than the predicate's one                  |
| Grayscale Tones                      | 10.3 bit: (1276 gradation)                                                                                                                                                                                                                                                                                                               | 10.3 bit: (1276 gradation)                                                                                                                                                                                                                                                                                                               | —                                                                            |
| Non-Uniformity<br>Compensation       | Digital Uniformity Correction System                                                                                                                                                                                                                                                                                                     | Digital Uniformity Correction System                                                                                                                                                                                                                                                                                                     | —                                                                            |
| Input Video Signal                   | DVI-D x1<br>DisplayPort x1                                                                                                                                                                                                                                                                                                               | DVI-D x1<br>DisplayPort x1                                                                                                                                                                                                                                                                                                               | —                                                                            |
| Scanning Frequency                   | DVI<br>46.6KHz, Vertical: 30Hz (Landscape)<br>61.9KHz, Vertical: 30Hz (Portrait)<br>93.1KHz, Vertical: 60Hz (Landscape)<br>123.9KHz, Vertical: 60Hz (Portrait)<br>DisplayPort<br>47.4KHz, Vertical: 30Hz (Landscape)<br>63.2KHz, Vertical: 30Hz (Portrait)<br>94.8KHz, Vertical: 60Hz (Landscape)<br>126.3KHz, Vertical: 60Hz (Portrait) | DVI<br>46.6KHz, Vertical: 30Hz (Landscape)<br>61.9KHz, Vertical: 30Hz (Portrait)<br>93.1KHz, Vertical: 60Hz (Landscape)<br>123.9KHz, Vertical: 60Hz (Portrait)<br>DisplayPort<br>47.4KHz, Vertical: 30Hz (Landscape)<br>63.2KHz, Vertical: 30Hz (Portrait)<br>94.8KHz, Vertical: 60Hz (Landscape)<br>126.3KHz, Vertical: 60Hz (Portrait) | —                                                                            |
| Dot Clock                            | 216 MHz                                                                                                                                                                                                                                                                                                                                  | 216 MHz                                                                                                                                                                                                                                                                                                                                  | —                                                                            |
| Power Requirements                   | AC100-240V, 50/60Hz                                                                                                                                                                                                                                                                                                                      | AC100-240V, 50/60Hz                                                                                                                                                                                                                                                                                                                      | —                                                                            |
| Power Consumption                    | 60W<br>Less than 2W                                                                                                                                                                                                                                                                                                                      | 55W<br>Less than 2W                                                                                                                                                                                                                                                                                                                      | Power consumption on MS-S300 is less than the predicate's one.               |
| Power Management                     | DVI DMPM,<br>DisplayPort 1.1a                                                                                                                                                                                                                                                                                                            | DVI DMPM,<br>DisplayPort 1.2a                                                                                                                                                                                                                                                                                                            | —                                                                            |
| QA Software                          | Medivisor NX<br>F-CAL                                                                                                                                                                                                                                                                                                                    | QA Medivisor Agent<br>F-CAL                                                                                                                                                                                                                                                                                                              | "QA Medivosior Agent" is the later software of "Medivisor NX".               |
| Sensor                               | Built- in Front Sensor<br>Built-in ambient Light Sensor                                                                                                                                                                                                                                                                                  | Built-in front sensor<br>Built-in ambient light sensor<br>Built-in human presence sensor                                                                                                                                                                                                                                                 | Used for the power save function.                                            |
| USB Ports / Standard                 | 1 upstream,<br>2 downstream / Rev. 2.0                                                                                                                                                                                                                                                                                                   | 1 upstream,<br>2 downstream / Rev. 2.0                                                                                                                                                                                                                                                                                                   | —                                                                            |
| Dimensions with Stand<br>(W x H x D) | 367.0 x 522/583 x 220.0 mm                                                                                                                                                                                                                                                                                                               | 361.5 x 517.0 / 612.0 x 196.5 mm                                                                                                                                                                                                                                                                                                         | —                                                                            |

{7}------------------------------------------------

{8}------------------------------------------------

### Recommended Physical Laboratory Tests

| Measurements<br>Guidance                                                                                         | MS-S300<br>Measurements                                                                                                          |
|------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| a. Spatial resolution                                                                                            | MTF measurement method that uses a bar-pattern<br>image.<br>(Rectangle chart method)                                             |
| b. Pixel defects<br>(maximum counts, allowed defect<br>types, and locations)                                     | ISO13406-2 and Flat Panel Display Measurement<br>Standard [VESA 2001] were used.                                                 |
| c. Artifacts                                                                                                     | Artifacts (phase or clock, ringing, ghosting, image<br>sticking, etc.)                                                           |
| d. Temporal response                                                                                             | JVCKENWOOD uses typical data provided by the<br>panel manufacturer.<br>(5-95%, 10-90%, 40-60%)                                   |
| e. Luminance<br>(maximum, minimum, achievable,<br>and recommended)                                               | Lmin, and Lmax on the calibrated luminance was<br>confirmed by the requirement of the "Luminance<br>Response" test of AAPM-TG18. |
| f. Conformance to a grayscale-to-<br>luminance function<br>(e.g., DICOM GSDF)                                    | Luminance Response at 256 digital values by<br>AAPM-TG18                                                                         |
| g. Luminance at 30° and 45° in<br>diagonal, horizontal, and vertical<br>directions at center and four<br>corners | Angular dependency of luminance by AAPM-TG18                                                                                     |
| h. Luminance uniformity or Mura<br>test                                                                          | Luminance Uniformity by AAPM-TG18<br>Chromaticity by AAPM-TG18                                                                   |
| i. Stability of luminance and<br>chromaticity response with<br>temperature and time of operation<br>or on-time   | 0°C, 25°C, and 40°C on Luminance response by<br>AAPM-TG18<br>(Power On Drift.)                                                   |
| j. Spatial noise                                                                                                 | Noise Power Spectrum                                                                                                             |
| k. Reflection coefficient                                                                                        | Specular reflection and Diffuse reflection by<br>AAPM-TG18                                                                       |
| I. Veiling glare or small-spot<br>contrast                                                                       | Veiling Glare test by AAPM-TG18                                                                                                  |
| Other                                                                                                            | Pixel fill factor<br>(Pixel structure and aperture ratio of pixel.                                                               |

---

**Source:** [https://fda.innolitics.com/device/K200161](https://fda.innolitics.com/device/K200161)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).