The LightForce Orthodontic System is a treatment planning software (TPS) and orthodontic appliance system used to correct malocclusions in orthodontic patients using patient-matched orthodontic appliances.
Device Story
System comprises treatment planning software (TPS), patient-specific ceramic brackets, and placement jigs. Input: patient digital impression (3D files). Process: orthodontist uses TPS to view, measure, and modify tooth setup/bracket placement; LFO manufactures patient-specific components via additive manufacturing. Output: custom brackets and placement jigs. Used in orthodontic clinics by orthodontists. Orthodontist bonds brackets to teeth using jigs and ligates wires. Benefits: patient-matched appliances for malocclusion correction. Software update (v3.1 to v4.0) improves architecture, hosting, and UI; adds movement accuracy around jaw/TMJ axes.
Clinical Evidence
No clinical performance testing conducted. Substantial equivalence supported by bench testing of software validation and comparison of device features/functions to predicates.
Technological Characteristics
System includes patient-specific ceramic brackets, placement jigs, and TPS. Additive manufacturing used for components. Connectivity: HTTPS transmission of impression data. Software: TPS v4.0 for case planning, viewing, and measurement. No patient-specific arch wires, ligatures, or adhesives included.
Indications for Use
Indicated for orthodontic patients requiring correction of malocclusions using patient-matched orthodontic appliances.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font below.
February 21, 2020
LightForce Orthodontics % Prithul Bom Responsible Third Party Official Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114
Re: K200148
Trade/Device Name: LightForce Orthodontic System (LFO System) Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic plastic bracket Regulatory Class: Class II Product Code: PNN, NJM Dated: January 21, 2020 Received: January 22, 2020
Dear Prithul Bom:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
gencies. You must comply with all the Act's
Page 2
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known)
#### K200148
Device Name
LightForce Orthodontic System
Indications for Use (Describe)
The LightForce Orthodontic System is a treatment planning software (TPS) and Orthodontic appliance system used to correct malocclusions in orthodontic patients using patient-matched orthodontic appliances.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:14px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:14px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### TAB 6 - 510(k) Summary
# K200148 - 510(k) SUMMARY
#### LightForce Orthodontics' LightForce Orthodontic System
#### Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
LightForce Orthodontics 1035 Cambridge Street Cambridge, MA 02141
Phone: 800-481-0185 Facsimile: N/A Email: amos@lightforceortho.com
Contact Person: Amos N Benninga
Date Prepared: February 20, 2020
#### Name of Device and Name/Address of Sponsor
LightForce Orthodontic System LightForce Orthodontics 1035 Cambridge Street Cambridge, MA 02141
Trade/Proprietary Name of Device: LightForce Orthodontic System Common or Usual Name: Orthodontic Ceramic Bracket and Accessory Classification Name: Orthodontic Ceramic Bracket, 21CFR§872.5470 Regulatory Class: II Product Code: NJM
Primary Predicate: LightForce Orthodontics' LightForce Orthodontic System (K183542) Reference Device: LightForce Orthodontics' LightForce Orthodontic System (K181271)
(Note: Previously known as Signature Orthodontic System)
{5}------------------------------------------------
#### Device Description and Justification
The LightForce Orthodontic System (LFO System) is a treatment planning software (TPS) and orthodontic appliance system used to correct malocclusions in orthodontic patients using patient-matched orthodontic appliances. The LFO System consists of patient-specific ceramic brackets, patient-specific bracket placement jigs, arch wire templates, and a TPS for viewing, measuring,modifying cases and submitting orders. LightForce Orthodontics' (LFO) operators and the orthodontists use the TPS to generate a prescription of their choosing. LFO then manufactures the patient-specific brackets and placement jigs using proprietary additive manufacturing techniques. The orthodontist then bonds the brackets to the teeth using the optional placement jig and ligates wires to enable tooth movement. The LFO System does not contain commercially-available or patient-specific shaped arch wires, ligatures, or adhesives that affixes the brackets to the teeth.
The change is to upgrade LFO System's Treatment Planning Software (TPS) from version 3.1 to version 4.0. TPS version 3.1 was originally cleared in K183542 and the LFO System was originally cleared in K181271. TPS version 4.0 provides improved software architecture, updated hosting infrastructure and improvements to the user interface to better match the intended use of the product based on customer feedback and validation input.
A comparison of TPS 3.1 and TPS 4.0 shows that both software provide the same features and functional workflows is evidence of substantial equivalence. The replacement software is identical in performance and function to the previously used software.
#### Indications for Use
The LightForce Orthodontic System is a treatment planning software and orthodontic appliance system used to correct malocclusions in orthodontic patient-matched orthodontic appliances.
#### Comparison to Primary Predicate and Reference Device
The LightForce Orthodontic System in the primary predicate (K183542) is a treatment planning software and orthodontic appliance system used to correct malocclusions in orthodontic patients using patient-matched orthodontic appliances. This current submission provides software updates to the software delivered in the primary predicate.
The LightForce Orthodontic System in the reference predicate (K181271) is a treatment planning software and orthodontic appliance system used to correct malocclusions in orthodontic patients using patient-matched orthodontic appliances.
{6}------------------------------------------------
Note: this current submission (the subject device) is leveraging performance testing such as shear bond strength, torque strength, tensile strength, wire drag force, etc. of the brackets in the reference device, which remain identical in the subject's predicate device, and the subject device. The current submission only differs from the reference device in the software component (Treatment Planning Software / TPS) which has changed. The changes are described and documented in this submission.
#### Non-clinical Performance Testing
Validation testing of the TPS was performed in accordance with LFO's design control activities for software and to the software's Test Plan. Validation results, Table 6-1, show that the version of the TPS tested performed equivalent to the software component of the primary predicate (K183542).
| Table 6-1 - LFO System TPS Performance Testing | | | | |
|--------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|-----------------------------|--|
| LFO System<br>(K181271) | LFO System TPS<br>3.1<br>(K183542) | LFO System TPS<br>4.0 (current<br>submission) | Equivalence Result | |
| 4.1 Diagnosis -<br>viewing of patient's<br>digital impression | 4.1 Diagnosis -<br>viewing of patient's<br>digital impression | 4.1 Diagnosis -<br>viewing of patient's<br>digital impression | Equivalent (Test<br>Report) | |
| 4.1 Diagnosis -<br>successful diagnosis<br>of patient's<br>malocclusion | 4.1 Diagnosis -<br>successful diagnosis<br>of patient's<br>malocclusion | 4.1 Diagnosis -<br>successful diagnosis<br>of patient's<br>malocclusion | Equivalent (Test<br>Report) | |
| 4.2 Planning -<br>successful movement<br>of patient's teeth<br>(within the software) | 4.2 Planning -<br>successful movement<br>of patient's teeth<br>(within the software) | 4.2 Planning -<br>successful movement<br>of patient's teeth<br>(within the software) | Equivalent (Test<br>Report) | |
| 4.3 Planning -<br>successful movement<br>of brackets | 4.3 Planning -<br>successful movement<br>of brackets | 4.3 Planning -<br>successful movement<br>of brackets | Equivalent (Test<br>Report) | |
| 4.4 Data Handling -<br>case delivered<br>securely and<br>uncorrupted | 4.4 Data Handling -<br>case delivered<br>securely and<br>uncorrupted | 4.4 Data Handling -<br>case delivered<br>securely and<br>uncorrupted | Equivalent (Test<br>Report) | |
{7}------------------------------------------------
# Clinical Performance Testing
No clinical performance testing was conducted on LFO System brackets.
## Device Comparison Tables
The following tables, Table 6-2 and 6-3, provide a comparison of the device features, functions, and performance characteristics between the LightForce Orthodontic System (LFO System) and its predicate devices.
| Table 6-2 - LFO System TPS Feature Comparison | | | | | |
|---------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|------------------------------------------------------------------------------------|
| Workflow<br>Function | LFO System<br>(K181271) | LFO System v3.1<br>(K183542) | LFO System v4.0<br>(current<br>submission) | Similarities | Differences |
| Indications<br>for Use | The LightForce<br>Orthodontic<br>System (K183542)<br>is a treatment<br>planning software<br>and orthodontic<br>appliance system<br>used to correct<br>malocclusions in<br>orthodontic<br>patients using<br>patient-matched<br>orthodontic<br>appliances. | The LightForce<br>Orthodontic<br>System (K183542)<br>is a treatment<br>planning software<br>and orthodontic<br>appliance system<br>used to correct<br>malocclusions in<br>orthodontic<br>patients using<br>patient-matched<br>orthodontic<br>appliances. | The LightForce<br>Orthodontic<br>System is a<br>treatment planning<br>software and<br>orthodontic<br>appliance system<br>used to correct<br>malocclusions in<br>orthodontic<br>patients using<br>patient-matched<br>orthodontic<br>appliances. | Identical | |
| 4.1 Diagnosis<br>- viewing<br>patient's<br>digital<br>impression | Rendering of<br>impression using<br>3D files | Rendering of<br>impression using<br>3D files | Rendering of<br>impression using<br>3D files | Identical | |
| 4.1 Diagnosis<br>- viewing<br>patient's<br>digital | Rotate impression<br>display | Rotate impression<br>display | Rotate impression<br>display | Identical | |
| impression | | | | | |
| 4.1 Diagnosis<br>- viewing<br>patient's<br>digital<br>impression | Zoom impression<br>display | Zoom impression<br>display | Zoom impression<br>display | Identical | |
| 4.1 Diagnosis<br>- viewing<br>patient's<br>digital<br>impression | Pan impression<br>display | Pan impression<br>display | Pan impression<br>display | Identical | |
| 4.1 Diagnosis<br>- successful<br>diagnosis of<br>patient's<br>malocclusion | Hide/Show<br>individual arches | Hide/Show<br>individual arches | Hide/Show<br>individual arches | Identical | |
| 4.2 Planning -<br>move teeth<br>based on<br>desired final<br>location | Rotate & translate<br>around tooth X,Y,Z | Rotate & translate<br>around tooth X,Y,Z | Rotate & translate<br>around tooth X,Y,Z<br>as well as Jaw and<br>TMJ axis | | Improved<br>movemen<br>t<br>accuracy<br>by<br>moving<br>around<br>desired<br>axis. |
| 4.3 Planning -<br>move<br>brackets<br>based on<br>desired final<br>location | Move brackets<br>controls | Move brackets<br>controls | Move brackets<br>controls | Identical | |
| 4.4 Data<br>Handling -<br>case data<br>delivered<br>securely and<br>uncorrupted | Impression data<br>transmitted over<br>HTTPS | Impression data<br>transmitted over<br>HTTPS | Impression data<br>transmitted over<br>HTTPS | Identical | |
{8}------------------------------------------------
{9}------------------------------------------------
| Table 6-3 - Device Comparison | | | | |
|-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Item | Reference Predicate<br>K181271 LightForce<br>Orthodontic System; | Primary Predicate<br>K183542 LightForce<br>Orthodontic System;<br>TPS 3.1 | LightForce<br>Orthodontic System;<br>TPS 4.0 | Comments |
| Description | The LightForce<br>Orthodontic System<br>(LFO System) is a<br>treatment planning<br>software (TPS) and<br>orthodontic appliance<br>system used to correct<br>malocclusions in<br>orthodontic patients<br>using patient-matched<br>orthodontic appliances.<br>The SO System<br>consists of<br>patient-specific ceramic<br>brackets, patient-<br>specific bracket<br>placement jigs, arch<br>wire templates, and a<br>TPS for viewing,<br>measuring, and<br>modifying cases.<br>LightForce<br>Orthodontics' (SO)<br>operators and the<br>orthodontists use the<br>TPS to generate a<br>prescription of their<br>choosing. SO then<br>manufactures the<br>patient-specific<br>brackets and placement<br>jigs using proprietary<br>additive manufacturing<br>techniques. The<br>orthodontist then bonds<br>the brackets to the<br>teeth using the optional<br>placement jig and<br>ligates wires to enable | The LightForce<br>Orthodontic System<br>(LFO System) is a<br>treatment planning<br>software (TPS) and<br>orthodontic appliance<br>system used to correct<br>malocclusions in<br>orthodontic patients<br>using patient-matched<br>orthodontic appliances.<br>The SO System<br>consists of<br>patient-specific ceramic<br>brackets, patient-<br>specific bracket<br>placement jigs, arch<br>wire templates, and a<br>TPS for viewing,<br>measuring, and<br>modifying cases.<br>LightForce<br>Orthodontics' (SO)<br>operators and the<br>orthodontists use the<br>TPS to generate a<br>prescription of their<br>choosing. SO then<br>manufactures the<br>patient-specific brackets<br>and placement jigs<br>using proprietary<br>additive manufacturing<br>techniques. The<br>orthodontist then bonds<br>the brackets to the teeth<br>using the optional<br>placement jig and<br>ligates wires to enable | The LightForce<br>Orthodontic System<br>(LFO System) is a<br>treatment planning<br>software (TPS) and<br>orthodontic appliance<br>system used to correct<br>malocclusions in<br>orthodontic patients<br>using patient-matched<br>orthodontic appliances.<br>The SO System<br>consists of<br>patient-specific ceramic<br>brackets, patient-<br>specific bracket<br>placement jigs, arch<br>wire templates, and a<br>TPS for viewing,<br>measuring, and<br>modifying cases.<br>LightForce<br>Orthodontics' (SO)<br>operators and the<br>orthodontists use the<br>TPS to generate a<br>prescription of their<br>choosing. SO then<br>manufactures the<br>patient-specific brackets<br>and placement jigs<br>using proprietary<br>additive manufacturing<br>techniques. The<br>orthodontist then bonds<br>the brackets to the teeth<br>using the optional<br>placement jig and<br>ligates wires to enable | Identical |
| | tooth movement. The SO System does not contain commercially-available or patient-specific shaped arch wires, ligatures, or adhesives that affixes the brackets to the teeth. | tooth movement. The SO System does not contain commercially-available or patient-specific shaped arch wires, ligatures, or adhesives that affixes the brackets to the teeth. | tooth movement. The SO System does not contain commercially-available or patient-specific shaped arch wires, ligatures, or adhesives that affixes the brackets to the teeth. | |
| Product Codes /<br>Regulations | NJM (Orthodontic Ceramic Bracket, 21CFR§872.5470) | NJM (Orthodontic Ceramic Bracket, 21CFR§872.5470) | NJM (Orthodontic Ceramic Bracket, 21CFR§872.5470) | Identical |
| Indications for Use | The LightForce Orthodontic System is a treatment planning software and orthodontic appliance system used to correct malocclusions in orthodontic patients using patient-matched orthodontic appliances. | The LightForce Orthodontic System is a treatment planning software and orthodontic appliance system used to correct malocclusions in orthodontic patients using patient-matched orthodontic appliances. | The LightForce Orthodontic System is a treatment planning software and orthodontic appliance system used to correct malocclusions in orthodontic patients using patient-matched orthodontic appliances. | Identical |
| Sequence of Treatment Plan or Mode of Use | LightForce Orthodontics (LFO) receives the patient's digital data via commercially available communications tools from the DDS/DMD. LFO prepares a final tooth setup and bracket placement plan and returns the digital data back to the DDS/DMD. The orthodontist can then view, measure and modify the case using the TPS. Once approved, brackets and jigs are manufactured and shipped to the DDS/DMD. | LightForce Orthodontics (LFO) receives the patient's digital data via commercially available communications tools from the DDS/DMD. LFO prepares a final tooth setup and bracket placement plan and returns the digital data back to the DDS/DMD. The orthodontist can then view, measure and modify the case using the TPS. Once approved, brackets and jigs are manufactured and shipped to the DDS/DMD. | LightForce Orthodontics (LFO) receives the patient's digital data via commercially available communications tools from the DDS/DMD. LFO prepares a final tooth setup and bracket placement plan and returns the digital data back to the DDS/DMD. The orthodontist can then view, measure and modify the case using the TPS. Once approved, brackets and jigs are manufactured and shipped to the DDS/DMD. | Identical |
| Manufac<br>turing<br>Method | Final desired<br>arrangement of teeth,<br>brackets, wires and jigs<br>are designed with the<br>guidance of LFO<br>internal computer<br>software using a 3-D<br>model of the patient.<br>The Treatment<br>Planning Software<br>allows the clinician to<br>review, measure, and<br>modify the case.<br>Internal LFO software<br>generates the 3D<br>image file that<br>proprietary additive<br>manufacturing<br>equipment uses to<br>create the brackets and<br>indirect bonding (IDB)<br>tray. | Final desired<br>arrangement of teeth,<br>brackets, wires and jigs<br>are designed with the<br>guidance of LFO<br>internal computer<br>software using a 3-D<br>model of the patient.<br>The Treatment Planning<br>Software allows the<br>clinician to review,<br>measure, and modify<br>the case. Internal LFO<br>software generates the<br>3D image file that<br>proprietary additive<br>manufacturing<br>equipment uses to<br>create the brackets and<br>indirect bonding (IDB)<br>tray. | Final desired<br>arrangement of teeth,<br>brackets, wires and jigs<br>are designed with the<br>guidance of LFO<br>internal computer<br>software using a 3-D<br>model of the patient.<br>The Treatment Planning<br>Software allows the<br>clinician to review,<br>measure, and modify<br>the case. Internal LFO<br>software generates the<br>3D image file that<br>proprietary additive<br>manufacturing<br>equipment uses to<br>create the brackets and<br>indirect bonding (IDB)<br>tray. | Identical |
{10}------------------------------------------------
{11}------------------------------------------------
### Conclusion
The LightForce Orthodontics' LightForce Orthodontics System presented in this 510(k) submission and its primary predicate the LightForce Orthodontic System (K183542) and its reference predicate the LightForce Orthodontics System (K181271) are substantially equivalent in intended use and technological characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.