hCG Urine Test Strip, hCG Urine Test Cassette, hCG Urine Test Midstream

K200133 · Safecare Biotech (Hangzhou) Co., Ltd. · LCX · Aug 13, 2020 · Clinical Chemistry

Device Facts

Record IDK200133
Device NamehCG Urine Test Strip, hCG Urine Test Cassette, hCG Urine Test Midstream
ApplicantSafecare Biotech (Hangzhou) Co., Ltd.
Product CodeLCX · Clinical Chemistry
Decision DateAug 13, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1155
Device ClassClass 2

Intended Use

The hCG Urine Test Strip is a rapid one step assay designed for qualitative detection of human chorionic gonadotropin (hCG) in urine for early detection of pregnancy. The hCG Urine Test Cassette is a rapid one step assay designed for qualitative detection of human chorionic gonadotropin (hCG) in urine for early detection of pregnancy. The hCG Urine Test Midstream is a rapid one step assay designed for qualitative detection of human chorionic gonadotropin (hCG) in urine for early detection of pregnancy.

Device Story

Lateral flow sandwich immunochromatographic assay for qualitative hCG detection in urine; early pregnancy indicator. Available in strip, cassette, and midstream formats. Device contains membrane coated with mouse monoclonal anti-βhCG antibody and pad with mouse monoclonal anti-α-hCG antibody colloidal gold conjugate; goat anti-rabbit IgG used for control. User applies urine sample (dip or midstream); results read visually within 3-5 minutes. Used by lay persons in home settings or professionals in clinical settings. Provides qualitative positive/negative result; aids in pregnancy confirmation. Benefits include rapid, accessible, and easy-to-interpret pregnancy testing.

Clinical Evidence

Bench testing only. Precision/reproducibility study (n=30 per concentration/lot) confirmed 25 mIU/mL sensitivity. Interference/cross-reactivity studies showed no interference from common substances, hCG beta-core fragments, pH (3-9), or specific gravity (1.001-1.039). Hook effect absent up to 850,000 mIU/mL. Method comparison study (n=120) showed 100% agreement with predicate. Lay-user study (n=360) demonstrated 100% agreement between lay-user results and professional-user results across all formats.

Technological Characteristics

Lateral flow sandwich immunoassay. Components: nitrocellulose membrane coated with mouse monoclonal anti-beta-hCG antibody; pad with mouse monoclonal anti-alpha-hCG antibody-colloidal gold conjugate. Control line: goat anti-rabbit IgG. Formats: strip, cassette, midstream. Qualitative visual readout. Shelf life: 30 months at 4-30°C. Traceable to WHO 5th IS.

Indications for Use

Indicated for qualitative detection of hCG in urine for early pregnancy detection in women. Over-the-counter use.

Regulatory Classification

Identification

A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. August 13, 2020 Safecare Biotech (Hangzhou) Co., Ltd. Selina Zhang Regulatory Affair 18 Haishu Road, Yuhang District Hangzhou, China #### Re: K200133 Trade/Device Name: hCG Urine Test Strip, hCG Urine Test Cassette, hCG Urine Test Midstream Regulation Number: 21 CFR 862.1155 Regulation Name: Human Chorionic Gonadotropin (HCG) Test System Regulatory Class: Class II Product Code: LCX Dated: June 29, 2020 Received: June 29, 2020 #### Dear Selina Zhang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Marianela Perez- Torres, Ph.D. Acting Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K200133 Device Name hCG Urine Test Strip hCG Urine Test Cassette hCG Urine Test Midstream Indications for Use (Describe) The hCG Urine Test Strip is a rapid one step assay designed for qualitative detection of human chorionic gonadotropin (hCG) in urine for early detection of pregnancy. The hCG Urine Test Cassette is a rapid one step assay designed for qualitative detection of human chorionic gonadotropin (hCG) in urine for early detection of pregnancy. The hCG Urine Test Midstream is a rapid one step assay designed for qualitative detection of human chorionic gonadotropin (hCG) in urine for early detection of pregnancy. Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # Section 5: # 510(K) Summary Date of Preparation: August 12, 2020 # A. Applicant Name: Safecare Biotech (Hangzhou) Co., Ltd. Address: 18 Haishu Road, Yuhang District, Hangzhou, China Official Contact Person Information Name: Selina Zhang Tel: 0086-571-89712897 Mail: selinazhang@safecare.com.cn #### B. Device Trade name: hCG Urine Test Strip hCG Urine Test Cassette hCG Urine Test Midstream | Common name: | KIT, TEST, PREGNANCY, HCG, OVER THE COUNTER | |-------------------------------------------------|---------------------------------------------| | Classification name: | KIT, TEST, PREGNANCY, HCG, OVER THE COUNTER | | Regulation Medical Specialty Clinical Chemistry | | | Regulation Number | 862.1155 | | Product Code | LCX | | Classification | Class II | # C. Predicate device 510 (K) Number: k043443 One Step HCG Urine Pregnancy Test Produced by Guangzhou Wondfo Biotech Co., Ltd. # D. Intended use of the device Intended use(s): The hCG Urine Test Strip is a rapid one step assay designed for qualitative detection of human chorionic gonadotropin (hCG) in urine for early detection of pregnancy. The hCG Urine Test Cassette is a rapid one step assay designed for qualitative detection of human chorionic gonadotropin (hCG) in urine for early detection of pregnancy. The hCG Urine Test Midstream is a rapid one step assay designed for qualitative detection of human chorionic gonadotropin (hCG) in urine for early detection of pregnancy. {4}------------------------------------------------ #### E. Device Description The hCG Urine Test will be sold in three formats: cassette, test strip, and midstream. The test strip and midstream kits consist of one test device and a package insert. The cassette kit consists of one test device and a disposable plastic dropper, and a package insert. Each test device contains mouse monoclonal anti-βhCG antibody coated membrane and a pad containing mouse monoclonal anti-α-hCG antibody colloidal gold conjugate. The control antibodies are goat anti-rabbit IgG. #### F. Comparison with predicate A summary comparison of the HCG Urine Test and the predicate devices is provided in the following Table: | Device | New Device | Predicate (K043443) | Device Comparison | |----------------|---------------------------------------|------------------------------|-------------------| | Manufacturer | Safecare Biotech (Hangzhou) Co., Ltd. | Guangzhou Biotech Co., Ltd. | NA | | Intended use | early detection of pregnancy | early detection of pregnancy | Same | | Specimen | Urine | Urine | Same | | Sensitivity | 25mIU/mL | 25mIU/mL | Same | | Indications | Over the Counter (OTC) | Over the Counter (OTC) | Same | | Read time | 3-5 minutes | 5 minutes | Similar | | Test Principle | Immunochromatographic assay | Immunochromatographic assay | Same | | Format | Strip, cassette, midstream | Strip, cassette, midstream | Same | | Result | Qualitative | Qualitative | Same | There are no differences identified between new device and the predicate device for intended use, clinical cut-off, test principle, test format, etc. There is a minor difference in the reading time, the new device has a 3-5 minutes reading time while the predicate device has 5 minutes of reading time. However, this minor difference will not impact the safety or performance of the device. The HCG Urine Test Strip/cassette/midstream, and the predicate device use the same chemistry with essentially the same test design, thus they are substantial equivalent. #### G. Standard/Guidance Document Referenced (if applicable) Guidance for Over-the Counter (OTC) Human Chorionic Gonadotropin (hCG) 510(k)s #### H. Test Principle This device operates by detecting human chorionic gonadotropin (hCG), the hormone produced during pregnancy in urine, using a lateral flow sandwich immunochromatographic assay. #### Performance characteristic l. - 1. Analytical performance - a. Precision/Reproducibility: {5}------------------------------------------------ Fresh urine samples were took from normal, nonpregnant females spiked with the HCG (traceable to WHO 5th IS ) at different concentration 0mlU/ml, 18.75mlU/ml, 25mlU/ml, 50mlU/ml, 100mlU/ml. The study was conducted 3 runs / day and lasted 10 days and was conducted by 3 operators. There are 3 batches HCG Urine Test of three formats and each operator should conduct one batch separately. The midstream format was performed with both of these midstream test sample application methods (simulated midstream and dip). the result was recorded as the following: #### The results of precision (strip) | HCG | LOT 1 | | LOT 2 | | LOT 3 | | |---------------|----------|----------|----------|----------|----------|----------| | concentration | positive | negative | positive | negative | positive | negative | | 0mIU/ml | 0 | 30 | 0 | 30 | 0 | 30 | | 12.5mIU/ml | 0 | 30 | 0 | 30 | 0 | 30 | | 18.75mIU/ml | 0 | 30 | 0 | 30 | 0 | 30 | | 25mIU/ml | 30 | 0 | 30 | 0 | 30 | 0 | | 50mIU/ml | 30 | 0 | 30 | 0 | 30 | 0 | | 100mIU/ml | 30 | 0 | 30 | 0 | 30 | 0 | #### The results of precision (cassette) | HCG<br>concentration | LOT 1 | | LOT 2 | | LOT 3 | | |----------------------|----------|----------|----------|----------|----------|----------| | | positive | negative | positive | negative | positive | negative | | 0mlU/ml | 0 | 30 | 0 | 30 | 0 | 30 | | 12.5mlU/ml | 0 | 30 | 0 | 30 | 0 | 30 | | 18.75mlU/ml | 0 | 30 | 0 | 30 | 0 | 30 | | 25mlU/ml | 30 | 0 | 30 | 0 | 30 | 0 | | 50mIU/ml | 30 | 0 | 30 | 0 | 30 | 0 | | 100mlU/ml | 30 | 0 | 30 | 0 | 30 | 0 | {6}------------------------------------------------ #### The results of precision (midstream, using the dip method) | HCG<br>concentration | LOT 1 | | LOT 2 | | LOT 3 | | |----------------------|----------|----------|----------|----------|----------|----------| | | positive | negative | positive | negative | positive | negative | | 0mIU/ml | 0 | 30 | 0 | 30 | 0 | 30 | | 12.5mIU/ml | 0 | 30 | 0 | 30 | 0 | 30 | | 18.75mIU/ml | 0 | 30 | 0 | 30 | 0 | 30 | | 25mIU/ml | 30 | 0 | 30 | 0 | 30 | 0 | | 50mIU/ml | 30 | 0 | 30 | 0 | 30 | 0 | | 100mIU/ml | 30 | 0 | 30 | 0 | 30 | 0 | #### The results of precision (midstream, using the simulated midstream method) | HCG<br>concentration | LOT 1 | | LOT 2 | | LOT 3 | | |----------------------|----------|----------|----------|----------|----------|----------| | | positive | negative | positive | negative | positive | negative | | 0mIU/ml | 0 | 30 | 0 | 30 | 0 | 30 | | 12.5mIU/ml | 0 | 30 | 0 | 30 | 0 | 30 | | 18.75mIU/ml | 0 | 30 | 0 | 30 | 0 | 30 | | 25mIU/ml | 30 | 0 | 30 | 0 | 30 | 0 | | 50mIU/ml | 30 | 0 | 30 | 0 | 30 | 0 | | 100mIU/ml | 30 | 0 | 30 | 0 | 30 | 0 | The results show that the precision of HCG Urine Test in 3 batches of different formats are good. HCG Urine Test exhibited reproducibility of results. Based on the above results, the sensitivity is demonstrated to be 25 mIU/mL. b. Linearity/assay reportable range: Not applicable. This is a qualitative assay. - c. Traceability, Stability, Expected values (controls, calibrators, or methods): {7}------------------------------------------------ #### Traceability: The 510(k) describes traceability of the assay to the WHO 5th IS material. Stability testing was provided to support the claimed shelf life (30months). #### Stability: The stability data supports that the products have the shelf life of 30 months when stored at 4-30℃. #### d. Detection limit/Sensitivity study: See Precision section above. According to the results of precision data above, the sensitivity of HCG Urine Test is 25mlU/ml. #### e. Analytical specificity: e.1) To determine if the hCG Test Strip/cassette/midstream was affected (i.e. cross-reacted) by a relevant challenge dose of closely-related human lutieinzing hormone (hLH), human follicle stimulating hormone (hFSH) or human thyroid stimulating hormone (hTSH). The results demonstrated no cross reaction with LH at 500mlU/ml, FSH at 1000 mIU/mL, and TSH at 1000 μIU/mL. e.2) To evaluate the potential for interference by certain exogenous compounds and potentially interfering clinical conditions. Each substance was prepared by diluting stock interference material to the desired concentration. Normal, nonpregnant female's urine specimens containing 0 and 25mlU/ml HCG were spiked with the interferents to obtain the desired test concentration. Three batches of each format were tested. The results show that no interferences were observed from substance at the following concentrations for both negative and positive HCG urine samples. | Interfering substances | Substances concentration | |------------------------|--------------------------| | Acetaminophen | 20mg/mL | | Aspirin | 20mg/dL | | Ascorbic acid | 20mg/mL | | Caffeine | 20mg/dL | | Glucose | 200mg/dL | | Hemoglobin | 1mg/mL | | Tetracycline | 20mg/dL | | Ampicillin | 20mg/dL | | Albumin | 2000mg/dL | | Bilirubin | 2mg/mL | | Erythrocytes | >250/uL | | Leukocyte | >500/ul | | Uric acid | 450mMol/L | | Ketone | > 80mg/dL | | Ethanol | 1% | {8}------------------------------------------------ e.3) To evaluate the effects of the HCG ß-core fragment normal nonpregnant female urine specimens containing 0 and 25mlU/ml HCG were spiked with the HCG ß-core fragment (traceable to WHO reference reagent 99/708) at the concentration of 125000, 50000, 500000 and 1000000 pmol/mL. Three batches of each format were tested. The data shows that there's no interference in the test result when the HCG ß-core fragment at the highest levels at which it is likely to be found on patient samples. #### e.4) PH study The PH of an aliguot negative urine pool is adjusted to a PH range of 3 to 9 in 1 PH unit increments and spiked with HCG at 25mlU/ml and 0mIU/ml and 3 batches of HCG Urine Test were tested repeatedly. The result demonstrates that varying ranged of PH do not interfere with the performance of the test. #### e.5) Specific gravity Negative specimen 0 mIU/ml and specimen with HCG 25mlU/ml were formulated into the solution with specific gravity at 1.001, 1.020, 1.031, 1.039separately. Three lots of HCG Urine Test were tested. The data show that there's no interference in the test result when the specific gravity is between 1.001-1.039. #### e.6) HOOK effect study The test was evaluated for high dose hook effect. 30 HCG free specimens spiked with the HCG at different concentration containing 5,000mlU/ml, 10,000mlU/ml, 500,000mlU/ml, 650,000mIU/ml, 850,000mIU/ml,950,000mIU/ml. Three lot of tests for each format were tested. The result show that HCG Urine Test can get the positive result when the HCG concentration is range from 5,000 to 850,000 mIU/ml. #### 2. Comparison studies: #### a. Professional method comparison Urine samples were collected from 120 women at hospital laboratory to test for pregnancy. The data show that the agreement of hCG Test Strip/cassette/midstream with the predicate device was 100%. | Candidate device | Predicate device professional | | | |------------------|-------------------------------|----------|----------| | | | Positive | Negative | | Professional A | Positive | 53 | 0 | | | Negative | 0 | 67 | | Professional B | Positive | 53 | 0 | | | Negative | 0 | 67 | # The results of professional method comparison (strip) {9}------------------------------------------------ | Candidate device | Predicate device professional | | | |------------------|-------------------------------|----------|----| | | Positive | Negative | | | Professional A | Positive | 53 | 0 | | | Negative | 0 | 67 | | Professional B | Positive | 53 | 0 | | | Negative | 0 | 67 | #### The results of professional method comparison (cassette) #### The results of professional method comparison (midstream, using dip method) | Candidate device | Predicate device professional | | |------------------|-------------------------------|----------| | | Positive | Negative | | Professional A | 53 | 0 | | | 0 | 67 | | Professional B | 53 | 0 | | | 0 | 67 | #### The results of professional method comparison (midstream, using the simulated method) | Candidate device | Predicate device professional | | | |------------------|-------------------------------|----------|----| | | Positive | Negative | | | Professional A | Positive | 53 | 0 | | | Negative | 0 | 67 | | Professional B | Positive | 53 | 0 | | | Negative | 0 | 67 | #### b. The lay user method comparison: Urine samples were collected from 360 women at hospital laboratory to test for pregnancy. Ages were from 18 to 45 years. Out of the 360 women, 84 have an education background of middle school, 103 have an education background of high school, 84 have an education background of college degree, 59 have an education background of bachelor degree, 30 have an education background of Master & PhD degree. The conformity between the user interpretation of the and the professional interpretation of the hCG Test Strip/cassette/midstream is 100%. #### The results of the lay user method comparison (strip) | Safecare device | Safecare device professional | | Total | |-----------------|------------------------------|----------|-------| | | Positive | Negative | | {10}------------------------------------------------ #### Safecare Biotech (Hangzhou) Co. Ltd. 18 Haishu Road, Yuhang District Hangzhou, China | Lay<br>users | Positive | 33 | 0 | 33 | |--------------|----------|----|----|----| | | Negative | 0 | 57 | 57 | | | Total | 33 | 57 | 90 | #### The results of the lay user method comparison (cassette) | Safecare device | Safecare device professional | | Total | | |-----------------|------------------------------|----------|-------|----| | | Positive | Negative | | | | Lay<br>users | Positive | 29 | 0 | 29 | | | Negative | 0 | 61 | 61 | | Total | | 29 | 61 | 90 | #### The results of the lay user method comparison (midstream, using dip method) | Safecare device | Safecare device professional | | Total | | |-----------------|------------------------------|----------|-------|----| | | Positive | Negative | | | | Lay<br>users | Positive | 23 | 0 | 23 | | | Negative | 0 | 67 | 67 | | Total | | 23 | 67 | 90 | #### The results of the lay user method comparison (midstream, using the actual midstream method) | Safecare device | Safecare device professional | | Total | | |-----------------|------------------------------|----------|-------|----| | | Positive | Negative | | | | Lay<br>users | Positive | 26 | 0 | 26 | | | Negative | 0 | 64 | 64 | | | Total | 26 | 64 | 90 | #### c. The performance tested by OTC user: To evaluate its suitability to be used by the home use consumers (lay persons), spiked urine samples were tested by the lay persons and the results were compared with professional laboratory results. The results show that hCG Test Strip/cassette/midstream can be used by the untrained operator and get the correct results. #### Results of performance tested by OTC user | | | Masked spiked sample | | |---------|----------------------|--------------------------------------------|--| | Formats | Masked spiked sample | Masked spiked sample<br>Professional users | | {11}------------------------------------------------ #### Safecare Biotech (Hangzhou) Co. Ltd. 18 Haishu Road, Yuhang District Hangzhou, China | | | | +(31.25mIU/ml) | -(18.75mIU/ml) | |-------------------------------------------------------|-----------|----------------|----------------|----------------| | strip | Lay users | +(31.25mIU/ml) | 30 | 0 | | | | -(18.75mIU/ml) | 0 | 30 | | Cassette | Lay users | +(31.25mIU/ml) | 30 | 0 | | | | -(18.75mIU/ml) | 0 | 30 | | Midstream, using dip<br>method | Lay users | +(31.25mIU/ml) | 30 | 0 | | | | -(18.75mIU/ml) | 0 | 30 | | Midstream, using the<br>simulated midstream<br>method | Lay users | +(31.25mIU/ml) | 30 | 0 | | | | -(18.75mIU/ml) | 0 | 30 | After recording their results, participants were asked to evaluate the test. All participants thought the test was either "very easy" or "easy" to read and interpret (on a scale ranging from very difficult to very easy). d. Matrix comparison: Not applicable. 3. Clinical studies: a. Clinical Sensitivity: Not applicable. b. Clinical specificity: Not applicable. c. Other clinical supportive data (when a. and b. are not applicable): 4. Clinical cut-off: 2SmIU/mL 5. Expected values/Reference range: Not applicable #### N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. # O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...