← Product Code [FRC](/productcode/FRC) · K200126

# VERIFY Assert Self-Contained Biological Indicator (K200126)

_STERIS Corporation · FRC · May 22, 2020 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K200126

## Device Facts

- **Applicant:** STERIS Corporation
- **Product Code:** [FRC](/productcode/FRC.md)
- **Decision Date:** May 22, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.2800
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The VERIFY Assert Self-Contained Biological Indicator (SCBI) is for routine monitoring, qualification testing, load monitoring and product testing of the following steam sterilization processes. Cycle Type Temperature Time Dynamic Air Removal 270°F (132°C) 4 minutes Dynamic Air Removal 275°F (135°C) 3 minutes Gravity 250°F (121°C) 30 minutes Gravity 270°F (132°C) 15 minutes. When used in conjunction with the Reader for the VERIFY Assert Self-Contained Biological Indicator, the VERIFY Assert Self-Contained Indicator provides a fluorescent result within 40 minutes

## Device Story

Biological indicator for steam sterilization monitoring; contains Geobacillus stearothermophilus spores and nutrient media. Device placed in sterilizer; after cycle, incubated in proprietary reader. Reporter enzyme produced by surviving spores reacts with fluorogenic substrate to produce fluorescent signal. Reader detects fluorescence within 40 minutes to confirm sterilization efficacy. Used in clinical settings for routine monitoring, load monitoring, and qualification testing. Output informs healthcare providers of sterilization process success; failure to kill spores indicates sterilization cycle failure, preventing use of non-sterile instruments.

## Clinical Evidence

Bench testing only. Testing included Reduced Incubation Time (RIT) validation (>97% alignment with 7-day conventional method), spore population enumeration (2.7-3.7 x 10^6 spores/BI), resistance testing (D121 ≥ 2.13 min), signal generation validation, hold time stability (up to 72 hours), and simulated use with AAMI 16-towel pack reference load.

## Technological Characteristics

Self-contained biological indicator (SCBI) in plastic vial. Contains Geobacillus stearothermophilus spores (1.0-4.0 x 10^6/BI) and fluorogenic nutrient media. Sensing principle: enzymatic fluorescence detection. Incubation: 55-59°C. Shelf-life: 13 months. No exogenous DNA/plasmids. Compatible with STERIS proprietary reader.

## Regulatory Identification

Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.

## Predicate Devices

- Celerity 20 Steam Biological Indicator ([K173634](/device/K173634.md))

## Reference Devices

- VERIFY Assert Self-Contained Biological Indicator ([K162701](/device/K162701.md))
- STERIS STEAM II ([K112256](/device/K112256.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

May 22, 2020

STERIS Corporation Anthony Piotrkowski Regulatory Affairs Director 5960 Heisley Road Mentor, Ohio 44060

Re: K200126

Trade/Device Name: VERIFY Assert Self-Contained Biological Indicator Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: FRC Dated: April 23, 2020 Received: April 24, 2020

Dear Anthony Piotrkowski:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For CAPT Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

### K200126

### Device Name

VERIFY Assert Self-Contained Biological Indicator

Indications for Use (Describe)

The VERIFY Assert Self-Contained Biological Indicator (SCBI) is for routine monitoring, qualification testing, load monitoring and product testing of the following steam sterilization processes.

| Cycle Type          | Temperature   | Time       |
|---------------------|---------------|------------|
| Dynamic Air Removal | 270°F (132°C) | 4 minutes  |
| Dynamic Air Removal | 275°F (135°C) | 3 minutes  |
| Gravity             | 250°F (121°C) | 30 minutes |
| Gravity             | 270°F (132°C) | 15 minutes |

When used in conjunction with the Reader for the VERIFY Assert Self-Contained Biological Indicator, the VERIFY Assert Self-Contained Indicator provides a fluorescent result within 40 minutes

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

__ Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary For VERIFY Assert Self-Contained Biological Indicator

### Sponsor Facility

STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 357-9198

### Manufacturing Facility

STERIS Corporation 9325 Pinecone Drive Mentor, OH 44060 Phone: (440) 392-7800 Fax No: (440) 392-7896

Contact: Tony Piotrkowski Director, Regulatory Affairs

Telephone: (440) 392-7437 Fax No: (440) 357-9198 e-mail: tpiotrko@steris.com

STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600

{4}------------------------------------------------

#### 1. Device Name

| Trade Name:            | VERIFY Assert Self-Contained Biological Indicator                    |
|------------------------|----------------------------------------------------------------------|
| Common/usual Name:     | Biological Indicator (BI, SCBI)                                      |
| Device Classification: | Class II                                                             |
| Classification Name:   | Biological Sterilization Process Indicator<br>(21 CFR 880.2800. FRC) |

### 2. Predicate Device

Celerity 20 Steam Biological Indicator, K173634, as modified under K181686. (21 CFR 880.2800. FRC)

### 3. Reference Device

VERIFY Assert Self-Contained Biological Indicator. K162701, as modified under K181422. (21 CFR 880.2805, OWP)

#### 4. Description of Device

The product is intended to monitor the critical parameters of steam sterilization cycles described in the indications for use by producing an optical change (signal) that is detected by the STERIS proprietary reader, VERIFY Incubator for Assert Self-Contained Biological Indicator, in 40 minutes to confirm the viability of the biological indicator at the end of a steam sterilization process. The product consists of a biological organism known to be resistant to steam (Geobacillus stearothermophilus) and a defined nutrient media. A reporter enzyme, which is produced by the organism, reacts with a fluorogenic substrate within the defined nutrient media to produce a fluorescent moiety.

#### 5. Intended Use/ Indications for Use

The VERIFY Assert Self-Contained Biological Indicator (SCBI) is for routine monitoring, qualification testing, load monitoring and product testing of the following steam sterilization processes.

| Cycle Type          | Temperature   | Time       |
|---------------------|---------------|------------|
| Dynamic Air Removal | 270°F (132°C) | 4 minutes  |
| Dynamic Air Removal | 275°F (135°C) | 3 minutes  |
| Gravity             | 250°F (121°C) | 30 minutes |
| Gravity             | 270°F (132°C) | 15 minutes |

{5}------------------------------------------------

When used in conjunction with the VERIFY Incubator for Assert Self-Contained Biological Indicator, the VERIFY Assert Self-Contained Indicator provides a fluorescent result within 40 minutes.

#### Summary of Technical Characteristics 6.

## Table 5-1.Comparison of SCBI Physical Description and Technological Properties to Reference

|                               | VERIFY Assert SCBI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | VERIFY Assert SCBI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                        |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|
| Feature                       | (modified)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Reference (K162701)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Comparison                                             |
| Intended Use                  | The VERIFY Assert Self-<br>Contained Biological<br>Indicator (SCBI) is for<br>routine monitoring,<br>qualification testing, load<br>monitoring and product<br>testing of the following<br>steam sterilization<br>processes:<br>270F, 4-minute dynamic<br>air removal; 275F, 3-<br>minute dynamic air<br>removal; 250 F, 30-<br>minute gravity; 270, 15-<br>minutes gravity.<br>When used in conjunction<br>with the VERIFY<br>Incubator for Assert Self-<br>Contained Biological<br>Indicator, the VERIFY<br>Assert Self-Contained<br>Indicator provides a<br>fluorescent result within<br>40 minutes. | The VERIFY Assert Self-<br>Contained Biological<br>Indicator (SCBI) is for<br>routine monitoring,<br>qualification testing and<br>product testing of the<br>following steam<br>sterilization processes:<br>270F, 4-minute dynamic<br>air removal; 275F, 3-<br>minute dynamic air<br>removal; 250 F, 30-<br>minute gravity; 270, 15-<br>minutes gravity.<br>When used in conjunction<br>with the VERIFY<br>Incubator for Assert Self-<br>Contained Biological<br>Indicator, the VERIFY<br>Assert Self-Contained<br>Indicator provides a<br>fluorescent result within<br>40 minutes. | Added load monitoring<br>to the indications for<br>use |
| Indicator<br>organism         | > 90% similarity to ATCC<br>7953 Geobacillus<br>stearothermophilus                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | > 90% similarity to ATCC<br>7953 Geobacillus<br>stearothermophilus                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Identical                                              |
| Mechanism<br>of action        | An enzyme, which is<br>produced by the indicator<br>organism, reacts with a<br>fluorogenic substrate<br>within the defined nutrient<br>media to produce a<br>fluorescent moiety.                                                                                                                                                                                                                                                                                                                                                                                                                       | An enzyme, which is<br>produced by the indicator<br>organism, reacts with a<br>fluorogenic substrate<br>within the defined nutrient<br>media to produce a<br>fluorescent moiety.                                                                                                                                                                                                                                                                                                                                                                                                   | Identical                                              |
| Accessories                   | Automated incubator /<br>reader                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Automated incubator /<br>reader                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Identical                                              |
| Viable spore<br>population    | 1.0 - 4.0 x 106 spore/BI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | 1.0 - 4.0 x 106 spore/BI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Identical                                              |
| Resistance<br>characteristics | D121 ≥ 1.5 min<br>D132 ≥ 10 s<br>D135 ≥ 8 s                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | D121 ≥ 1.5 min<br>D132 ≥ 10 s<br>D135 ≥ 8 s                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Identical                                              |
| Culture<br>Conditions         | 55- 59 °C, media included<br>in SCBI, 40-minute<br>incubation time.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | 55- 59 °C, media included<br>in SCBI, 40-minute<br>incubation time.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Identical                                              |
| Feature                       | VERIFY Assert SCBI<br>(modified)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | VERIFY Assert SCBI<br>Predicate (K162701)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Comparison                                             |
| Primary<br>Packaging          | Direct inoculum on plastic<br>vial, cap with recovery<br>media.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Direct inoculum on plastic<br>vial, cap with recovery<br>media.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Identical                                              |
| Process<br>indicator          | STERIS STEAM Π<br>(K112256)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | STERIS STEAM Π<br>(K112256)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Identical                                              |
| Label                         | Single-ply on cap edge                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Two-ply on top of cap                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | This modification was<br>cleared under K181442         |
| Shelf-life                    | 13 months                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 13 months                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Identical                                              |
| Exogenous<br>DNA              | None                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | plasmid pSTERIS-Q1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Plasmid has been<br>removed                            |

{6}------------------------------------------------

# Table 5-2. Comparison of SCBI Physical Description and Technological Properties to Predicate Device

| Feature                    | VERIFY Assert SCBI<br>(modified)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Celerity 20 Steam SCBI<br>Predicate (K173634)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Comparison                                                                                                                            |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Intended<br>Use            | The VERIFY Assert Self-<br>Contained Biological Indicator<br>(SCBI) is for routine monitoring,<br>qualification testing, load<br>monitoring and product testing<br>of the following steam<br>sterilization processes:<br>• 270F, 4-minute dynamic air<br>removal;<br>• 275F, 3- minute dynamic air<br>removal;<br>• 250 F, 30-minute gravity;<br>• 270, 15- minutes gravity.<br>When used in conjunction with<br>the VERIFY Incubator for Assert<br>Self- Contained Biological<br>Indicator, the VERIFY Assert<br>Self-Contained Indicator<br>provides a fluorescent result<br>within 40 minutes. | The Celerity 20 Steam<br>Biological Indicator is used for<br>routine monitoring,<br>qualification testing, load<br>monitoring and product testing<br>of the following steam<br>sterilization cycles:<br>• Dynamic Air Removal<br>270°F (132°C) 4 minutes<br>• Dynamic Air Removal<br>275°F (135°C) 3 minutes<br>• Gravity 250°F (121°C) 30<br>minutes<br>• Gravity 270°F (132°C) 15<br>minutes.<br>When used in conjunction<br>with the Celerity™ Steam<br>Incubator, the Incubator<br>provides a fluorescent result<br>within 20 minutes. | Predicate includes<br>an indication for<br>load monitoring<br>The only<br>differences are<br>product name, read<br>time and incubator |
| Indicator<br>organism      | > 90% similarity to ATCC<br>7953 Geobacillus<br>stearothermophilus                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | > 90% similarity to ATCC<br>7953 Geobacillus<br>stearothermophilus                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Identical                                                                                                                             |
| Mechanism<br>of action     | An enzyme, which is<br>produced by the indicator<br>organism, reacts with a<br>fluorogenic substrate within<br>the defined nutrient media to<br>produce a fluorescent moiety.                                                                                                                                                                                                                                                                                                                                                                                                                     | An enzyme, which is produced<br>by the indicator organism,<br>reacts with a fluorogenic<br>substrate within the defined<br>nutrient media to produce a<br>fluorescent moiety                                                                                                                                                                                                                                                                                                                                                               | Identical                                                                                                                             |
| Accessories                | Automated incubator / reader                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Automated incubator / reader                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Identical                                                                                                                             |
| Viable spore<br>population | 1.0 – 4.0 x 106 spore/BI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 1.0 - 4.0 x 106 spore/SCBI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Identical                                                                                                                             |
| Resistance                 | D121 ≥ 1.5 min<br>D132 ≥ 10 s<br>D135 ≥ 8 s                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | D121 ≥ 1.5 min<br>D132 ≥ 10 s<br>D135 ≥ 8 s                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Identical                                                                                                                             |

{7}------------------------------------------------

| Feature               | VERIFY Assert SCBI<br>(modified)                                    | Celerity 20 Steam SCBI<br>Predicate (K173634)                       | Comparison                                        |
|-----------------------|---------------------------------------------------------------------|---------------------------------------------------------------------|---------------------------------------------------|
| Culture<br>Conditions | 55- 59 °C, media included in<br>SCBI, 40-minute incubation<br>time. | 55- 59 °C, media included in<br>SCBI, 20-minute incubation<br>time. | Reference device<br>has a shorter read<br>time    |
| Primary<br>Packaging  | Direct inoculum on plastic vial,<br>cap with recovery media.        | Direct inoculum on plastic<br>vial, cap with recovery media.        | Identical                                         |
| Process<br>indicator  | STERIS STEAM Π (K112256)                                            | STERIS STEAM Π (K112256)                                            | Identical                                         |
| Label                 | Single-ply on cap edge                                              | Two-ply on top of cap                                               | This modification<br>was cleared under<br>K181442 |
| Shelf-life            | 13 months                                                           | 13 months                                                           | Identical                                         |
| Exogenous<br>DNA      | None                                                                | None                                                                | Identical                                         |

The modifications that have triggered this premarket notification include a change to remove the GMO component (plasmid pSTERIS-Q1) from the biological indicator spore and to revise the indications for use to include load monitoring.

### 7. Summary of Nonclinical Tests

Performance testing to demonstrate substantial equivalence to the predicate has been completed and is summarized in Table 5-3 below.

Table 5-3. Summary of Non-clinical Testing

| Test                                           | Purpose of the Test                                                                                        | Acceptance Criteria                                                                                                              | Conclusion                                                   |
|------------------------------------------------|------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|
| Reduced<br>Incubation<br>Time (RIT)<br>Testing | Validate the labeled<br>incubation time of the<br>SCBI                                                     | Meets FDA's requirement of ><br>97% alignment of the 40-minute<br>results with the conventional<br>incubation time of 7 days*, ‡ | PASS                                                         |
| Viable spore<br>population                     | Enumerate the number<br>of viable spores per<br>SCBI                                                       | $1.0 - 4.0 x 10^6 spore/SCBI**, ‡$                                                                                               | $2.7 - 3.7 x 10^6$<br>spore/SCBI                             |
| Resistance                                     | Determine the<br>resistance of the SCBI<br>to steam                                                        | $D_{121} ≥ 1.5 min *$<br>$D_{132} ≥ 10 s *$<br>$D_{135} ≥ 8 s *$                                                                 | $D_{121} ≥ 2.13 min$<br>$D_{132} ≥ 42 s$<br>$D_{135} ≥ 31 s$ |
| Signal<br>Generation                           | Validate that the SCBI<br>will produce signal in<br>its reader                                             | All unexposed SCBI show<br>positive growth signal in reader<br>within 40 minutes of incubation‡                                  | PASS                                                         |
| Hold Time                                      | Demonstrate that<br>performance of the<br>SCBI is not affected by<br>delaying incubation up<br>to 72 hours | Performance not affected if<br>incubated within 72 hours of<br>exposure to steam sterilization‡                                  | PASS                                                         |

{8}------------------------------------------------

| Test          | Purpose of the Test                                                                             | Acceptance Criteria                                                                | Conclusion                                           |
|---------------|-------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|------------------------------------------------------|
| Simulated Use | Demonstrate SCBI performance in a steam sterilizer with the AAMI reference load (16 towel pack) | Demonstrate growth when exposed to abbreviated cycle and all kill in a full cycle‡ | Abbreviated cycle – growth<br>Full cycle – no growth |

*Acceptance criteria based on recommendations in FDA Guidance Biological Indicator (BI) Premarket Notification [510(k)] Submissions

** Minimum acceptance criteria based on recommendations in FDA Guidance Biological Indicator (BI) Premarket Notification [510(k)] Submissions

* Acceptance criteria based on specifications of the reference device, VERIFY Assert Self-Contained BI

### 8. Conclusion

The VERIFY Assert Self-Contained Biological Indicator has met the established performance criteria. The conclusions drawn from the nonclinical and clinical tests demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed device, K173634 Class II (21 CFR 880.2800, Product code FRC).

---

**Source:** [https://fda.innolitics.com/device/K200126](https://fda.innolitics.com/device/K200126)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
