← Product Code [KTT](/productcode/KTT) · K200123

# Accessories for the SixFix® Hexapod Fixator (K200123)

_Arrowhead De, LLC · KTT · Sep 22, 2020 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K200123

## Device Facts

- **Applicant:** Arrowhead De, LLC
- **Product Code:** [KTT](/productcode/KTT.md)
- **Decision Date:** Sep 22, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The SixFix®Hexapod is intended to be used for post-traumatic joint contracture which has resulted in loss of range of motion; fractures and disease which generally may result in ioint contractures or loss of range of motion and fractures vrequiring distraction; open and closed fracture fixation; pseudo-arthrosis of long bones; limb lengthening by epiphyseal or metaphyseal distraction; correction of bony or soft tissue deformities; correction of bony or soft tissue defects; joint arthrodesis; infected fractures or nonunions.

## Device Story

SixFix Hexapod Fixator is a multilateral circular external fixation system; includes rings, arches, struts, threaded rods, reduction struts, and assembly accessories. Used for orthopedic stabilization, distraction, and deformity correction. Operates as a mechanical fixation device; provides structural support for bone segments. Used in clinical settings by orthopedic surgeons. Accessories (reduction struts, foot arches, clamps) facilitate adjustment and fixation of the hexapod frame. Output is physical stabilization of bone; enables healing of fractures or correction of deformities. Non-sterile, single-use components.

## Clinical Evidence

Bench testing only. Static axial compression testing performed per ASTM F1541 comparing SixFix Reduction Struts to predicate devices. Results demonstrated comparable performance and slightly greater stiffness for the subject device.

## Technological Characteristics

Multilateral circular external fixation system. Components: rings, arches, struts, threaded rods, reduction struts, clamps. Materials: metallic bone fixation components. Mechanical principle: external fixation/distraction. Non-sterile, single-use. Testing standard: ASTM F1541.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- SixFix Hexapod Fixator ([K190069](/device/K190069.md))
- TL-HEX TRUELOK External Fixation System ([K141078](/device/K141078.md))
- Hoffman LRF System ([K161753](/device/K161753.md))
- Hoffman LRF System ([K163656](/device/K163656.md))
- Hoffman LRF System ([K182968](/device/K182968.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

September 22, 2020

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Thomas Twardzik VP - Marketing and Operations 328 Poplar View Lane East, Suite 2 Collierville, Tennessee 38017

Re: K200123

Trade/Device Name: Accessories for the SixFix® Hexapod Fixator Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: KTT Dated: August 18, 2020 Received: August 19, 2020

Dear Thomas Twardzik:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm product submissions. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or

{1}------------------------------------------------

safety reporting (21) CFR CFR 4, Subpart B) for combination products postmarketing (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and a CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatorythe DICE assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ting Song Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K200123

Device Name Accessories for the SixFix® Hexapod Fixator

#### Indications for Use (Describe)

INDICATIONS for Accessories for the SixFix Hexapod Fixator (Reduction Struts, Foot Arches with Locking Hinges, Motion Hinge, Variable Angle Clamps)

The SixFix® Hexapod Fixator is intended to be used for post-traumatic joint contracture which has resulted in loss of range of motion; fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction; open and closed fracture fixation; pseudo-arthrosis of long bones; limb lengthening by epiphyseal or metaphyseal distraction of bony or soft tissue deformities; correction of bony or soft tissue defects; joint arthrodesis; infected fractures or nonunions.

Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary K200123 Accessories for the SixFix® Hexapod Fixator

# I. Submitter:

Arrowhead Medical Device Technologies, LLC 328 Poplar View Lane East, Suite 2 Collierville, TN 38017

Contact Person: Thomas J. Twardzik Vice President, Marketing and Operations Office: (901) 853-4366 Fax: (206) 222-9173 Email: INFO@ArrowheadDevices.com

Date of Summary: September 17, 2020

## II: Device

Proprietary Name: Accessories for the SixFix® Hexapod Fixator Common Name: External Fixator Regulatory Class: Class II Regulation: 21 CFR 888.3030 Single/multiple component metallic bone fixation appliances and

accessories

Device Product Codes: KTT

Panel: Orthopedic

### III. Predicate Devices

| Device                                     | Manufacturer | 510(k) No. | Clearance<br>Date |
|--------------------------------------------|--------------|------------|-------------------|
| Primary Predicate                          |              |            |                   |
| SixFix Hexapod Fixator                     | AMDT         | K190069    | May 14, 2019      |
| Secondary Predicates                       |              |            |                   |
| TL-HEX TRUELOK External<br>Fixation System | Orthofix Srl | K141078    | September 2, 2014 |
| Hoffman LRF System                         | Stryker GmbH | K161753    | November 15, 2016 |
|                                            |              | K163656    | April 14, 2017    |
|                                            |              | K182968    | January 25, 2019  |

# IV. Device Description

The SixFix Hexapod Fixator is a multilateral circular external fixation system. The system includes the following external fixator elements: rings, arches, struts, threaded rods, reduction struts, and assembly accessories. All the elements are provided non-sterile and are for single use only.

{4}------------------------------------------------

### 510(k) Summary K200123 Accessories for the SixFix® Hexapod Fixator

### V. Intended Use

INDICATIONS for Accessories for the SixFix Hexapod Fixator (Reduction Struts, Foot Arches with Locking Hinge, Variable Angle Clamps)

The SixFix®Hexapod is intended to be used for post-traumatic joint contracture which has resulted in loss of range of motion; fractures and disease which generally may result in ioint contractures or loss of range of motion and fractures vrequiring distraction; open and closed fracture fixation; pseudo-arthrosis of long bones; limb lengthening by epiphyseal or metaphyseal distraction; correction of bony or soft tissue deformities; correction of bony or soft tissue defects; joint arthrodesis; infected fractures or nonunions.

### VI. Comparison of Technological Characteristics with the Predicate Devices

The accessories used with the SixFix Hexapod Fixator are technologically substantially equivalent to predicate devices in terms of intended use, material, design, mechanical performance, and safety. Comparisons confirmed that the accessories used with the SixFix Hexapod Fixator are substantially equivalent when compared to those accessories used with the predicate device. The design characteristics of the accessories used with the subject system do not raise any new types of questions of safety or effectiveness. From the evidence submitted in this 510(k), the subject devices can be expected to perform at least as well as the predicate device.

### VII. Performance Data

The SixFix Reduction Struts were compared to the predicate device in accordance to ASTM F1541 static axial compression testing. The SixFix Reduction Struts exhibited results comparable to the predicate device in addition to possessing a slightly greater stiffness than the predicate device. Because there are no new concerns related to safety and effectiveness, it can be determined that the SixFix Reduction Struts are substantially equivalent to the predicate devices.

### VIII. Conclusions

A review of the device indications, material composition, external element design, and technological characteristics confirmed that the accessories used with the SixFix Hexapod Fixator are substantially equivalent to the predicate devices. While the accessories used with the SixFix Hexapod Fixator are not identical to the predicate devices, comparisons of the subject and predicate device confirmed that any differences between the subject device and predicate do not render the device NSE as there is not a new intended use; and any differences in technological characteristics do not raise different questions of safety and effectiveness than the predicate device. Therefore, it is concluded that the accessories used with the SixFix Hexapod Fixator are substantially equivalent to the predicate devices.

---

**Source:** [https://fda.innolitics.com/device/K200123](https://fda.innolitics.com/device/K200123)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com