Zimmer Hip Joint Replacement Prostheses are hip replacement system components. A femoral stem component is used in conjunction with a femoral head component for replacement of the proximal femur in total hip arthroplasty. Femoral stems are available in different designs, materials, sizes, neck lengths and taper sizes. A taper is incorporated in the design of the stem to interlock it with the femoral head. Femoral head components are available in a variety of sizes and neck lengths for reconstruction of the natural head center of the femur.
Device Story
Zimmer Hip Joint Replacement system consists of modular femoral stems and femoral head components used in total or hemi-hip arthroplasty. Components are implanted by surgeons in clinical settings to replace the proximal femur. The device functions as a mechanical prosthesis to restore joint function and alleviate pain. This submission updates labeling to include MR Conditional information for 1.5T and 3.0T environments. No changes were made to device materials, design, dimensions, or manufacturing processes. The update ensures patient safety during MRI procedures by providing standardized safety markings and data regarding RF-induced heating and image artifacts.
Clinical Evidence
No clinical data was required or provided for this submission. Substantial equivalence is supported by non-clinical performance testing regarding MRI compatibility.
Technological Characteristics
Modular hip prosthesis components (femoral stems and heads). Materials, dimensions, and design features remain unchanged from predicate devices. MR Conditional per ASTM F2503-13. Evaluated for 1.5T and 3.0T MRI environments.
Indications for Use
Indicated for patients with noninflammatory degenerative joint disease (e.g., avascular necrosis, osteoarthritis), inflammatory joint disease (e.g., rheumatoid arthritis), failed previous hip surgery, revision of failed hip arthroplasty, acute traumatic femoral head/neck fractures, congenital hip dysplasia, protrusio acetabuli, slipped capital femoral epiphysis, or disability due to previous fusion.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
{0}------------------------------------------------
April 10, 2020
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the logo is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Zimmer GmbH Anne-Kathrin Born Regulatory Affairs Senior Specialist Sulzer Allee 6 Winterthur. 8404 Ch
## Re: K200112
Trade/Device Name: Zimmer Hip Joint Replacement Regulation Number: 21 CFR 888.3353 Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis Regulatory Class: Class II Product Code: LZO, LPH, KWZ, JDI, KWL, KWY, LWJ, MEH Dated: February 6, 2020 Received: February 12, 2020
Dear Anne-Kathrin Born:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
FOR Vesa Vuniqi Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known) K200112
Device Name
Alloclassic® Zweymüller® SL / SL Offset Femoral Stems
Indications for Use (Describe)
Alloclassic® Zweymüller® SL/ SL Offset Femoral Stems:
· Noninflammatory degenerative joint disease (NID), e.g., avascular necrosis, osteoarthritis, and inflammatory joint disease (IJD), e.g., rheumatoid arthritis.
- · Failed previous surgery where pain, deformity, or dysfunction persists.
· Revision of previously failed hip arthroplasty
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|--------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="text-decoration: overline;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{3}------------------------------------------------
510(k) Number (if known) K200112
Device Name Avenir® Cemented Hip Stem
Indications for Use (Describe)
The product is intended for total or hemi hip arthroplasty with cemented applications for rehabilitating hips damaged as a result of:
- · Advanced wear of the joint due to degenerative, post-traumatic, or rheumatic diseases
- · Failed previous hip surgery including joint reconstruction (osteotomy), arthrodesis, or total hip replacement (THR)
- · Acute traumatic fracture of the femoral head or neck
- · Avascular necrosis of the femoral head.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
|X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{4}------------------------------------------------
510(k) Number (if known) K200112
Device Name Avenir® Müller Stem
## Indications for Use (Describe)
- · Advanced wear of the joint due to degenerative, post-traumatic or rheumatic diseases.
- · Failed previous hip surgery (not THA) where pain, deformity, or dysfunction persists.
- · Optional use in revision: in some medical conditions (e.g., early revision when healthy and good bone stock exists), the surgeon may opt to use primary implants in a revision procedure.
- · Acute traumatic fracture of the femoral head or neck.
- · Avascular necrosis of the femoral head.
Avenir® Müller Stems are for cementless use only.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{5}------------------------------------------------
510(k) Number (if known) K200112
Device Name BIOLOX® delta Ceramic Femoral Head
#### Indications for Use (Describe)
The BIOLOX delta Ceramic Femoral Heads are modular components used in total hip arthroplasty and are indicated for the following:
· Patients suffering from severe hip pain and disability due to rheumatoid arthritis, traumatic arthritis,
polyarthritis, collagen disorders, avascular necrosis of the femoral head, and nonunion of previous fractures of the femur. · Patients with congenital hip dysplasia, protrusio acetabuli, or slipped capital femoral epiphysis.
· Patients suffering from disability due to previous fusion; patients with previously failed endoprostheses and/or total hip components in the operative extremity.
· Patients with acute femoral neck fractures.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{6}------------------------------------------------
510(k) Number (if known) K200112
#### Device Name
BIOLOX® OPTION Ceramic Femoral Head System
The BIOLOX OPTION Ceramic Femoral Head System is comprised of modular components used in primary or revision total hip arthroplasty and is indicated for the following:
· Patients suffering from severe hip pain and disability due to rheumatoid arthritis, traumatic arthritis,
polyarthritis, collagen disorders, avascular necrosis of the femoral head, and nonunion of previous fractures of the femur.
- · Patients with congenital hip dysplasia, protrusio acetabuli, or slipped capital femoral epiphysis.
- · Patients suffering from disability due to previous fusion.
- · Patients with previously failed endoprostheses and/or total hip components in the operative extremity.
- · Patients with acute neck fractures.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{7}------------------------------------------------
510(k) Number (if known) K200112
Device Name Fitmore® Hip Stem
| Indications for Use (Describe) |
|--------------------------------|
|--------------------------------|
- This femoral stem is for total hip or hemi-hip arthroplasty and is indicated for the following conditions:
- Noninflammatory degenerative joint disease (NID), e. g., avascular necrosis, osteoarthritis and inflammatory joint disease (IJD), e. g., rheumatoid arthritis.
- · Failed previous hip surgery (not THA) where pain, deformity or dysfunction persists.
- · Optional use in revision: in some medical conditions (e. g., early revision when healthy and good bone stock exists) the surgeon may opt to use primary implants in a revision procedure.
This stem is for uncemented use only.
Type of Use (*Select one or both, as applicable*)X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{8}------------------------------------------------
510(k) Number (if known) K200112
Device Name Wagner Cone Prosthesis
• Noninflammatory degenerative joint disease (NID), e.g., avascular necrosis, osteoarthritis, and inflammatory joint disease (IJD), e.g., rheumatoid arthritis.
· Failed previous surgery where pain, deformity, or dysfunction persists.
· Optional use in revision: in some medical conditions (e.g., early revision when healthy and good bone stock exists), the surgeon may opt to use primary implants in a revision procedure.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
| <span style="font-family:Wingdings;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family:Wingdings;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{9}------------------------------------------------
510(k) Number (if known) K200112
Device Name Wagner SL Revision Stem Lateral
Indications for Use (Describe)
· Revision of previously failed hip arthroplasty
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the Zimmer Biomet logo. The logo consists of a blue circle with a white "Z" inside it, followed by the text "ZIMMER BIOMET" in a sans-serif font. The text is also blue, and the logo is clean and modern.
# 510(k) Summary
| Sponsor: | Zimmer GmbH<br>Sulzerallee 8, P.O. Box<br>8404 Winterthur, Switzerland |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Anne-Kathrin Born, Regulatory Affairs<br>Telephone: +41 58 854 86 19<br>Fax: + 41 52 244 86 58 |
| Date: | April 02, 2020 |
| Trade Name: | Zimmer Hip Joint Replacement |
| Common Name: | Hip Prosthesis |
| Classification Name: | LZO - Prosthesis, hip, semi-constrained, metal/ceramic/polymer,<br>cemented or non-porous, uncemented (21 CFR §888.3353).<br>LPH - Prosthesis, hip, semi-constrained, metal/polymer, porous<br>uncemented (21 CFR §888.3358).<br>KWZ - Prosthesis, hip, constrained, cemented or uncemented,<br>metal/polymer (21 CFR §888.3310).<br>JDI - Prosthesis, hip, semi-constrained, metal/polymer, cemented<br>(21 CFR §888.3350).<br>KWL - Prosthesis, hip, hemi-, femoral, metal (21 CFR §888.3360).<br>KWY - Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or<br>uncemented (21 CFR §888.3390).<br>LWJ - Prosthesis, hip, semi-constrained, metal/polymer,<br>uncemented (21 CFR §888.3360).<br>MEH - Prosthesis, Hip, Semi-Constrained, Uncemented,<br>Metal/Polymer, Non-Porous, Calcium Phosphate (21 CFR<br>§888.3353). |
| Predicate Devices: | Wagner SL Revision Stem Lateral, manufactured by Zimmer<br>GmbH, K191781, cleared August 06, 2019<br>Wagner Cone Prosthesis System, manufactured by Zimmer<br>GmbH, K191781, cleared August 06, 2019<br>Biolox Delta Ceramic Femoral Heads, manufactured by Zimmer<br>GmbH, K192416, cleared October 01, 2019<br>Biolox Option Ceramic Femoral Head System, manufactured by<br>Zimmer GmbH, K192416, cleared October 01, 2019<br>Fitmore® Hip Stem, manufactured by Zimmer GmbH, K192236,<br>cleared November 5, 2019 |
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the Zimmer Biomet logo. The logo consists of a blue circle with a stylized "Z" inside, followed by the text "ZIMMER BIOMET" in a sans-serif font. The text is also in blue, matching the color of the circle and "Z" symbol. The logo appears to be clean and modern, representing the company's brand identity.
| | Avenir® Cemented Hip Stem, manufactured by Zimmer GmbH, K193030, cleared December 6, 2019 |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Avenir® Müller Stem, manufactured by Zimmer GmbH, K193030, cleared December 6, 2019 |
| | Alloclassic® Zweymüller® SL Femoral Stem, manufactured by Zimmer GmbH, K193050, cleared December 26, 2019 |
| | Alloclassic® Zweymüller® SL Offset Femoral Stem, manufactured by Zimmer GmbH, K193050, cleared December 26, 2019 |
| Device Description: | Zimmer Hip Joint Replacement Prostheses are hip replacement system components. A femoral stem component is used in conjunction with a femoral head component for replacement of the proximal femur in total hip arthroplasty. Femoral stems are available in different designs, materials, sizes, neck lengths and taper sizes. A taper is incorporated in the design of the stem to interlock it with the femoral head. Femoral head components are available in a variety of sizes and neck lengths for reconstruction of the natural head center of the femur. The purpose of this submission is the addition of MR conditional information to the labeling for the predicate devices. The addition of MR labeling to the subject devices does not impact indications, materials, design features or dimensions of the components, compatibility, packaging, or sterilization. |
| Indications for Use: | Wagner Cone Prosthesis are indicated for: <ul><li>Noninflammatory degenerative joint disease (NIDJD), e.g., avascular necrosis, osteoarthritis, and inflammatory joint disease (IJD), e.g., rheumatoid arthritis.</li><li>Failed previous surgery where pain, deformity, or dysfunction persists.</li><li>Optional use in revision: in some medical conditions (e.g., early revision when healthy and good bone stock exists), the surgeon may opt to use primary implants in a revision procedure.</li></ul> |
| | Wagner SL Revision Stem Lateral is indicated for: <ul><li>Revision of previously failed hip arthroplasty.</li></ul> |
| | The BIOLOX delta Ceramic Femoral Heads are modular components used in total hip arthroplasty and are indicated for the following: <ul><li>Patients suffering from severe hip pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head, and nonunion of previous fractures of the femur</li></ul> |
femur.
- Patients with congenital hip dysplasia, protrusio acetabuli, .
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the logo for Zimmer Biomet. The logo consists of a blue circle with a stylized "Z" inside, followed by the text "ZIMMER BIOMET" in a sans-serif font. The text is also in blue, matching the color of the circle and the "Z" symbol.
or slipped capital femoral epiphysis.
- . Patients suffering from disability due to previous fusion; patients with previously failed endoprostheses and/or total hip components in the operative extremity.
- . Patients with acute femoral neck fractures.
The BIOLOX OPTION Ceramic Femoral Head System is comprised of modular components used in primary or revision total hip arthroplasty and is indicated for the following:
- Patients suffering from severe hip pain and disability due to . rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head, and nonunion of previous fractures of the femur.
- o Patients with congenital hip dysplasia, protrusio acetabuli, or slipped capital femoral epiphysis.
- . Patients suffering from disability due to previous fusion.
- Patients with previously failed endoprostheses and/or total . hip components in the operative extremity.
- . Patients with acute neck fractures.
The Fitmore® Hip Stem is for total hip or hemi-hip arthroplasty and is indicated for the following conditions:
- Noninflammatory degenerative joint disease (NIDJD), e. . g., avascular necrosis, osteoarthritis and inflammatory joint disease (IJD), e. g., rheumatoid arthritis.
- . Failed previous hip surgery (not THA) where pain, deformity or dysfunction persists.
- Optional use in revision: in some medical conditions (e. g., o early revision when healthy and good bone stock exists) the surgeon may opt to use primary implants in a revision procedure.
- This stem is for uncemented use only.
Avenir® Müller Stem is intended for:
- o Advanced wear of the joint due to degenerative, posttraumatic or rheumatic diseases.
- . Failed previous hip surgery (not THA) where pain, deformity, or dysfunction persists.
- . Optional use in revision: in some medical conditions (e.g., early revision when healthy and good bone stock exists), the surgeon may opt to use primary implants in a revision procedure.
- . Acute traumatic fracture of the femoral head or neck.
- . Avascular necrosis of the femoral head.
Avenir® Müller Stems are for cementless use only.
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image contains the logo for Zimmer Biomet. The logo consists of a blue circle with a stylized "Z" inside, followed by the text "ZIMMER BIOMET" in a sans-serif font. The text is also in blue, matching the color of the circle and the "Z" symbol.
The Avenir® Cemented Hip Stem is intended for total or hemi hip arthroplasty with cemented applications for rehabilitating hips damaged as a result of:
- Advanced wear of the joint due to degenerative, post-. traumatic, or rheumatic diseases
- . Failed previous hip surgery including ioint reconstruction (osteotomy), arthrodesis, hemiarthroplasty or total hip replacement (THR)
- . Acute traumatic fracture of the femoral head or neck
- o Avascular necrosis of the femoral head.
### Alloclassic® Zweymüller® SL/ SL Offset Femoral Stems are indicated for:
- . Noninflammatory degenerative joint disease (NIDJD), e.g., avascular necrosis, osteoarthritis, and inflammatory joint disease (IJD), e.q., rheumatoid arthritis.
- . Failed previous surgery where pain, deformity, or dysfunction persists.
- . Revision of previously failed hip arthroplasty.
Comparison to Predicate Device: The subject devices have the same intended use and are substantially equivalent to the legally marketed predicated devices. Performance Data (Nonclinical Non-Clinical Performance and Conclusions: and/or Clinical): The interactions of Zimmer Hip Joint Replacement implants with static and time varying magnetic fields during Magnetic Resonance Imaging procedures have been evaluated and the subject devices have been determined to be MR Conditional in accordance to ASTM F2503-13 Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment. Data and scientific rationalizations to quantify the
> RF-induced heating, static magnetic field interactions, and image artifact generation of the Zimmer Hip Joint Replacement implants in the 1.5 Tesla (T) and 3.0 T MRI clinically relevant environments have been generated.
Clinical Performance and Conclusions:
Clinical data and conclusions were not needed for this device.
Conclusion: The subject devices have the same intended use and similar indications for use as the predicate devices. The subject devices use the same operating principle, incorporate the same basic design and labeling and are manufactured and sterilized using the same materials and processes as the predicate devices. Except for the modifications described in this submission the
subject devices are identical to the predicate devices, and the performance data and analyses demonstrate that:
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the logo for Zimmer Biomet. The logo consists of a blue circle with a stylized "Z" inside, followed by the text "ZIMMER BIOMET" in a sans-serif font. The color of the text is a dark gray or blue, matching the color of the "Z" symbol.
- any differences do not raise new questions of safety and ● effectiveness as established with performance testing; and
- . the subject devices are at least as safe and effective as the legally marketed predicate devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.