s-Clean SQ-SL Implant System Mini is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple unit restorations including; cemented retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. This system is dedicated for one and two stage surgical procedures. This system is intended for delayed loading.
Device Story
Endosseous dental implant system; consists of fixture and abutments. Fixture placed in alveolar bone to replace missing tooth function; internal hex connection to upper prosthesis. Fixture surface treated with SLA (sandblasted, large-grit, acid-etched). Abutments include Sole Abutment S-Line and TiN-coated variants; cover screw protects inner hole during healing. Used in dental clinics by clinicians; surgical placement in mandible or maxilla. Fixtures and cover screws provided sterile; abutments non-sterile (steam sterilized by user). System supports tooth restoration; provides stable foundation for prosthetic devices.
Clinical Evidence
Bench testing only. Evidence includes sterilization validation (ISO 11137, ISO 17665), shelf-life testing (ASTM F1980), biocompatibility (ISO 10993 series), and bacterial endotoxin testing (ANSI/AAMI ST72, USP <161>, USP <85>).
Technological Characteristics
Materials: CP Titanium (ASTM F67) for fixtures/screws; Ti-6Al-4V ELI (ASTM F136) for abutments. Surface: SLA (fixtures), TiN coating (select abutments). Connection: Internal Hex. Sterilization: Gamma (fixtures), Steam (abutments). Dimensions: Fixtures Ø 3.7-4.1mm; Abutments Ø 4.0-5.0mm.
Indications for Use
Indicated for partially or fully edentulous mandibles and maxillae patients requiring single or multiple unit restorations (cemented or overdenture) or fixed bridgework support. Dedicated for one and two-stage surgical procedures; intended for delayed loading.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the symbol of the Department of Health & Human Services on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out next to it.
April 13, 2020
Dentis Co., Ltd. April Lee Consultant Withus Group Inc. 106 Superior Irvine, California 92620
Re: K200099
Trade/Device Name: s-Clean SQ-SL Implant System Mini Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE Dated: January 10, 2020 Received: January 16, 2020
Dear April Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or post marketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for
Srinivas Nandkumar, Ph.D. Director Implantable Dental Devices Team DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K200099
Device Name s-Clean SQ-SL Implant System Mini
#### Indications for Use (Describe)
s-Clean SQ-SL Implant System Mini is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple unit restorations including; cemented retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. This system is dedicated for one and two stage surgical procedures. This system is intended for delayed loading.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid 0MB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the word "DENTIS" in bold, black letters. To the left of the word is an orange graphic that appears to be a stylized flame or abstract design. The text is clear and legible, suggesting it may be a logo or brand name.
99, Seongseoseo-ro, Dalseo-gu, Daegu, Korea Tel : 82.53.582.2804 / Fax : 82.53.583.2806
#### 510(K) Summary K200099
#### Submitter
Dentis Co., Ltd. Gyu Ri Kim 99, Seongseoseo-ro, Dalseo-gu Daegu, 42718 Korea Email: kgr1026@dentis.co.kr Tel. +82-53-589-3541 Fax. +82-53-289-7922
### Official Correspondent
Withus Group Inc. April Lee 106 Superior, Irvine, CA 92620 USA Email: withus6664@gmail.com Phone: 1-909-274-9971 Fax: 1-909-460-8122
#### Device Information
- Trade Name: s-Clean SO-SL Implant System Mini ●
- Common Name: Dental Implant System
- Classification Name: implant, endosseous, root-form
- Product Code: DZE
- Secondary Product Code: NHA
- Panel: Dental
- Regulation Number: 872.3640
- Device Class: Class II
- . Date Prepared: 01/08/2020
#### Predicate Devices:
The subject device is substantially equivalent to the following predicate devices:
Primary Predicate
- K153639, OneQ-SL s-Clean Implant System by Dentis Co., Ltd. ●
#### Reference Predicates
- K161689, OSSTEM Implant System Abutment by Osstem Implant Co. Ltd ●
- K171027, Dentis Dental Implant System by Dentis Co., Ltd. ●
- K171694, s-Clean TiN Coating Abutments ●
#### Indication for Use:
s-Clean SQ-SL Implant System Mini is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple unit restorations including; cemented retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. This system is dedicated for one and two stage surgical procedures. This system is intended for delayed loading.
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the text "Dentis Co., Ltd." in a bold, sans-serif font. The text is arranged on a single line, with the company name followed by the abbreviation "Co." and "Ltd.". The text is black against a white background.
Image /page/4/Picture/1 description: The image shows the logo for DENTIS. The logo features an orange abstract shape on the left, resembling a stylized flame or a curved shape with a circle at the top. To the right of the shape, the word "DENTIS" is written in a bold, sans-serif font. The overall design is clean and modern.
99, Seongseoseo-ro, Dalseo-gu, Daegu, Korea Tel : 82.53.582.2804 / Fax : 82.53.583.2806
#### Device Description:
s-Clean SQ-SL Implant System Mini is composed of Fixture and Abutments. s-Clean SQ-SL Implant System Mini is a thread type implant made of Pure titanium according to ASTM F67 which will be placed in the alveolar bone to replace the function of the missing tooth. This device has connection between the upper prosthesis and the internal Hex.
Fixture's surface is treated with SLA (Sandblasted with Large-grit and Acid-etching). It is only part to be implanted into bone, and to provide connection of prosthetic devices or other components of a dental implant set with human body (mandibular or maxillary bone). The dimensions are as following:
| No. | Device Name | Dimension Ranges |
|-----|-----------------------|-------------------------------------------------------|
| 1 | s-Clean SQ-SL Fixture | Ø 3.7, 4.1mm (D) X 7.0, 7.5, 9.5, 11.5 and 13.5mm (L) |
Tolerance of dimension shall be within ± 1% range.
The s-Clean SQ-SL Implant System Mini Abutments are composed as below;
s-Clean Sole Abutment S-Line, s-Clean TiN Half Coating Sole Abutment S-Line, and s-Clean Mini Fixture Cover Screw
The dimensions of abutments are as following:
| No. | Device Name | Dimension Ranges | Angulation |
|-----|-----------------------------------------------|----------------------------------------------|------------|
| 1 | s-Clean Sole Abutment S-Line | Ø 4.0 and 5.0mm (D) X 11.2, 12.2, 13.2mm (L) | 0° |
| 2 | s-Clean TiN Half Coating Sole Abutment S-Line | Ø 4.0 and 5.0mm (D) X 11.2, 12.2, 13.2mm (L) | 0° |
| 3 | s-Clean Mini Fixture Cover Screw | Ø 3.1mm (D) X 5.4mm (L) | 0° |
The Abutments have below featured:
| Name | Uses | Surface | Connection |
|--------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|-------------------|
| s-Clean Sole Abutment S-Line | | N/A | Internal Hex |
| s-Clean TiN Half Coating Sole<br>Abutment S-Line | The Abutment is connected with fixture and it<br>supports prosthesis which restores tooth function. | TiN-Coating | Internal Hex |
| s-Clean Mini Fixture Cover<br>Screw | It is used for protecting inner hole and connecting<br>part with exposed upper part of structure during<br>the healing period after inserting dental implant<br>fixture | N/A | Screw<br>Retained |
Tolerance of dimension for Abutments shall be within ± 1% range.
s-Clean SQ-SL Fixture and s-Clean Mini Fixture Cover Screw are provided sterilized. And the other Abutments are provided non-sterilized.
s-Clean SO-SL Fixture is enclosed with s-Clean Mini Fixture Cover Screw in a set packing. s-Clean Mini Fixture Cover Screw is also provided separately.
#### Materials:
- The fixtures and s-Clean Mini Fixture Cover Screw are fabricated from Pure titanium of ASTM F67 ●
- . The s-Clean Sole Abutment S-Line and s-Clean TiN Half Coating Sole Abutment S-Line are fabricated from Ti-6Al-4V of ASTM F136
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "DENTIS" in bold, black letters. To the left of the word is an orange graphic that resembles a curved check mark with a circle at the top. The logo is simple and professional, likely representing a dental-related business.
99, Seongseoseo-ro, Dalseo-gu, Daegu, Korea
## Summaries of Technological Characteristics
| s-Clean SQ-SL Fixture | Subject Device | Primary Predicate Device | |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| 510(k) Number | N/A | K153639 | |
| Trade Name | s-Clean SQ-SL Implant System Mini | OneQ-SL s-Clean Implant System | |
| Manufacturer | Dentis Co., Ltd. | Dentis Co., Ltd. | |
| Design<br>(Fixture Type) | Image: s-Clean SQ-SL Implant System Mini<br>- Internal Hex-connected<br>- Submerged Fixture<br>- Bone level<br>- Tapered body<br>- 3 sided cutting edge with self-tapping | Image: OneQ-SL s-Clean Implant System<br>- Internal Hex-connected<br>- Submerged Fixture<br>- Bone Level<br>- Tapered & straight body<br>- 3 sided cutting edge with self-tapping | |
| Fixture<br>Diameter(Ø) | 3.7, 4.1mm | 3.7, 3.9, 4.2, 4.7, 5.2, 6.0, 7.0, 8.0mm | |
| Fixture Length | 7, 7.5, 9.5, 11.5, 13.5mm | 7, 8, 9, 10, 12, 14mm | |
| Indication for<br>Use | s-Clean SQ-SL Implant System Mini is<br>indicated for use in partially or fully edentulous<br>mandibles and maxillae, in support of single or<br>multiple unit restorations including; cemented<br>retained, screw retained, or overdenture<br>restorations, and terminal or intermediate<br>abutment support for fixed bridgework. This<br>system is dedicated for one and two stage<br>surgical procedures. This system is intended<br>for delayed loading. | The OneQ-SL s-Clean Implant System is<br>indicated for use in partially or fully edentulous<br>mandibles and maxillae, in support of single or<br>multiple unit restoration including;<br>cemented retained, screw retained, or<br>overdenture restoration, and terminal or<br>intermediate abutment support for fixed<br>bridgework. This system is dedicated for one<br>and two stage surgical procedures and not<br>dedicated for immediate loading. This system<br>is intended for delayed loading. | |
| Surface<br>Treatment | SLA | SLA | |
| Material | CP Titanium Grade 4<br>(ASTM F67) | CP Titanium Grade 4<br>(ASTM F67) | |
| Sterilization | Gamma Sterilization | Gamma Sterilization | |
| Product Code | DZE, NHA | DZE | |
| Shelf Life | 8 years | | |
| Brief<br>Comparison | s-Clean SQ-SL Implant System Mini has same device characteristics with the Primary<br>predicate devices, OneQ-SL s-Clean Implant System (K153639) such as indications for<br>Use, material, functions, general shape (Design), manufacturing process, dimensions,<br>structure, shelf life and surface treatment.<br>Any differences do not affect the clinical performance and efficacy. | | |
# s-Clean SO-SL Fixture
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "DENTIS" in bold, black font. To the left of the word is an orange graphic. The graphic is a curved shape with a circle at the top.
99, Seongseoseo-ro, Dalseo-gu, Daegu, Korea
## s-Clean Sole Abutment S-Line
| | Subject Device | Reference Device | Reference Device |
|------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) | N/A | K171027 | K161689 |
| Trade Name | s-Clean SQ-SL Implant<br>System Mini | Dentis Dental Implant<br>System s-Clean Sole<br>Abutment | OSSTEM Implant System<br>- Abutment |
| Manufacturer | Dentis Co., Ltd. | Dentis Co., Ltd. | OSSTEM Implant Co.,<br>Ltd. |
| Product Name | s-Clean Sole Abutment S-<br>Line | s-Clean Sole Abutment | Rigid Abutment |
| Design | Image: s-Clean SQ-SL Implant System Mini | Image: Dentis Dental Implant System s-Clean Sole Abutment | Image: OSSTEM Implant System - Abutment |
| Dimension | Ø 4.0, 5.0mm (D) X 11.2,<br>12.2, 13.2mm (L) | Ø 4.5, 4.8, 5.5, 6.0, 6.5mm<br>(D) X<br>12.5,13.0,13.5,14.,15.0,16.0,17.0mm (L) | Ø 4.0, 4.6, 5.0, 6.0, 7.0mm<br>(D) X 10, 10.4, 11, 11.4,<br>11.5, 11.9, 12, 12.4, 12.5,<br>12.9, 13, 13.4, 13.5, 13.9,<br>14, 14.4, 14.5, 14.9, 15,<br>15.4, 15.5, 15.9, 16, 16.4,<br>17, 17.4mm (L) |
| Angulation | N/A | N/A | N/A |
| Material | Ti-6Al-4V ELI<br>(ASTM F136) | Ti-6Al-4V ELI<br>(ASTM F136) | Titanium Alloy<br>(Ti-6Al-4V) |
| Sterilization | Steam sterilization by user | Steam sterilization by user | Steam sterilization by user |
| Product Code | NHA | NHA | NHA |
| Brief Comparison | The subject device has no difference from the predicate device that affects the clinical<br>performance and efficacy. Dimensional specification of the subject device falls into<br>the range of the reference device. | | |
#### s-Clean TiN Half Coating Sole Abutment S-Line
| | Subject Device | Reference Device | Reference Device |
|--------------|------------------------------------------------------|--------------------------------------------------|-------------------------------------|
| 510(k) | N/A | K171694 | K161689 |
| Trade Name | s-Clean SQ-SL Implant<br>System Mini | s-Clean TiN Coating<br>Abutments | OSSTEM Implant System -<br>Abutment |
| Manufacturer | Dentis Co., Ltd. | Dentis Co., Ltd. | OSSTEM Implant Co., Ltd. |
| Product Name | s-Clean TiN Half Coating<br>Sole Abutment S-Line | s-Clean TiN Partial Coating<br>Sole Abutment | Rigid Abutment |
| Design | Image: s-Clean TiN Half Coating Sole Abutment S-Line | Image: s-Clean TiN Partial Coating Sole Abutment | Image: Rigid Abutment |
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the word "DENTIS" in bold, black font. To the left of the word is an orange graphic that resembles a flame or a stylized letter "D". The graphic is a gradient, with the bottom being a darker orange and the top being a lighter orange.
99, Seongseoseo-ro, Dalseo-gu, Daegu, Korea
| Dimension | Ø 4.0, 5.0mm (D) X 11.2, 12.2, 13.2mm (L) | Ø 4.5, 5.5, 6.0 and 6.5mm (D) x 11, 11.5, 12.0, 12.5, 13.5, 14.5, 15.5mm (L) | Ø 4.0, 4.6, 5.0, 6.0, 7.0mm (D) X 10, 10.4, 11, 11.4, 11.5, 11.9, 12, 12.4, 12.5, 12.9, 13, 13.4, 13.5, 13.9, 14, 14.4, 14.5, 14.9, 15, 15.4, 15.5, 15.9, 16, 16.4, 17, 17.4mm (L) |
|-------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Angulation | N/A | N/A | N/A |
| Material | Ti-6Al-4V ELI (ASTM F136) | Ti-6Al-4V ELI (ASTM F136) | Titanium Alloy (Ti-6Al-4V) |
| Surface Treatment | TiN Coating | TiN Coating | Non-Coating |
| Sterilization | Steam sterilization by user | Steam sterilization by user | Steam sterilization by user |
| Product Code | NHA | NHA | NHA |
| Brief Comparison | The subject device has no difference from the predicate device that affects the clinical performance and efficacy. Dimensional specification of the subject device falls into the range of the reference device. | | |
### s-C lean Mini Fixture Cover Screw
| | Subject Device | Primary Predicate deviece |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|
| 510(k) | N/A | K153639 |
| Trade Name | s-Clean SQ-SL Implant System Mini | OneQ-SL s-Clean Implant System |
| Manufacturer | Dentis Co., Ltd. | Dentis Co., Ltd. |
| Design | Image: Subject Device Design | Image: Primary Predicate Device Design |
| Dimension | Ø 3.1mm (D) X 5.4mm (L) | Ø 3.6mm (D) X 5.9mm (L) |
| Angulation | N/A | N/A |
| Material | CP Titanium Grade 4<br>(ASTM F67) | CP Titanium Grade 4<br>(ASTM F67) |
| Sterilization | Gamma Sterilization | Gamma Sterilization |
| Product Code | NHA | NHA |
| Brief<br>Comparison | The subject device is a bit smaller than the predicate device in its diameter and length.<br>This device (cover screw) is to seal the fixture opening from the exposure to the<br>surrounding tissue during the healing process. Thus, the measurement is not an<br>important factor to the device performance but the fitness accuracy to the fixture<br>opening is. The dimensional difference is ignorable. | |
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the word "DENTIS" in bold, black letters. To the left of the word is an orange graphic that appears to be a stylized flame or swirl. The text and graphic are aligned horizontally. The overall impression is that of a company logo.
#### Non-Clinical Test Data
Below tests were performed for predicate devices and leveraged for the subject device:
- . Sterilization Validation Test on Fixtures according to ISO 11137-1,2,3 referenced in K153639
- . End User Sterilization Validation Test Report on Abutments according to ANSVAAMI ST79, ISO 17665-1, ISO 17665-2, ISO 11737-1, ISO 11737-2, and ISO 11138-1 referenced in K171027
- . Shelf Life Test on Fixtures according to ASTM F1980 referenced in K153639
- . Biocompatibility testing on fixtures according to ISO 10993-1:2009. ISO 10993-3:2014. ISO 10993-5:2009, ISO 10993-6:2007, ISO 10993-10:2010 and ISO 10993-11:2006 referenced in K153639
- Biocompatibility testing on Abutments according to ISO 10993-1:2009, ISO 10993-3:2014, ISO 10993-5:2009, ISO 10993-6:2007, ISO 10993-10:2010 and ISO 10993-11:2006 referenced in K171027
- . Biocompatibility testing on TiN Coating Abutments according to ISO 10993-1:2009, ISO 10993-5:2009, ISO 10993-10:2010 and ISO 10993-11:2006 referenced in K171694
- . Bacterial Endotoxin Test Report on Fixtures according to ANSI/AAMI ST72:2011, USP <161>, and USP <85> referenced in K192688
The results of the above tests have met the criteria of the standards, and demonstrated the substantial equivalence with the predicate device. The non-clinical testing results demonstrate that the subject device is substantially equivalent to the predicate device.
### Conclusion
s-Clean SQ-SL Implant System Mini constitutes a substantially equivalent medical device, meeting all the declared requirements of its intended use. This system has the same intended use and fundamental scientific technology as its predicate devices. Therefore, s-Clean SQ-SL Implant System Mini and its predicates are substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.