F3D Cervical Stand-Alone Interbody Fusion System
K200087 · Corelink, LLC · OVE · Jun 15, 2020 · Orthopedic
Device Facts
| Record ID | K200087 |
| Device Name | F3D Cervical Stand-Alone Interbody Fusion System |
| Applicant | Corelink, LLC |
| Product Code | OVE · Orthopedic |
| Decision Date | Jun 15, 2020 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3080 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The F3D Cervical Stand-Alone System is a stand-alone anterior cervical interbody fusion system indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one or two contiguous disc levels. DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. The F3D Cervical Stand-Alone System is used to facilitate intervertebral body fusion in the cervical spine and is placed via an anterior approach at one- or two-disc levels (C2-T1) using autograft bone. Patients should have at least six (6) weeks of non-operative treatment prior to treatment. The F3D Cervical Stand-Alone Interbody Fusion System is intended to be used with the bone screw fixation provided and requires no additional fixation.
Device Story
F3D Cervical Stand-Alone Interbody Fusion System; implantable spinal hardware for anterior cervical interbody fusion. System comprises additively manufactured titanium alloy (Ti-6Al-4V) interbody spacers and machined titanium alloy bone screws. Spacer features trapezoidal column shape with vertical cavity for autograft bone packing. Device provides surgical stabilization of spine; requires no additional fixation. Used by surgeons in clinical settings to treat degenerative disc disease. Output is mechanical stabilization of vertebral segments to facilitate fusion. Benefits include restoration of disc height and spinal stability.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included static and dynamic axial compression, compression shear, and torsion per ASTM F2077, as well as subsidence testing per ASTM F2267 and expulsion testing.
Technological Characteristics
Materials: Ti-6Al-4V (ASTM F3001, ASTM F136, ISO 5832-3). Form factor: Trapezoidal column spacer with integrated bone screw fixation. Manufacturing: Additive (spacers) and subtractive (screws).
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C2-T1) with radicular symptoms at one or two contiguous levels. Requires failure of at least six weeks of non-operative treatment. Used with autograft bone.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
- Matrixx Stand Alone Cervical System (K190546)
- COALITION® MIS SPACER (K173115)
- Shoreline™ ACS System (K190655)
- Unison-C Anterior Cervical Fixation System (K152793)
- Acapella Cervical Spacer System (K171489)
Reference Devices
- Foundation™ 3D Interbody (K162496)
Related Devices
- K200543 — NEXXT MATRIXX Stand Alone Cervical-Turn Lock (-TL) System · Nexxt Spine, LLC · Mar 23, 2020
- K212904 — SeaSpine WaveForm C Interbody System · SeaSpine Orthopedics Corporation · Dec 14, 2021
- K252737 — DeGen Medical Latitude-C AM Cervical Interbody Fusion System · Degen Medical · Oct 10, 2025
- K213359 — SeaSpine WaveForm, C Interbody System · SeaSpine Orthopedics Corporation · Dec 9, 2021
- K191489 — Genesys Spine 3DP Cervical Interbody System · Genesys Spine · Jan 8, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
June 15, 2020
CoreLink, LLC % Nathan Wright, MS Engineer & Regulatory Specialist Empirical Testing Corp. 4628 Northpark Drive Colorado Springs, Colorado 80918
Re: K200087
Trade/Device Name: F3D Cervical Stand-Alone Interbody Fusion System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: OVE Dated: May 12, 2020 Received: May 15, 2020
Dear Mr. Wright:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Brent Showalter, Ph.D. Acting Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement on last page.
-
510(k) Number (if known) K200087
Device Name
F3D Cervical Stand-Alone Interbody Fusion System
Indications for Use (Describe)
The F3D Cervical Stand-Alone System is a stand-alone anterior cervical interbody fusion system indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one or two contiguous disc levels. DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. The F3D Cervical Stand-Alone System is used to facilitate intervertebral body fusion in the cervical spine and is placed via an anterior approach at one- or two-disc levels (C2-T1) using autograft bone. Patients should have at least six (6) weeks of non-operative treatment prior to treatment.
The F3D Cervical Stand-Alone Interbody Fusion System is intended to be used with the bone screw fixation provided and requires no additional fixation.
{3}------------------------------------------------
# 510(K) SUMMARY
| Submitter's Name: | CoreLink, LLC |
|----------------------------|------------------------------------------------------------------------------------------|
| Submitter's Address: | 2072 Fenton Logistics Park Blvd.<br>St. Louis, Missouri 63026 |
| Submitter's Telephone: | 888-349-7808 |
| Contact Person: | Nathan Wright MS<br>Empirical Testing Corp.<br>719-351-0248<br>nwright@empiricaltech.com |
| Date Summary was Prepared: | 14-Jan-2020 |
| Trade or Proprietary Name: | F3D Cervical Stand-Alone Interbody Fusion System |
| Common or Usual Name: | Intervertebral Fusion Device with Integrated Fixation, Cervical |
| Classification: | Class II per 21 CFR §888.3080 |
| Product Code: | OVE |
| Classification Panel: | Division of Orthopedic Devices |
### DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:
The CoreLink F3D Cervical Stand-Alone Interbody Fusion System is a collection of additively and subtractively manufactured implants and associated instruments for surgical site preparation and implantation. The subject cages are additively manufactured from Ti-6Al-4V per ASTM F3001. The subject screws are machined from Ti-6Al-4V per ASTM F136. The F3D Cervical Stand-Alone Interbody Fusion System includes additively manufactured interbody spacers. The spacer and screw components are available in an assortment of dimensional combinations to accommodate the individual anatomic and clinical circumstances of each patient. The basic shape of the spacer is a trapezoidal column to provide surgical stabilization of the spine. The inferior/superior aspects of the spacer incorporate a vertical cavity which can be packed with bone graft.
#### INDICATIONS FOR USE
The F3D Cervical Stand-Alone System is a stand-alone anterior cervical interbody fusion system indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one or two contiguous disc levels. DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. The F3D Cervical Stand-Alone System is used to facilitate intervertebral body fusion in the cervical spine and is placed via an anterior approach at one- or two-disc levels (C2-T1) using autograft bone. Patients should have at least six (6) weeks of nonoperative treatment prior to treatment.
The F3D Cervical Stand-Alone Interbody Fusion System is intended to be used with the bone screw fixation provided and requires no additional fixation.
{4}------------------------------------------------
### TECHNOLOGICAL CHARACTERISTICS
The F3D Cervical Stand-Alone Interbody Fusion System is made from titanium alloy that conforms to ASTM F3001, ISO 5832-3, and ASTM F136. The subject and predicate devices have nearly identical technological characteristics and the minor differences do not raise any new issues of safety and effectiveness. Specifically, the following characteristics are identical between the subject and predicates:
- Indications for Use
- Materials of manufacture ●
- Structural support mechanism ●
| 510k Number | Trade or Proprietary or<br>Model Name | Manufacturer | Predicate<br>Type |
|-------------|-----------------------------------------------|-----------------------------------------|-------------------|
| K190546 | Matrixx Stand Alone<br>Cervical System | Nexxt Spine LLC | Primary |
| K173115 | COALITION® MIS<br>SPACER | Globus Medical Inc. | Additional |
| K190655 | Shoreline™ ACS System | SeaSpine®<br>Orthopedics<br>Corporation | Additional |
| K152793 | Unison-C Anterior Cervical<br>Fixation System | RTI Surgical, Inc. | Additional |
| K171489 | Acapella Cervical Spacer<br>System | Choice Spine, LP. | Additional |
| K162496 | Foundation™ 3D Interbody | CoreLink, LLC | Reference |
## Performance Data
The F3D Cervical Stand-Alone Interbody Fusion System has been tested in the following test modes:
- Static and dynamic axial compression per ASTM F2077 ●
- Static and dynamic compression shear per ASTM F2077
- Static and dynamic torsion per ASTM F2077 ●
- Subsidence per ASTM F2267 ●
- . Expulsion
The results of this non-clinical testing show that the strength of the F3D Cervical Stand-Alone Interbody Fusion System is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices.
{5}------------------------------------------------
### CONCLUSION
The overall technology characteristics and mechanical performance data lead to the conclusion that the F3D Cervical Stand-Alone Interbody Fusion System is substantially equivalent to the predicate device.