K200077 · Dr. Collins, Inc. · LBH · Nov 6, 2020 · Dental
Device Facts
Record ID
K200077
Device Name
BioMin Restore Plus
Applicant
Dr. Collins, Inc.
Product Code
LBH · Dental
Decision Date
Nov 6, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3260
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Prescription fluoride dentifrice for use in the prevention and treatment of dental sensitivity.
Device Story
BioMin Restore Plus is a prescription fluoride dentifrice used for dental sensitivity. Active ingredient: BioMin F (calcium fluoro-phosphosilicate bioactive glass). Mechanism: upon exposure to aqueous environment (saliva), bioactive glass undergoes surface reaction to form fluorapatite, which physically occludes open dentin tubules. Applied to teeth via brush or similar applicator. Benefits: reduces dentinal hypersensitivity by mimicking natural tooth mineral structure. Used in clinical or home settings under prescription.
Clinical Evidence
Bench testing only. Dentine tubule occlusion effectiveness demonstrated to be equivalent to primary predicate. Relative Dentin Abrasivity (RDA) measured and found well below ADA safety limit (<250). Biocompatibility testing performed per ISO 10993; biological evaluation and toxicological risk assessment indicate no hazardous effects.
Technological Characteristics
Active ingredient: Calcium Fluoro-Phosphosilicate (bioactive glass). Form factor: Paste/gel. Principle: Physical occlusion of dentin tubules via fluorapatite layer formation. Biocompatibility: ISO 10993 compliant.
Indications for Use
Indicated for patients requiring prevention and treatment of dental sensitivity.
Regulatory Classification
Identification
Cavity varnish is a device that consists of a compound intended to coat a prepared cavity of a tooth before insertion of restorative materials. The device is intended to prevent penetration of restorative materials, such as amalgam, into the dentinal tissue.
Special Controls
*Classification.* Class II (special controls). The device, when it is an external cleaning solution, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 6, 2020
Dr. Collins. Inc. Jiahe Li Regulatory Affair Associate 4330 Barranca Pkwy. Ste #100 Irvine, California 92604
Re: K200077
Trade/Device Name: BioMin Restore Plus Regulation Number: 21 CFR 872.3260 Regulation Name: Cavity Varnish Regulatory Class: Class II Product Code: LBH Dated: July 19, 2020 Received: August 10, 2020
Dear Jiahe Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Section IV
# Indications for Use Statement
510(k) Number (if known): K200077
Device Name: _BioMin® Restore Plus
Indications for Use:
Prescription fluoride dentifrice for use in the prevention and treatment of dental sensitivity.
Prescription Use _____ X ____ OR Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
{3}------------------------------------------------
# Section V
# 510(k) Summary
### Submitter:
Dr. Collins, Inc. 4330 Barranca Pkwy. Ste #100 Irvine, CA 92604 Contact Person: Jiahe Li Regulatory Affairs Manager Lijh0919@gmail.com
#### Date Summary Prepared:
November 4th, 2020
DEVICE NAME
| TRADE NAME: | BioMin® Restore Plus |
|------------------------------|------------------------|
| COMMON NAME: | Dentifrice, Toothpaste |
| DEVICE CLASS: | Class II |
| CLASSIFICATION NAME: | Cavity Varnish |
| CLASSIFICATION PRODUCT CODE: | LBH |
| 510(K) NUMBER: | K200077 |
| REGULATION NUMBER: | 21 CFR 872.3260. |
## PREDICATE DEVICE
Primary Predicate Decice: Restore Toothpaste - Dr. Collins, Inc. - K181965
{4}------------------------------------------------
Reference Device: Voco Paste - VOCO GMBH - K101104
#### DESCRIPTION OF DEVICE
BioMin® Restore Plus is a daily-use, prescription fluoride dentifrice use in the prevention and treatment of dental sensitivity through physical occlusion of dentin tubules. The formulation incorporates BioMin™ F (Calcium Fluoro-Phosphosilicate, a particular form of bioactive glass) as its active ingredient. When exposed to an aqueous environment, BioMin™ F undergoes a rapid surface reaction to form Fluorapatite, which is chemically similar to natural tooth mineral, allowing it to physically occlude tubules.
#### INDICATIONS FOR USE
Prescription fluoride dentifrice for use in the prevention and treatment of dental sensitivity.
#### Technological Characteristics
The fundamental principle and mode of action of BioMin® Restore Plus is the same as the predicate devices. All three devices are designed to reduce dentinal hypersensitivity through the physical occlusion of open dentin tubules by the formation of an apatite layer.
BioMin™ F (Calcium Fluoro-Phosphosilicate, a particular form of bioactive glass) is the active ingredient in BioMin® Restore Plus's formulation. When exposed to an aqueous environment, BioMin™ F undergoes a rapid surface reaction to form Fluorapatite, which is chemically and structurally similar to natural tooth mineral, allowing it to physically occlude tubules.
#### Performance Tests
The formula of BioMin® Restore Plus was tested for its ability to reduce dentine hypersensitivity through physical occlusion of dentine tubule and its relative abrasion level of dentine and enamel.
The dentine tubule occlusion test result indicates that BioMin® Restore Plus is as effective as the primary predicate device at dentine tubule occlusion. Meanwhile, the Relative Dentin Abrasivity (RDA) value is similar to the primary predicate and is well under the limit considered safe for daily use (<250) established by American Dental Association (ADA).
#### Biocompatibility
Different biocompatibility tests in accordance with ISO10993 have been performed on the subject
{5}------------------------------------------------
device. Meanwhile, a biological evaluation and a toxicological risk assessment have also been conducted on the subject device. The results of these tests and studies indicate there is no evidence of any hazardous effects and the subject device is safe for its intended use.
### Device Comparison Table
| Element<br>of<br>Comparison | Subject<br>Device<br>BioMin®<br>Restore Plus | Primary<br>Predicate<br>Device<br>Restore<br>Toothpaste -<br>K181965 | Reference<br>Device<br>Voco Paste<br>K101104 | Substanti<br>ally<br>Equivalen<br>t | Remarks |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | For dental<br>hypersensitivit<br>y relief | For dental<br>hypersensitivit<br>y relief | For dental<br>hypersensitivity<br>relief | Yes | -- |
| Indications for<br>Use | Prescription<br>fluoride<br>dentifrice for<br>use in the<br>prevention<br>and treatment<br>of dental<br>sensitivity. | A daily use<br>over-the-<br>counter<br>toothpaste<br>formulated to<br>provide rapid<br>and continual<br>tooth<br>sensitivity<br>relief. | Intended to be<br>used after<br>professional<br>tooth whitening,<br>professional<br>tooth cleaning<br>and for<br>prevention and<br>control of<br>hypersensitivitie<br>ട. | Yes | The only difference is<br>that the subject device<br>and reference device<br>are for prescription<br>use, while the primary<br>predicate device is for<br>OTC use. |
| Design | Paste/gel that<br>is applied to<br>the tooth<br>using an<br>applicator<br>brush or<br>similar<br>application. | Paste/gel that<br>is applied to<br>the tooth<br>using an<br>applicator<br>brush or<br>similar<br>application. | Paste/gel that is<br>applied to the<br>tooth using an<br>applicator brush<br>or similar<br>application. | Yes | -- |
| Composition of<br>Material | Apatite<br>forming<br>ingredient,<br>filler,<br>surfactant,<br>thickening<br>agent,<br>colorant, | Apatite<br>forming<br>ingredient,<br>filler,<br>surfactant,<br>thickening<br>agent,<br>colorant, | Apatite forming<br>ingredient &<br>hydroxyapatite,<br>filler, surfactant,<br>thickening<br>agent, colorant,<br>sweetener &<br>flavor enhancer. | Yes | The subject device<br>and the primary<br>predicate device have<br>exact the same<br>composing<br>ingredients, except for<br>the active ingredients,<br>which are also |
{6}------------------------------------------------
| Technological<br>Characteristics | Mode of<br>Action | sweetener &<br>flavor<br>enhancer.<br>Physical<br>occlusion of<br>dentin tubules | sweetener &<br>flavor<br>enhancer.<br>Physical<br>occlusion of<br>dentin tubules | Physical<br>occlusion of<br>dentin tubules | Yes | compositionally and<br>structurally very<br>similar. |
|----------------------------------|---------------------------|--------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Principle<br>of<br>Action | Occlude<br>dentin tubules<br>through<br>reaction with<br>saliva and<br>subsequent<br>formation of<br>apatite layer | Occlude<br>dentin tubules<br>through<br>reaction with<br>saliva and<br>subsequent<br>formation of<br>apatite layer | Occlude dentin<br>tubules through<br>reaction with<br>saliva and<br>subsequent<br>formation of<br>apatite layer<br>and direct<br>occlusion with<br>pre-formed<br>hydroxyapatite<br>particles | Yes | The technological<br>characteristics<br>between the subject<br>device and primary<br>predicate are the<br>same.<br>The reference device<br>differs slightly in that it<br>also contains some<br>pre-formed<br>hydroxyapatite<br>particles which can<br>occlude dentine tubule<br>directly. |
| | Biocompatibility | Yes | Yes | Yes | Yes | -- |
## Conclusion
Based on the comparison of the compositions and actions, as well as the testing data provided, BioMin® Restore Plus is considered to be substantially equivalent to the identified legally marketed predicate devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.