K193622 · Perfection Aligner System Hong Kong Limited · NXC · Jul 6, 2021 · Dental
Device Facts
Record ID
K193622
Device Name
TRIOCLEAR System
Applicant
Perfection Aligner System Hong Kong Limited
Product Code
NXC · Dental
Decision Date
Jul 6, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.5470
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TRIOCLEAR System is a series of clear, lightweight, plastic appliances indicated for the treatment of tooth malocclusions in patients with permanent dentition (i.e. all second molars). Utilizing a series of incremental tooth movement, it sequentially positions teeth by way of continuous gentle force.
Device Story
TRIOCLEAR System consists of clear, removable, thermoformed plastic aligners for orthodontic treatment. Dental professional takes physical or optical measurements of patient teeth; sends data and prescription to manufacturer. Manufacturer uses standard dental software to design series of plastic trays for incremental tooth movement; prescribing physician reviews and approves model scheme. Trays produced via thermoforming; provided to patient by dental professional. Patient wears trays sequentially to apply continuous gentle force to teeth, moving them to desired positions. Treatment monitored by dental professional. System intended for single patient use; removable by patient.
Clinical Evidence
Bench testing only. Physical properties (tensile strength, modulus, elongation) tested per ISO 527-2; hardness per ASTM D785; humidity absorption tested. Digital/3D printed models and thermoformed aligners validated via CAD overlay and RMS calculations. Biocompatibility testing performed per ISO 10993-1 (cytotoxicity, sensitization, irritation, acute systemic toxicity, pyrogenicity, subacute toxicity, implantation).
Technological Characteristics
Thermoplastic material; thermoformed clear aligners. Sensing/actuation via mechanical force application. Dimensions/form factor: custom-fitted to patient dentition. Connectivity: data transfer for design. Sterilization: non-sterile. Software: standard dental alignment software used for design.
Indications for Use
Indicated for treatment of tooth malocclusions in patients with permanent dentition (all second molars).
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".
July 06, 2021
Perfection Aligner System Hong Kong Limited Alwin Ngai Director Flat/Rm 4B, 4/F, Hang Fat Industrial Building, 550-556 Castle Peak Road Kowloon, HONG KONG
Re: K193622
Trade/Device Name: TRIOCLEAR System Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NXC Dated: May 10, 2021 Received: June 2, 2021
Dear Alwin Ngai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193622
Device Name TRIOCLEAR System
Indications for Use (Describe)
The TRIOCLEAR System is a series of clear, lightweight, plastic appliances indicated for the treatment of tooth malocclusions in patients with permanent dentition (i.e. all second molars). Utilizing a series of incremental tooth movement, it sequentially positions teeth by way of continuous gentle force.
Type of Use (*Select one or both, as applicable*):
| <span>☒</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------|
| <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## K193622
# Section 5: 510(k) Summary
### 1. Submitter Information
| Company Name: | Perfection Aligner System Hong Kong Limited |
|------------------|-----------------------------------------------------------------------------------------------|
| Company Address: | Flat/Rm 4B, 4/F, Hang Fat Industrial Building<br>550-556 Castle Peak Road, Kowloon, Hong Kong |
| Company Phone: | (852) 3766 0737 |
| Company Fax: | (852) 2577 7511 |
| Contact Person: | Mr. Alwin Ngai |
| Email: | alwin.ngai@moderndentalgp.com |
| Date Prepared | June 29, 2021 |
### 2. Device Identification
| Device Model Name: | TRIOCLEAR System |
|----------------------|-----------------------------|
| Classification Name: | Orthodontic Plastic Bracket |
| Regulation Number: | 872.5470 |
| Product Code: | NXC |
| Class | II |
| Panel | Dental |
### 3. Predicated Devices
| Primary Predicate: | 3MTM Clear Tray Aligner, K163689 |
|--------------------|----------------------------------|
| Reference Device: | eCligner®, K143499 |
### 4. Device Description
The TRIOCLEAR System is a removable, non-sterile device intended for single patient use. TRIOCLEAR is a series of clear plastic aligners that offer a solution for patients who want an aesthetic orthodontic treatment by utilizing sets of removable aligners to correct tooth malocclusions without the use of conventional wire and bracket orthodontic technology.
### 5. Indications for Use
The TRIOCLEAR System is a series of clear, lightweight, plastic appliances
510(k) Files: Section 5
Page 1 of 7
{5}------------------------------------------------
indicated for the treatment of tooth malocclusions in patients with permanent dentition (i.e. all second molars). Utilizing a series of incremental tooth movement, it sequentially positions teeth by way of continuous gentle force.
## 6. Mechanism of Action
The mechanism of action is similar to the predicate devices and supports a determination of substantial equivalence. Orthodontic tooth movement occurs through forces applied by the appliance to the dentition as each tooth follows the programmed displacement based on a dental health professional's prescription.
# 7. Technological Characteristics
A dental health care professional (e.g. orthodontist or dentist), prescribes the TRIOCLEAR System based on an assessment of the patient's teeth, determines a course of treatment with the system, takes physical or optical measurements of the patient's teeth and completes a prescription from. The measurements and prescription are sent to Perfection.
Utilizing standard dental software used for tooth alignment. Perfection designs a series of plastic trays intended to gradually realign the patient's teeth in accordance with the physician's prescription. The prescribing physician reviews and approves the model scheme before the molds are produced. Once approved, Perfection produces the trays, which are formed of clear, thin thermoformed plastic. The trays are sent back to the dental health care professional, who then provides them to the patient, confirming fit and design. Over a period, additional trays are provided sequentially to the patient by the dental health care professional to gradually move the target teeth to the desired position.
The dental care professional monitors treatment from the moment the first aligner is delivered to when treatment is completed. The trays are held in place by pressure and can be removed by the patient at any time.
This technology is essentially identical to that used by a number of sequential alignment systems, including the predicated referenced below.
# 8. Performance Testing
510(k) Files: Section 5
{6}------------------------------------------------
Physical properties testing was conducted on the device. Tensile properties (tensile strength, modulus and elongation) were tested according to ISO 527-2. Hardness was tested according to ASTM D785. Humidity absorption was tested by immersion of specimen in distilled water during 24 hours at 23°C. All results of the above tests met the acceptance criteria.
The substantial equivalence of the device is supported by non-clinical testing. Validation was performed on the digital and 3D printed models, as well as the thermoformed aligners and the differences were calculated using a CAD overlay method and by using RMS calculations. All aligners fabricated met the pre-defined acceptance criteria and received an A grade for form and fit of the thermoformed aligners.
# 9. Biocompatibility Testing
The biocompatibility evaluation for the final finished device was conducted in accordance with the US FDA CDRH Guidance Document Number 1811 "Use of International Standard ISO 10993-1, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process - Guidance for Industry and Food and Drug Administration Staff" as recognized by FDA.
The TRIOCLEAR System is considered mucosal membrane direct contacting for a duration of greater than 30 days. The battery of testing included following tests:
- Cytotoxicity ●
- Sensitization ●
- Irritation
- Acute Systemic Toxicity
- Pyrogenicity
- Subacute Toxicity
- Implantation
The result of the testing met the requirements of study plan and the TRIOCLEAR System is considered non-cytotoxic, non-sensitizing, nonirritant, no acute systemic toxicity, non-pyrogenic, no subacute repeated exposure systemic toxicity, and nonirritant when compared to implanted negative control article.
# 10. Substantial Equivalence Comparison
510(k) Files: Section 5
{7}------------------------------------------------
The following table compares the TRIOCLEAR System to the predicate devices 3M™ Clear Tray Aligner and eCligner® with respect to indications for use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence.
{8}------------------------------------------------
| Elements of<br>Comparison | Subject Device | Primary Predicate Device | Reference Device |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | TRIOCLEAR System | 3M™ Clear Tray Aligner | eCligner® |
| 510(k) Number | K193622 | K163689 | K143499 |
| Manufacturer | Perfection Aligner System Hong Kong<br>Limited | 3M Unitek Corporation | eClear International Co., Ltd. |
| Regulation<br>Number | 21 CFR 872.5470 | 21 CFR 872.5470 | 21 CFR 872.5470 |
| Device<br>Classification<br>Name | Orthodontic plastic bracket | Orthodontic plastic bracket | Orthodontic plastic bracket |
| Classification<br>Product Code | NXC | NXC | NXC |
| Device Class | Class II | Class II | Class II |
| Indications for<br>Use | The TRIOCLEAR System is a series of clear,<br>lightweight, plastic appliances indicated for<br>the treatment of tooth malocclusions in<br>patients with permanent dentition (i.e. all<br>second molars). Utilizing a series of<br>incremental tooth movement, it sequentially<br>positions teeth by way of continuous gentle<br>force. | 3M™ Clear Tray Aligner is a series of clear,<br>lightweight, plastic appliances indicated for the<br>treatment of tooth malocclusions in patients<br>with permanent dentition (i.e. all second<br>molars). Utilizing a series of incremental tooth<br>movement, it sequentially positions teeth by<br>way of continuous gentle force. | eCligner® is a series of clear, lightweight,<br>plastic appliances indicated for the<br>correction of dental malocclusion in<br>patients with permanent dentition (i.e. all<br>second molars). The eCligner® System is<br>intended for minor anterior tooth<br>movement by way of continuous gentle<br>force. |
| Material of<br>Fabrication | Thermoplastic | Thermoplastic | Thermoplastic |
| Material<br>Properties | Demonstrates sufficient tensile strength,<br>ductility, chemical resistance, and clarity for<br>sue as a clear tray aligner. | Demonstrates sufficient tensile strength,<br>ductility, chemical resistance, and clarity for<br>sue as a clear tray aligner. | Demonstrates sufficient tensile strength,<br>ductility, chemical resistance, and clarity<br>for sue as a clear tray aligner. |
| Design | Image: Clear aligners arranged in a loose S shape | Image: Clear aligners arranged in a loose S shape | Image: Clear aligner for the lower teeth |
| Mode of Action | Orthodontic tooth movement occurs through<br>forces applied by the appliance to the<br>dentition as each tooth follows the<br>programmed displacement based on doctor's<br>prescription.<br>Three steps aligners will be worn by patient<br>for each set. | Orthodontic tooth movement occurs through<br>forces applied by the appliance to the dentition<br>as each tooth follows the programmed<br>displacement based on doctor's prescription.<br>One single aligner will be worn by patient for<br>each set. | Orthodontic tooth movement occurs<br>through forces applied by the appliance<br>to the dentition as each tooth follows the<br>programmed displacement based on<br>doctor's prescription.<br>Three steps aligners will be worn by<br>patient for each set. |
| Technological<br>Characteristics | Treatment of tooth malocclusion is via a<br>series of plastic appliances that incrementally<br>moves teeth to a desired end-state. | Treatment of tooth malocclusion is via a series<br>of plastic appliances that incrementally moves<br>teeth to a desired end-state. | Treatment of tooth malocclusion is via a<br>series of plastic appliances that<br>incrementally moves teeth to a desired<br>end-state. |
| OTC or Rx | Rx | Rx | Rx |
510(k) Files: Section 5
{9}------------------------------------------------
{10}------------------------------------------------
# 11. Substantial Equivalence Conclusion
Based on the indications for use and performance information provided in this submission, demonstrates that the TRIOCLEAR System is as safe, as effective, and is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.