← Product Code [HSB](/productcode/HSB) · K193617

# Precice® Screws (K193617)

_Nuvasive Specialized Orthopedics, Inc. · HSB · May 8, 2020 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K193617

## Device Facts

- **Applicant:** Nuvasive Specialized Orthopedics, Inc.
- **Product Code:** [HSB](/productcode/HSB.md)
- **Decision Date:** May 8, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3020
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The Precice® Screws are provided sterile and intended to be used with the systems below: The Precice® System is indicated for limb lengthening, open and closed fracture fixation, pseudoarthrosis, mal-unions, non-unions, or bone transport of long bones. The Precice® Stryde System is indicated for limb lengthening, open and closed fracture fixation, pseudoarthrosis, mal-unions, non-unions, or bone transport of long bones. The Precice® Bone Transport System is indicated for limb lengthening, open and closed fracture fixation, pseudoarthrosis, mal-unions, non-unions, or bone transport of long bones. The Precice® Plate System is indicated for limb lengthening, open and closed fracture fixation, pseudoarthrosis, mal-unions and non-unions of long bones for pediatrics and small stature adult patients.

## Device Story

Precice Screws are bone fixation fasteners used as accessories with previously cleared Precice systems (System, Stryde, Bone Transport, Plate). Available in lengths 10-100mm with various thread styles (pegs, partially/fully threaded). Primary purpose of submission is to provide these screws in a sterile format. Used by orthopedic surgeons in clinical settings for bone stabilization and lengthening procedures. Benefits include reduced preparation time for sterile components in the OR.

## Clinical Evidence

Bench testing only. Validation performed for gamma sterilization, sterile packaging, and sterile barrier integrity over time.

## Technological Characteristics

Metallic bone fixation fasteners (screws/pegs). Lengths 10-100mm. Sterile, gamma-sterilized. No software or electronic components.

## Regulatory Identification

An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.

## Predicate Devices

- Precice® System ([K172628](/device/K172628.md))
- Precice® Stryde System ([K180503](/device/K180503.md))
- Precice® Bone Transport ([K182170](/device/K182170.md))
- Precice® Plate System ([K192181](/device/K192181.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 8, 2020

NuVasive Specialized Orthopedics, Inc. Madison Heffron Regulatory Affairs Specialist 101 Enterprise. Suite 100 Aliso Viejo, California 92656

Re: K193617

Trade/Device Name: Precice Screws Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: Class II Product Code: HSB, KTT, HRS, HWC Dated: April 7, 2020 Received: April 9, 2020

Dear Madison Heffron:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K193617

Device Name

The Precice® Screws

### Indications for Use (Describe)

The Precice® Screws are provided sterile and intended to be used with the systems below:

The Precice® Systemis indicated for limb lengthening, open and closed fracture fixation, pseudoarthrosis, mal-unions, non-unions, or bone transport of long bones.

The Precice® Stryde System is indicated for limb lengthening, open and closed fracture fixation, pseudoarthrosis, mal-unions, non-unions, or bone transport of long bones.

The Precice® Bone Transport Systemis indicated for limb lengthening, open and closed fracture fixation, pseudoarthrosis, mal-unions, non-unions, or bone transport of long bones.

The Precice® Plate System is indicated for limb lengthening, open and closed fracture fixation, pseudoarthrosis, mal-unions and non-unions of long bones for pediatrics and small stature adult patients.

| Type of Use (Select one or both, as applicable)                                                                                                                     |                                                                                                                                                                    |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span> <div style="display:inline-block;"> <div style="display:inline-block;">☒</div>           Prescription Use (Part 21 CFR 801 Subpart D)         </div> </span> | <span> <div style="display:inline-block;"> <div style="display:inline-block;">☐</div>           Over-The-Counter Use (21 CFR 801 Subpart C)         </div> </span> |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Nuvasive Specialized Orthopedics, Inc. The logo features a stylized purple and gray graphic to the left of the company name, "NUVASIVE," in gray, sans-serif font. Below the company name is the text "SPECIALIZED ORTHOPEDICS, INC." in a smaller, purple, sans-serif font.

# Precice® Screws 510(k) Summary

In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR §807.92. the following summary of information is provided:

#### A. Submitted by:

Madison Heffron Regulatory Affairs Specialist NuVasive Specialized Orthopedics, Inc. 101 Enterprise, Suite 100 Aliso Viejo. CA 92656 Telephone: (949) 532-7868

Date Prepared: May 7th, 2020

#### B. Device Name

| Trade or Proprietary Name: | Precice® Screws                                    |
|----------------------------|----------------------------------------------------|
| Common or Usual Name:      | Screw, Fixation, Bone                              |
| Classification Name:       | Smooth or threaded metallic bone fixation fastener |
| Device Class:              | Class II                                           |
| Classification:            | 21 CFR § 888.3020, 888.3030, and 888.3040          |
| Product Code:              | HSB, KTT & HRS, and HWC                            |

#### C. Predicate Devices

The subject Precice® Screws are substantially equivalent to the predicate device the Precice® System (K172628) and additional predicates, Precice® Stryde System (K180503), Precice® Bone Transport (K182170), and Precice® Plate System (K192181).

#### D. Device Description

The Precice Screws accompanying previously cleared Precice System (K172628) and additional predicates, Precice Stryde System (K180503), Precice Bone Transport (K182170), and Precice Plate System (K192181) are available in a variety of lengths, materials and thread styles. The screws are offered in lengths between 10-100mm and styles including pegs, partially threaded and fully threaded shaft designs.

The purpose of this submission is to introduce a sterile option of previously cleared screws as part of the Precice System (K172628) and additional predicates, Precice Stryde System (K180503), Precice Bone Transport (K182170), and Precice Plate System (K192181).

#### E. Indications for Use

The Precice® Screws are provided sterile and intended to be used with the systems below:

{4}------------------------------------------------

The Precice® System is indicated for limb lengthening, open and closed fracture fixation. pseudoarthrosis, mal-unions, non-unions, or bone transport of long bones.

The Precice® Stryde System is indicated for limb lengthening, open and closed fracture fixation, pseudoarthrosis, mal-unions, non-unions, or bone transport of long bones.

The Precice® Bone Transport System is indicated for limb lengthening, open and closed fracture fixation, pseudoarthrosis, mal-unions, non-unions, or bone transport of long bones.

The Precice® Plate System is indicated for limb lengthening, open and closed fracture fixation, pseudoarthrosis, mal-unions and non-unions of long bones for pediatrics and small stature adult patients.

#### Technological Characteristics F.

As was established in this submission, the subject Precice Screws are substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States.

The subject device was shown to be substantially equivalent and have equivalent technological characteristics to its predicate device through comparison in areas including design, labeling/intended use, material composition, and function.

#### G. Performance Data

Gamma sterilization validation, sterile packaging validation, and integrity of the sterile barrier over time validation are performed to qualify new packaging and sterilization method for the Precice Screws.

The results demonstrate that the subject Precice Screws are substantially equivalent to the predicate.

#### H. Conclusions

The subject Precice Screws have been shown to be substantially equivalent to the legally marketed predicate devices.

---

**Source:** [https://fda.innolitics.com/device/K193617](https://fda.innolitics.com/device/K193617)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
