Viveve 2.0 System

K193611 · Viveve Medical, Inc. · GEI · Jan 16, 2020 · General, Plastic Surgery

Device Facts

Record IDK193611
Device NameViveve 2.0 System
ApplicantViveve Medical, Inc.
Product CodeGEI · General, Plastic Surgery
Decision DateJan 16, 2020
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Viveve 2.0 System is indicated for use in general surgical procedures for electrocoagulation and hemostasis.

Device Story

The Viveve 2.0 System is an electrosurgical device using monopolar radiofrequency (RF) energy to heat tissue beneath the surface while simultaneously delivering cryogen to the treatment tip to cool surface tissue (reverse thermal gradient). Components include an RF generator with integrated cooling module, handpiece, footswitch, and sterile disposable treatment tips (5cm or 8cm). The system includes a Contact Quality Monitoring (CQM) feature that monitors return pad contact via a contact map on the user interface to ensure safe energy dissipation. Used in clinical settings for electrocoagulation and hemostasis. The healthcare provider operates the system, monitoring the interface for contact quality and system status. The device benefits patients by providing controlled thermal coagulation during surgical procedures.

Clinical Evidence

Bench testing only. No clinical data presented. Verification testing included hardware/software verification of the CQM circuit (response time <20ms, resistance measurement accuracy), electrical safety (AAMI ES60601-1, CSA C22.2#60601-1), and EMC (IEC 60601-1-2). Validation testing using phantom tissue (pork belly) confirmed the CQM interface correctly detects return pad contact and triggers system faults (E63 error) upon loss of contact.

Technological Characteristics

Monopolar electrosurgical system; RF energy (6.78 MHz ±15%); cryogen cooling; 50W max power; impedance range 20–185 Ω. Includes Contact Quality Monitoring (CQM) circuit. Sterile disposable treatment tips (ETO sterilized). Connectivity includes return pad and cable. Software-controlled UI with contact map. Complies with AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-2, and IEC 60601-1-6.

Indications for Use

Indicated for use in general surgical procedures for electrocoagulation and hemostasis. Patient population not specified.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the agency's name, "U.S. Food & Drug Administration." January 16, 2020 Viveve Medical, Inc. Kevin Robison Regulatory Affairs Specialist 345 Inverness Drive South, Building B, Suite 250 Englewood, Colorado 80112 Re: K193611 Trade/Device Name: Viveve 2.0 System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: December 23, 2019 Received: December 26, 2019 Dear Kevin Robison: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {2}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Long Chen, PhD Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K193611 Device Name Viveve 2.0 System Indications for Use (Describe) The Viveve 2.0 System is indicated for use in general surgical procedures for electrocoagulation and hemostasis. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the word "viveve" in a sans-serif font. The color of the text is a light green. The letters are evenly spaced and the word is horizontally oriented. The background is white. ## SECTION 5 ### 510(k) SUMMARY #### 5.1 REGULATORY AUTHORITY Safe Medical Devices Act of 1990, 21 CFR 807.92 #### 5.2 APPLICANT INFORMATION - Viveve Inc. Applicant: 345 Inverness Drive South Building B, Suite 250 Englewood, CO 80112 - Contact: Kevin Robison Regulatory Affairs Specialist krobison@viveve.com C: 317-435-8898 F: 720-696-8199 Date Prepared: December 23, 2019 ### 5.3 SUBJECT DEVICE INFORMATION | Trade Name: | Viveve® 2.0 System | |------------------------|----------------------------------------------------------------------------------| | Common Name: | Electrosurgical System | | Product Code: | GEI | | Classification Name: | Electrosurgical Cutting and Coagulation Device and Accessories (21 CFR 878.4400) | | Device Panel: | General Surgery/Restorative Device | | Device Classification: | Class II | ### 5.4 PREDICATE DEVICE Viveve 2.0 System (K190422) ### 5.5 DEVICE DESCRIPTION The Viveve 2.0 System utilizes monopolar radiofrequency (RF) energy to selectively heat a given volume of tissue beneath the surface, while cryogen is delivered to the inside of the treatment tip to cool the surface tissue. The generator delivers energy to the treatment tip to create an electric field under the treatment tip (electrode). The mechanism of action is the application of RF energy to the tissue causing coagulation and/or hemostasis. The Viveve® 2.0 System consists of four (4) primary components: - . An RF Generator to provide the heating energy. The Generator incorporates the Cooling Module to supply coolant which provides the cooling energy. - . A hand piece that couples the cooling and heating energy to the tissue through the treatment tip. - A footswitch that allows the user to turn the RF Energy on or off. - . 5cm or 8cm Sterile Disposable Treatment Tips. {5}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the word "viveve" in a sans-serif font. The color of the text is a light green. The text is all lowercase. Accessories include: - Coupling Fluid - Cryogen ● - Return Cable ● - Return Pad ● - . Power Cord # 5.6 INDICATIONS FOR USE The Viveve 2.0 System is indicated for use in general surgical procedures for electrocoagulation and hemostasis. ## 5.7 COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE The technological characteristics of the subject device Viveve 2.0 System are substantially equivalent to the predicate device, Viveve 2.0 System (K190422). The Viveve 2.0 System is an electrosurgical device that delivers radiofrequency (RF) energy to selectively heat a given area of tissue, while cryogen is delivered to the inside of the treatment tip to cool the surface tissue at the end of energy deposition. The application of RF energy causes the tissue to coagulate and/or become hemostatic. The subject device differs from the predicate device in the addition of Contact Quality Monitoring (CQM). The device software monitors the return pad to determine if it is adequately secured to the patient's body in order to dissipate any residual radiofrequency energy. The display module software differs from that of the predicate device only in the addition of a Contact Map which indicates the amount of contact between the return pad and the patient's skin. In all other regards the display module software is identical. The design specifications of the subject device are identical to the predicate device specifications with the sole exception of the addition of CQM and the modification to the user interface. This submission application confirms the continued conformance to applicable technical design specifications and performance requirements, including requirements associated with industry safety and performance standards. ## 5.8 BASIS FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE The Viveve 2.0 System is substantially equivalent to the predicate device listed in K190422. The principle of operation between the predicate device and the subject device remain the same as do all output parameters to tissue. Specifications of the Viveve 2.0 System are discussed in further detailed in Section 09: Substantial Equivalence Discussion of this Premarket Notification. A comparison of the technical characteristics of Viveve 2.0 System are compared to those of the predicate device, Viveve 2.0 System, in Table 5-1 below. {6}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the word "viveve" in a sans-serif font. The color of the text is a light green. The text is slightly blurred, giving it a soft appearance. | Item | Viveve 2.0 System<br>(Subject Device) | Viveve 2.0 System<br>(Predicate Device K190422) | |----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) Number | Subject device | K190422 | | Legal Manufacturer | Viveve, Inc. | Viveve, Inc. | | Contract Manufacturer | Sparton Medical Systems Colorado, LLC<br>•<br>Cirtec Medical (Treatment Tips)<br>• | Sparton Medical Systems Colorado, LLC<br>•<br>Cirtec Medical (Treatment Tips)<br>• | | Indication for Use | The Viveve 2.0 System is indicated for use in general surgical procedures for electrocoagulation and hemostasis. | The Viveve 2.0 System is indicated for use in general surgical procedures for electrocoagulation and hemostasis. | | FDA Classification | Class II | Class II | | CFR/Product Code | 21 CFR 878.4400/GEI | 21 CFR 878.4400/GEI | | Invasiveness of Treatment | Non-invasive. Device applies to the surface. | Non-invasive. Device applies to the surface. | | Principles of Operation | Radiofrequency (RF) energy selectively heats a given volume of tissue beneath the surface, while cryogen is delivered to the inside of the Treatment Tip to cool the surface tissue. The Treatment Tip is placed on the surface of the skin and the internal tissues are heated while the surface tissue is protected. (Reverse thermal gradient) | Radiofrequency (RF) energy selectively heats a given volume of tissue beneath the surface, while cryogen is delivered to the inside of the Treatment Tip to cool the surface tissue. The Treatment Tip is placed on the surface of the skin and the internal tissues are heated while the surface tissue is protected. (Reverse thermal gradient) | | Energy | RF | RF | | Contact Quality Monitoring | Yes | No | | Treatment Type | Monopolar | Monopolar | | Main Input<br>(Input<br>voltage/Current/Frequency) | 100 - 120 Vac / 10A / 50/60 Hz<br>220 - 240 Vac / 5A / 50/60 Hz | 100 - 120 Vac / 10A / 50/60 Hz<br>220 - 240 Vac / 5A / 50/60 Hz | | Maximum Power<br>(generator) | 50 Watts | 50 Watts | | Operating Frequency | 6.78 MHz ±15%. | 6.78 MHz ±15%. | | Voltage Waveform | 6.78 MHz continuous sinusoidal waveforms | 6.78 MHz continuous sinusoidal waveforms | | Electrode Probe | Monopolar | Monopolar | | Item | Viveve 2.0 System<br>(Subject Device) | Viveve 2.0 System<br>(Predicate Device K190422) | | Impedance Range | 20 – 185 Ω | 20 – 185 Ω | | Tip | 5cm and 8cm Treatment<br>Tips | 5cm and 8cm Treatment Tips | | Packaging | Tyvek pouch | Tyvek pouch | | Sterility | ETO | ETO | | Cooling Solution | Cryogen | Cryogen | ## Table 5-1: Comparison of Technological Characteristics of Subject Viveve 2.0 System and previously cleared Viveve 2.0 System (K190422) {7}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the word "viveve" in a sans-serif font. The color of the text is a light green. The letters are evenly spaced and the word is horizontally oriented. There is a small circle to the right of the last 'e'. ## 5.9 PERFORMANCE DATA Verification testing, including electrical safety/electromagnetic compatibility, software verification/validation were performed for the addition of the Contact Quality Monitoring software and the revised display module software to the Viveve 2.0 System. - 1. Sparton Medical performed software and hardware verifications to ensure the Contact Quality Monitoring worked appropriately. The following verifications were performed as part of the Sparton Medical Hardware Verification Test Report 140154: - The RF Generator was tested to ensure CQM contact resistance was working ● appropriately. The contact resistance measurement range was 0 to 200 Ω with ±10% (100 Ω - 200 Ω range) and ± 10 Ω accuracy (< 100 Ω). - The RF Generator shall have a CQM circuit response time constant of <20 ● ms.). - The RFG hardware shall support detection and provide means to disable the RF output for the following operating condition error: Loss of patient contact. - 2. Intertek performed the following IEC and EMC testing on the Viveve 2.0 System with CQM added: - Constructional Data Report according to: ● - o Medical Electrical Equipment Part 1: General Requirements For Basic Safety And Essential Performance (R2012) [AAMI ES60601-1:2005 + C1;A2} - Medical Electrical Equipment Part 1: General Requirements for O Basic Safety and Essential Performance (R2013) {CSA C22.2#60601-1:2008 Ed.2 + C2] - Medical Electrical Equipment Part 1-6: General Requirements O For Basic Safety and Essential Performance - Collateral Standard: Usability [IEC 60601-1-1-6:2010 Ed.3+A1] - Medical Electrical Equipment Part 2-2: Particular Requirements O For The Basic Safety and Essential Performance of High Frequency Surgical Equipment and High Frequency Surgical Accessories [IEC 60601-2-2:2017 Ed.6] - EMC Test Report according to: - IEC 60601-1-2 ed. 4.0 (2014-02): Medical Electrical Equipment -O Part 1-2: General Requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - {8}------------------------------------------------ Requirements and tests (With EMC deviations per IEC 60601-2-2:2017:03) - Medical Electrical Equipment Part 1: General Requirements for Basic Safety and Essential Performance according to: - IEC 60601-1:2005, OR1:2006, COR2:2007, AMD1:2012 (or IEC o 60601-1:2012 reprint) - Medical Electrical Equipment Part 1-6: General Requirements for Safety -● Collateral Standard: Usability according to: - o IEC 60601-1-6:2010, AMD1:1013 for use in conjunction with IEC 62366:2007, AMD1:2014 and IEC 60601-1:2005, OR1:2006, COR2:2007, AMD1: 2012 or equivalent consolidated version IEC 60601-1:2012 (Edition 3.1) - Medical Electrical Equipment Part 2: Particular Requirements for the Basic Safety and Essential Performance of High Frequency Surgical Equipment and High Frequency Surgical Accessories according to: - IEC 60601-2-2: 2017 for use in conjunction with IEC 60601- O 1:2005, COR1:2006, COR 2:2007, AMD1:2012 or IEC 60601-1:2012 All the above Intertek test reports are on file at Viveve and are available in full upon request. - 3. Viveve performed a master Viveve 2.0 System Validation to ensure the CQM software performs appropriately. The following parameters were tested: - Ensure the CONTACT MAP graphic within the Graphic User Interface is not ● illuminated when the Return Pad is not connected to the phantom tissue (pork belly) and is illuminated green when the Return Pad is connected to the phantom tissue. - . When the Return Pad is not adequately connected to the phantom tissue, the RF Generator will fault and result in an E63 - Check Tip Contact error. Acknowledge the error by selecting OK. Place the Return Pad securely on the phantom tissue so that the CONTACT MAP is illuminated and ensure a pulse is delivered. The Verifications, EMC and IEC Testing, and Validation ensure that the CQM addition to the Viveve 2.0 System are performing to the correct specifications for safety and efficacy of the device. # 5.10 CONCLUSION The following results were achieved referencing the tests performed in 1-3 above: - 1. The RF Generator performed as expected when the contact resistance was measured at the appropriate values. The CQM reading was as expected between 50 Ω± 10 Ω with the RF Activated. The voltage on channel 1 of the oscilloscope transitioned 63% of the way from low too high in <= 20ms. The RF Generator did fault when the CQM circuit was not complete resulting in radiofrequency interruption to the patient. All of the above parameters were met and were recorded as PASS results. - 2. Intertek performed the above tests according to the applicable standards. All of the above parameters were met and were recorded as PASS results. {9}------------------------------------------------ Image /page/8/Picture/1 description: The image shows the word "viveve" in a sans-serif font. The color of the text is a light green. There is a small circle to the right of the last "e" in the word. - 3. During the execution of the protocol, all tested functions performed as expected. Test results indicated that the CQM updates to the Viveve 2.0 System have met the impacted customer requirements, identified within ENG-017 Impact Analysis for CQM Updates. The Viveve 2.0 System successfully passed all functional testing performed in the protocol. Noting all testing performed above, considering the risk and safety of the device, the Viveve 2.0 System performed as expected after the addition of the CQM Software. The changes to the Viveve 2.0 System Software do not raise questions of safety or efficacy of the overall system; therefore, the Viveve 2.0 System is substantially equivalent to the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...