K193586 · Breas Medical AB · NOU · Jan 27, 2021 · Anesthesiology
Device Facts
Record ID
K193586
Device Name
Vivo 45 LS
Applicant
Breas Medical AB
Product Code
NOU · Anesthesiology
Decision Date
Jan 27, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Vivo 45 LS ventilator (with or without the SpO2 and CO2 sensors) is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. Specifically, the ventilator is applicable for pediatric through adult patients weighing more than 5 kg (11 lbs.), however, the mouthpiece ventilation modes are for adult patients only. The Vivo 45 LS with the SpO2 sensor is intended to measure functional oxygen saturation of arterial hemoglobin (%SpO2) and pulse rate. The Vivo 45 LS with the CO2 sensor is intended to measure CO2 in the inspiratory and expiratory gas. The device is intended to be used in home, institution, hospitals and portable applications such as wheelchairs and gurneys. It may be used for both invasive and non-invasive ventilation. The Vivo 45 LS is not intended to be used as an emergency transport or critical care ventilator.
Device Story
Portable, microprocessor-controlled, turbine-based ventilator; provides pressure support, pressure control, or volume-controlled ventilation. Inputs: ambient air via inlet; patient effort via flow/pressure sensors; optional SpO2/CO2 sensor data. Operation: turbine generates pressure/flow; air delivered via single-limb circuit (active valve or passive leak). Output: ventilatory support; real-time graphs/numbers for pressure, flow, volume on LCD screen. Used in home, hospital, or portable settings (wheelchairs/gurneys) by clinicians or lay caregivers. Integrated user support via 'Information' button; alarm management via dedicated LEDs/buttons. Benefits: provides continuous/intermittent respiratory support; improves patient mobility via portable design/battery power. Healthcare providers use monitored parameters to adjust therapy settings.
Clinical Evidence
Bench testing only. Includes electrical safety (ANSI/AAMI ES60601-1), EMC (IEC 60601-1-2), usability (IEC 60601-1-6), alarm systems (IEC 60601-1-8), and respiratory performance (ISO 80601-2-12, 2-55, 2-61, 2-72). Biocompatibility evaluated per ISO 18562 for VOCs, particulates, CO, CO2, and ozone. Summative usability/human factors testing conducted with respiratory therapists, nurses, and lay caregivers. Cybersecurity and software V&V performed at unit, integration, and system levels.
Technological Characteristics
Turbine-based ventilator; microprocessor-controlled. Materials: gas pathway biocompatible per ISO 18562. Power: AC/DC and integrated/optional click-in battery. Connectivity: data connectivity features. Standards: ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-8, IEC 60601-1-11, ISO 80601-2-12, ISO 80601-2-55, ISO 80601-2-61, ISO 80601-2-72, IEC 62133, ISO 18562-1/2/3. Software: rule-based control algorithms derived from Vivo 60.
Indications for Use
Indicated for pediatric through adult patients weighing >5 kg requiring mechanical ventilation; mouthpiece ventilation modes for adult patients only. Used for invasive and non-invasive support in home, institutional, and portable settings.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 27, 2021
Breas Medical AB % Maureen O'Connell President O'Connell Regulatory Consultants, Inc. 44 Oak Street Stoneham, Massachusetts 02180
Re: K193586
Trade/Device Name: Vivo 45 LS Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: Class II Product Code: NOU. CBK. DOA. CCK Dated: December 23, 2020 Received: December 28, 2020
Dear Maureen O'Connell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Brandon L. Blakely, Ph.D. Acting Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193586
Device Name Vivo 45 LS
#### Indications for Use (Describe)
The Vivo 45 LS ventilator (without the SpO2 and CO2 sensors) is intended to provide continuous or internittent ventilatory support for the care of individuals who require mechanical ventilation. Specifically, the ventilator is applicable for pediatric through adult patients weighing more than 5 kg (1 1bs.), however, the mouthpiece ventilation modes are for adult patients only.
The Vivo 45 LS with the SpO2 sensor is intended to measure function of arterial hemoglobin (% SpO2) and pulse rate.
The Vivo 45 LS with the CO2 sensor is intended to measure CO2 in the inspiratory and expiratory gas.
The device is intended to be used in home, institution, hospitals applications such as wheelchairs and gurneys. It may be used for both invasive and non-invasive ventilation. The Vivo 45 LS is not intended to be used as an emergency transport or critical care ventilator.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) SUMMARY
Prepared in accordance with 21 CFR § 807.92
| Date Summary Prepared: | January 20, 2021 | |
|----------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|---------------------------------------|
| Submitter Information:<br>Company Name:<br>Company Address: | Breas Medical AB<br>Företagsvagen 1<br>SE-453 33 Mölnlycke<br>SWEDEN | |
| Contact Person: | Maureen O'Connell<br>O'Connell Regulatory Consultants, Inc.<br>44 Oak Street<br>Stoneham, MA 02180<br>Telephone: 978-207-1245 | |
| Device Information:<br>Trade Name:<br>Common Name:<br>Classification Name: | Vivo 45 LS<br>Portable Ventilator<br>Continuous Ventilator<br>Product code: NOU<br>21 CFR §868.5895 | |
| | Additional product codes:<br>Continuous Ventilator<br>Product code: CBK<br>21 C.F.R. §868.5895 | |
| | Oximeters<br>Product code: DQA<br>21 C.F.R. §870.2700 | |
| | Carbon Dioxide Gas Analyzer<br>Product code CCK<br>21 C.F.R. §868.1400 | |
| Device Class: | Class II | |
| Predicate Device<br>(Primary): | Device:<br>510(k) Number:<br>Manufacturer: | Vivo 60<br>K160481<br>Breas Medical |
| Predicate Device<br>(Secondary): | Device:<br>510(k) Number:<br>Manufacturer: | Trilogy Evo<br>K181166<br>Respironics |
{4}------------------------------------------------
### Intended Use:
To provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation.
#### Indications for Use:
The Vivo 45 LS ventilator (with or without the SpO2 and CO2 sensors) is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. Specifically, the ventilator is applicable for pediatric through adult patients weighing more than 5 kg (11 lbs.), however, the mouthpiece ventilation modes are for adult patients only.
The Vivo 45 LS with the SpO2 sensor is intended to measure functional oxygen saturation of arterial hemoglobin (%SpO2) and pulse rate.
The Vivo 45 LS with the CO2 sensor is intended to measure CO2 in the inspiratory and expiratory gas.
The device is intended to be used in home, institution, hospitals and portable applications such as wheelchairs and gurneys. It may be used for both invasive and non-invasive ventilation. The Vivo 45 LS is not intended to be used as an emergency transport or critical care ventilator.
### Device Description:
The Vivo 45 LS Ventilator is a portable, microprocessor controlled turbine based pressure support, pressure control or volume controlled ventilator intended for the care of individuals who require mechanical ventilation.
Flow and pressure are read using flow and pressure sensors. Essential parameters such as pressure, flow and volume are presented on the ventilator screen, both in the form as graphs and numbers.
Operator actions are performed via the front panel where the buttons and an LCD screen are located (and two dedicated buttons on the top of the ventilator control starting/stopping treatment and pausing the alarm audio). There are dedicated LEDs and buttons for managing alarm conditions and an Information button which provides integrated user support.
The Vivo 45 LS can be operated by external AC or DC power supply and contains an integrated battery as well as an optional click in battery.
The Vivo 45 LS can be used with two types of patient circuits: single limb patient circuits including an active exhalation valve and single limb patient circuits including a passive leakage port.
## Technological Characteristics Compared to Predicate:
The Vivo 45 LS is a modification to the Vivo 60 ventilator cleared in K160481. The Vivo 45 LS is physically smaller than the Vivo 60 and includes some added features. The Vivo 45 LS user interface design is the same as the Vivo 60, with the same layout of displays, data and menus on the LCD screen, and the same layout of physical buttons for the user to take actions such as starting and stopping treatment, navigating the screen, and selecting parameters. The Vivo 45 LS software originated from the Vivo 60 software, and as such shares the same algorithms and remains very similar. The 45 LS electronics design also originated from the Vivo 60 and remains very similar. The blower inside the Vivo 45 LS is a slightly smaller version of the blower inside the Vivo 60 but has the same mechanical and electrical design.
{5}------------------------------------------------
The Breas Vivo 45 LS has the same intended use and similar technological characteristics to the two predicate devices; the primary predicate device is the Vivo 45 LS (cleared in K160481) and the secondary predicate is the Respironics Trilogy Evo (cleared in K181166). Breas Medical believes that the Vivo 45 LS described in this notification and for use under the conditions of the proposed labeling is substantially equivalent to legally marketed predicate devices that are also Class II medical devices.
The intended use of the Vivo 45 LS it to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation which is the same as the intended use for the Vivo 60. Although the Vivo 45 LS is a modification of the Vivo 60 and has many shared features, there are also technological characteristics which are different. Both the Vivo 45 LS and the Vivo 60 are software controlled devices that receive ambient air through an inlet, utilize a turbine to generate the required pressures and flows, and pass the air to the patient via an outlet. Both devices provide a bleed-in connection for a low pressure, low flow supplemental oxygen supply. The majority of the ventilation modes are identical between the Vivo 45 LS and the Vivo 60 with one additional mode, mouthpiece ventilation, provided in the Vivo 45 LS. Additionally the ventilation parameter settings are largely unchanged as are the alarm settings. The accuracy of controls and accuracy of monitored values are the same between the devices. The user interfaces, power management features, environmental characteristics and data connectivity are all substantially equivalent between the devices. The optional accessories are either cleared in another 510(k) or substantially equivalent to those cleared in the predicate devices.
The Vivo 45 LS includes mouthpiece ventilation which is not included in the Vivo 60 but is a feature of the secondary predicate device, the Trilogy Evo. Both devices have intended use which is to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. The Vivo 45 LS and the Trilogy Evo have the same technological characteristics in regards to mouthpiece ventilation with minor differences in the exact performance specifications.
The Vivo 45 LS is substantially equivalent to the combination of the Vivo 60 and Trilogy Evo both of which are legally marketed predicate devices.
## Performance Testing:
The Vivo 45 LS was subjected to performance testing which verified conformance with all requirements specifications and applicable standards, and which included comparative testing with the Vivo 60 predicate device which supported substantial equivalence. Additional testing was performed to support the substantial equivalence to the secondary predicate device, the Trilogy Evo for the mouthpiece ventilation.
| Performance Testing to Standards | |
|------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Electrical Safety | ANSI/AAMI ES60601-1:2005 Medical electrical equipment-Part<br>1: General requirements for basic safety and essential<br>performance |
| Electromagnetic<br>compatibility | IEC 60601-1-2: 2014 Medical electrical equipment-Part 1-2:<br>General requirements for basic safety and essential performance-<br>Collateral standard: Electromagnetic disturbances-Requirements<br>and tests |
| Usability | IEC 60601-1-6: 2010+A1:2013 Medical electrical equipment-Part |
| | |
| Alarms systems | 1-6: General requirements for basic safety and essential<br>performance-Collateral standard: Usability<br>IEC 60601-1-8: 2006 (Second edition) + Am. 1: 2012 Medical<br>electrical equipment-Part 1-8: General requirements for basic<br>safety and essential performance-Collateral standard: General<br>requirements, tests and guidance for alarm systems in medical<br>electrical equipment and medical electrical systems |
| Medical equipment used in<br>home healthcare<br>environment | IEC 60601-1-11: 2015 Medical electrical equipment-Part 1-11:<br>General requirements for basic safety and essential performance-<br>Collateral standard: Requirements for medical electrical<br>equipment systems used in the home healthcare environment |
| Critical care ventilators | ISO 80601-2-12:2011 Medical electrical equipment Part 2-12:<br>Particular requirements for safety of lung ventilators-Critical care<br>ventilators |
| Respiratory gas monitors | ISO 80601-2-55: 2018 Medical electrical equipment Part 2:<br>Particular requirements for the basic safety and essential<br>performance of respiratory gas monitors |
| Pulse oximeter equipment | ISO 80601-2-61:2017 Medical electrical equipment Part 2-61:<br>Particular requirements for basic safety and essential performance<br>of pulse oximeter equipment |
| Home healthcare<br>environment ventilators for<br>ventilator-dependent<br>patients | ISO 80601-2-72:2015 Medical electrical equipment Part 2-72:<br>Particular requirements for basic safety and essential performance<br>of home healthcare environment ventilators for ventilator-<br>dependent patients |
| Battery testing | IEC 62133: 2012 (2nd Ed) Secondary cells and batteries<br>containing alkaline or other non-acid electrolytes – Safety<br>requirements for portable sealed secondary cells, and for batteries<br>made from them, for use in portable applications |
| | Biocompatibility: The materials which are in the gas pathway have been evaluated via gas<br>emission VOC, inorganic gases (CO, CO2, and Ozone) and PM2.5/PM10 testing with a risk<br>based assessment. The materials were found to be biocompatible for the intended use, intended<br>population and type of patient contact. Further details are provided below which support<br>substantial equivalence. |
| VOC testing | ISO 18562-3: 2017 Biocompatibility evaluation of breathing gas<br>pathways in healthcare applications-Part 3: Tests for emissions of<br>volatile organic compounds. No VOC compounds were observed<br>in quantities that represent a toxicological risk to the intended<br>patient population. |
| Particulates testing | ISO 18562-2: 2017 Biocompatibility evaluation of breathing gas<br>pathways in healthcare applications-Part 2: Tests for emissions of<br>particulate matter. Particulate quantities were well below<br>acceptable limits of exposure for all patient populations. |
| Carbon monoxide testing | The Vivo 45 LS was tested for generation of inorganic gases<br>including carbon monoxide per the recommendations of ISO<br>18562-1. The Vivo 45 LS was found not to generate carbon<br>monoxide. |
| Carbon dioxide testing | The Vivo 45 LS was tested for generation of inorganic gases<br>including carbon dioxide per the recommendations of ISO 18562-<br>1. The Vivo 45 LS was found not to generate carbon dioxide. |
| Ozone testing | The Vivo 45 LS was tested for generation of inorganic gases |
| including ozone per the recommendations of ISO 18562-1. The | |
| Vivo 45 LS was found not to generate ozone. | |
Performance testing included testing to the standards and procedures listed below:
{6}------------------------------------------------
{7}------------------------------------------------
## Performance Testing
Waveform performance testing was conducted comparing the Vivo 45 LS to Vivo 60 (and to the Trilogy EVO for MPV modes). Characteristics tested included flow, pressure and volume waveforms. The comparison of the recorded waveforms supports the claim that Vivo 45 LS is substantially equivalent to the predicate devices.
Triggering testing of Vivo 45 LS was performed which showed that the Vivo 45 LS performed as intended, detecting each patient effort within the permissible trigger delay without false-triggers.
Testing of the Vivo 45 LS was performed to confirm accuracy of controls and monitored values. The testing confirmed that the Vivo 45 LS meets its accuracy specifications.
Testing of the mouthpiece ventilation (MPV) and synchronized intermittent mandatory ventilation (SIMV) modes was performed and the Vivo 45 LS passed all tests.
Alarms testing of the Vivo 45 LS was performed which confirmed proper operation of physiologic and technical alarms.
Power management testing confirmed proper operation of the Vivo 45 LS power management system including transitioning between the different internal and external power sources, power source alarms, and battery operating time.
Treatment and alarm settings testing confirmed the range and operation of settings for all treatment and alarm parameters conform to specifications.
Cybersecurity testing conformance with all cybersecurity specifications.
Software verification and validation were performed at the unit, integration, and system level according to plans and protocols with predetermined pass/fail criteria. All tests passed.
Summative usability / human factors testing was performed including respiratory therapists, registered nurses, and lay caregivers, and the Vivo 45 LS was found to be safe and effective for the intended users, uses and use environments.
Cleaning validation was performed to ensure no physical or performance degradation occurred.
RFID immunity testing was performed to the Association for Automatic Identification and Mobility (AIM) standard 7351731includng 134 kHz and 13.56 MHz RFID sources. The Vivo 45 LS passed all tests.
The testing described confirms that the Vivo 45 LS meets all requirements specifications and complies with the relevant standards, and is therefore substantially equivalent to the predicate devices.
## Conclusion:
The Vivo 45 LS is substantially equivalent to the predicate devices, as the devices share a common intended use and technological characteristics as demonstrated through performance testing.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.