Infinitt RT PACS

K193498 · Infinitt Healthcare Co., Ltd. · LLZ · Feb 13, 2020 · Radiology

Device Facts

Record IDK193498
Device NameInfinitt RT PACS
ApplicantInfinitt Healthcare Co., Ltd.
Product CodeLLZ · Radiology
Decision DateFeb 13, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesSoftware as a Medical Device

Intended Use

INFINITT RT PACS is a software medical device intended to be used for reviewing and assessment of DICOM based datasets which may occur in a radiotherapy environment. It is used by qualified specialist to load and display data generated by different DICOM devices including: RT-STRUCTURE, RT-PLAN, RT-DOSE, RT-IMAGE, CR, CT, MR and PET. The software can connect to a HIS and provides the radiotherapy specialist with the necessary visualization and image manipulation tools to review, present or compare DICOM, RT related datasets including the geometrical super imposition of 3D plan data. INFINITT RT PACS, is intended for reviewing purposes only and is not capable to generate new or modified RT plan data. Users of INFINITT RT PACS, should be trained medical professionals including, radiologists, oncologists, physicians, medical technologists, and dosimetrists. Users should be familiar with the different sources of input data (such as images, structure sets, treatment plans, and calculated dose) as well as how to understand and interpret derived metrics (e.g., dose-volume histograms). Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretation. Mammographic images may only be interpreted using an FDA cleared monitor that offers at least 5 Megapixel resolution and meets other technical specifications reviewed and accepted by FDA. INFINITT RT PACS, is not intended for diagnostic image review on mobile devices.

Device Story

INFINITT RT PACS is a radiotherapy image management software solution for oncology departments. It ingests DICOM-compliant data (CT, RT-PLAN, RT-STRUCTURE, RT-DOSE, DRR) via C-Store from external treatment planning systems and HIS. The software provides visualization tools including multi-planar reformation (axial, coronal, sagittal), 3D volume rendering, and dose distribution analysis (DVH, isodose levels). It is operated by clinicians (oncologists, dosimetrists) in a clinical environment to review, present, and compare radiotherapy datasets. The system facilitates geometric superimposition of 3D plan data and supports image/data transfer to EMR via HL7 or database links. It does not generate or modify RT plans. The output assists clinicians in assessing treatment plans and dose distributions, supporting clinical decision-making in radiotherapy workflows.

Clinical Evidence

Bench testing only. Validation testing confirmed that the software meets predetermined acceptance criteria for all output functions and risk analysis requirements. No clinical data provided.

Technological Characteristics

Software-based PACS; runs on Windows 7 or higher. Supports DICOM 3.1 networking standards. Features include multi-planar reformation, 3D volume rendering, and dose distribution visualization. Complies with ISO 14971 (risk management), IEC 62304 (software lifecycle), and NEMA PS 3.1-3.20 (DICOM).

Indications for Use

Indicated for use by trained medical professionals (radiologists, oncologists, physicians, medical technologists, dosimetrists) for reviewing and assessing DICOM-based radiotherapy datasets (RT-STRUCTURE, RT-PLAN, RT-DOSE, RT-IMAGE, CR, CT, MR, PET). Contraindicated for primary interpretation of lossy compressed mammographic or digitized film screen images; not for diagnostic review on mobile devices.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/10 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services-USA. To the right of the symbol, the text "FDA U.S. FOOD & DRUG ADMINISTRATION" is displayed in blue. The word "FDA" is in a larger, bolder font compared to the rest of the text. INFINITT Healthcare Co., Ltd. % Mr. Carl Alletto Consultant OTech Inc. 8317 Belew Drive MCKINNEY TX 75071 February 13, 2020 ## Re: K193498 Trade/Device Name: INFINITT RT PACS Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: October 31, 2019 Received: December 17, 2019 Dear Mr. Alletto: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for {1}------------------------------------------------ devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. For Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K193498 Device Name Infinitt RT PACS #### Indications for Use (Describe) INFINITT RT PACS is a software medical device intended to be used for reviewing and assessment of DICOM based datasets which may occur in a radiotherapy environment. It is used by qualified specialist to load and display data generated by different DICOM devices including: RT-STRUCTURE, RT-PLAN, RT-DOSE, RT-IMAGE, CR, CT, MR and PET. The software can connect to a HIS and provides the radiotherapy specialist with the necessary visualization and image manipulation tools to review, present or compare DICOM, RT related datasets including the geometrical super imposition of 3D plan data. INFINITT RT PACS, is intended for reviewing purposes only and is not capable to generate new or modified RT plan data. Users of INFINITT RT PACS, should be trained medical professionals including, radiologists, oncologists, physicians, medical technologists, and dosimetrists. Users should be familiar with the different sources of input data (such as images, structure sets, treatment plans, and calculated dose) as well as how to understand and interpret derived metrics (e.g., dose-volume histograms). Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretation. Mammographic images may only be interpreted using an FDA cleared monitor that offers at least 5 Megapixel resolution and meets other technical specifications reviewed and accepted by FDA. INFINITT RT PACS, is not intended for diagnostic image review on mobile devices. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| | | <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | |--|------------------------------------------------------------------------------------------| | | <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image contains the logo for INFINITT Healthcare. The logo consists of a blue circular graphic on the left, followed by the word "INFINITT" in large, bold, blue letters. Below the word "INFINITT" is the word "Healthcare" in smaller, gray letters. The logo is simple and professional, and it is likely used to represent the company's brand. K193498 This 510(k) Summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92. #### I. SUBMITTER Mr. Sang Wook Cho Chief Quality Officer, Research and Development Center INFINITT Healthcare Co. Ltd. 12F Daerung Post Tower III, 27 Digital-ro 34-gil, Guro-gu, Seoul, 08378, South Korea Tel: +82-2-2194-1631 Fax: +82-2-6969-5455 Email: bigmouse@infinitt.com Date Prepared: December 9, 2019 #### II. DEVICE Name of Device: Infinitt RT PACS Common or Usual Name: Picture Archiving Communications System Classification Name: system, image processing, radiological (21 CFR 892.2050) Requlatory Class: II Product Code: LLZ ### III. PREDICATE DEVICES Primary Predicate Device: ProKnow DS (K182855), by ProKnow LLC, CFR 892.2050, Product Code LLZ. ### IV. DEVICE DESCRIPTION Infinitt RT PACS, is a radiation therapy image Management solution dedicated to the oncology department. By using Infinitt RT PACS, the treatment planning system of various vendors can be stored and managed. It also transfers DICOM RT type plan images and information to EMR through HL7or database links. Infinitt RT PACS, provides the function to acquire and view DICOM data generated from RTP (Radiotherapy Planning) System in the form of DICOM C-Store. Supported data includes CT (Computerized Tomography) DICOM image, DICOM RT PLAN data, DRR (Digitally Reconstruct Radiography) image and DICOM RT-Structure, dose information of DICOM RT-Dose and DVH (Dose-Volume Histogram) data. In addition, it supports Axial, Coronal, Sagittal view to check the location of the corresponding Axial CT image and dose distribution to Coronal, Sagittal View through scout line. Infinitt RT PACS, provides DICOM conformant services for both short and long-term image storage and retrieval of digital image data. The application uses DICOM as the interface to the external world and enables interoperability between different devices. Infinitt RT PACS, accepts DICOM association requests for the purpose of storing images and for query and retrieval of images. It also initiates DICOM association requests for the purpose of sending images to an external application entity. #### V. INDICATIONS FOR USE INFINITT RT PACS is a software medical device intended to be used for reviewing and assessment of DICOM based datasets which may occur in a radiotherapy environment. It is used by qualified specialist to load and display data generated by different DICOM devices including: RT-STRUCTURE. RT-PLAN. RT-DOSE. RT-IMAGE. CR. CT. MR and PET. The software can connect to a HIS and provides the radiotherapy specialist with the necessary visualization and image manipulation tools to review, present or compare DICOM, RT related datasets including the geometrical super imposition of 3D plan data. {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for INFINITT Healthcare. The logo features a blue sphere on the left side, with the word "INFINITT" in large, bold, blue letters to the right of the sphere. Below the word "INFINITT" is the word "Healthcare" in a smaller font size. INFINITT RT PACS, is intended for reviewing purposes only and is not capable to generate new or modified RT plan data. Users of INFINITT RT PACS, should be trained medical professionals including, radiologists, oncologists, physicians, medical technologists, and dosimetrists. Users should be familiar with the different sources of input data (such as images, structure sets, treatment plans, and calculated dose) as well as how to understand and interpret derived metrics (e.g., dose-volume histograms). Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretation. Mammographic images may only be interpreted using an FDA cleared monitor that offers at least 5 Mega-pixel resolution and meets other technical specifications reviewed and accepted by FDA. INFINITT RT PACS, is not intended for diagnostic image review on mobile devices. ### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES The subject device and predicate are both PACS, which are indicated for medical image management, review, and data distribution. Both systems have been developed to replace traditional film handling in radiology. The subject device and the predicate device are substantially equivalent in the areas of general function, application, and intended use. Any differences between the predicate and the subject device has no negative impact on the device safety or efficacy and does not raise any new potential or increased safety risks and is equivalent in performance to existing legally marketed devices. | Functionality | Subject Device<br>INFINITT RT PACS | Predicate Device<br>ProKnow DS<br>(K182855) | If different, Impact on Safety and or<br>Efficacy | |---------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Hardware<br>Requirements | OS: Windows 7 or higher<br>Processor: Intel Core i5<br>or higher<br>Memory: 4GM or more<br>Graphic card:<br>1920x1080, 32bpp | OS: Windows 7, 10,<br>and macOS High<br>Sierra<br>Processor: Dual<br>Core2.2+ GHz Intel or<br>equivalent processor<br>Memory: 8GB<br>Graphic card: WebGL<br>compatible graphic<br>card<br>1280x800 effective<br>display resolution<br>(1900x1080<br>recommended) | Yes, there are differences in the computer<br>requirements which are dependent upon<br>software development and implementation.<br>The differences do not raise any new<br>potential safety risks and therefore, we<br>believe there is no impact on safety or<br>efficacy for the subject device. | | Enterprise<br>distribution of<br>images and data<br>via Internet or<br>Intranet | Yes | Yes | No differences | | Networking<br>Communications<br>Protocol - DICOM<br>3.1 | Yes | No | Yes, there are differences. The subject<br>device uses the DICOM 3.1 FDA recognized<br>standard for network communications and<br>device interoperability. The differences do<br>not raise any new potential safety risks and<br>therefore, we believe there is no impact on<br>safety or efficacy for the subject device. | | Functionality | Subject Device<br>INFINITT RT PACS | Predicate Device<br>ProKnow DS<br>(K182855) | If different, Impact on Safety and or<br>Efficacy | | Image<br>Compression | No compression | No compression | No differences | | Supported DICOM<br>Standard SOP<br>Classes | CT Image Storage<br>MR Image Storage<br>PET Image Storage<br>RT Dose Storage<br>RT Structure Set Storage<br>RT Plan Storage<br>RT Ion Plan Storage<br>RT Image Storage<br>Secondary Capture<br>Image Storage | CT Image Storage<br>MR Image Storage<br>RT Dose Storage<br>RT Structure Set<br>Storage<br>RT Plan Storage<br>RT Ion Plan Storage<br>Spatial Registration | Yes, there are differences in the supported<br>DICOM SOP Classes. The subject device<br>complies with the FDA recognized DICOM<br>Standard and has been tested. The product<br>functions are explained in labeling and does<br>not raise any new potential safety risks and<br>therefore, we believe there is no impact on<br>safety or efficacy for the subject device. | | Worklist | Search Filter<br>Search Criteria<br>Examined Patient History<br>Patient List<br>Latest Plan List | Patient List<br>Filter Workspace<br>Filter Patient<br>Patient Collection | Yes, the difference is how the data is<br>retrieved. This does not raise any new<br>potential safety risks and therefore, we<br>believe there is no impact on safety or<br>efficacy for the subject device. | | Image Display<br>Function | Slice Navigation<br>Window/Level<br>Zoom<br>Zoom to Selection<br>Pan<br>Reset | Slice Navigation<br>Window/Level<br>Zoom<br>Zoom to Selection<br>Zoom to Fit<br>Pan<br>Probe | There are differences where the subject<br>device does not have the exact features of<br>the predicate. The subject device has been<br>validated and features and functions are<br>documented in the device labeling. The<br>differences do not raise any new potential<br>safety risks and therefore, we believe there<br>is no impact on safety or efficacy for the subject device. | | Annotation/Measu<br>rement | Yes | No | The subject device supports<br>annotation/measurement. The subject<br>device has been validated and features and<br>functions are documented in the device<br>labeling. The difference does not raise any<br>new potential safety risks and therefore, we<br>believe there is no impact on safety or<br>efficacy for the subject device. | | Multi-Planar<br>Reformation | Axial, Sagittal, Coronal | Axial, Sagittal, Coronal | No differences | | 3D Volume<br>Rendering | Yes | No | The subject device supports 3D Volume<br>Rendering, has been validated and features<br>and functions are documented in the device<br>labeling. The difference does not raise any<br>new potential safety risks and therefore, we<br>believe there is no impact on safety or<br>efficacy for the subject device. | | Image Capture<br>Tool | Yes | No | The subject device supports an Image<br>Capture Too, has been validated, and<br>features and functions are documented in<br>the device labeling. The difference does not<br>raise any new potential safety risks and | | Functionality | Subject Device<br>INFINITT RT PACS | Predicate Device<br>ProKnow DS<br>(K182855) | If different, Impact on Safety and or<br>Efficacy | | Contouring Tool | No | Yes | therefore, we believe there is no impact on<br>safety or efficacy for the subject device.<br>The subject device does not support a<br>Contouring Tool. The subject device has<br>been validated and features and functions<br>are documented in the device labeling. The<br>difference does not raise any new potential<br>safety risks and therefore, we believe there<br>is no impact on safety or efficacy for the<br>subject device. | | Display ROI | Yes | Yes | No differences | | Plan Information | Yes | Yes | No differences | | Beam Information | Yes | Yes | No differences | | Display Dose<br>Distribution | Yes | Yes | No differences | | Isodose Level<br>Setting | Yes | Yes | No differences | | Dose Volume<br>Histograms | Yes | Yes | No differences | | Scorecards | Yes | Yes | No differences | | Plan Summation | Dose Composition based<br>on the image fusion | Dose Composition<br>based on the spatial<br>registration | The difference on how each device handles<br>Plan Summation. The subject device has<br>been validated and the difference does not<br>raise any new potential safety risks and<br>therefore, we believe there is no impact on<br>safety or efficacy for the subject device. | | Plan Comparison | Yes | No | The subject device supports Plan<br>Comparison, has been validated and<br>features and functions are documented in<br>the device labeling. The difference does not<br>raise any new potential safety risks and<br>therefore, we believe there is no impact on<br>safety or efficacy for the subject device. | | Display<br>DRR/Portal/Simul<br>ator Image | Yes | No | The subject device supports<br>DRR/Portal/Simulator Image, has been<br>validated and features and functions are<br>documented in the device labeling. The<br>difference does not raise any new potential<br>safety risks and therefore, we believe there<br>is no impact on safety or efficacy for the<br>subject device. | | HIS Interface | Yes | No | The subject device supports HIS Interface.<br>The subject device has been validated and<br>features and functions are documented in<br>the device labeling. The difference does not<br>raise any new potential safety risks and<br>therefore, we believe there is no impact on<br>safety or efficacy for the subject device. | {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for INFINITT Healthcare. The logo consists of a blue circular emblem on the left, followed by the word "INFINITT" in large, bold, blue letters. Below "INFINITT" is the word "Healthcare" in a smaller, lighter font. {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for INFINITT Healthcare. The logo consists of a blue circular graphic on the left, followed by the word "INFINITT" in large, bold, blue letters. Below the word "INFINITT" is the word "Healthcare" in smaller, lighter-colored letters. {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for Infinit Healthcare. The logo consists of a blue sphere on the left, with the word "INFINITT" in large, bold, blue letters to the right of the sphere. Below the word "INFINITT" is the word "Healthcare" in smaller, lighter blue letters. #### VII. PERFORMANCE DATA #### Nonclinical Testing: The Infinitt RT PACS, has been assessed and tested at the factory and has passed all predetermined testing criteria. The Validation Test Plan was designed to evaluate output functions, and actions performed by Infinitt RT PACS, and followed the process documented in the Validation Test Plan. Nonclinical testing results are provided in the 510(k). Validation testing indicated that as required by the risk analysis, designated individuals performed all verification and validation activities and that the results demonstrated that the predetermined acceptance criteria were met. #### Summary: Based on the performance as documented in the Validation Testing, Infinitt RT PACS, was found to have a safe and effectiveness profile that is similar to the predicate device. The following Standards were used to develop Infinitt RT PACS, and the device has met all the requirements listed in the Standards except for inapplicable requirements: - ISO14971:2007/(R)2010 (Corrected 4 October 2007), Medical devices Applications . of risk management to medical devices, FDA FR Recognition # 5-40. - NEMA PS 3.1 3.20 (2016, Digital Imaging and Communications in Medicine ● (DICOM) Set, FDA FR Recognition # 12-300. - IEC 62304:2006, Medical device software Software life cycle processes, FDA FR ● Recognition # 13-32. - FDA Guidance on Cyber Security: Content of Premarket Submissions for Management of Cybersecurity in Medical Devices Document Issued on: October 2, 2014 - . FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document issued on: May 11, 2005 #### VIII. CONCLUSIONS The 510(k) Pre-Market Notification for Infinitt RT PACS, contains adequate information, data, and nonclinical test results to enable FDA - CDRH to determine substantial equivalence to the predicate device. The subject device and predicate device are substantially equivalent in the areas of technical characteristics, general function, application, and intended use does not raise any new potential safety risks and is equivalent in performance to existing legally marketed devices. Nonclinical tests demonstrate that the device is as safe, as effective, and performs comparably to the predicate devices.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%