← Product Code [FMI](/productcode/FMI) · K193422

# Easydrip Pen Needle, Easydrip Plus Pen Needle (K193422)

_Sandstone Medical (Suzhou), Inc. · FMI · Feb 2, 2021 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K193422

## Device Facts

- **Applicant:** Sandstone Medical (Suzhou), Inc.
- **Product Code:** [FMI](/productcode/FMI.md)
- **Decision Date:** Feb 2, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5570
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

The Pen Needle is intended for use with pen injector device for subcutaneous injection of insulin.

## Device Story

Single-use sterile pen needles for subcutaneous insulin delivery; compatible with various insulin pen injectors. Device consists of 304 stainless steel needle tubing, needle base, needle tip shield, container, and sealing paper. User attaches needle to pen injector manually; performs injection. Differences from predicate include needle tip shield material (polyethylene vs polypropylene) and additional 12.7mm length option. Sterilized via irradiation. Benefits include reliable insulin administration via standard pen systems.

## Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing, including biocompatibility (ISO 10993), sterility, and mechanical performance (ISO 11608-2).

## Technological Characteristics

Materials: 304 stainless steel needle, polyethylene/polypropylene shields, polypropylene base. Sterilization: Irradiation (SAL 10^-6). Compliance: ISO 7864, ISO 9626, ISO 11608-2, ISO 10993. Dimensions: 29G-33G, 4mm-12.7mm lengths. Manual operation.

## Regulatory Identification

A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.

## Predicate Devices

- Verifine® Common Type Insulin Pen Needle ([K161950](/device/K161950.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.

February 2, 2021

Sandstone Medical (Suzhou) Inc. % Elly Xu Consultant Manager Shenzhen Joyantech Consulting Co., Ltd 1713A, 17th Floor, Block A, Zhongguan Times Square, Nanshan District Shenzhen, 518000 Cn

Re: K193422

Trade/Device Name: Easydrip Pen Needle, Easydrip Plus Pen Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Dated: November 27, 2020 Received: December 28, 2020

Dear Elly Xu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Rumi Young Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K193422

Device Name

Easydrip Pen Needle, Easydrip Plus Pen Needle

Indications for Use (Describe)

The Pen Needle is intended for use with pen injector device for subcutaneous injection of insulin.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

# 510(k) Number: K193422

## 1. Contact Details

- 1.1 Applicant information

| Applicant Name          | Sandstone Medical (Suzhou) Inc.                            |  |  |
|-------------------------|------------------------------------------------------------|--|--|
| Address                 | No.168 Pu TuoShan Road, New District, Suzhou 215153, China |  |  |
| Phone No.               | 86-0512-65799368                                           |  |  |
| Fax No.                 | 86-0512-65799368                                           |  |  |
| Contact person          | Juanjuan Sun                                               |  |  |
| Contact person's e-mail | juanjuans@sandstonemed.com                                 |  |  |
| Company e-mail          | juanjuans@sandstonemed.com                                 |  |  |
| Date Prepared           | November 29, 2019                                          |  |  |
| Website                 | www.sandstone.com                                          |  |  |

### 1.2 Consultant information

| Name                    | Shenzhen Joyantech Consulting Co., Ltd                              |
|-------------------------|---------------------------------------------------------------------|
|                         |                                                                     |
| 皇 in 大                  |                                                                     |
| Address                 | 1713A, 17th Floor, Block A, Zhongguan Times Square, Liuxian Avenue, |
|                         | Xili Town, Nanshan District, Shenzhen, Guangdong Province, China.   |
| Phone No.               | +86-755-86069197                                                    |
| Contact person          | Joyce Yang, Field Fu                                                |
| Contact person's e-mail | joyce@cefda.com; field@cefda.com                                    |
| Website                 | http://www.cefda.com                                                |

### 2. Device information

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

| Trade name          | Easydrip Pen Needle, Easydrip Plus Pen Needle |
|---------------------|-----------------------------------------------|
| Common name         | Insulin Pen Needle                            |
| Model               | 29G/30G/31G/32G/33G                           |
| Classification      | II                                            |
| Classification name | Needle, Hypodermic, Single Lumen              |
| Product code        | FMI                                           |
| Regulation No.      | 21 CFR 880.5570                               |

#### Legally Marketed Predicate Device 3.

| Trade Name    | Verifine® Common Type Insulin Pen Needle |
|---------------|------------------------------------------|
| 510(k) Number | K161950                                  |
| Product Code  | FMI                                      |
| Manufacturer  | Promisemed Hangzhou Meditech Co., Ltd.   |

{4}------------------------------------------------

#### 4. Device Description

The proposed devices, Pen Needles, are a single-use device, which is designed for use with a pen injector for the subcutaneous injection of insulin. The proposed devices include two models, i.e. Easydrip Pen Needle and Easydrip Plus Pen Needle. Both models have the same construction, operation methods and similar materials. Both models consist of needle tubing, needle tip shield, needle base, needle container, sealing dialysis paper, glue and silicone oil. The needle base of the device can be screwed onto compatible insulin pens. The only minor differences between the Easydrip and Easydrip Plus models are the outer shape and the material of the Needle tip shield. The Easydrip Needle tip shield is slightly shorter in length and more tapered than the Easydrip Plus Needle tip shield. The material of construction of the Easydrip Needle tip shield is polyethylene, while the Easydrip Plus Needle tip shield is polypropylene.

The Pen Needles are offered in various gauge sizes and needle lengths. The proposed devices are irradiation sterilized to achieve a Sterility Assurance Level (SAL) of 10-9.

| Type          | Gauge(G) | Needle OD.(mm) | Needle available length(mm) |
|---------------|----------|----------------|-----------------------------|
| Easydrip      | 29       | 0.33           | 10                          |
|               | 29       | 0.33           | 12                          |
|               | 29       | 0.33           | 12.7                        |
|               | 30       | 0.30           | 6                           |
|               | 30       | 0.30           | 8                           |
|               | 30       | 0.30           | 10                          |
| 31            | 0.25     | 5              |                             |
| 31            | 0.25     | 6              |                             |
| 31            | 0.25     | 8              |                             |
| 32            | 0.23     | 4              |                             |
| 32            | 0.23     | 5              |                             |
| 33            | 0.20     | 4              |                             |
| 33            | 0.20     | 5              |                             |
| Easydrip Plus | 29       | 0.33           | 10                          |
|               | 29       | 0.33           | 12                          |
|               | 29       | 0.33           | 12.7                        |
|               | 30       | 0.30           | 6                           |
|               | 30       | 0.30           | 8                           |
|               | 30       | 0.30           | 10                          |
| 31            | 0.25     | 5              |                             |
| 31            | 0.25     | 6              |                             |
| 31            | 0.25     | 8              |                             |
| 32            | 0.23     | 4              |                             |
| 32            | 0.23     | 5              |                             |
| 33            | 0.20     | 4              |                             |
| 33            | 0.20     | 5              |                             |

The dimension for both Easydrip Pen Needle and Easydrip Plus Pen Needle are shown as below.

{5}------------------------------------------------

#### Intended Use/Indication for Use 5.

The Pen Needle is intended for use with pen injector device for subcutaneous injection of insulin.

#### 6. Product compatibility

The Pen Needle may be used with Type A compatible insulin delivery system, including:

| Pen name       | Pen Manufacturer      |
|----------------|-----------------------|
| NovoPen 5      | Novo Nordisk          |
| AUTOPEN 24     | Owen mumford          |
| Gansulin Pen   | Tonghua Dongbao       |
| Ypsopen        | Ypsomed AG            |
| UNIPEN         | United Laboratories   |
| XiuLin Pen     | Gan & Lee             |
| HumaPen Ergoll | EILILILLY             |
| Byetta         | Baxter Pharmaceutical |
| Victoza Pen    | Novo nordisk          |

| Item                                                                | Proposed Device:<br>Easydrip Pen Needle                                                                        | Proposed Device:<br>Easydrip Plus Pen<br>Needle                                                                | Predicate Device:<br>Common Type Insulin<br>Pen Needle<br>(K161950)                                                            | Comments         |
|---------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|------------------|
| Product Code                                                        | FMI                                                                                                            | FMI                                                                                                            | FMI                                                                                                                            | Same             |
| Indications for<br>Use                                              | The Pen Needle is<br>intended for use with<br>pen injector device for<br>subcutaneous<br>injection of insulin. | The Pen Needle is<br>intended for use with<br>pen injector device for<br>subcutaneous<br>injection of insulin. | The Insulin Pen<br>Needle is intended for<br>use with pen injector<br>devices for the<br>subcutaneous<br>injection of insulin. | Same             |
| Configuration<br>and Material                                       | Needle tubing (Needle<br>Tube):304 Stainless<br>steel                                                          | Needle tubing (Needle<br>Tube):304 Stainless<br>steel                                                          | Needle tubing (Needle<br>Tube):304 Stainless<br>steel,                                                                         | Same             |
|                                                                     | Needle tip shield<br>(Tube Sheath):<br>Polyethylene                                                            | Needle tip shield<br>(Tube Sheath):<br>Polypropylene                                                           | Needle tip shield<br>(Tube Sheath):<br>Polypropylene                                                                           | Difference<br>#1 |
|                                                                     | Needle base (Hub):<br>Polypropylene                                                                            | Needle base (Hub):<br>Polypropylene                                                                            | Needle hub:<br>Polypropylene                                                                                                   | Same             |
|                                                                     | Needle container (Hub<br>Sheath): Polyethylene                                                                 | Needle container (Hub<br>Sheath): Polyethylene                                                                 | Needle container (Hub<br>Sheath): Polyethylene                                                                                 | Same             |
| Item                                                                | Proposed Device:<br>Easydrip Pen Needle                                                                        | Proposed Device:<br>Easydrip Plus Pen Needle                                                                   | Predicate Device:<br>Common Type Insulin Pen Needle<br>(K161950)                                                               | Comments         |
| Sealing dialysis paper<br>(Sealed Paper):<br>Sealing dialysis paper | Sealing dialysis paper<br>(Sealed Paper):<br>Sealing dialysis paper                                            | Sealing dialysis paper<br>(Sealed Paper):<br>Sealing dialysis paper                                            | Sealing dialysis paper<br>(Sealed Paper):<br>Sealing dialysis paper                                                            | Same             |
| Operation mode                                                      | Manual                                                                                                         | Manual                                                                                                         | Manual                                                                                                                         | Same             |
| Needle Gauge                                                        | 29G/30G/31G/32G/33G                                                                                            | 29G/30G/31G/32G/33G                                                                                            | 29G/30G/31G/32G/33G                                                                                                            | Same             |
| Needle length                                                       | 4mm, 5mm, 6mm,<br>8mm, 10mm,12mm,<br>12.7mm                                                                    | 4mm, 5mm, 6mm,<br>8mm, 10mm,12mm,<br>12.7mm                                                                    | 4mm, 5mm, 6mm,<br>8mm,12mm                                                                                                     | Difference #2    |
| Design of the<br>needle base                                        | Image: Easydrip Pen Needle base                                                                                | Image: Easydrip Plus Pen Needle base                                                                           | Image: Common Type Insulin Pen Needle base                                                                                     | Same             |
| Sterilization                                                       | SAL: 10-6<br>Method: Irradiation<br>Sterilized                                                                 | SAL: 10-6<br>Method: Irradiation<br>Sterilized                                                                 | SAL: 10-6<br>Method: EO Sterilized                                                                                             | Difference #3    |
| Performance                                                         | Complied with ISO<br>7864, ISO 9626, and<br>ISO 11608-2                                                        | Complied with ISO<br>7864, ISO 9626, and<br>ISO 11608-2                                                        | Complied with ISO<br>7864, ISO 9626, and<br>ISO 11608-2                                                                        | Same             |
| Shelf Life                                                          | 5 years                                                                                                        | 5 years                                                                                                        | 5 years                                                                                                                        | Same             |
| Single Use                                                          | Yes                                                                                                            | Yes                                                                                                            | Yes                                                                                                                            | Same             |
| Biocompatibility                                                    | Complied with ISO<br>10993 series<br>standards                                                                 | Complied with ISO<br>10993 series<br>standards                                                                 | Complied with ISO<br>10993 series<br>standards                                                                                 | Same             |
| Cytotoxicity                                                        | No cytotoxicity                                                                                                | No cytotoxicity                                                                                                | No cytotoxicity                                                                                                                | Same             |
| Skin Irritation                                                     | No evidence of skin<br>irritation                                                                              | No evidence of skin<br>irritation                                                                              | No evidence of skin<br>irritation                                                                                              | Same             |
| Skin Sensitization                                                  | No evidence of<br>sensitization                                                                                | No evidence of<br>sensitization                                                                                | No evidence of<br>sensitization                                                                                                | Same             |
| Acute Systemic Toxicity                                             | No systemic toxicity                                                                                           | No systemic toxicity                                                                                           | No systemic toxicity                                                                                                           | Same             |
| Hemolysis                                                           | No evidence of<br>hemolysis                                                                                    | No evidence of<br>hemolysis                                                                                    | No evidence of<br>hemolysis                                                                                                    | Same             |
| Pyrogen                                                             | No pyrogen                                                                                                     | No pyrogen                                                                                                     | No pyrogen                                                                                                                     | Same             |

### 7. Substantial Equivalence Comparison

{6}------------------------------------------------

{7}------------------------------------------------

#### 8. Substantial Equivalence Discussion

### Difference #1 - Needle tip shield material

The needle tip shield of the Easydrip Pen Needle is different from that of the predicate. However, the difference in material is addressed through biocompatibility testing according to ISO 10993 series standards. The biocompatibility testing demonstrates that the difference in material does not affect the safety and effectiveness of the device.

### Difference #2 - Needle length

Compared to the predicate device, the proposed device has an additional 12.7mm needle length. The bench tests of the needle with 12.7mm needle length demonstrated conformance to ISO 7864, ISO 9626, and ISO 11608-2. Therefore, the additional needle length does not affect the safety and effectiveness of the device.

### Difference #3 - Sterilization method

The subject device is irradiation sterilized while the predicate device is Ethylene Oxide sterilized. While the sterilization methods between the proposed device and the predicate device are different, validation testing was performed to validate the sterilization method. The sterilization report showed that the sterilization effect of the proposed device can achieve a Sterility Assurance Level (SAL) of 10°, and the radiation dose audit test reports showed that the minimum gamma radiation dose to achieve a 10° SAL was acceptable according to ISO 11737-2:2012. Therefore, the difference in sterilization does not affect the safety and effectiveness of the device.

#### Non-clinical Testing 9.

All non-clinical testing performed on the subject devices is to demonstrate substantial equivalence to the predicate device. Tests setup and execution are performed in accordance with applicable standards.

| Test                                                                                                              | Requirements                                                                                                                                                                                        | Results                                                                                                                                                                                                 |        |
|-------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Materials                                                                                                         | The needle shall be made of tubing materials<br>specified in ISO 9626.                                                                                                                              | Passed                                                                                                                                                                                                  |        |
| Dimensions                                                                                                        | The needles shall fit the test apparatus<br>specified in item 7.2 of ISO 11608-2.                                                                                                                   | Passed                                                                                                                                                                                                  |        |
| Flow rate through the needle                                                                                      | The needles were tested in accordance with<br>Annex A to ISO 11608-2.                                                                                                                               | Passed                                                                                                                                                                                                  |        |
| Binding force between needle base and<br>needle tubing                                                            | The union of the hub and needle tube shall not<br>break when tested in accordance with Clause 9<br>of ISO 11608-2.                                                                                  | Passed                                                                                                                                                                                                  |        |
| Test                                                                                                              | Requirements                                                                                                                                                                                        | Results                                                                                                                                                                                                 |        |
| Needle tip appearance, needle tubing<br>flawlessness, size of inside and outside<br>diameter and puncturing force | The needle tip appearance shall fulfil the 4.5 of<br>ISO 11608-2.<br>The needle tubing flawlessness shall fulfil the<br>requirements of ISO 7864:1993, 11.3.                                        | Passed                                                                                                                                                                                                  |        |
| Dislocation of measuring point patient<br>end                                                                     | Dislocation of the cannula point at the patient<br>end shall be in accordance with Table 2 (ISO<br>11688-2) when tested in accordance with<br>Clause 8 of ISO 11608-2.                              | Passed                                                                                                                                                                                                  |        |
| Functional compatibility with<br>needle-based injected systems                                                    | Compatibility with any NIS shall be claimed<br>only after testing in accordance with Clause 11<br>of ISO 11608-2.                                                                                   | Passed                                                                                                                                                                                                  |        |
| Ease of assemble and disassembly                                                                                  | Attachment of the needle shall be possible<br>without removing the needle from its opened<br>unit packaging. Compliance is checked<br>according to the requirements of Clause 11 of<br>ISO 11608-2. | Passed                                                                                                                                                                                                  |        |
| Cytotoxicity                                                                                                      | ISO 10093-5 Biological evaluation of medical<br>devices -Part 5: Tests for in vitro cytotoxicity                                                                                                    | Passed                                                                                                                                                                                                  |        |
| Sensitization                                                                                                     | ISO 10993-10 Biological evaluation of medical<br>devices -Part 10:Tests for irritation and skin<br>sensitization                                                                                    | Passed                                                                                                                                                                                                  |        |
| Irritation                                                                                                        | ISO 10993-10 Biological evaluation of medical<br>devices -Part 10:Tests for irritation and skin<br>sensitization                                                                                    | Passed                                                                                                                                                                                                  |        |
| Biocompatibility                                                                                                  | Hemocompatibility                                                                                                                                                                                   | ISO 10993-4 Biological evaluation of medical<br>devices Part 4: Selection of tests for<br>interactions with blood<br>ASTM F765 Standard Practice for Assessment<br>of Hemolytic Properties of Materials | Passed |
|                                                                                                                   | System toxicity<br>(acute)                                                                                                                                                                          | ISO 10993-11:2006/(R)2010, Biological<br>evaluation of medical devices - Part 11: Tests<br>for systemic toxicity.                                                                                       | Passed |
|                                                                                                                   | Pyrogen                                                                                                                                                                                             | USP 40.NF 35 <151> Pyrogen Test.                                                                                                                                                                        | Passed |
|                                                                                                                   | Particulate Matter                                                                                                                                                                                  | USP <788> Particulate Matter in Injections,<br>Method 1: Light Obscuration Particle Count Test                                                                                                          | Passed |

The following performance data was provided to support a substantial equivalence determination.

{8}------------------------------------------------

### 10. Clinical Testing

No clinical test data was included in this submission.

### 11. Conclusions

Based on device comparison information and non-clinical bench testing, the Pen Needle and the predicate device have the same indication for use, similar materials of construction, and similar specifications. The bench tests and biocompatibility tests support that the proposed device is as safe and effective as the predicate device, and the differences between them do not raise any new 

{9}------------------------------------------------

questions of safety and effectiveness. Therefore, the proposed device is determined to be

{10}------------------------------------------------

Substantially Equivalent to legally marketed predicate device.

---

**Source:** [https://fda.innolitics.com/device/K193422](https://fda.innolitics.com/device/K193422)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
