MBT Sepsityper

K193419 · Bruker Daltonik GmbH · QNJ · Dec 27, 2020 · Microbiology

Device Facts

Record IDK193419
Device NameMBT Sepsityper
ApplicantBruker Daltonik GmbH
Product CodeQNJ · Microbiology
Decision DateDec 27, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3378
Device ClassClass 2

Indications for Use

The MBT Sepsityper is a qualitative in vitro diagnostic device consisting of a MBT-CA (Sepsityper) software extension and a reagent kit (MBT Sepsityper Kit US IVD) for use in conjunction with other clinical and laboratory findings to aid in the early diagnosis of bacterial and yeast infections from positively flagged blood cultures using the MALDI Biotyper CA System. The MBT Sepsityper Kit US IVD is a disposable blood culture processing device that includes associated reagents that are intended to concentrate and purify microbial cells from blood culture samples identified as positive by a continuous monitoring blood culture system and confirmed to demonstrate the presence of a single organism as determined by Gram stain. This sample preparation manual method is performed by laboratory health professionals in a clinical diagnostic setting. Subculturing of positive blood cultures is necessary to recover organisms for identification of organisms not identified by the MBT-CA System, for susceptibility testing and for differentiation of mixed growth. Positive MBT Sepsityper results do not rule out co-infection with organisms that may not be detected by the MBT-CA System. Results of the MBT Sepsityper should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Results of the MBT Sepsityper should be correlated with Gram stain results and used in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial and yeast bloodstream infections.

Device Story

MBT Sepsityper is an in vitro diagnostic system for rapid microbial identification from positive blood cultures. Input: positive blood culture samples (confirmed single organism via Gram stain). Process: manual sample preparation using MBT Sepsityper Kit reagents to concentrate/purify microbial cells; subsequent analysis via MALDI Biotyper CA System (matrix-assisted laser desorption/ionization time-of-flight mass spectrometry). Output: microbial identification results provided to laboratory personnel. Context: clinical diagnostic laboratory; used by trained technicians. Clinical utility: aids early diagnosis of bacterial/yeast infections; facilitates faster clinical decision-making compared to traditional subculturing. Benefit: rapid identification of bloodstream pathogens, though subculturing remains necessary for susceptibility testing and mixed growth differentiation.

Clinical Evidence

No clinical data provided in the document. The document focuses on the regulatory clearance of the MBT Sepsityper system for use with the MALDI Biotyper CA System.

Technological Characteristics

System consists of a reagent kit for sample preparation and a software extension (MBT-CA) for the MALDI Biotyper CA System. Employs MALDI-TOF mass spectrometry for microbial identification. Manual sample preparation method. Software-based identification of bacterial and yeast species.

Indications for Use

Indicated for use in clinical laboratory settings to aid in the early diagnosis of bacterial and yeast bloodstream infections in patients with positively flagged blood cultures. Requires confirmation of single organism presence via Gram stain. Not for use as sole diagnostic basis; results must be correlated with clinical findings and Gram stain.

Regulatory Classification

Identification

A clinical mass spectrometry microorganism identification and differentiation system is a qualitative in vitro diagnostic device intended for the identification and differentiation of microorganisms from processed human specimens. The system acquires, processes, and analyzes spectra to generate data specific to a microorganism(s). The device is indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial and fungal infection.

Special Controls

Clinical mass spectrometry microorganism identification and differentiation system must comply with the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) The intended use statement must include a detailed description of what the device detects, the type of results provided to the user, the clinical indications appropriate for test use, and the specific population(s) for which the device is intended, when applicable. (2) Any sample collection device used must be FDA-cleared, -approved, or -classified as 510(k) exempt with an indication for in vitro diagnostic use. (3) The labeling required under § 809.10(b) of this chapter must include: (i) A detailed device description, including all device components, control elements incorporated into the test procedure, instrument requirements, ancillary reagents required but not provided, and a detailed explanation of the methodology and all pre-analytical methods for processing of specimens, and algorithm used to generate a final result. This must include a description of validated inactivation procedure(s) that are confirmed through a viability testing protocol, as applicable. (ii) Performance characteristics for all claimed sample types from clinical studies with clinical specimens that include prospective samples and/or, if appropriate, characterized samples. (iii) Performance characteristics of the device for all claimed sample types based on analytical studies, including limit of detection, inclusivity, reproducibility, interference, cross-reactivity, interfering substances, carryover/cross-contamination, sample stability, and additional studies regarding processed specimen type and intended use claims, as applicable. (iv) A detailed explanation of the interpretation of test results for clinical specimens and acceptance criteria for any quality control testing. (4) The device's labeling must include a prominent hyperlink to the manufacturer's website where the manufacturer must make available their most recent version of the device's labeling required under § 809.10(b) of this chapter, which must reflect any changes in the performance characteristics of the device. FDA must have unrestricted access to this website, or manufacturers must provide this information to FDA through an alternative method that is considered and determined by FDA to be acceptable and appropriate. (5) Design verification and validation must include: (i) Any clinical studies must be performed with samples representative of the intended use population and compare the device performance to results obtained from an FDA-accepted reference method and/or FDA-accepted comparator method, as appropriate. Documentation from the clinical studies must include the clinical study protocol (including predefined statistical analysis plan, if applicable), clinical study report, and results of all statistical analyses. (ii) Performance characteristics for analytical and clinical studies for specific identification processes for the following, as appropriate: (A) Bacteria, (B) Yeasts, (C) Molds, (D) Mycobacteria, (E) Nocardia, (F) Direct sample testing ( *e.g.,* blood culture),(G) Antibiotic resistance markers, and (H) Select agents ( *e.g.,* pathogens of high consequence).(iii) Documentation that the manufacturer's risk mitigation strategy ensures that their device does not prevent any device(s) with which it is indicated for use, including incorporated device(s), from achieving their intended use ( *e.g.,* safety and effectiveness of the functions of the indicated device(s) remain unaffected).(iv) A detailed device description, including the following: (A) Overall device design, including all device components and all control elements incorporated into the testing procedure. (B) Algorithm used to generate a final result from raw data ( *e.g.,* how raw signals are converted into a reported result).(C) A detailed description of device software, including validation activities and outcomes. (D) Acquisition parameters ( *e.g.,* mass range, laser power, laser profile and number of laser shots per profile, raster scan, signal-to-noise threshold) used to generate data specific to a microorganism.(E) Implementation methodology, construction parameters, and quality assurance protocols, including the standard operating protocol for generation of reference entries for the device. (F) For each claimed microorganism characteristic, a minimum of five reference entries for each organism (including the type strain for microorganism identification), or, if there are fewer reference entries, a clinical and/or technical justification, determined by FDA to be acceptable and appropriate, for why five reference entries are not needed. (G) DNA sequence analysis characterizing all type strains and at least 20 percent of the non-type strains of a species detected by the device, or, if there are fewer strain sequences, then a clinical and/or technical justification, determined by FDA to be acceptable and appropriate, must be provided for the reduced number of strains sequenced. (H) As part of the risk management activities, an appropriate end user device training program, which must be offered as an effort to mitigate the risk of failure from user error.

Submission Summary (Full Text)

{1} FDA U.S. FOOD & DRUG ADMINISTRATION # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT ## I Background Information: A 510(k) Number K193419 B Applicant Bruker Daltonik GmbH C Proprietary and Established Names MBT Sepsityper D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | QNJ | Class II | 21 CFR 866.3378 - Clinical Mass Spectrometry Microorganism Identification and Differentiation System | MI | ## II Submission/Device Overview: A Purpose for Submission: 1. Addition of MBT Sepsityper consisting of a MBT-CA (Sepsityper) software extension and a reagent kit (MBT Sepsityper Kit US IVD) 2. Update of MBT-CA to Windows 10 3. Addition of MBT Galaxy System – US IVD 4. Addition of MBT Pilot System – US IVD B Measurand: See Indications for Use (III.B below) Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov {2} 2 C Type of Test: The MBT Sepsityper is an in vitro, qualitative device for species identification of Gram positive and negative bacteria and yeasts in positive blood cultures. The MBT Sepsityper device comprises a sample preparation kit (MBT Sepsityper Kit US IVD) and a MBT-CA (Sepsityper) software extension to be used in conjunction with the MBT-CA automated mass spectrometry system. The MBT Sepsityper uses the same MBT-CA Library as the MBT-CA colony identification. Methods of spotting include MBT Sepsityper “Rapid Sepsityper DT” (RS_DT), “Rapid Sepsityper eDT” (RS_eDT) and “Full Sepsityper” (Ext); analogous to the MBT-CA (DT, eDT and Ext). III Intended Use/Indications for Use: A Intended Use(s): See Indications for Use. B Indication(s) for Use: The MBT Sepsityper is a qualitative in vitro diagnostic device consisting of a MBT-CA (Sepsityper) software extension and a reagent kit (MBT Sepsityper Kit US IVD) for use in conjunction with other clinical and laboratory findings to aid in the early diagnosis of bacterial and yeast infections from positively flagged blood cultures using the MALDI Biotyper CA System. The MBT Sepsityper Kit US IVD is a disposable blood culture processing device that includes associated reagents that are intended to concentrate and purify microbial cells from blood culture samples identified as positive by a continuous monitoring blood culture system and confirmed to demonstrate the presence of a single organism as determined by Gram stain. This sample preparation manual method is performed by laboratory health professionals in a clinical diagnostic setting. Subculturing of positive blood cultures is necessary to recover organisms for identification of organisms not identified by the MBT-CA System, for susceptibility testing and for differentiation of mixed growth. Positive MBT Sepsityper results do not rule out co-infection with organisms that may not be detected by the MBT-CA System. Results of the MBT Sepsityper should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Results of the MBT Sepsityper should be correlated with Gram stain results and used in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial and yeast bloodstream infections. Organisms recovered from positive blood culture bottles that are suitable for identification using the MBT Sepsityper include: Abiotrophia defectiva Achromobacter xylosoxidans Acinetobacter baumannii / nosocomialis group {3} 3 Acinetobacter calcoaceticus Acinetobacter haemolyticus Acinetobacter johnsonii Acinetobacter junii Acinetobacter lwoffii Acinetobacter pittii Acinetobacter radioresistens Acinetobacter ursingii Actinomyces europaeus Actinomyces funkei Actinomyces graevenitzii Actinomyces hyovaginalis Actinomyces meyeri Actinomyces neuii Actinomyces odontolyticus Actinomyces oris Actinomyces radingae Actinomyces turicensis Actinomyces urogenitalis Actinotignum schaalii group Aerococcus sanguinicola Aerococcus urinae Aerococcus viridans Aeromonas hydrophila / caviae group Aeromonas salmonicida Aggregatibacter actinomycetemcomitans Aggregatibacter aphrophilus Aggregatibacter segnis Alcaligenes faecalis Alloiococcus otitis Alloscardovia omnicolens Anaerococcus murdochii Anaerococcus vaginalis Arthrobacter cumminsii Bacteroides caccae Bacteroides fragilis Bacteroides nordii Bacteroides ovatus group Bacteroides pyogenes Bacteroides salyersiae Bacteroides stercoris group Bacteroides thetaiotaomicron group Bacteroides uniformis Bacteroides vulgatus group Bifidobacterium breve Bordetella hinzii Bordetella pertussis / bronchiseptica / parapertussis Brevibacterium casei Brevundimonas diminuta group Burkholderia cepacia complex Burkholderia gladioli Burkholderia multivorans Campylobacter coli Campylobacter jejuni Campylobacter ureolyticus Candida albicans Candida auris Candida boidinii Candida dubliniensis Candida duobushaemulonii Candida famata Candida glabrata Candida guilliermondii Candida haemulonis Candida inconspicua Candida intermedia Candida kefyr Candida krusei Candida lambica Candida lipolytica Candida lusitaniae Candida metapsilosis Candida norvegensis Candida orthopsilosis Candida parapsilosis Candida pararugosa Candida pelliculosa Candida tropicalis Candida valida Candida zeylanoides Capnocytophaga ochracea Capnocytophaga sputigena Chryseobacterium gleum Chryseobacterium indologenes Citrobacter amalonaticus complex Citrobacter freundii complex Citrobacter koseri Clostridium beijerinckii Clostridium bifermentans Clostridium butyricum Clostridium clostridioforme group {4} 4 Clostridium difficile Clostridium innocuum Clostridium paraputrificum Clostridium perfringens Clostridium ramosum Clostridium septicum Clostridium sordellii Clostridium sporogenes / Clostridium botulinum (group I) Clostridium tertium Corynebacterium accolens Corynebacterium afermentans group Corynebacterium amycolatum Corynebacterium aurimucosum group Corynebacterium bovis Corynebacterium coyleae Corynebacterium diphtheriae Corynebacterium frenesi Corynebacterium glucuronolyticum Corynebacterium glutamicum Corynebacterium jeikeium Corynebacterium kroppenstedtii Corynebacterium macginleyi Corynebacterium minutissimum Corynebacterium mucifaciens / ureicelerivorans group Corynebacterium propinquum Corynebacterium pseudodiphtheriticum Corynebacterium pseudotuberculosis Corynebacterium resistens Corynebacterium riegelii Corynebacterium striatum group Corynebacterium tuberculostearicum Corynebacterium ulcerans Corynebacterium urealyticum Corynebacterium xerosis Cronobacter sakazakii group Cryptococcus gattii Cryptococcus neoformans var grubii Cryptococcus neoformans var neoformans Cupriavidus pauculus group Cyberlindnera jadinii Delftia acidovorans group Dermabacter hominis Dermacoccus nishinomiyaensis Edwardsiella tarda Eikenella corrodens Elizabethkingia meningoseptica group Enterobacter aerogenes Enterobacter amnigenus Enterobacter cloacae complex Enterococcus avium Enterococcus casseliflavus Enterococcus durans Enterococcus faecalis Enterococcus faecium Enterococcus gallinarum Enterococcus hirae Enterococcus mundtii Enterococcus raffinosus Escherichia coli Escherichia hermannii Escherichia vulneris Ewingella americana Facklamia hominis Finegoldia magna Fluoribacter bozemanae Fusobacterium canifelinum Fusobacterium necrophorum Fusobacterium nucleatum Gardnerella vaginalis Gemella haemolysans Gemella morbillorum Gemella sanguinis Geotrichum candidum Geotrichum capitatum Granulicatella adiacens Haemophilus haemolyticus Haemophilus influenzae Haemophilus parahaemolyticus group Haemophilus parainfluenzae Hafnia alvei Helcococcus kunzii Kingella denitrificans Kingella kingae Klebsiella oxytoca / Raoultella ornithinolytica Klebsiella pneumoniae Klebsiella variicola Kloeckera apiculata Kocuria kristinae Kytococcus sedentarius {5} 5 Lactobacillus gasseri Lactobacillus jensenii Lactobacillus rhamnosus Lactococcus garvieae Lactococcus lactis Leclercia adecarboxylata Legionella longbeachae Legionella pneumophila Leuconostoc citreum Leuconostoc mesenteroides Leuconostoc pseudomesenteroides Listeria monocytogenes Macrococcus caseolyticus Malassezia furfur Malassezia pachydermatis Mannheimia haemolytica group Micrococcus luteus Micrococcus lylae Mobiluncus curtisii Moraxella sg Branhamella catarrhalis Moraxella sg Moraxella nonliquefaciens Moraxella sg Moraxella osloensis Morganella morganii Myroides odoratimimus Myroides odoratus Neisseria bacilliformis Neisseria cinerea Neisseria elongata Neisseria flavescens / subflava group Neisseria gonorrhoeae Neisseria lactamica Neisseria meningitidis Neisseria sicca group Neisseria weaveri Nocardia brasiliensis Nocardia cyriacigeorgica Nocardia farcinica group Nocardia nova Nocardia otitidiscaviarum Ochrobactrum anthropi Oligella ureolytica Oligella urethralis Pantoea agglomerans Parabacteroides distasonis Parabacteroides goldsteinii Parabacteroides johnsonii / merdae group Parvimonas micra Pasteurella multocida Pediococcus acidilactici Pediococcus pentosaceus Peptoniphilus harei group Peptostreptococcus anaerobius Pichia ohmeri Plesiomonas shigelloides Pluralibacter gergoviae Porphyromonas gingivalis Porphyromonas somerae Prevotella bivia Prevotella buccae Prevotella denticola Prevotella intermedia Prevotella melaninogenica Propionibacterium acnes Proteus mirabilis Proteus vulgaris group Providencia rettgeri Providencia stuartii Pseudomonas aeruginosa Pseudomonas fluorescens group Pseudomonas oryzihabitans Pseudomonas putida group Pseudomonas stutzeri Ralstonia pickettii Rhizobium radiobacter Rhodotorula mucilaginosa Rothia aeria Rothia dentocariosa Rothia mucilaginosa Saccharomyces cerevisiae Salmonella sp Serratia fonticola Serratia liquefaciens Serratia marcescens Serratia odorifera Serratia plymuthica Serratia rubidaea Sphingobacterium multivorum Sphingobacterium spiritivorum Sphingomonas paucimobilis group Staphylococcus aureus Staphylococcus auricularis Staphylococcus capitis {6} 6 Staphylococcus caprae Staphylococcus carnosus Staphylococcus cohnii Staphylococcus delphini Staphylococcus epidermidis Staphylococcus equorum Staphylococcuss felis Staphylococcus haemolyticus Staphylococcus hominis Staphylococcus intermedius Staphylococcus lentus Staphylococcus lugdunensis Staphylococcus pasteuri Staphylococcus pettenkoferi Staphylococcus pseudintermedius Staphylococcus saccharolyticus Staphylococcus saprophyticus Staphylococcus schleiferi Staphylococcus sciuri Staphylococcus simulans Staphylococcus vitulinus Staphylococcus warneri Staphylococcus xylosus Stenotrophomonas maltophilia Streptococcus agalactiae Streptococcus anginosus Streptococcus canis Streptococcus constellatus Streptococcus dysgalactiae Streptococcus equi Streptococcus gallolyticus Streptococcus gordonii Streptococcus intermedius Streptococcus lutetiensis Streptococcus mitis / oralis group Streptococcus mutans Streptococcus parasanguinis Streptococcus pneumoniae Streptococcus pyogenes Streptococcus salivarius / vestibularis group Streptococcus sanguinis Streptococcus sobrinus Streptococcus thermophilus Sutterella wadsworthensis Trichosporon asahii Trichosporon inkin Trichosporon mucoides group Trueperella bernardiae Turicella otitidis Vagococcus fluvialis Veillonella parvula group Vibrio parahaemolyticus Vibrio vulnificus Weeksella virosa Yersinia enterocolitica Yersinia frederiksenii Yersinia intermedia Yersinia kristensenii Yersinia pseudotuberculosis ## C Special Conditions for Use Statement(s): Rx - For Prescription Use Only ## D Special Instrument Requirements: MBT-CA System including: - MALDI Biotyper CA System - MALDI Biotyper smart CA System - MALDI Biotyper sirius CA System Optional Instrument Requirements: - MBT Galaxy System - US IVD - MBT Pilot System - US IVD ## IV Device/System Characteristics: {7} 7 # A Device Description: The MBT Sepsityper is an in vitro, qualitative device for species identification of Gram positive and negative bacteria as well as yeasts by matrix-assisted laser desorption/ionization - time-of-flight mass spectrometry (MALDI-TOF MS) from positive blood cultures. The MBT Sepsityper device comprises a sample preparation kit (MBT Sepsityper Kit US IVD) and a MBT-CA (Sepsityper) software extension that is used with the MBT-CA automated mass spectrometry system (DEN170081). Methods of spotting include Rapid Sepsityper DT (RS_DT), Rapid Sepsityper eDT (RS_eDT) and Full Sepsityper (Ext); analogous to the MBT-CA spotting methods (DT, eDT and Ext). The device differs from the predicate device in the use of the MBT Sepsityper Kit US IVD which contains lysis and wash buffers for the lysis of blood cells and enrichment of microorganisms from positive blood culture bottles and in the use of the MBT-CA (Sepsityper) software extension. The MBT-CA (Sepsityper) software extension is considered an extension of the claimed MBT-CA System software. The algorithm remains the same except for the mass range. Due to noise of red blood cells between 3,000-4,000 m/z, the peak picking mass range is limited to 4,000-15,000 m/z instead from 3,000-15,000 m/z for isolated colonies. The MBT Sepsityper performs a standard Main Spectrum reference Pattern (MSP) match of the peak list from the unidentified spectrum against the MBT-CA MSP Reference Library. The MSP Library is the same for the MBT-CA and MBT Sepsityper; however, a different confidence score range is used to denote high confidence (1.8 to 3), low confidence (1.6 to 1.79), and no identification (0 to 1.59). The MBT Sepsityper is for use with BACTEC (BD), BacT/ALERT without charcoal (bioMérieux), and VersaTREK (Thermo Scientific) positive blood culture bottles. The MBT Sepsityper also provides mixed culture hints (See IV.A.4). # MBT Sepsityper Device The MBT Sepsityper is comprised of: - MBT-CA (Sepsityper) software extension - MBT Sepsityper Kit US IVD Required Materials Supplied by Bruker: - A Bruker MALDI-TOF Mass Spectrometer instrument connected to a computer with dedicated MBT-CA software installed: - Bruker microflex LT/SH (benchtop) - Bruker microflex LT/SH smart (benchtop) - Bruker MBT sirius (benchtop) - US IVD 48 Spot Target [P/N: 8604532] - MBT Biotarget 96 US IVD [P/N: 1840380] - MSP Biotarget Adapter [P/N: 8267615] - US IVD BTS (Bacterial Test Standard) [P/N: 8604530] - US IVD HCCA portioned [P/N: 8604531] Required Materials Not Supplied by Bruker: - Standard solvent - Acetonitrile {8} - HPLC-grade water - Formic Acid (FA) - Absolute Ethanol (EtOH) - Trifluoroacetic acid (TFA) - Other common laboratory supplies ## B Principle of Operation: **MBT Sepsityper Kit US IVD Procedures:** Note: US IVD HCCA solution must be added within 30 minutes after sample spots have dried, otherwise the spotting must be performed again. 1. Harvesting blood culture fluid a. Disinfect the septum of the blood culture bottle with 70% ethanol or similar. b. Collect sufficient blood culture fluid. Preparing MBT Sepsityper samples using the MBT Sepsityper Kit US IVD: c. Transfer 1 mL blood culture fluid to a microcentrifuge tube. d. Add 200 µL Lysis Buffer and mix by vortexing (full speed) for 10 (± 5) seconds. e. Centrifuge the tube for 2 minutes at 13,000-15,000 rpm at room temperature. f. Remove the supernatant by pipetting and discard. g. Add 1 mL Washing Buffer and resuspend the pellet by pipetting up and down. h. Centrifuge the tube for 1 minute at 13,000-15,000 rpm at room temperature. i. Remove the supernatant from the pellet (in the following 'Sepsityper Pellet') by pipetting and discard. 2. Performing the Rapid Sepsityper Workflow using Direct Transfer (RS_DT) and extended Direct Transfer (RS_eDT) sample preparation procedures a. Using a transfer device such as a toothpick, transfer some Sepsityper Pellet material onto two (2) MALDI target plate positions. b. Overlay the 2nd spot with 1 µL 70% formic acid and dry at room temperature (RS-eDT). c. For validation of the IVD system, inoculate US IVD BTS onto two target plate positions following Instructions for Use of US IVD BTS. d. Overlay each sample spot and US IVD BTS spots with 1 µL US IVD HCCA solution. **Note:** If one of the two Rapid Sepsityper samples (RS_DT or RS_eDT) show high confidence results the Sepsityper procedure ends up here. If low confidence results or no ID results were observed continue with Full Sepsityper. 3. Performing the Full Sepsityper Workflow using extraction (Ext) sample preparation procedures a. Add 300 µL HPLC-grade water and resuspend the Sepsityper Pellet by pipetting up and down. b. Add 900 µL ethanol and mix the suspension with a vortex mixer (full speed) for 10 (± 5) seconds. c. Centrifuge the tube for 2 minutes at 13,000-15,000 rpm at room temperature. d. Remove the supernatant by pipetting and discard. e. Centrifuge the tube for 2 minutes at 13,000-15,000 rpm at room temperature. f. Remove the residual ethanol by pipetting and discard. {9} g. Allow the pellet to dry for 5 (±1) minutes at room temperature. h. Add 2-50 µL (depending on the pellet size) 70% formic acid and thoroughly resuspend the pellet by pipetting up and down. i. Add an equal volume of acetonitrile and mix the suspension by pipetting up and down two to three times. j. Centrifuge the tube for 2 minutes at 13,000-15,000 rpm at room temperature. k. Pipette 1 µL of the supernatant onto an unoccupied MALDI target plate position and dry at room temperature. l. For validation of the IVD system, inoculate US IVD BTS onto two target plate positions following Instructions for Use of US IVD BTS. m. Overlay each sample spot and US IVD BTS spots with 1 µL US IVD HCCA solution. C Instrument Description Information: 1. Instrument Name: MBT Sepsityper 2. Specimen Identification: The user manually enters the specimen identification information into the MALDI Biotyper CA System. The user first defines active sample positions and US IVD BTS control positions. The defined sample positions are required to have a user entered and valid sample identifier in the Id column along with an optional description. All inoculated positions must contain a valid ID before the run can proceed. After all sample positions have been input, at least two US IVD BTS control positions must be defined. It is recommended that the cross-joint positions on the US IVD 48 Spot Target are used as quality control positions. Alternatively, the user can use an optional Honeywell (Hyperion 1300g) Barcode Reader that is connected via a USB cable to the MALDI Biotyper CA System computer. The barcode reader scans the unique ten-digit target ID which appears in the Target ID box on the target plate. After the target ID has been entered, a new Run page opens, and the ten-digit target ID appears as the Plate Id and is appended to the Run name. Sample identifications are entered into the computer corresponding to the target plate position for that run. The user may also use the optional MBT Pilot System - US IVD. The MBT Pilot System - US IVD is a tool providing visual support for the manual preparation of biological samples on MALDI target plates in the MALDI Biotyper CA System workflow. The MBT Pilot System - US IVD indicates the correct MALDI target plate positions by means of projected cross hairs by providing guided MALDI target plate preparation, using patented micro-projection technology. Traceability is achieved with a barcode linking sample projects to specific MALDI target plates (i.e., US IVD 48 Spot Target (Bruker #8604532) and MBT Biotarget 96 US IVD (Bruker #1840380). A USB connection is utilized for data exchange. Usability studies were conducted and found acceptable. 3. Specimen Sampling and Handling: In addition to See IV.C.2 above, the MBT Galaxy System - US IVD may be utilized. {10} The MBT Galaxy System - US IVD device, as part of the MBT-CA System deposits matrix droplets (in DT, Ext workflows) or FA+matrix droplets (in eDT workflows) onto sample spot positions on MALDI targets. The MBT Galaxy System - US IVD reads the target barcode and loads the associated worklist from the MBT-CA server. Validation for the MBT Galaxy System - US IVD included droplet positioning test, carry over - cross-contamination studies, repeatability and reproducibility, sample stability of automatically overlaid with HCCA matrix solution or FA/HCCA matrix solution on test organism set at two temperatures [21(±2)°C (lower range) and at 25(±2)°C (upper range)], equivalence (manual vs automated sample prep) and spectra re-evaluation. All results were found acceptable and equivalent to manual methods. No carry over - cross-contamination was observed. 4. Calibration: Calibration is achieved using the US IVD Bacterial Test Standard (BTS), an in-vitro-diagnostic product used for quality control and validation. US IVD BTS contains a manufactured extract of Escherichia coli DH5alpha that demonstrates a characteristic peptide and protein profile mass spectrum, when tested on the MALDI Biotyper CA System. US IVD BTS is spiked with two additional proteins that extend the upper boundary of the mass range of the US IVD BTS. The overall mass range covered by US IVD BTS is 3.6 to 17 kDa. If US IVD BTS does not meet all required performance specifications, the test run will be invalid. 5. Quality Control: The user will follow local, state and/or federal regulations for Quality Control requirements. V Substantial Equivalence Information: A Predicate Device Name(s): MALDI Biotyper CA System B Predicate 510(k) Number(s): DEN170081 {11} C Comparison with Predicate(s): | Similarities | | | | --- | --- | --- | | Characteristic | NEW DEVICE MBT-CA System with MBT Sepsityper (K193419) | PRIMARY PREDICATE DEVICE MBT-CA System (DEN170081) | | Product Codes | QNJ | QBN | | Indications for Use | The MBT Sepsityper is a qualitative in vitro diagnostic device consisting of a MBT-CA (Sepsityper) software extension and a reagent kit (MBT Sepsityper Kit US IVD) for use in conjunction with other clinical and laboratory findings to aid in the early diagnosis of bacterial and yeast infections from positively flagged blood cultures using the MALDI Biotyper CA System.The MBT Sepsityper Kit US IVD is a disposable blood culture processing device that includes associated reagents that are intended to concentrate and purify microbial cells from blood culture samples identified as positive by a continuous monitoring blood culture system and confirmed to demonstrate the presence of a single organism as determined by Gram stain. This sample preparation manual method is performed by laboratory health professionals in a clinical diagnostic setting.Subculturing of positive blood cultures is necessary to recover organisms for identification of organisms not identified by the MBT-CA System, for susceptibility testing and for differentiation of mixed growth.Positive MBT Sepsityper results do not rule out co-infection with organisms that may not be detected by the MBT-CA System. Results of the MBT Sepsityper should not be used as the sole basis for diagnosis, treatment, or | The MALDI Biotyper CA System is a mass spectrometer system using matrix-assisted laser desorption/ionization - time-of-flight (MALDI-TOF) for the identification and differentiation of microorganisms cultured from human specimens.The MALDI Biotyper CA System is a qualitative in vitro diagnostic device indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial and fungal infections.For measurand, see DEN170081 | {12} | | other patient management decisions. Results of the MBT Sepsityper should be correlated with Gram stain results and used in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial and yeast bloodstream infections. Organisms recovered from positive blood culture bottles that are suitable for identification using the MBT Sepsityper include (see III.B above) | | | --- | --- | --- | | Type of Test | Automated Mass Spectrometry System | Automated Mass Spectrometry System | | Matrix | α-Cyano-4-hydroxycinnamic acid | α-Cyano-4-hydroxycinnamic acid | | Matching Algorithm | Calculates matches by comparing a new spectrum against each single reference entry of a reference database. | Calculates matches by comparing a new spectrum against each single reference entry of a reference database. | | Instrument Control Software | flexControl and AutoX | flexControl and AutoX | | Calibration and Quality Control | Bruker US IVD Bacterial Test Standard (BTS) | Bruker US IVD Bacterial Test Standard (BTS) | | MALDI Matrix | US IVD HCCA portioned | US IVD HCCA portioned | | Library | MALDI Biotyper for Clinical Applications (MBT-CA) | MALDI Biotyper for Clinical Applications (MBT-CA) | | MALDI Target Plate | US IVD 48 Spot Target (48 positions reusable steel targets) | US IVD 48 Spot Target (48 positions reusable steel targets) | | | MBT Biotarget 96 US IVD (96 positions disposable targets) | MBT Biotarget 96 US IVD (96 positions disposable targets) | | MALDI-TOF MS Instruments | Bruker microflex LT/SH (benchtop) Bruker microflex LT/SH smart (benchtop) Bruker MBT sirius (benchtop) | Bruker microflex LT/SH (benchtop) Bruker microflex LT/SH smart (benchtop) Bruker MBT sirius (benchtop) | {13} | Differences | | | | --- | --- | --- | | Characteristic | NEW DEVICE MBT-CA System with MBT Sepsityper (K193419) | PRIMARY PREDICATE DEVICE MBT-CA System (DEN170081) | | Test Sample | **Bottles of the BD BACTEC (Becton Dickinson) System:** • BD BACTEC™ Standard - Aerobic / Anaerobic • BD BACTEC™ PLUS - Aerobic / Anaerobic • BD BACTEC™ - Lytic Anaerobic • BD BACTEC™ - BD Peds Plus™ • BD BACTEC™ - Myco/F Lytic • BD BACTEC™ - Mycosis IC **Bottles without charcoal of the BacT/ALERT® (bioMérieux) System:** • BacT/ALERT® SA Standard Aerobic • BacT/ALERT® SN Standard Anaerobic • BacT/ALERT® FA Plus • BacT/ALERT® FN Plus • BacT/ALERT® PF Plus **Bottles of the VersaTREK® (Thermo Scientific) System:** • VersaTREK® REDOX 1 • VersaTREK® REDOX 2 | Isolated colony from any patient sample source. Acceptable media: • Columbia blood agar with 5% sheep blood (Gram negative bacteria) • Trypticase soy agar with 5% sheep blood (Gram negative bacteria, Gram positive bacteria, yeasts) • Chocolate agar (Gram negative bacteria, Gram positive bacteria) • MacConkey Agar (Gram negative bacteria) • Columbia CNA agar with 5% sheep blood (Gram positive bacteria) • Brucella Agar with 5% horse blood (Gram negative anaerobic bacteria, Gram positive anaerobic bacteria) • CDC anaerobe agar with 5% sheep blood (Gram negative anaerobic bacteria, Gram positive anaerobic bacteria) • CDC anaerobe 5% sheep blood Agar with phenylethyl alcohol (Gram negative anaerobic bacteria, Gram positive anaerobic bacteria) • CDC anaerobe laked sheep blood Agar with kanamycin and vancomycin (Gram negative anaerobic bacteria) • Bacteroides bile esculin Agar with amikacin (Bacteroides species) • Clostridium difficile Agar with 7% sheep blood (Clostridium difficile) • Sabouraud-Dextrose Agar (Yeasts) • Brain Heart Infusion Agar (Yeasts) • Campylobacter Agar with 5 Antimicrobics and 10% Sheep Blood (Campylobacter species) • Bordet Gengou Agar with 15% sheep blood (Bordetella species) • Buffered Charcoal Yeast Extract Agar (Legionella species) | {14} 14 | | | • Buffered Charcoal Yeast Extract Selective Agar with polymyxin, anisomycin and vancomycin (Nocardia species) • Modified Thayer-Martin Agar (Neisseria species) | | --- | --- | --- | | System Software | MBT-CA Software including the MBT-CA (Sepsityper) software extension | MBT-CA Software | | Sample Type | Positive blood cultures | Isolated colonies | | Sample Preparation Reagents | Lysis Buffer / Washing Buffer (MBT Sepsityper Kit US IVD) | No | | Workflow | The approved MBT-CA Workflow plus the MALDI Sepsityper Workflows [Rapid Sepsityper DT (RS_DT), Rapid Sepsityper eDT (RS_eDT) and Full Sepsityper (Ext)] | MBT-CA Workflow (DT, eDT, Ext) | | Analyzed Mass Range | 4,000 – 15,000 m/z (Sepsityper samples) | 3,000 – 15,000 m/z (Isolated colonies) | | log(score) Values | MBT Sepsityper Samples: • High Confidence ID: 1.80 - 3.00; • Low Confidence ID: 1.60 - 1.79; • No Organism ID Possible: 0.00 - 1.59 | Isolated colony from agar plate: • High Confidence ID: 2.00 - 3.00; • Low Confidence ID: 1.70 - 1.99; • No Organism ID Possible: 0.0 - 1.69 | | Mix Culture Hint | Yes | No | ## VI Standards/Guidance Documents Referenced: CLSI documents EP07-Ed3 (Interference Testing in Clinical Chemistry) ## VII Performance Characteristics (if/when applicable): ### A Analytical Performance: 1. Precision: Repeatability and Reproducibility: At each of the three sites (2 US; 1 Bruker) ten different samples were tested on five different days with two operators. Each site had two (2) operators test a panel of ten spiked samples (4x Gram positive, 4x Gram negative, 2x yeast) for five non-consecutive days representing two instrument runs a day. Each were tested in triplicate. No incorrect identifications were observed with any method sample preparation procedures (i.e., RS_DT, RS_eDT or Ext). A summary of combination of DT + eDT + Ext workflow is shown in Table 1 below. {15} Table 1: Repeatability and Reproducibility of the MBT Sepsityper | Site 1 | | | | | | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Sepsityper Result | Operator 1 | | | | | Operator 2 | | | | | Op 1 | Op 2 | | | d1 | d2 | d3 | d4 | d5 | d1 | d2 | d3 | d4 | d5 | all days | | | high confidence ID | 100% | 100% | 100% | 100% | 100% | 100% | 90% | 100% | 100% | 90% | 100% | 96% | | low confidence ID | 0% | 0% | 0% | 0% | 0% | 0% | 0% | 0% | 0% | 10% | 0% | 2% | | no ID | 0% | 0% | 0% | 0% | 0% | 0% | 10% | 0% | 0% | 0% | 0% | 2% | | Site 2 | | | | | | | | | | | | | | Sepsityper Result | Operator 1 | | | | | Operator 2 | | | | | Op 1 | Op 2 | | | d1 | d2 | d3 | d4 | d5 | d1 | d2 | d3 | d4 | d5 | all days | | | high confidence ID | 100% | 100% | 90% | 100% | 100% | 90% | 90% | 100% | 90% | 100% | 98% | 94% | | low confidence ID | 0% | 0% | 0% | 0% | 0% | 0% | 0% | 0% | 10% | 0% | 0% | 2% | | no ID | 0% | 0% | 10% | 0% | 0% | 10% | 10% | 0% | 0% | 0% | 2% | 4% | | Site 3 | | | | | | | | | | | | | | Sepsityper Result | Operator 1 | | | | | Operator 2 | | | | | Op 1 | Op 2 | | | d1 | d2 | d3 | d4 | d5 | d1 | d2 | d3 | d4 | d5 | all days | | | high confidence ID | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% | | low confidence ID | 0% | 0% | 0% | 0% | 0% | 0% | 0% | 0% | 0% | 0% | 0% | 0% | | no ID | 0% | 0% | 0% | 0% | 0% | 0% | 0% | 0% | 0% | 0% | 0% | 0% | # 2. Sample Stability: Studies were performed to determine sample stability in various conditions: - Organism stability in blood culture prior Sepsityper processing: Positive blood culture bottles can be stored at room temperature up to 24 hours prior to Sepsityper processing. - Organism stability in extended incubation blood culture prior Sepsityper processing: Blood cultures which are flagged positive by the continuously monitoring blood culture instrument can remain on the instrument up to 12 hours prior to Sepsityper processing. - Sample Stability of the Sepsityper Pellet: The concentrated Sepsityper Pellet can be maintained at room temperature for 1 hour prior to spotting the target plate. - Sample Stability post matrix application: If HCCA matrix was applied to the ready prepared MBT Biotargets 96 or Steel Targets, the targets can be kept at room temperature and the measurement and identification can be performed within $24\mathrm{h}$ without reducing identification performance. # 3. Analytical Specificity/Interference: To assess the inhibitory effects of substances encountered in blood and blood culture media the substances listed in Table 2 were tested in the given concentration in blood cultures spiked with each one of the following species: Klebsiella pneumoniae, Staphylococcus aureus and Candida albicans. Each interferent - species combination was prepared with the three Sepsityper workflows (RS_DT, RS_eDT, Ext) and MALDI target preparation was performed in quadruplicates. No interference was found. {16} Table 2: Substances and concentrations tested for interference in positive blood cultures | Routine Interferents | | Drug Interferents | | | --- | --- | --- | --- | | Interferent | Test Concentration | Interferent | Test Concentration | | Triglyceride-rich lipoproteins | 10 g/L | Acetaminophen | 1.324 mmol/L | | Hemolysate (hemoglobin) | 5 g/L | Acetylcysteine | 10.2 mmol/L | | Protein | 120 g/L | Acetylsalicylic Acid | 3.62 mmol/L | | Conjugated Bilirubin | 200 mg/L | Cefoxitin (Na) | 1.55 mmol/L | | Unconjugated Bilirubin | 200 mg/L | Cyclosporine | 5 mg/L | | Further Interferents | Doxycycline (HCl) | 0.0675 mmol/L | | | | | Heparin | 3,000 U/L | | White blood cells | Five-fold increase (compared to usual concentration) | Ibuprofen | 2.425 mmol/L | | | | Metronidazole | 0.7 mmol/L | | Sodium Polyanethole Sulfonate | 0.5 mg/mL | Vancomycin | 0.069 mmol/L | # 4. Interference (Polymicrobial Samples) During the clinical studies polymicrobial samples were excluded via study design (Gram staining). However, polymicrobial samples may not always be detected by Gram stain and may be subsequently processed with the MBT Sepsityper. The mixed culture hint is a functionality of the MBT-CA (Sepsityper) software extension which is able to create a user warning about a possible polymicrobial infection. The mixed culture warning is based on the ranking list of an identification. First, the algorithm scans the ranking list of a MBT Sepsityper identification until a defined log(score) is reached ("parameter #1"). The scanned ranking list is longer than the typical 10 positions. Within this ranking list ideally only one single species appears until the log(score) threshold is reached, however, if species are closely related the ranking list can contain both (or even more) related species until the defined log(score) is reached. Additionally, if a mixed culture is present, the dominant species will appear in higher ranking list positions with the second species also present within the listing. The distance (or the MALDI based similarity) between two species (between two single MSPs of a species) can be calculated. The warning algorithm collects in the first step all unique species from the ranking list and checks in the second step the relation of all species in this list. For this second step a further log(score) ("parameter #2") is defined as a close relation between two species. If this log(score) is exceeded both species will be excluded from the mixed culture warning automatically. Only if the log(score) between two species is less than the threshold and a close relation is unlikely will both species be further processed in the warning generation. In these cases, if the parameter #2 went below the threshold both reference spectra (MSPs) of the two mentioned species are combined temporary and the match of the unknown spectrum will be compared against this temporary, artificial MSP. If the log(score) against this artificial MSP is higher than the highest log(score) in the normal ranking list the mixed culture warning is generated, and a second species will be reported. {17} For this study, the following combinations of polymicrobial samples, as listed in Table 3 below, were tested at different concentrations (i.e., 10:0, 10:1, 5:1, 1:1, 1:5, 1:10 and 0:10). Table 3: Test organisms for polymicrobial validation | Organism Combination | Organisms Tested | | --- | --- | | Two taxonomically distant cleared species | Escherichia coli : Pseudomonas aeruginosa Escherichia coli : Staphylococcus aureus Klebsiella pneumoniae : Staphylococcus aureus | | Two different cleared organisms (bacteria and yeast) | Escherichia coli : Candida albicans Klebsiella pneumoniae : Candida albicans Staphylococcus aureus : Candida albicans | | Two taxonomically closely related cleared species | Escherichia coli : Klebsiella pneumoniae Klebsiella pneumoniae : Klebsiella variicola | | Non-clinically validated species and cleared species (as part of the ncv MBT-CA Library) | Escherichia coli : Corynebacterium argentoratense Staphylococcus aureus : Corynebacterium argentoratense | | Cleared species with contaminants (other cleared species) | Escherichia coli : Corynebacterium glutamicum Escherichia coli : Propionibacterium acnes Klebsiella pneumoniae : Corynebacterium glutamicum Staphylococcus aureus : Corynebacterium glutamicum Staphylococcus aureus : Propionibacterium acnes Staphylococcus aureus : Staphylococcus epidermidis | | Cleared species with RUO Species | Escherichia coli : Corynebacterium glaucum Staphylococcus aureus : Corynebacterium glaucum | In all cases, clean ranking lists, mixed ranking lists and correct mixed culture warnings were observed. No incorrect identification was reported by mixing a “RUO species” to a blood culture bottle with a claimed organism. A mixed culture warning was not expected for all polymicrobial samples since different concentration of the mixed organisms leads to different output (i.e., a factor of LoD of testing). The generation of a mixed culture hint is a strong indication, but not a confirmation, that a sample contains a mixture of organisms. Conversely, a sample containing a mixture of organisms might not generate a mixed culture hint. Therefore, as with all other MBT-CA analyses, final results from MBT Sepsityper samples must be assessed by a professional experienced in clinical microbiology and positive blood cultures should be subcultured for isolation and identification of organisms not identified by the MBT-CA System, for susceptibility testing, and differentiation of mixed growth. 5. Assay Reportable Range: MBT Sepsityper: 4,000-15,000 m/z MBT Sepsityper Samples: - High Confidence ID: 1.80 - 3.00; - Low Confidence ID: 1.60 - 1.79; - No Organism ID Possible: 0.00 - 1.59 {18} 6. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods): See IV.C.4 above 7. Detection Limit: This study demonstrated the tolerance range (dynamic range) analogous to the limit of detection (LOD) to verify that the time of positivity call by the continuously monitoring blood culture instrument provided sufficient biological material for successful identification. Five organisms (Klebsiella pneumoniae, Pseudomonas aeruginosa, Staphylococcus aureus, Streptococcus pneumoniae and Candida albicans) were tested (10 replicates each) at time of bottle positivity and at 1:10 and 1:100 dilutions. At "timepoint zero" (time of positivity call) of all positively flagged blood culture delivered sufficient biological material for correct MBT Sepsityper identifications. A few replicates of Klebsiella pneumoniae, were incorrectly identified as $K$ variicola or $E$ . aerogenes and a few replicates of Streptococcus pneumoniae were identified as S.mitis/oralis group. Diluted samples did not always reliably provide an identification; however, these dilutions demonstrated that lower concentration of microorganisms did not produce any incorrect identifications with the exception of the 1:100 dilution of Klebsiella pneumoniae, in which a few replicates were incorrectly identified as $K$ . variicola or $E$ . aerogenes. These are addressed as limitations in the instructions for use. 8. Assay Cut-Off: The ability of the MBT Sepsityper to identify bacteria and yeast in positive blood cultures was evaluated for all 334 claimed species and compared to the isolated colony identification on the MBT-CA System. For all claimed species / species groups one representative strain of each species (group) was processed by each of the MBT Sepsityper workflows (i.e., RS_DT, RS_eDT and Ext). Some "low log(score) based" incorrect results were observed, however for organisms with low confidence scores, the user should perform further methods to confirm the MBT Sepsityper result. A total of 14,247 analyzed log(scores) were determined in the assay cut-off study with the following results: | ID Level | # / % samples | | --- | --- | | High Confidence | 7531 (52.1%) | | Low Confidence | 1760 (12.4%) | | No ID | 3807 (26.7%) | | No Peaks | 1249 (8.75%) | | Incorrect Identification | 207 (1.45%) | Discordant Results: Altogether 207 samples $(1.45\%)$ of all analyzed 14,247 samples were incorrectly identified within the Assay Cut-Off study. "Contaminations": the majority (136 spectra) of incorrect results were resolved as contaminations during processing. After removing the samples suspected of contamination, sufficient test samples were measured for the mentioned species in the Assay Cut-Off Study. The remaining spectra are shown in brackets in the columns "# of Unique Spectra (minus discordant)" in Table 4 to Table 6. "Limitations": from the remaining 71 $(0.5\%)$ discordant {19} results 41 (0.29%) results represented minor errors since closely related species were incorrectly reported by the MBT-CA System. Where possible whenever cross identification occurred; a limitation was added to the matching hint table. "True Discordants": the remaining 30 (0.21%) incorrect results were represented by 2 high confidence and 28 low confidence results. Some (9) of these incorrect results (all high confidence and 7 low confidence) were derived from a single "bad reference entry" in the library which was subsequently replaced due to its low quality. From the remaining 21 incorrect results 17 resulted from ncv reference entries and 4 incorrect results were derived from cleared species. Please see Table 4 to Table 6 for summary of discordant results. Table 4: Assay Cut-Off Study Discordant Results "Contaminations". | Claimed Species | # of Unique Spectra (minus discordant) | Discordant Results | | | | | --- | --- | --- | --- | --- | --- | | | | confidence (# spectra) | | species | Resolution | | | | high | low | | | | Acinetobacter johnsonii | 36 (33) | 3 | | Acinetobacter haemolyticus | contamination | | Aerococcus urinae | 45 (37) | 7 | 1 | Vagococcus fluvialis | contamination | | Aerococcus viridans | 45 (44) | 1 | | Escherichia coli | contamination | | Aggregatibacter actinomycetemcomitans | 90 (87) | | 3 | Yersinia enterocolitica | contamination | | Aggregatibacter segnis | 36 (29) | 6 | 1 | Strep. salivarius / vestibularis group | contamination | | Bacteroides nordii | 36 (33) | | 3 | Propionibacterium acnes | contamination | | Bifidobacterium breve | 27 (23) | 2 | 2 | Staphylococcus hominis | contamination | | Klebsiella pneumoniae | 36 (34) | 2 | | Escherichia coli | contamination | | Legionella longbeachae | 54 (52) | 2 | | Escherichia coli | contamination | | Neisseria elongata | 36 (34) | | 2 | Corynebacterium coyleae | contamination | | Pediococcus pentosaceus | 45 (36) | 2 | 7 | Staphylococcus haemolyticus | contamination | | Plesiomonas shigelloides | 36 (19) | 16 | 1 | Chryseobacterium indologenes | contamination | | Porphyromonas gingivalis | 27 (19) | 5 | 3 | Staphylococcus hominis | contamination | | Rothia aeria | 45 (40) | 4 | 1 | Staphylococcus epidermidis | contamination | | Staphylococcus vitulinus | 54 (49) | 1 | 4 | Staphylococcus capitis | contamination | | Stenotrophomonas maltophilia | 72 (69) | 3 | | Yersinia kristensenii | contamination | | Vagococcus fluvialis | 36 (24) | 8 | 4 | Aerococcus urinae | contamination | | Candida famata | 36 (24) | 9 | 3 | Candida parapsilosis | contamination | | Candida inconspicua | 63 (57) | 4 | 2 | Pediococcus pentosaceus | contamination | | Candida lambica | 54 (49) | 5 | | Pediococcus pentosaceus | contamination | | Candida pararugosa | 36 (24) | 10 | 2 | Pediococcus acidilactici | contamination | | Candida tropicalis | 54 (47) | 6 | 1 | Staphylococcus epidermidis | contamination | | 96 | 40 | | --- | --- | | 136 | | {20} Table 5: Assay Cut-Off Study Discordant Results "Limitations", the species marked with (ncv) were from the clinically non validated library. | Claimed Species | # of Unique Spectra (minus discordant) | Discordant Results | | | | | --- | --- | --- | --- | --- | --- | | | | confidence (# spectra) | | species | Resolution | | | | high | low | | | | Corynebacterium pseudotuberculosis | 63 (62) | 1 | | Corynebacterium ulcerans | limitation | | Enterococcus raffinosus | 36 (35) | | 1 | Enterococcus avium | limitation | | Gemella haemolysans | 54 (52) | 1 | 1 | Gemella sanguinis | limitation | | Listeria monocytogenes | 36 (35) | 1 | | Listeria innocua | limitation (ncv) | | Neisseria lactamica | 36 (35) | | 1 | Neisseria flavescens / subflava group | limitation | | Neisseria weaveri | 36 (35) | 1 | | Neisseria zoodegmatis | limitation (ncv) | | Ochrobactrum anthropi | 36 (35) | 1 | | Ochrobactrum intermedium | limitation (ncv) | | Salmonella sp | 45 (43) | 1 | 1 | Citrobacter freundii | limitation | | Serratia plymuthica | 36 (35) | | 1 | Serratia liquefaciens | limitation | | Staphylococcus delphini | 54 (53) | | 1 | Staphylococcus pseudintermedius | limitation | | Staphylococcus pasteuri | 54 (51) | | 3 | Staphylococcus warneri | limitation | | Staphylococcus pseudintermedius | 45 (40) | 4 | 1 | Staphylococcus intermedius | limitation | | Staphylococcus schleiferi | 54 (53) | | 1 | Staphylococcus pseudintermedius | limitation | | Staphylococcus xylosus | 72 (71) | 1 | | Staphylococcus saprophyticus | limitation | | Streptococcus dysgalactiae | 54 (43) | 11 | | Streptococcus canis | limitation | | Streptococcus lutetiensis | 36 (35) | 1 | | Streptococcus gallolyticus | limitation | | Streptococcus mitis / oralis group | 36 (31) | 3 | 2 | Streptococcus pneumoniae | limitation | | Yersinia intermedia | 36 (35) | 1 | | Yersinia enterocolitica | limitation | | Candida dubliniensis | 72 (71) | | 1 | Candida albicans | limitation | | 27 | 14 | | --- | --- | | 41 | | {21} Table 6: Assay Cut-Off Study Discordant Results "True Discordants", the species marked with (ncv) were from the clinically non validated library, the MSP marked with "bad MSP" was replaced in the library due to quality issues. | Claimed Species | # of Unique Spectra (minus discordant) | Discordant Results | | | | | --- | --- | --- | --- | --- | --- | | | | confidence (# spectra) | | species | Resolution | | | | high | low | | | | Acinetobacter ursingii | 72 (71) | | 1 | Actinomyces neuii (eDT, 1.61) | discordant | | Actinomyces hyovaginalis | 72 (71) | | 1 | Actinomyces dentalis | discordant (ncv) | | Actinomyces urogenitalis | 54 (53) | | 1 | Actinomyces graevenitzii (eDT, 1.60) | discordant | | Anaerococcus vaginalis | 36 (35) | | 1 | Alloiococcus otitis | discordant (bad MSP) | | Capnocytophaga sputigena | 54 (51) | | 3 | Actinomyces dentalis | discordant (ncv) | | Clostridium paraputrificum | 36 (35) | | 1 | Alloiococcus otitis | discordant (bad MSP) | | Corynebacterium amycolatum | 54 (52) | | 2 | Actinomyces bovis | discordant (ncv) | | Corynebacterium aurimucosum group | 36 (35) | | 1 | Arthrobacter gandavensis | discordant (ncv) | | Corynebacterium bovis | 36 (35) | | 1 | Lactobacillus plantarum | discordant (ncv) | | Corynebacterium minutissimum | 36 (35) | | 1 | Lactobacillus paracasei | discordant (ncv) | | Corynebacterium pseudodiphtheriticum | 36 (35) | | 1 | Alloiococcus otitis | discordant (bad MSP) | | Dermacoccus nishinomiyaensis | 36 (34) | | 2 | Actinomyces oris (DT, 1.70; eDT, 1.63) | discordant | | Finegoldia magna | 36 (35) | | 1 | Alloiococcus otitis | discordant (bad MSP) | | Pseudomonas putida group | 63 (62) | | 1 | Pseudomonas alcaligenes | discordant (ncv) | | Pseudomonas stutzeri | 54 (53) | | 1 | Alloiococcus otitis | discordant (bad MSP) | | Rothia mucilaginosa | 54 (50) | 2 | 2 | Alloiococcus otitis | discordant (bad MSP) | | Staphylococcus sciuri | 54 (53) | | 1 | Corynebacterium cochlearium | discordant (ncv) | | Streptococcus canis | 36 (35) | | 1 | Actinomyces dentalis | discordant (ncv) | | Vibrio vulnificus | 54 (52) | | 2 | Actinomyces dentalis | discordant (ncv) | | Yersinia kristensenii | 54 (53) | | 1 | Actinomyces dentalis | discordant (ncv) | | Candida zeylanoides | 36 (35) | | 1 | Actinomyces dentalis | discordant (ncv) | | Saccharomyces cerevisiae | 81 (80) | | 1 | Actinomyces dentalis | discordant (ncv) | | 2 | 28 | | --- | --- | | 30 | | {22} 22 9. Clearance of Further Blood Culture Bottles: Clinical studies were performed across three clinical sites including two sites in the U.S. and one site in Germany. The following blood culture bottles were investigated: - 187 bottles of the BD BACTEC™ (Becton Dickinson) - BD BACTEC™ Standard Aerobic / Anaerobic - BD BACTEC™ PLUS Aerobic / Anaerobic - 145 bottles of the BacT/ALERT® (bioMérieux) - BacT/ALERT® FA Standard Aerobic - BacT/ALERT® FN Standard Anaerobic - 47 bottles of the VersaTREK® (Thermo Scientific). - VersaTREK® REDOX 1 - VersaTREK® REDOX 2 In addition to the clinical site method comparison which used bioMérieux, Becton Dickinson and Thermo Scientific blood culture bottles, additional bottle types were tested internally. The following Blood Culture Media/Bottle Types were tested: - BD BACTEC™ Pediatric Plus - BD BACTEC™ Lytic/10 Anaerobic/F - BD BACTEC™ Mycosis IC/F Medium Culture Vials - BD BACTEC™ Myco/F Lytic Medium - bioMérieux BacT/ALERT® SA Standard Aerobic - bioMérieux BacT/ALERT® SN Standard Anaerobic - bioMérieux BacT/ALERT® PF Plus For bioMérieux BacT/ALERT bottles, three bottle types were assessed using ten different microorganism species (Gram positive, Gram negative and yeast). For BD BACTEC bottles, four bottle types were assessed using five different microorganisms All sample preparations were performed (RS_DT, RS_eDT and Ext) and the measurement was executed using two MALDI instruments. Each sample preparation was prepared 12 times on the MALDI target and measured once. Altogether about 3,600 spectra were acquired and analyzed during this study. No misidentifications were seen. NOTE: - Reduced performance was seen with the BACTEC Lytic/10 Anaerobic/F. - MBT Sepsityper should only be used with charcoal free bottles. 10. Carry-Over: Five different organisms were cultivated in blood culture bottles. Additionally, five blood culture bottles were inoculated with sheep blood as negative samples. All bottles were placed on a blood culture instrument and pulled when positive (negative bottles were pulled at 19 hours) The bottles {23} were processed using the MBT Sepsityper Kit US IVD and with negative and positive samples spotted alternately on a MTB Biotarget 96 and a US IVD 48 Spot Target. No carry over or cross contamination on the target was observed using the Rapid Sepsityper workflow. # 11. Stability of Blood Culture (Frozen, Cooled or Seeded) Five blood culture bottles (BD BACTEC™ PLUS Aerobic) were inoculated with five different organisms (Candida albicans DSM 11943, Klebsiella pneumoniae DSM $16358^{\mathrm{T}}$ , Pseudomonas aeruginosa DSM $50071^{\mathrm{T}}$ , Staphylococcus aureus DSM $20231^{\mathrm{T}}$ , and Streptococcus pneumoniae DSM $20566^{\mathrm{T}}$ ) with $10^{2}$ cells/ml and cultivated in the blood culture instrument till positivity. Additionally, five blood culture bottles were inoculated with the same five organisms with $10^{7}$ cells/ml (seeded) and were harvested immediately. (Seeded samples consisted of the medium from blood culture bottles, inoculated with the appropriate volume of blood, seeded with the desired number of microbial cells). Both types of prepared blood culture bottles were aliquoted in 1ml portions. The aliquots were stored at room temperature (RT), $4^{\circ}\mathrm{C}$ , and $-18^{\circ}\mathrm{C}$ . Subsequently, the Rapid Sepsityper workflow was performed with one set of each fraction at different time points to show the equivalence between fresh (RT), cooled ( $4^{\circ}\mathrm{C}$ ), frozen ( $-18^{\circ}\mathrm{C}$ ) and seeded samples. Equivalency was demonstrated for the four types of positive blood cultures such that each could be independently from each other in the analytical studies performed. # B Comparison Studies: 1. Challenge Panel: Each site received a panel of 50 of the most common blood culture organisms (Gram positive, Gram negative and Yeast) in frozen blood culture aliquots. All three sample preparation workflows were applied in parallel. No incorrect identifications were obtained with Challenge Panel testing. Table 7 summarizes results below. Table 7: Challenge Panel Summary Results | | site 1 | | | site 2 | | | site 3 | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | high | low | no | high | low | no | high | low | no | | | confidence | | ID | confidence | | ID | confidence | | ID | | DT | 53% | 17% | 30% | 62% | 13% | 25% | 38% | 26% | 36% | | eDT | 75% | 6% | 19% | 83% | 9% | 8% | 58% | 16% | 26% | | Ext | 81% | 13% | 6% | 85% | 4% | 11% | 84% | 6% | 10% | | DT → eDT → Ext | 88% | 6% | 6% | 94% | 2% | 4% | 88% | 6% | 6% | # 2. Method Comparison with Predicate Device: Samples for evaluation of the MBT Sepsityper performance were collected from three (3) clinical sites: Site 1 analyzed 187 samples, site 2 analyzed 145 samples and site 3 analyzed 47 samples. All samples were obtained from the sites daily routine samples without any pre-selection. During the clinical studies polymicrobial samples were excluded via study design (Gram staining). Overall 379 samples were evaluated. Sixty-seven different species were identified during the clinical tests. No misidentifications occurred. Organisms are summarized in Table 8 below. {24} Table 8: Breakdown of Clinical Study Organisms | 2 | Achromobacter xylosoxidans | | --- | --- | | 1 | Acinetobacter_baumannii_nosocomialis group | | 1 | Actinomyces oris | | 1 | Actinotignum_schaalii group | | 1 | Aerococcus urinae | | 1 | Bacteroides fragilis | | 1 | Brevibacillus centrosporus (NCV) | | 1 | Citrobacter_freundii complex | | 1 | Clostridium perfringens | | 1 | Corynebacterium_afementans group | | 1 | Corynebacterium_aurimucosum group | | 4 | Corynebacterium_striatum group | | 1 | Enterobacter aerogenes | | 6 | Enterobacter cloacae_complex | | 1 | Enterococcus casseliflavus | | 10 | Enterococcus faecalis | | 7 | Enterococcus faecium | | 1 | Enterococcus gallinarum | | 63 | Escherichia coli | | 1 | Granulicatella adiacens | | 2 | Haemophilus influenzae | | 1 | Klebsiella_oxytoca Raoultella_ornithinolytica | | 10 | Klebsiella pneumoniae | | --- | --- | | 2 | Klebsiella variicola | | 1 | Lactobacillus rhamnosus | | 1 | Lactobacillus salivarius | | 6 | Micrococcus luteus | | 2 | Morganella morganii | | 2 | Neisseria gonorrhoeae | | 1 | Neisseria meningitidis | | 1 | Parvimonas micra | | 1 | Pasteurella multocida | | 8 | Propionibacterium acnes | | 6 | Proteus mirabilis | | 13 | Pseudomonas aeruginosa | | 1 | Rhizobium radiobacter | | 2 | Serratia marcescens | | 59 | Staphylococcus aureus | | 8 | Staphylococcus capitis | | 1 | Staphylococcus caprae | | 2 | Staphylococcus cohnii | | 75 | Staphylococcus epidermidis | | 14 | Staphylococcus hominis | | 1 | Staphylococcus lugdunensis | | 1 | Staphylococcus pasteuri | | --- | --- | | 3 | Staphylococcus pettenkoferi | | 1 | Staphylococcus saccharolyticus | | 1 | Stenotrophomonas maltophilia | | 4 | Streptococcus agalactiae | | 1 | Streptococcus anginosus | | 4 | Streptococcus dysgalactiae | | 1 | Streptococcus equi | | 1 | Streptococcus gallolyticus | | 1 | Streptococcus gordonii | | 1 | Streptococcus intermedius | | 1 | Streptococcus lutetiensis | | 8 | Streptococcus_mitis_oralis group | | 1 | Streptococcus parasanguinis | | 6 | Streptococcus pneumoniae | | 3 | Streptococcus pyogenes | | 2 | Streptococcus sanguinis | | 1 | Streptococcus thermophilus | | 5 | Candida albicans | | 4 | Candida glabrata | | 1 | Candida lusitaniae | | 1 | Candida parapsilosis | | 1 | Candida tropicalis | Sixty-seven different species belonging to 31 genera were observed in the study. Fifty-two percent of all samples were identified as Staphylococcus epidermidis, Escherichia coli or Staphylococcus aureus. The two Gram positive species Staphylococcus aureus and Staphylococcus epidermidis appeared very frequently and were responsible for 134 out of 248 (54%) Gram positive samples. All three sample preparations were performed for all clinical samples in this study, i.e., "Rapid Sepsityper DT" (RS_DT), "Rapid Sepsityper eDT (RS_eDT) and Full Sepsityper (Ext). Colonies were identified using the DT/eDT result. For result interpretation, spotting methods were analyzed separately as well as by MBT Sepsityper workflow. See Table 9 and Table 10 below. No discordant high & low confidence results were obtained. The identification performance of the Rapid Sepsityper Workflows compared to the MBT Sepsityper Full extraction was lower for certain groups of organisms. The RS_DT sample preparation for Gram positive bacteria showed only $32\%$ "high confidence ID" log(scores) compared to $55\%$ scores for Gram negative organisms prepared with the same technique. The combination of RS_DT/RS_eDT showed nearly identical performance for both bacterial groups. For Gram negative bacteria, the combination of RS_DT and RS_eDT led to 71 successful identifications compared to 89 identifications by using the extraction workflow alone. For Gram positive bacteria the combined rapid workflows identified 145 samples compared to 198 samples with extraction. This means that for both bacterial groups the rapid workflows delivered a final identification result in about $75\%$ of cases compared to the "extraction only workflow". The general performance of Gram positive and Gram negative bacteria was found to be very similar. The rapid workflows combined identified about $60\%$ of all samples of Gram positive as well as Gram negative bacteria. The entire workflow $(\mathrm{DT} \rightarrow \mathrm{eDT} \rightarrow \mathrm{Ext})$ was slightly better for Gram positive bacteria (77% vs 84%). {25} Four percent of all positive blood culture samples were found to contain yeast and no false ID was observed. The identification performance for yeast was low with about $30\%$ of isolates identified with high or low confidence after applying the entire workflow (RS_DT $\rightarrow$ RS_eDT $\rightarrow$ Ext) and the remainder $(70\%)$ reported as "no ID". # Clinical Study Polymicrobial Blood Cultures If the positive blood culture Gram stain indicated a polymicrobial infection, the sample was excluded from the study. However, after the positive blood culture bottles were subcultured, plates were examined for purity and isolated colonies were identified using the MBT-CA System in accordance with instructions for use. If a polymicrobial sample (mixed culture) was observed, each organism was processed for identification on the MBT-CA System following instructions for use. Thirteen samples were confirmed as "polymicrobial samples" after subculture and overnight incubation on agar plates. Of these, five polymicrobial samples were correctly identified within the identification ranking lists and eight were not identified within the identification ranking lists. In both cases, not a single completely false ID was observed in any MBT Sepsityper identification. In all cases of polymicrobial samples one part of the two species was identified correctly or a "no identification" was reported. Table 9: Overall Identification Results (%) by Spotting Method. | Number of Identifications - Clinical Studies | | | | | | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | organisms | Number (%) of Samples | RS_DT | | | | RS_eDT | | | | Ext | | | | | | confidence | | no | | confidence | | no | | confidence | | no | | | | high | low | ID | peak | high | low | ID | peak | high | low | ID | | Gram negative bacteria | 119 (31.4) | 65 (54.6) | 8 (6.7) | 22 (18.5) | 24 (20.2) | 64 (53.8) | 8 (6.7) | 20 (16.8) | 27 (22.7) | 89 (74.8) | 4 (3.4) | 19 (16) | | Gram positive bacteria | 248 (65.4) | 80 (32.3) | 38 (15.3) | 87 (35.1) | 43 (17.3) | 132 (53.2) | 37 (14.9) | 44 (17.7) | 35 (14.1) | 198 (79.8) | 25 (10.1) | 23 (9.3) | | Yeast | 12 (3.2) | 0 (0) | 1 (8.3) | 8 (66.7) | 3 (25) | 1 (8.3) | 0 (0) | 7 (58.3) | 4 (33.3) | 2 (16.7) | 1 (8.3) | 8 (66.7) | | All samples | 379 (100) | 145 (38.3) | 47 (12.4) | 117 (30.9) | 70 (18.5) | 197 (52) | 45 (11.9) | 71 (18.7) | 66 (17.4) | 289 (76.3) | 30 (7.9) | 50 (13.2) | Table 10: Overall Results (%) by Spotting Workflow | Number of Identifications - Clinical Studies | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | | organisms | Number (%) of Samples | RS_DT PLUS RS_eDT | | | RS_DT ▷ RS_eDT ▷ Ext | | | | | | High Confidence | High & Low Confidence | No ID | High Confidence | High & Low Confidence | No ID | | Gram negative bacteria | 119(31.4) | 71(59.7) | 82(68.9) | 37(31.1) | 92(77.3) | 99(83.2) | 20(16.8) | | Gram positive bacteria | 248(65.4) | 145(58.5) | 185(74.6) | 63(25.4) | 209(84.3) | 228(91.9) | 20(8.1) | | Yeast | 12(3.2) | 1(8.3) | 2(16.7) | 10(83.3) | 2(16.7) | 4(33.3) | 8(66.7) | | All samples | 379(100) | 217(57.3) | 269(71) | 110(29) | 303(79.9) | 331(87.3) | 48(12.7) | {26} 3. Matrix Comparison: N/A C Clinical Studies: 1. Clinical Sensitivity: See B.1 Method Comparison with Predicate Device: 2. Clinical Specificity: See B.1 Method Comparison with Predicate Device: 3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable): N/A D Clinical Cut-Off: See Assay Reportable Range VII.A.5 above. E Expected Values/Reference Range: See Assay Reportable Range VII.A.5 above. F Other Supportive Instrument Performance Characteristics Data: N/A VIII Proposed Labeling: The labeling supports the finding of substantial equivalence for this device. IX Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. As all blood culture bottles are subcultured and growth will be assessed as compared to MBT Sepsityper identification, reduced performance for MBT Sepsityper blood culture identification is acceptable as compared to isolated colony identification. Users should follow the Instructions for Use (IFU) which indicates all results should be reviewed by a trained microbiologist and final organism identification should be based on all relevant information available. 26
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...