The Simplant® Digital Guide is a surgical guide intended for use in assisting in the placement of dental implants. The Simplant® Digital Guide is designed centrally by the Simplant® planning software and manufactured locally using Formlabs Dental SG resin, Loctite 4304 adhesive, and Formlabs equipment: Form 2 printer, FormWash and FormCure washing and post-curing equipment.
Device Story
Simplant® Digital Guide is a patient-specific surgical template for dental implant placement. Input: digital pre-operative plan generated by Simplant® planning software. Transformation: clinician/lab receives digital design file; fabricates guide locally via stereolithography using Formlabs Form 2 printer, Dental SG resin, and FormWash/FormCure equipment. Output: physical surgical guide that seats on patient teeth/mucosa. Usage: clinician uses guide in clinic to direct surgical instruments according to pre-operative plan. Benefit: improved accuracy of implant placement relative to planned position. Device is non-sterile; requires user sterilization.
Clinical Evidence
No clinical data included. Substantial equivalence supported by bench testing only, including tube fixation (push-out and torque), strength testing per ISO 20795-1:2013, angulation/positional deviation tests, and vertical fit tests. Additive manufacturing process validated across multiple build runs. Biocompatibility confirmed per ISO 10993-1 and ISO 7405. Sterilization validated per ISO 17665-1/2.
Technological Characteristics
Patient-specific surgical guide; stereolithography (SLA) additive manufacturing. Materials: Formlabs Dental SG acrylic resin, titanium alloy sleeves, Loctite 4304 adhesive. Minimum wall thickness 2.5mm. Connectivity: digital design file transfer. Sterilization: user-performed moist heat (steam). Bench testing standards: ISO 20795-1:2013 (denture base polymers), ISO 10993-1, ISO 7405, ISO 17665-1/2.
Indications for Use
Indicated for use as a surgical guide to assist in the placement of dental implants in patients requiring dental implant surgery.
Regulatory Classification
Identification
Endosseous dental implant accessories are manually powered devices intended to aid in the placement or removal of endosseous dental implants and abutments, prepare the site for placement of endosseous dental implants or abutments, aid in the fitting of endosseous dental implants or abutments, aid in the fabrication of dental prosthetics, and be used as an accessory with endosseous dental implants when tissue contact will last less than 1 hour. These devices include drill bits, screwdrivers, countertorque devices, placement and removal tools, laboratory pieces used for fabrication of dental prosthetics, and trial abutments.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath in a smaller font.
September 2, 2020
Dentsply Sirona Karl Nittinger Vice President Corporate Regulatory Affairs 221 West Philadelphia Street, Suite 60W York, Pennsylvania 17401
# Re: K193405
Trade/Device Name: Simplant® Digital Guide Regulation Number: 21 CFR 872.3980 Regulation Name: Endosseous Dental Implant Accessories Regulatory Class: Class I Product Code: NDP Dated: July 30, 2020 Received: July 31, 2020
### Dear Karl Nittinger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193405
Device Name Simplant® Digital Guide
Indications for Use (Describe)
The Simplant® Digital Guide is a surgical guide intended for use in assisting in the placement of dental implants. The Simplant® Digital Guide is designed centrally by the Simplant® planning software and manufactured locally using Formlabs Dental SG resin, Locite 4304 adhesive, and Formlabs equipment: Form 2 printer, FormWash and FormCure washing and post-curing equipment.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------|--|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | |
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Dentsply Sirona 221 West Philadelphia Street Suite 60W York, PA 17401
Image /page/3/Picture/2 description: The image contains the logo for Dentsply Sirona. The logo consists of a stylized leaf-like shape on the left, followed by the text "Dentsply" on the top line and "Sirona" on the bottom line. The text is in a simple, sans-serif font and is aligned to the right of the leaf-like shape.
### 510(k) SUMMARY
for K193405 Simplant® Digital Guide
Submitter Information: Dentsply Sirona
> 221 West Philadelphia Street Suite 60W York, PA 17401
| Contact Person:<br>Telephone Number:<br>Fax Number:<br>Email: | Karl Nittinger<br>717-849-4424<br>717-849-4343<br>karl.nittinger@dentsplysirona.com |
|---------------------------------------------------------------|-------------------------------------------------------------------------------------|
| Date Prepared: | 02-September-2020 |
| Device Name:<br>• Proprietary Name: | Simplant® Digital Guide |
| • Classification Name:<br>• CFR Number:<br>• Device Class: | Endosseous Dental ImplantAccessories<br>• 21 CFR 872.3980<br>• Class I |
NDP
# Predicate Device:
.
Product Code:
| Predicate Device<br>Name | 510(k) | Company Name |
|--------------------------|---------|-----------------|
| Simplant® Guide | K170849 | Dentsply Sirona |
# Description of Device:
The Simplant® Digital Guide is intended for use in assisting placement of dental implants.
The subject Simplant® Digital Guide is a patient specific surgical template, which is designed according to the digital pre-operative plan of the dental implant positions. The clinician, or dental lab, orders a digital design file of the surgical quide (i.e. Simplant® Guide File) which is developed under the same process that is utilized to produce the design of the predicate Simplant® Guide (K170849). The subject Simplant® Digital Guide is then fabricated locally by the clinician or dental lab, utilizing stereolithography fabrication method. The subject Simplant® Digital Guide is intended to be fabricated using Dental SG acrylic resin, Form 2 SLA 3D printer, and FormWash and FormCure washing and post-curing equipment manufactured by Formlabs, Inc.
{4}------------------------------------------------
The Simplant® Digital Guide sits on the patient's oral anatomy, i.e. teeth, mucosa or combination thereof. Aided by the Simplant® Digital Guide, the implant sites can be prepared, and the dental implants can be placed in the predetermined locations according to the previously pre-operative dental implant treatment plan.
#### Indications for Use:
The Simplant® Digital Guide is a surgical quide intended for use in assisting in the placement of dental implants. The Simplant® Digital Guide is designed centrally by the Simplant® planning software and manufactured locally using Formlabs Dental SG resin, Loctite 4304 adhesive, and Formlabs equipment: Form 2 printer, FormWash and FormCure washing and post-curing equipment.
### Substantial Equivalence:
Technological Characteristics.
An overview of the similarities and differences between the subject and predicate devices are given in Table 5.1.
{5}------------------------------------------------
| Table 5.1-Similarities and Differences | | | |
|-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Element | Subiect Device<br>Simplant® Digital Guide | Predicate Device<br>Simplant® Guide<br>(K170849) | Similarities and Differences |
| Indications for<br>Use | The Simplant® Digital Guide is a<br>surgical guide intended for use in<br>assisting in the placement of<br>dental implants. The Simplant®<br>Digital Guide is designed centrally<br>by the Simplant® planning<br>software and manufactured<br>locally using Formlabs Dental SG<br>resin, Loctite 4304 adhesive, and<br>Formlabs equipment: Form 2<br>printer, FormWash and FormCure<br>washing and post-curing<br>equipment. | The Simplant® Guide is intended<br>for use in assisting placement of<br>dental implants. | Both the subject Simplant® Digital Guide and<br>predicate Simplant® Guide (K170849) are<br>intended for use in assisting placement of dental<br>implants. However, predicate Simplant® Guide<br>(K170849) is fabricated locally by Dentsply<br>Sirona and subject Simplant® Digital Guide is<br>fabricated locally by the clinician or dental lab. |
| Function and<br>Principle of<br>Operation | Supporting surface:<br>teeth and mucosa Function:<br>Guidance of instruments<br>according to pre-operative<br>plan Surgical procedure:<br>Sequence of instruments<br>according to manufacturer's<br>instructions. | Supporting surface:<br>teeth, mucosa and bone Function:<br>Guidance of instruments<br>according to pre-operative<br>plan Surgical procedure:<br>Sequence of instruments<br>according to manufacturer's<br>instructions. | The function and principle of operation of the<br>subject Simplant® Digital Guide are almost<br>identical to the function and principle of<br>operation of the predicate Simplant® Guide<br>(K170849). The subject Simplant® Digital Guide<br>is limited to seating on teeth, mucosa and<br>combination thereof. Predicate Simplant® Guide<br>(K170849) can also be seated directly on the<br>bone. The fabrication process of subject<br>Simplant® Digital Guide is not adapted to<br>fabrication of bone models, which are needed<br>for quality control verification of bone supported<br>surgical guides. Therefore, subject Simplant®<br>Digital Guide is not intended as a bone<br>supported variant. |
| Element | Subject Device<br>Simplant® Digital Guide | Predicate<br>Device<br>Simplant® Guide | Similarities and Differences |
| Technology | Guide design by Dentsply<br>Sirona or by clinician and review<br>by Dentsply Sirona.<br>Local fabrication by<br>stereolithography.<br>Print:<br>- Software: PreForm<br>(manufacturer: Formlabs)<br>- Equipment: Form 2<br>(manufacturer: Formlabs) Wash:<br>Equipment: Form Wash<br>(manufacturer: Formlabs) Post-cure:<br>Equipment: Form Cure<br>(manufacturer: Formlabs) | Guide design by Dentsply<br>Sirona or by clinician and review<br>by Dentsply Sirona.<br>Internal (Dentsply Sirona)<br>fabrication by stereolithography. | The design process for the subject Simplant®<br>Digital Guide and predicate Simplant® Guide<br>(K170849) are the same.<br>The technology applied to create the subject<br>Simplant® Digital Guide is different from<br>predicate Simplant® Guide (K170849). Both<br>subject and predicate device are created by<br>stereolithography, however, the fabrication of<br>predicate Simplant® Guide (K170849) is<br>executed centrally by Dentsply Sirona and the<br>fabrication of subject Simplant® Digital Guide is<br>performed locally by the clinician or lab utilizing<br>the Formlabs printer and acrylic resin.<br>Bench and performance testing is included to<br>address technological differences between the<br>subject and predicate device in support of<br>substantial equivalence. |
| Material | Body:<br>Acrylic resin: Dental SG<br>(manufacturer: Formlabs) Metal tube/sleeve:<br>Titanium alloy Adhesive:<br>Loctite 4304 | Body:<br>Acrylic resin Metal tube/sleeve:<br>Titanium alloy or stainless<br>steel Adhesive | The material composition of the subject<br>Simplant® Digital Guide is different than that of<br>the predicate Simplant® Guide (K170849). While<br>both are fabricated from acrylic resin in an<br>additive manufacturing process, a different<br>specific acrylic resin is used for the subject<br>device. The material for tubes/sleeves is limited<br>to titanium alloy and the glue is the same.<br>Assessment of the biocompatibility of the<br>subject device is included to address the<br>material and process differences between the<br>subject and predicate device in support of<br>substantial equivalence. |
| Element | Subject Device<br>Simplant® Digital Guide | Predicate<br>Device<br>Simplant® Guide<br>(K170849) | Similarities and Differences |
| Design and<br>Features | Body:<br>- Surface following<br>- Wall thickness: minimum<br>2.5 mm Metal tube/sleeve:<br>Dimensions and position<br>depending on pre-operative<br>plan | Body:<br>- Surface following<br>- Wall thickness: minimum<br>2.5 mm Metal tube/sleeve:<br>Dimensions and position<br>depending on pre-operative<br>plan | The fundamental design parameters and metal<br>tubes of the subject Simplant® Digital Guide are<br>identical as compared to predicate Simplant®<br>Guide (K170849). |
| Sterilization<br>Method | Sterilization by the user: | Sterilization by the user: | The sterilization method is the same. . |
| Shelf Life | 2 weeks | 2 months | The shelf life of the subject Simplant® Digital<br>Guide is reduced as compared to predicate<br>Simplant® Guide (K170849). This difference is<br>supported by the fact that the local manufacture<br>of the subject device lends itself to use of the<br>device within a short time period after fabrication. |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
### Non-Clinical Performance Data.
Non-clinical testing data submitted, referenced, or relied upon to demonstrate substantial equivalence include data from the following tests:
#### Bench Testing:
- . Tube fixation test - push out.
This test is executed to ensure that the subject device withstands typical vertical loads that may be applied during surgery.
- . Tube fixation test - torque.
This test is executed to ensure that the subject device withstands occasional torque loads that may be applied during surgery.
- . Strength test.
This test is executed to ensure that the subject device complies with the international standard related to dental materials, ISO 20795-1:2013-1 Dentistry - Base polymers - Part 1: Denture base polymers.
- Angulation deviation test.
This test is executed to ensure that the subject device is manufactured according to its pre-operative plan regarding the angular position of the quide sleeve.
- . Position deviation test.
This test is executed to ensure that the subject device is manufactured according to its pre-operative plan regarding the spatial position of the quide sleeve.
- . Vertical fit test.
This test is executed to ensure that the subject device is manufactured according to its pre-operative plan regarding the vertical position. As a result of the angulation deviation test, the centrally controlled (by Dentsply Sirona) design process is updated. For the subject device the safety margin between the planned implant and the lingual bone plate is slightly increased.
Development of bench testing to support the assessment of the additive manufacturing process variables was conducted with reference to the December, 2017 Guidance for Industry and Food and Drug Administration Staff: Technical Considerations for Additive Manufactured Medical Devices.
- . The additive manufacturing process workflow was validated considering the critical additive manufacturing process variables that could affect the final product specifications. Considerations for build location, build orientation, print layer thickness, and potential for raw material reuse were made. Strength test, tube fixation (pushout and torque), angulation deviation, position deviation, and vertical fit tests were conducted on samples from at least 3 build runs to confirm that the process variables do not affect the finished device's conformance to its predetermined specifications.
{9}------------------------------------------------
#### Biocompatibility Testing:
Biocompatibility assessment and relevant testing based on the requirements of ISO 10993-1:2009 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process and ISO 7405:2008 Dentistry - Evaluation of biocompatibility of medical devices used in dentistry was performed for the subject Simplant® Digital Guide. It is concluded that the subject Simplant® Digital Guide meets the criteria for biocompatibility according to these standards.
#### Sterilization Validation:
The updated steam sterilization parameters were validated according to ISO 17665-1:2006 Sterilization of health care products - Moist heat - Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices and ISO 17665-2:2009 Sterilization of health care products - Moist heat - Part 2: Guidance on the application of ISO 17665-1 demonstrating a sterility assurance level (SAL) of 10-9.
### Shelf Life Testing:
The shelf life of the subject Simplant® Digital Guide is 2 weeks. To ensure the performance of subject Simplant® Digital through the claimed shelf life, all the performed bench testing on subject device was executed on samples that were sterilized and subjected to real-time aging for 2 weeks.
The results from the non-clinical performance testing, in combination with an updated design process, support substantial equivalence of the subject device Simplant® Digital Guide to the predicate device Simplant® Guide (K170849).
# Clinical Performance Data.
No data from human clinical studies has been included to support the substantial equivalence of the Simplant® Digital Guide.
# Conclusion Regarding Substantial Equivalence
The subject Simplant® Digital Guide is a patient specific surgical template which is intended for use in assisting placement of dental implants. The Simplant® Digital Guide has identical Indications for Use, exhibits identical fundamental design features and similar principles of operation as the predicate Simplant® Guide cleared under premarket notification K170849. Test data to verify the performance of the Simplant® Digital Guide has been included. The results of this testing, and the comparison with the predicate device, support substantial equivalence.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.