The Miru 1day UpSide (midafilcon A) SPHERICAL Lens is indicated for daily wear single use only for the optical correction of refractive ametropia (myopia and hyperopia) in aphakic and non-aphakic people with disease free eyes who may have 1.50 diopter (D) or less of astigmatism. The Miru 1day UpSide (midafilcon A) TORIC Lens is indicated for daily wear single use only for the optical correction of refractive ametropia (myopia or hyperopia with astigmatism) in aphakic and non-aphakic people with disease free eyes with 3.00 diopter (D) or less of refractive astigmatism. The Miru 1day UpSide (midafilcon A) MULTIFOCAL Lens is indicated for daily wear single use only for the optical correction of refractive ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic and non-aphakic people with disease free eyes who may require a reading addition of 3.00 diopter (D) or less and who may have 1.50 diopter (D) or less of astigmatism. The Miru 1day UpSide (midafilcon A) should only be worn once and then discarded at the end of each wearing period on a daily basis. The patient should start the next wearing period with fresh lenses should not be cleaned or disinfected and should be discarded after a single use.
Device Story
Miru 1day UpSide is a hydrophilic, silicone hydrogel soft contact lens; available in spherical, toric, and multifocal designs. Material is midafilcon A (random co-polymer with polydimethyl siloxane macromonomer); 54% water content. Includes benzotriazole UV-absorbing monomer. Manufactured via cast molding; sterilized by moist heat. Intended for daily wear, single-use; discarded after one use without cleaning or disinfection. Prescribed by eye care practitioners for patients with refractive ametropia. Provides optical correction; improves visual acuity. Benefits include convenience of daily disposal and UV protection.
Clinical Evidence
Three-month randomized controlled clinical study; 148 subjects; 1:1 randomization (test vs. control). Primary endpoints: visual acuity, safety (adverse event rates, biomicroscopy, subjective acceptance). Results: test lens found substantially equivalent to control for safety and efficacy.
Technological Characteristics
Material: midafilcon A (silicone hydrogel, FDA SiHy class 5). Water content: 54%. Manufacturing: cast molded. Sterilization: moist heat. Packaging: blister pack. UV protection: benzotriazole monomer (transmittance <5% UVB, <50% UVA). Connectivity: none. Software: none.
Indications for Use
Indicated for daily wear single-use optical correction of refractive ametropia (myopia, hyperopia, presbyopia) in aphakic and non-aphakic patients with disease-free eyes. Spherical/Multifocal versions accommodate up to 1.50D astigmatism; Toric version accommodates up to 3.00D astigmatism.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 17, 2020
Menicon Co., Ltd. Li Haosheng, Ph.D. International Regulatory Affairs 3-21-19. Aoi. Naka-ku Nagoya, 460-0006 Japan
Re: K193399
Trade/Device Name: Miru 1day UpSide (midafilcon A) Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL, MVN Dated: May 15, 2020 Received: May 18, 2020
Dear Li Haosheng,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{2}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
J. Angelo Green, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
510(k) Number (if known)
K193399
Device Name
Miru 1day UpSide (midafilcon A)
#### Indications for Use (Describe)
The Miru 1day UpSide (midafilcon A) SPHERICAL Lens is indicated for daily wear single use only for the optical correction of refractive ametropia (myperopia) in aphakic and non-aphakic people with disease free eyes who may have 1.50 diopter (D) or less of astigmatism.
The Miru 1day UpSide (midafilcon A) TORIC Lens is indicated for daily wear single use only for the optical correction of refractive ametropia (myopia or hyperopia with astigmatism) in aphakic people with disease free eyes with 3.00 diopter (D) or less of refractive astigmatism.
The Miru 1day UpSide (midafilcon A) MULTIFOCAL Lens is indicated for daily wear single use only for the optical correction of refractive ametropia (myopia) and emmetropia with presbyopia in aphakic and non-aphakic people with disease free eyes who may require a reading addition of 3.00 diopter (D) or less and who may have 1.50 diopter (D) or less of astigmatism.
The Miru 1day UpSide (midafilcon A) should only be worn once and then discarded at the end of each wearing period on a daily basis. The patient should start the next wearing period with fresh lenses should not be cleaned or disinfected and should be discarded after a single use.
| Type of Use (Select one or both, as applicable) |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:inline-block;">☒ Prescription Use (Part 21 CFR 801 Subpart D)</div> <div style="display:inline-block;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the text "Menicon Co., Ltd." The text is written in a simple, sans-serif font. The text is centered in the image and is the only element present.
# 510(k) Summary
# Miru 1day UpSide (midafilcon A)
#### 1. Applicant Information
Menicon Co., Ltd. 21-19, Aoi 3, Naka-ku, Nagoya, 460-0006 JAPAN
| Contact Person: | Li Haosheng, Ph.D. |
|-----------------|---------------------------|
| Telephone No.: | +81-52-935-1676 |
| Fax No.: | +81-52-935-1633 |
| E-mail: | haosheng-li@menicon.co.jp |
| Date Prepared: | June 16, 2020 |
#### 2. Device Information
| Classification name: | Soft (hydrophilic) Contact Lens |
|------------------------|---------------------------------|
| Device classification: | Class II |
| Regulation number: | 21 CFR 886.5925 |
| Product code: | LPL, MVN |
| Proprietary name: | Miru 1day UpSide (midafilcon A) |
#### 3. Predicate Device
Menicon claims substantial equivalence to K160344 Visco Si-Hy (olifilcon B) Silicone Hydrogel Soft Contact Lens and K131378 CooperVision MyDay (stenfilcon A) Soft Contact Lens.
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the alphanumeric string "K193399". The string is written in a simple, sans-serif font. The characters are all the same size and are evenly spaced.
#### 4. Description of Device
Menicon Co., Ltd.
The Miru 1day UpSide (midafilcon A) is a hydrophilic contact lens which is available as a spherical, toric and multifocal lens. The lens is indicated for daily wear disposable single use.
The non-ionic lens material (midafilcon A) is a random co-polymer containing polydimethyl siloxane macromonomer. It consists of 46% midafilcon A and 54% water by weight when immersed in a buffered saline solution. The lens is available with a pale blue visibility handling tint.
The lens contains a benzotriazole UV absorbing monomer which is used to block UV radiation. The transmittance characteristics for the lens (-3.00D) are less than 5% of UVB radiation and less than 50% of UVA radiation.
#### 5. Indications for Use
The Miru 1day UpSide (midafilcon A) SPHERICAL Lens is indicated for daily wear single use only for the optical correction of refractive ametropia (myopia and hyperopia) in aphakic and non-aphakic people with disease free eyes who may have 1.50 diopter (D) or less of astigmatism.
The Miru 1day UpSide (midafilcon A) TORIC Lens is indicated for daily wear single use only for the optical correction of refractive ametropia (myopia or hyperopia with astigmatism) in aphakic and non-aphakic people with disease free eyes with 3.00 diopter (D) or less of refractive astigmatism.
The Miru 1day UpSide (midafilcon A) MULTIFOCAL Lens is indicated for daily wear single use only for the optical correction of refractive ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic and non-aphakic people with disease free eyes who may require a reading addition of 3.00 diopter (D) or less and who may have 1.50 diopter (D) or less of astigmatism.
The Miru 1day UpSide (midafilcon A) should only be worn once and then discarded at the end of each wearing period on a daily basis. The patient should start the next wearing period with fresh lenses. The lenses should not be cleaned or disinfected and should be discarded after a single use.
{6}------------------------------------------------
## 6. Performance Data
#### Non-Clinical Data
A series of in-vitro and in-vivo preclinical tests were performed to assess the safety and effectiveness of the Miru 1day UpSide (midafilcon A). All tests were conducted in accordance with the GLP regulation (21 CFR Part 58) or according to valid scientific protocols.
The results of the non-clinical testing and evaluation demonstrate that the lens material/extracts are non-toxic, non-irritating and non-sensitizing under the experimental conditions, and the material properties are consistent with the predicate lens.
#### Clinical Data
A three-month randomized controlled clinical study was completed to evaluate the safety and efficacy of the Miru 1day UpSide (midafilcon A) contact lenses for daily wear, single use.
The study evaluated 148 male and female subjects who were randomized and dispensed lenses in a 1:1 ratio into the test and control groups. The efficacy outcome measures were lens visual acuity comparisons between the test and the control contact lenses. The safety outcome measures included adverse event rates, biomicroscopy and subjective acceptance.
The test contact lens was found to be substantially equivalent to the control contact lens for safety and efficacy.
#### Conclusion
Based upon the data presented, the Miru 1day UpSide (midafilcon A) is as safe, as effective and performs as well as the predicate devices.
{7}------------------------------------------------
## 7. Substantial equivalence
Menicon Co., Ltd.
The claim of substantial equivalence to the previously cleared devices is supported by the following tables of Comparison of Characteristics.
| | Miru 1day UpSide | Visco Si-Hy | MyDay |
|-----------------------------------------------------|---------------------------------------------|-----------------------------------------------|-------------------------------------------------------|
| 510(k) | K193399 | K160344 | K131378 |
| USAN | midafilcon A | olifilcon B | stenfilcon A |
| Product Code | LPL, MVN | LPL, MVN | LPL, MVN |
| Modality | Daily wear,<br>Single Use | Daily wear,<br>Single Use | Daily wear,<br>Single Use |
| FDA Lens Group | SiHy class 5 | SiHy class 5 | SiHy class 5 |
| Device Classification | II | II | II |
| Manufacturing Method | Cast Molded | Cast Molded | Cast Molded |
| Sterilization | Moist Heat | Moist Heat | Moist Heat |
| Packaging | Blister Pack | Blister Pack | Blister Pack |
| Visibility Tint | Reactive Blue #246<br>Reactive Blue #247 | Reactive Blue #19 | Reactive Blue #246 |
| Light Transmittance | > 92 % | 94% | 96% |
| Water Content | 54% (USAN specification) | 47% | 54% |
| Refractive Index | 1.403 | 1.410 | 1.401 |
| Dk<br>× 10-11 [(cm²/sec) × (mL O₂)/(mL × mm Hg)] | 64 | 120 | 80 |
| Powers | - 20.00 to +20.00 D | - 20.00 to +20.00 D | - 20.00 to +20.00 D |
| Diameter | 13.0 to 18.0 mm | 13.0 to 15.0 mm | 13.0 to 15.5 mm |
| Base Curve | 8.0 to 9.0 mm | 8.0 to 9.2 mm | 8.4 and 8.7 mm |
| Optical Design | Spherical, Toric,<br>Multifocal | Spherical, Toric,<br>Multifocal | Aspherical, Toric,<br>Multifocal,<br>Multifocal Toric |
| Comparison of Characteristics (Indications for Use) | | | |
| Miru 1day UpSide | Visco Si-Hy | MyDay | |
| The Miru 1day UpSide (midafilcon A) | The Si-Hy Spherical Silicone Hydrogel Soft | MyDay ASPHERE Soft Contact lenses are | |
| SPHERICAL Lens is indicated for daily | Contact Lenses are indicated as daily wear | indicated for the correctionof ametropia | |
| wear single use only for the optical | single use soft contact lens for the | (myopia and hyperopia) in aphakic and | |
| correction of refractive ametropia (myopia | correction of refractive ametropia (myopia | non-aphakic persons with non-diseased | |
| and hyperopia) in aphakic and non-aphakic | and hyperopia) in phakic or aphakic | eyes in powers from -20.00D to +20.00D | |
| people with disease free eyes who may | persons with non-diseased eyes who exhibit | diopters. The lenses may be worn by | |
| have 1.50 diopter (D) or less of | refractive astigmatism of 1.00D or less | persons who exhibit astigmatism of -2.00 | |
| astigmatism. | where the astigmatism does not interfere | diopters or less that does not interfere with | |
| The Miru 1day UpSide (midafilcon A) | with visual acuity. | visual acuity. | |
| TORIC Lens is indicated for daily wear | The Si-Hy Toric Silicone Hydrogel Soft | Toric: MyDay (stenfilcon A) Toric Soft | |
| single use only for the optical correction of | Contact Lenses are indicated as daily wear | Contact lenses are indicated for the | |
| refractive ametropia (myopia or hyperopia | for the correction of ametropia (myopia and | correction of ametropia (myopia or | |
| with astigmatism) in aphakic and non- | hyperopia) with astigmatism in aphakic and | hyperopia with astigmatism) in aphakic and | |
| aphakic people with disease free eyes with | non-aphakic persons with non-diseased | non-aphakic persons with non-diseased | |
| 3.00 diopter (D) or less of refractive | eyes and whose powers are from -20.00 to | eyes in powers from -20.00 to +20.00 | |
| astigmatism. | +20.00 diopters and astigmatic corrections | diopters and astigmatic corrections from - | |
| The Miru 1day UpSide (midafilcon A) | are from -0.75 to -2.25 diopters. | 0.25 to -10.00 diopters. | |
| MULTIFOCAL Lens is indicated for daily | The Si-Hy Multifocal Silicone Hydrogel | Multifocal: MyDay (stenfilcon A) | |
| wear single use only for the optical | Soft Contact lenses are indicated as daily | MULTIFOCAL Soft Contact lenses are | |
| correction of refractive ametropia (myopia | wear for the correction of ametropia | indicated for the correction of refractive | |
| and hyperopia) and emmetropia with | (myopia and hyperopia) and emmetropia | ametropia (myopia and hyperopia) and | |
| presbyopia in aphakic and non-aphakic | with presbyopia in aphakic and non-aphakic | emmetropia with presbyopia in aphakic and | |
| people with disease free eyes who may | persons with non-diseased eyes and whose | non-aphakic persons with non-diseased | |
| require a reading addition of 3.00 diopter | powers are from -20.00 to +20.00 diopters | eyes. The lenses may be worn by persons | |
| (D) or less and who may have 1.50 diopter | with add powers from +0.75 to +2.75 | who exhibit astigmatism of -2.00 diopters | |
| (D) or less of astigmatism. | diopters. The lenses may be worn by | or less that does not interfere with visual | |
| The Miru 1day UpSide (midafilcon A) | persons who exhibit astigmatism of 1.00 | acuity. | |
| should only be worn once and then | diopters or less where the astigmatism does | Multifocal Toric: MyDay (stenfilcon A) | |
| discarded at the end of each wearing period | not interfere with visual acuity. | MULTIFOCAL TORIC Soft Contact lenses | |
| on a daily basis. The patient should start the | Eye care practitioners may prescribe the | are indicated for the optical correction of | |
| next wearing period with fresh lenses. The | lens for daily wear (disposable use) single | distance and near vision in presbyopic | |
| lenses should not be cleaned or disinfected | use. The lenses are to be discarded upon | phakic or aphakic persons with non- | |
| and should be discarded after a single use. | removal. Therefore, no cleaning or | diseased eyes who may have -10.00 | |
| | disinfecting is required. | diopters of astigmatism or less. | |
### Comparison of Characteristics (General Information)
Image /page/6/Picture/7 description: The image shows a logo that is divided into two halves. The top half is red, and the bottom half is green. A white line runs through the middle of the logo, creating a stylized letter 'f' shape. The logo is simple and uses contrasting colors to create a visually appealing design.
{8}------------------------------------------------
Image /page/7/Picture/2 description: The image shows the logo for Menicon Co., Ltd. The logo consists of a red and green circular design with a white stylized "M" in the center. To the right of the logo is the text "Menicon Co., Ltd." in a simple, sans-serif font.
#### C 10 10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.