CS Model+ is intended for use as a front-end software tool for management of orthodontic models, detailed analysis, treatment simulation and virtual appliance design options, including dental casts and orthodontic appliances. These applications are based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment. It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives. The use of CS Model+ requires the user to have the necessary training and domain knowledge in the practice of orthodontics.
Device Story
CS Model+ is a standalone software tool for orthodontic practitioners. It ingests 3D dental models from intraoral scanners or extraoral 3D imaging devices. The software provides tools for displaying, measuring, and analyzing 3D dentition models; performing treatment simulations; and designing virtual orthodontic appliances or dental casts. Outputs include digital 2D/3D views, orthodontic reports, and STL/PLY files for external fabrication of aligners or retainers. Used in clinical settings by orthodontists to plan treatment, monitor progress, and evaluate outcomes. The software automates detection of teeth cervical margins, numbering, and orientation, and generates virtual setups for aligner treatment. It assists clinicians in decision-making by providing quantitative analysis and visual simulation of treatment objectives.
Clinical Evidence
No clinical data; bench testing only. Software verification and validation were performed in accordance with FDA guidance to ensure stability and performance.
Technological Characteristics
Standalone software for Windows 10 64-bit. Requires 8 GB RAM, 1 GB video memory, and 10 GB HDD. Supports DICOM, PNG, JPG, and STL file formats. Features include 3D model visualization, automated tooth margin/numbering detection, measurement tools, and virtual setup generation. Connectivity via internet for data management.
Indications for Use
Indicated for orthodontic patients requiring management of dental models, analysis, treatment simulation, and virtual appliance design. Requires users to have orthodontic training and domain knowledge.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 2, 2020
Trophy Marie-Pierre Labat-Camy Global Regulatory Affairs Senior Manager 4 Rue F. Pelloutier Croissy-Beaubourg, Marne La Vallee Cedex 2, 77435 France
Re: K193390
Trade/Device Name: CS Model+ Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: PNN Dated: March 4, 2020 Received: March 6, 2020
Dear Marie-Pierre Labat-Camy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K193390
Device Name CS Model+
#### Indications for Use (Describe)
CS Model+ is intended for use as a front-end software tool for management of orthodontic models, detailed analysis, treatment simulation and virtual appliance design options, including dental casts and orthodontic appliances. These applications are based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment. It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives.
The use of CS Model+ requires the user to have the necessary training and domain knowledge in the practice of orthodontics.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(K) SUMMARY
(As per 21 CFR 807.92)
# 1) Date 510(k) Summary prepared
April 2nd, 2020
# 2) Contact information
# a) Submitter information
Contact: Ms. Marie-Pierre LABAT-CAMY Phone: +33 (0) 1 64 80 85 26 Address: Trophy 4 Rue F. Pelloutier - Croissy-Beaubourg 77435 Marne La Vallée, Cedex 2 France
# b) United Stated Sales Representative (U.S. Designated Agent)
Contact: Mr. Sonny T. Nguyen Phone: +1 (470) 481-4619 Address: Carestream Dental LLC 3625 Cumberland Boulevard, Suite 700, Atlanta, GA USA 30339
### 3) Device Identification
Trade/Device Name: CS Model+ Regulation Number: 21 CFR 872.5470 Common Name: Orthodontic Plastic Brackets Regulatory Class: II Product code: PNN
# 4) Primary Predicate
We consider the CS Model+ to be similar in design, composition and function to the following device, introduced into commercial distribution after May 28, 1976:
Trade/Device Name: Ortho System™ 510(k) Number: K171634 510(k) Submitter/holder: 3Shape A/S
Ortho System™ has been found to be substantially equivalent through the 510(k) premarket notification process.
{5}------------------------------------------------
# Traditional 510(k) Submission for CS Model+
# 5) Device description
CS Model+ is a software that allows practitioners to apply software imaging tools for management of orthodontic models, detailed analysis, treatment simulation and virtual appliance design. CS Model+ allows design of virtual models, also known as virtual dental casts. Those can be exported as STL or PLY files which may be used for fabrication of orthodontic models (dental casts). Then, which may be used to fabricate sequential aligner trays or retainers. The production of physical models and the other fabrication processes, such as thermoforming of aligner, is out of scope of Carestream Dental's manufacturing process. Those fabrication processes are done by the user and is under their sole control and responsibility.
It is the responsibility of the user to ensure:
- that all production machines and materials used are suitable for producing models or । appliances for orthodontic purposes.
- that all production machines and materials used are FDA cleared. -
CS Model+ Indication for Use is the following:
CS Model+ is intended for use as a front-end software tool for management of orthodontic models, detailed analysis, treatment simulation and virtual appliance design options, including dental casts and orthodontic appliances. These applications are based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment. It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives.
The use of CS Model+ requires the user to have the necessary training and domain knowledge in the practice of orthodontics.
CS Model+ is intended to be used with the following Carestream Dental's digital imaging devices including:
- -Intraoral scanners;
- -Extraoral devices with 3D object modality.
CS Model+ allows healthcare professionals to do the following:
- -Display digital 2D or 3D views of digital 3D models;
- Adjust the color and opacity of digital 3D models; -
- -Make indicative measurements of digital 3D models;
- Print and export images of digital 3D models; -
- -In a digital model, automatically detect the teeth cervical margins, teeth numbering and mesio distal orientations;
- -Generate an orthodontic report;
- Create virtual setups; -
- -Automatically generate virtual setups suitable for aligners treatment;
- Make automatic and customizable export of virtual models; -
- -Manage the intermediate steps;
- -Make refinement during treatment.
{6}------------------------------------------------
# 6) Indications for Use
CS Model+ is intended for use as a front-end software tool for management of orthodontic models, detailed analysis, treatment simulation and virtual appliance design options, including dental casts and orthodontic appliances. These applications are based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment. It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives.
The use of CS Model+ requires the user to have the necessary training and domain knowledge in the practice of orthodontics.
### Comparison of Indications for Use
The indication for use for the CS Model+ is similar to the primary predicate device Ortho System™ (K171634); the differences do not alter the intended use of the device. Therefore, the CS Model+ software can be considered substantially equivalent to its predicate device.
# 7) Substantial Equivalence
### a) Summary of the technological characteristics
CS Model+ is a standalone software which has the following PC/laptop hardware requirements:
| PC/ Laptop hardware requirements comparison | | |
|--------------------------------------------------------------------|-----------------------------|-----------------------------|
| Comparison Areas | CS Model+ | Ortho System™ |
| OS | Windows 10 64-bit | Windows 7, 8 or 10 64-bit |
| RAM | 8 GB | 8 GB |
| Monitor Resolution | 1024x768 minimum | 1280x800 or similar |
| Video Card Memory | 1 GB | 1 GB |
| Available HDD Space | 10 GB | 250 GB |
| CPU | 2,4 GHz Intel Duo Core | Intel Core i5 or equivalent |
| Network | Network Internet connection | Network Internet connection |
| Mouse | With the wheel button | With the wheel button |
| Intended use and technical characteristics comparison | | |
| Comparison Areas | CS Model+ | Ortho System™ |
| Destination | | |
| Supported anatomic areas | Maxilla / Mandible | Maxilla / Mandible |
| Intended use | | |
| Managing patient and case base data | Yes | Yes |
| Collection of study material | Yes | Yes |
| Alignment of study material | Yes | Yes |
| Measuring study material | Yes | Yes |
| Analyzing study material | Yes | Yes |
| Treatment simulation | Yes | Yes |
| Virtual appliance design | Yes | Yes |
| Supported PC formats | | |
| Operating System | Windows | Windows |
| Managing patient and case base data | | |
| Creating, editing, deleting and copying patient data | Yes | Yes |
| Creating, editing, deleting and copying case data | Yes | Yes |
| Collection of study material /<br>Acquisition Data Formats Support | | |
| Standalone software module | Yes | Yes |
| Surface scan from intra-oral scanner | Yes | Yes |
| Image data | DICOM, PNG, JPG, STL | DICOM, PNG, JPG, STL |
| Alignment of study material | | |
| Alignment of 2D overlays (e.g. ideal arch) | Yes | Yes |
| Measuring study material | | |
| 2D measurement toolbox | Yes | Yes |
| 3D measurement toolbox | Yes | Yes |
| Analyzing study material | | |
| Arch shape | Yes | Yes |
| Tooth width | Yes | Yes |
| Bolton | Yes | Yes |
| Space analysis | Yes | Yes |
| Overjet/overbite | Yes | Yes |
| Occlusion map | Yes | Yes |
| Treatment simulation | | |
| 3D simulation | Yes | Yes |
| Virtual appliance design | | |
| Orthodontic appliance virtual preparation | Yes | Yes |
| Orthodontic appliance design | Yes | Yes |
{7}------------------------------------------------
# Traditional 510(k) Submission for CS Model+
CS Model+ has the same intended use and technical characteristics as Ortho System™ (K171634) from 3Shape A/S:
{8}------------------------------------------------
# b) Nonclinical testing
Prior to release, software verification and validation testing of the CS Model+ have been performed in accordance with the FDA Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (Issued on May 11, 2005). Testing confirmed that CS Model+ is stable and operating as designed.
The risk management procedures have been applied to the development of the software and the risks identified have been mitigated and reduced to acceptable levels.
All test results have been reviewed and approved, showing the CS Model+ software to be substantially equivalent to the primary predicate Ortho System™ (K171634).
# c) Clinical testing
Clinical testing is not a requirement and has not been performed.
# 8) Conclusion
Based on a comparison of intended use, indications, principle of operations, features and technical data, and the test results, CS Model+ is found to be substantially equivalent to the primary predicate Ortho System™ (K171634).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.