Wrist Electronic Sphygmomanometer is non-invasive blood measurement of arterial blood pressure values in adults. It is intended to measure the diastolic, systolic blood pressures and pulse rate through an inflatable cuff wrapped around the wrist. It can be used by medical professionals or at home. The cuff circumference is limited to 13.5-19.5 cm.
Device Story
Battery-powered automatic wrist-worn sphygmomanometer; uses oscillometric technique to measure systolic/diastolic blood pressure and pulse rate. Device inflates cuff to 180-230mmHg; deflates at 3-6 mmHg/sec. Transducer measures cuff pressure and minute oscillations. Algorithm detects pulsations; performs peak matching to reject artifacts from patient movement. Systolic pressure identified at 0.45x reference amplitude (pre-peak); diastolic at 0.75x reference amplitude (post-peak). Results displayed on LCD. Used by clinicians or patients at home for blood pressure monitoring; aids in hypertension management and cardiovascular health assessment.
Clinical Evidence
Bench testing only. Evaluated per IEC 60601-1, IEC 60601-1-11, IEC 80601-2-30 (electrical safety), IEC 60601-1-2 (EMC), IEC 60601-1-6 (usability), AAMI/ANSI/ISO 81060-2 (performance), IEC 62133 (battery), and ISO 10993-5/10 (biocompatibility). Software verified and validated per FDA guidance.
Technological Characteristics
Oscillometric non-invasive blood pressure measurement. Powered by 2x AAA alkaline batteries or 3.7V 400mAh Li-ion battery. Wrist-worn form factor with LCD. Cuff circumference 13.5-19.5 cm. Pressure range 0-299 mmHg; pulse range 40-199 bpm. Accuracy: ±3 mmHg (pressure), ±5% (pulse). Biocompatible materials per ISO 10993.
Indications for Use
Indicated for non-invasive measurement of arterial blood pressure (systolic/diastolic) and pulse rate in adults. Cuff circumference 13.5-19.5 cm. Suitable for use by medical professionals or home users.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
Med-link Wrist Digital Blood Pressure Monitor (K181154)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 27, 2021
Shenzhen Changkun Technology Co., Ltd. % Cassie Lee Manager Guangzhou GLOMED Biological Technology Co., Ltd. 2231, Building 1, Rui Feng Center, Kaichuang Road, Huangpu District Guangzhou, 510000 China
Re: K193386
Trade/Device Name: Wrist Electronic Sphygmomanometer Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: December 28, 2020 Received: December 31, 2020
Dear Cassie Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for LT Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193386
#### Device Name
Wrist Electronic Sphygmomanometer
#### Indications for Use (Describe)
Wrist Electronic Sphygmomanometer is non-invasive blood measurement of arterial blood pressure values in adults. It is intended to measure the diastolic, systolic blood pressures and pulse rate through an inflatable cuff wrapped around the wrist. It can be used by medical professionals or at home. The cuff circumference is limited to 13.5-19.5 cm.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
|X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Sponsor: | ShenZhen Changkun Technology CO., LTD. |
|-----------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject Device: | Wrist Electronic Sphygmomanometer, Models: CK-101, CK-101S, CK-102S, CK-W118, CK-W132, CK-W133, CK-W135, CK-W175, CK-W176, CK-W355, CK-W356, CK-W358 |
| Document Name: | FDA 510(k) Submission Report |
# 510(k) Summary
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
# 1. Submitter's Information
510(k) Owner's Name: ShenZhen Changkun Technology CO., LTD. Establishment Registration Number: Applying Address: 801, 3 floor 4floor 5floor 7floor, B building, NO.69, zhenbi road, biling community, biling street, pingshan district, Shenzhen city, Guangdong, China Tel: +86-0755-29100487 Fax: +86-0755-23463480 Contact Person (including title): Steve Li (Manger Representative) E-mail: changkunkj@163.com
### Application Correspondent:
Contact Person: Cassie Lee Guangzhou GLOMED Biological Technology Co., Ltd. Address: 2231, Building 1, Rui Feng Center, Kaichuang Road, Huangpu District, Guangzhou, Guangdong, China Tel: +86 20 8266 2446 Email: requlatory(@glomed-info.com
# 2. Subject Device Information
Type of 510(k): Traditional Common Name: Noninvasive blood pressure measurement systems Classification Name: system, measurement, blood-pressure, non-invasive Trade Name: Wrist Electronic Sphygmomanometer Model Name: CK-101, CK-102S, CK-W118, CK-W132, CK-W133, CK-W135, CK-W135, CK-W175, CK-W176, CK-W355, CK-W356, CK-W358 Review Panel: Cardiovascular Product Code: DXN Regulation Number: 870.1130 Regulatory Class: 2
#### 3. Predicate Device Information
| Sponsor | Shenzhen Med-link Electronics Tech Co., Ltd. | |
|---------------------|----------------------------------------------|----------------------------------------------------------------|
| Device<br>and Model | Name | Med-link Wrist Digital Blood Pressure Monitor<br>Model: ESM101 |
| 510(k) Number | K181154 | |
{4}------------------------------------------------
| Sponsor: | ShenZhen Changkun Technology CO., LTD. |
|-----------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject Device: | Wrist Electronic Sphygmomanometer, Models: CK-101, CK-101S, CK-102S, CK-W118, CK-W132, CK-W133, CK-W135, CK-W175, CK-W176, CK-W355, CK-W356, CK-W358 |
| Document Name: | FDA 510(k) Submission Report |
| Product Code | DXN |
|----------------------|----------|
| Regulation<br>Number | 870.1130 |
| Regulation<br>Class | II |
### 4. Device Description
The Wrist Electronic Sphygmomanometer is a battery driven automatic non-invasive blood pressure meter. It can automatically conduct the inflation and measurement, which can measure systolic and diastolic blood pressure as well as the pulse rate of adult at wrist within its claimed range and accuracy via the oscillometric technique.
The device uses the oscillometric method to determine blood pressure. This is accomplished by means of a sensitive transducer, which measures cuff pressure and minute pressure oscillations within the cuff. The first determination sequence initially pumps up to a cuff pressure of about 180-230mmHg for adults. After inflating the cuff, the Wrist Electronic Sphygmomanometer begins to deflate it at a rate of 3-6 mmHq/sec, and measures systolic, mean, and diastolic pressure. The Wrist Electronic Sphygmomanometer deflates the cuff one step each time it detects two pulsations of relatively equal amplitude ("peak matching"). The time between deflation steps depends on the frequency of these matched pulses (pulse rate of the patient). However, if the Wrist Electronic Sphygmomanometer is unable to find any pulse within several seconds, it will deflate to the process of finding two matched pulses at each step provides artifact rejection due to patient movement and greatly enhances accuracy.
The device will obtain the pressure point, which is before the maximum peak and whose amplitude of peak is 0.45 times the reference amplitude, is determined to be the systolic pressure; and the pressure point, which is after the maximum peak and whose amplitude of peak is 0.75 times the reference amplitude, is determine to be the diastolic. When the diastolic pressure has been determined, the NIBP Meter finishes deflating the cuff and updates the screen.
# 5. Intended Use / Indications for Use
Wrist Electronic Sphygmomanometer is non-invasive blood measurement of arterial blood pressure values in adults. It is intended to measure the diastolic, systolic blood pressures and pulse rate through an inflatable cuff wrapped around the wrist. It can be used by medical professionals or at home. The cuff circumference is limited to 13.5-19.5 cm.
#### 6. Test Summary
Wrist Electronic Sphygmomanometer has been evaluated the safety and performance by lab bench testing as following:
- � Electrical safety test according to IEC 60601-1, IEC 60601-1-11 and IEC 80601-2-30 standards
- � Electromagnetic compatibility test according to IEC 60601-1-2 standard
- � Electromagnetic compatibility test according to IEC 60601-1-6 standard
- � Performance according to AAMI / ANSI / ISO 81060-2 standard
- � Lithium battery report in accordance with IEC62133
- � Biocompatibility test according to ISO 10993-5 and ISO 10993-10 standards
- � Usability test according to IEC62366 standard
- � Software verification and validation test according to the requirements of the FDA "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"
{5}------------------------------------------------
| Sponsor: | ShenZhen Changkun Technology CO., LTD. |
|-----------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject Device: | Wrist Electronic Sphygmomanometer, Models: CK-101, CK-101S, CK-102S, CK-W118, CK-W132, CK-W133, CK-W135, CK-W175, CK-W176, CK-W355, CK-W356, CK-W358 |
| Document Name: | FDA 510(k) Submission Report |
# 7. Comparison to predicate device and conclusion
The technological characteristics, features, specifications, materials, mode of operation, and intended use of Wrist Electronic Sphygmomanometer is substantially equivalent to the predicate devices quoted above. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.
| Elements of<br>Comparison | Subject Device | Predicate Device | Verdict |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Company | ShenZhen Changkun<br>Technology CO., LTD. | Shenzhen Med-link Electronics Tech<br>Co., Ltd. | -- |
| Product Name | Wrist Electronic<br>Sphygmomanometer | Med-link Wrist Digital Blood<br>Pressure Monitor | -- |
| Model Name | CK-101, CK-101S, CK-102S,<br>CK-W118, CK-W132, CK-W133,<br>CK-W135, CK-W175, CK-W176,<br>CK-W355, CK-W356, CK-W358 | ESM101 | -- |
| Classification<br>Name | System, Measurement, Blood-<br>pressure, Non-invasive | System, Measurement, Blood-<br>pressure, Non-invasive | SE |
| 510(k) Number | Applying | K181154 | -- |
| Product Code | DXN | DXN | SE |
| Intended Use and Indications for Use | | | |
| Intended Use /<br>Indications for<br>Use | Wrist Electronic<br>Sphygmomanometer is non-<br>invasive blood measurement of<br>arterial blood pressure values in<br>adults. It is intended to measure<br>the diastolic, systolic blood<br>pressures and pulse rate through<br>an inflatable cuff wrapped<br>around the wrist. It can be used<br>by medical professionals or at<br>home. The cuff circumference is<br>limited to 13.5-19.5 cm. | Med-link Wrist Digital Blood<br>Pressure Monitor is non-invasive<br>blood measurement of arterial blood<br>pressure values in adults. It is<br>intended to measure the diastolic,<br>systolic blood pressures and pulse<br>rate through an inflatable cuff<br>wrapped around the wrist. It can be<br>used by medical professionals or at<br>home. The cuff circumference is<br>limited to 13.5-19.5 cm. | SE |
| Electrical requirement | | | |
| Power Supply | For model CK-101, CK-101S,<br>CK-102S, CK-W118, CK-W132,<br>CK-W133, CK-W135, CK-<br>W176: 2 x 1.5V "AAA" alkaline<br>batteries;<br>For model CK-W175, CK-W355,<br>CK-W356, CK-W358: 3.7V<br>400mAh Li-ion Battery | DC 3.0V(2xAAA alkaline battery) | SE<br>Note 1 |
| Performance specification | | | |
| Measurement<br>Site | Wrist | Wrist | SE |
{6}------------------------------------------------
| Sponsor: | ShenZhen Changkun Technology CO., LTD. |
|-----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject Device: | Wrist Electronic Sphygmomanometer, Models: CK-101, CK-101S, CK-102S, CK-<br>W118, CK-W132, CK-W133, CK-W135, CK-W175, CK-W176, CK-W355, CK-W356,<br>CK-W358 |
| Document Name: | FDA 510(k) Submission Report |
| Elements of<br>Comparison | Subject Device | Predicate Device | Verdict |
|-----------------------------------|---------------------------------------------------------------------------------------------|--------------------------------------------------------------|---------|
| Measuring | Pressure:0~299mmHg | Pressure:0~280mmHg | SE |
| range | Pulse 40~199 times/min | Pulse:40~199beats/minute | Note 2 |
| Pressure<br>resolution | 1 mmHg or 0.1kPa | 1 mmHg or 0.1kPa | SE |
| Measuring | Pressure: ± 3mmHg | Pressure: ± 3mmHg | SE |
| accuracy | Pulse: ±5% | Pulse: ±5% | SE |
| Cuff<br>Circumference | 13.5-19.5 cm | 13.5-19.5 cm | SE |
| Inflation and<br>Deflation | Automatic | Automatic | SE |
| Measuring<br>Method | Non-invasive Oscillometric | Non-invasive Oscillometric | SE |
| Patient<br>Population | Adult | Adult | SE |
| Display | LCD | LCD | SE |
| Operation | Temperature: +5 to +40 °C,<br>Humidity: 15 to 90%,<br>Atmospheric Pressure: 80 kPa~106 kPa | Temperature: +5°C~+40°C,<br>Humidity: 15%~85%RH | SE |
| condition | | | |
| Transport/stora<br>ge environment | Temperature: -25°C to +70°C;<br>Humidity: ≤93%R.H.;<br>Atmospheric Pressure: 70 kPa~106 kPa | Temperature: -20°C~+55°C<br>Humidity: ≤95%RH, non-condensing | SE |
| Safety | IEC 60601-1<br>IEC 60601-1-11<br>IEC 80601-2-30 | IEC 60601-1<br>IEC 80601-2-30 | SE |
| EMC | IEC 60601-1-2 | IEC 60601-1-2 | SE |
| Biocompatibility | ISO 10993 series | ISO 10993 series | SE |
#### Comparison
#### Note 1:
Although the "Power Supply" of the subject device is a little different from the predicate device, they all complied with the requirements of safety standards as IEC 60601-1, IEC 60601-1-2, IEC 80601-2-30 required. So this difference will not raise any safety or effectiveness issue.
# Note 2:
Although "Measuring scope" of the subject device are a little different from the predicate devices, but the difference of subject device is very similar with the K181154, and both they meet the requirements as safety and performance standards required. So these parameters' differences will not raise any safety or effectiveness issue.
#### Final Conclusion:
The subject device Non-invasive Sphygmomanometer (Wrist Electronic Sphygmomanometer) (Model:CK-101, CK-101S, CK-102S, CK-W118, CK-W132, CK-W133, CK-W135, CK-W175, CK-W176, CK-W355,
{7}------------------------------------------------
| Sponsor: | ShenZhen Changkun Technology CO., LTD. |
|-----------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject Device: | Wrist Electronic Sphygmomanometer, Models: CK-101, CK-101S, CK-102S, CK-W118, CK-W132, CK-W133, CK-W135, CK-W175, CK-W176, CK-W355, CK-W356, CK-W358 |
| Document Name: | FDA 510(k) Submission Report |
CK-W356, CK-W358) has all features of the predicate device. The few differences do not affect the safety and effectiveness of the subject device. Thus, the subject device is substantially equivalent to the predicate device.
# 8. Date of the summary prepared: December 28, 2020
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.