iCTmotor (WL-1) is intended for use in dental surgery and implantology. The main control unit is designed to operate a specific dental micro motor that drives dental handpieces to cut hard and soft tissues in the mouth and screw dental implants. iCTmotor (WL-1) is compatible with a handpiece equipped with connection according to ISO 3964.
Device Story
iCTmotor (WL-1) is a software-driven dental micromotor control system for implantology and surgery. Input: user-programmed parameters via wireless foot pedal. Transformation: main control unit regulates speed and torque of a micromotor to drive ISO 3964-compliant handpieces. Output: mechanical rotation for cutting tissue and implant placement. Used in dental clinics by dental professionals. Benefits: provides controlled, programmable motor operation for surgical procedures. Healthcare providers use the system to manage surgical parameters, directly affecting the precision of tissue cutting and implant insertion.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical bench testing, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), software validation, reprocessing validation, and performance testing (ISO 14457:2012) for speed, torque, and irrigation.
Technological Characteristics
Main control unit with wireless foot pedal, micromotor, and cable. Materials: Stainless Steel (SUS 304). Power: 100-240V. Motor: Surgical collector type, 36 VDC, 400-40,000 rpm. Connectivity: Wireless foot pedal (2.405-2.480 GHz). Sterilization: Sterilizable components. Software: Embedded control logic.
Indications for Use
Indicated for use in dental surgery and implantology for cutting hard and soft oral tissues and screwing dental implants. Compatible with ISO 3964-compliant handpieces.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 4, 2020
Dentium Co., Ltd % Dave Kim Medical Device Regulatory Affairs Mtech Group 7707 Fannin St. Ste 200, V111 Houston, Texas 77054
Re: K193341
Trade/Device Name: iCTmotor (WL-1) Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece And Accessories Regulatory Class: Class I. reserved Product Code: EBW Dated: September 3, 2020 Received: September 3, 2020
Dear Dave Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For Srinivas "Nandu" Nandkumar Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193341
Device Name iCTmotor (WL-1)
### Indications for Use (Describe)
iCTmotor (WL-1) is intended for use in dental surgery and implantology. The main control unit is designed to operate a specific dental micro motor that drives dental handpieces to cut hard and soft tissues in the mouth and screw dental implants. iCTmotor (WL-1) is compatible with a handpiece equipped with connection according to ISO 3964.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div><span style="font-size: 10pt;"> <input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</span></div> |
|-------------------------------------------------------------------------------------------------------------------------|
| <div><span style="font-size: 10pt;"><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</span></div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K193341
This summary of 510(k) information is being submitted in accordance with requirements of 21 CFR Part 807.92.
#### Date 510k summary prepared: September 2, 2020
#### I. SUBMITTER
| Submitter's Name | Dentium Co., Ltd (ICT Branch) |
|------------------------|---------------------------------------------------------------------------------------------|
| Submitter's Address | 76, Changnyong-daero 256beon-gil, Yeongtong-gu<br>Suwon-si, Gyeonggi-do, Korea (Zip. 16229) |
| Submitter's Telephone | +82 (70) 7098-4027 |
| Contact person | Sangwoo Lee / RA Manager<br>swlee1@dentium.com |
| Official Correspondent | Dave Kim (davekim@mtech-inc.net) |
| Address | Mtech Group<br>7707 Fannin St. Ste 200-V111<br>Houston, TX 77054 |
| Telephone | +713-467-2607 |
#### II. DEVICE
| Trade/proprietary Name | iCTmotor (WL-1) |
|------------------------|---------------------------------------|
| Common Name | Controller, Foot, Handpieces and Cord |
| Regulation Name | Dental Handpieces and Accessories |
| Regulation Number | 21 CFR 872.4200 |
| Product Code | EBW |
| Regulatory Class | Class I |
#### PRIMARY PREDICATE DEVICE (K140308) III.
| Primary Manufacturer | Kaltenbach & Voigt GmbH |
|----------------------|---------------------------------------|
| Device Name | MASTERsurg / EXPERTsurg |
| Common Name | Controller, Foot, Handpieces and Cord |
| Regulation Name | Dental Handpieces and Accessories |
| Regulation Number | 21 CFR 872.4200 |
| Product Code | EBW, EGS |
| Regulatory Class | Class I |
{4}------------------------------------------------
#### IV. DEVICE DESCRIPTION
iCTmotor (WL-1) is a software based driving engine that controls the speed of a specific dental micromotor. This device is optimized for dental implant procedures and user programmable parameters operate and control a dental handpiece for dental implant surgery. iCTmotor (WL-1) consists of a main controller unit, a charger, a foot controller, micro motor, cable, a water holder, tube holder, and micro motor holder. The main control unit operates the speed and torque of a dental micromotor that drives dental handpiece to cut tissues in the mouth and to screw dental implants. The main control unit is operated via a wireless foot pedal. The holders are used for placement of a water bag, a micro motor and a handpiece. The power cord delivers electric power to the main control unit.
#### V. INDICATIONS FOR USE:
iCTmotor (WL-1) is intended for use in dental surgery and implantology. The main control unit is designed to operate a specific dental micro motor that drives dental handpieces to cut hard and soft tissues and screw dental implants. iCTmotor (WL-1) is compatible with a handpiece equipped with connection according to ISO 3964.
| Device name | iCTmotor (WL-1) | EXPERTsurg + INTRA LUXS600 LED | Remarks |
|-------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|
| 510K Number | K193341 | K140308 | |
| Main body image | Image: iCTmotor (WL-1) | Image: EXPERTsurg + INTRA LUXS600 LED | |
| | | | |
| Foot switch image | Image: Foot switch | Image: Wireless type | Wireless<br>type (same) |
| Micromotor | Image: Micromotor | Image: Micromotor | Similar |
| Indication for use | iCTmotor (WL-1) is intended<br>for use in dental surgery and<br>implantology. The main control<br>unit is designed to operate a<br>specific dental micro motor that<br>drives dental handpieces to cut<br>hard and soft tissues in the<br>mouth and screw dental<br>implants. iCTmotor (WL-1) is<br>compatible with a handpiece<br>equipped with connection<br>according to ISO 3964. | Unit: This KaVo product is intended<br>for surgery to expose and dissect oral<br>tissue structures or endodontic<br>treatments (e.g. periodontal gap,<br>gingiva, bone, aw, extractions and<br>implantations). Motor: The medical<br>device is intended to drive / operate<br>a dental handpiece / contra-angle<br>handpiece equipped with a<br>handpiece connection according to<br>ISO 3964. This medical device is an<br>electrical low-voltage motor for<br>dental purposes according to ISO<br>11498 type 2 and classified as a type<br>B application part. | Similar |
| Compliance to<br>Standards | Handpiece in compliance with<br>ISO 3964 | Handpiece in compliance with ISO | Same |
| Dimensions | 278 x 175 x 134 (mm) | 265 x 255 x 100 mm | Similar |
| Electrical<br>Specification | 100 – 240 V | 100 – 240 V | Same |
| Main Weight | 3kg | 2.0kg | Similar |
| Foot Weight | 0.7kg | 1.1kg | Similar |
| Motor | 135g | 125g | Similar |
| Max output power | Max 70W | Max 150W | Different |
| Motor Type | Surgical Motor (Collector) | Surgical Motor (Collector) | Same |
| Handpiece<br>Connection | INTRAmatic Coupling System<br>(ISO 3964) | INTRAmatic Coupling System<br>(ISO 3964) | Same |
| Electrical<br>Specification<br>Motor (Voltage) | 36 VDC | 22 V AC | Different |
| Electrical<br>Specification<br>Motor (20:1<br>Torque) | 70Ncm | 55 Ncm | Different |
| Electrical<br>Specification<br>Motor (Torque) | 2.3mNm | 5.5Ncm | Different |
| Micro motor LED<br>Input Voltage | DC 3.3V | DC 3.0-3.6V | Similar |
| Micro motor LED<br>Input<br>Current(Max) | 150mA | 150mA | Same |
| Electrical<br>Specification<br>Motor (Rotation) | Clockwise, Counter Clockwise | Clockwise, Counter Clockwise | Same |
| Electrical<br>Specification<br>Motor (Speed) | 400- 40,000rpm | 300 - 40,000 rpm | Similar |
| Pump Delivery<br>rate | 40-60ml/min | 30-110ml/min | Different |
| Handpiece<br>rotational speed<br>(20:1 ratio) | 20 ~ 2,000 rpm | 15 ~ 2,000 rpm | Similar |
| Wireless foot<br>Emitted power | Max 8.6dBm(e.i.r.p.) | Max 3dBm(e.i.r.p.) | Different |
| Wireless foot<br>Frequency bend | 2.405~2.480GHZ | 2.4GHZ | Similar |
| Length of motor<br>cable | 2m | 2m | Same |
| Operating mode | 3min ON/10min OFF | 30sec. of operation/9min.pause | Different |
| Handpiece gearing 20:1<br>ratio | 20:1 | | Same |
| Sterilization | Sterilizable | Sterilizable | Same |
| Storage<br>Ambient<br>temperature | 0°C-+60°C | -20°C-+50°C | Different |
| Operating<br>Ambient<br>temperature | 10°C-+35°C | 10°C-+35°C | Same |
| Storage<br>Relative humidity | Max 90% | 5%-95% | Similar |
| Operating<br>Relative humidity | Max 80% | 15%-80% | Similar |
| Atmospheric<br>pressure Range | 700hPa-1,060hpa | 700hPa-1,060hpa | Same |
| Foot switch<br>control | - Speed key<br>- Program key<br>- Motor | - Direction of motor rotation key<br>- Speed key<br>- Program key<br>- Pump key | Similar |
| Electrical<br>Specification<br>Motor (Speed) | Up to 40,000rpm | Up to 40,000 rpm | Same |
| Handpiece<br>rotational speed | 20 ~ 2,000 rpm | 15 ~ 2,000 rpm | Similar |
| Handpiece gearing | 20:1 | 20:1 | Same |
| Sterilization | Sterilizable | Sterilizable | Same |
#### VI. PREDICATE COMPARISON
{5}------------------------------------------------
{6}------------------------------------------------
K193341
{7}------------------------------------------------
### K193341
The proposed iCTmotor (WL-1) is similar in the indications for use, overall design and function to the EXPERTsurg + INTRA LUXS600 LED (K140308). Both the proposed subject device and the primary predicate device consist of a control unit with a micromotor, connecting cable and a wireless foot pedal.
The subject device does not include a straight or contra angle handpiece. An irrigation tube is not included neither.
ICTmotor(WL-1) motor is compatible with any contra-angle handpiece in compliance with EN ISO 3964:2016 and ICTmotor (WL-1) motor is also compatible with irrigation tubes in compliance with EN ISO 7405
The differences do not render the device NSE because the performance tests demonstrate that the differences in technological characteristics do not raise new safety concerns.
{8}------------------------------------------------
#### VII. SUMMARY OF NON-CLINICAL TESTS
iCTmotor (WL-1) complies with voluntary standards for electrical safety, EMC testing, and use in the dental clinic. The following data were provided to support the substantial equivalence determination:
Electrical Safety:
Testing was conducted in accordance with IEC 60601-1:2012 Medical electrical equipment, Part 1: General requirements for basic safety and essential performance
Electromagnetic Compatibility:
Testing was conducted in accordance with IEC 60601-1-2:2015 Medical Electrical Equipment -Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests
### Software:
Software verification and validation testing as recommended in FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." (May 11, 2005)
### Reprocessing:
Cleaning and sterilization validation test report subject to "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling" (March 17, 2015) are provided.
### Biocompatibility:
The subject device is not in direct contact with a patient's skin. The user is instructed to wear gloves while operating the iCTmotor (WL-1). Therefore, biocompatibility test is not considered.
The manufacturer also provided a material/chemical composition and manufacturing based rationale as to how your final finished device is substantially equivalent to the identified primary predicate device:
"The [Stainless Steel: SUS 304] of iCTmotor (K193341) in its final finished form is identical to the [Stainless Steel: SUS 304] of other micro motor manufactured by Maxon motor ag and marketed in the US in formulation, processing, sterilization, and geometry, and no other chemicals have been added (e.g., plasticizers, fillers, additives, cleaning agents, mold release agents)."
" iCTmotor micro motor in its final finished form is identical to other micro motor manufactured by Maxon motor ag and marketed in the US as private labels in formulation, processing, sterilization, and geometry and no other chemicals have been added (e.g., plasticizers, fillers, additives, cleaning agents, mold release agents)."
{9}------------------------------------------------
### K193341
Performance testing was conducted cording to ISO 14457:2012 to show that the device meets its design requirements and performs as intended. The specifications were met for:
- Rotating speed of micromotor
- . Torque of micromotor
- Stop of micromotor
- . Rotating direction of micro motor
- Irrigation amount
## VIII. SUMMARY OF CLINICAL TESTS
Clinical testing was not required to demonstrate the substantial equivalence of iCTmotor (WL-1) to its predicate device.
#### CONCLUSIONS IX.
Based on the information above, iCTmotor (WL-1) is substantially equivalent to the predicate device. Based on the performance testing results including the nonclinical tests, the subject device is as safe, as effective, and performs as well as or better than the legally marketed device predicate (21 CFR 807.92(b)(3))
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.