← Product Code [EMA](/productcode/EMA) · K193260

# U-Cem Premium & MAZIC Cem (K193260)

_Vericom Co., Ltd. · EMA · Aug 21, 2020 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K193260

## Device Facts

- **Applicant:** Vericom Co., Ltd.
- **Product Code:** [EMA](/productcode/EMA.md)
- **Decision Date:** Aug 21, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3275
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

1. Luting metal crowns, bridge, inlay and including porcelain-fused-to-metal and composite-to-metal varieties 2. Luting resin crowns, bridges, inlays, onlays and veneers 3. Luting metal or non-metal/fiber posts 4. Luting orthodontic appliances 5. Luting porcelain inlays, onlays, crowns and veneers (includes alumina and zirconia)

## Device Story

U-Cem Premium and MAZIC Cem are dual-cured, self-adhesive, fluoride-releasing, radiopaque resin cements. Used by dental professionals in clinical settings for cementation of various dental restorations including crowns, bridges, inlays, onlays, veneers, posts, and orthodontic appliances. Available in automix and double dispenser delivery systems; offered in three shades (Clear, Universal, Opaque). Device functions by chemically bonding to tooth structure and restorative materials without requiring a separate bonding agent. Healthcare providers apply the paste to the restoration, which then sets to secure the prosthesis. Faster setting time compared to predicate; provides mechanical retention and esthetic options for dental procedures.

## Clinical Evidence

Bench testing only. Performance evaluated per ISO 4049 (polymer-based restorative materials), ISO 9917-1 (water-based cements), and ISO/TS 11405 (adhesion to tooth structure). Biocompatibility testing conducted per ISO 10993 series (parts 1, 3, 5, 10, 11). Results demonstrate physical properties (compressive strength 236.35 MPa, flexural strength 135.87 MPa, film thickness 11 μm) and shear bond strengths comparable or superior to the predicate device. No clinical data provided.

## Technological Characteristics

Dual-cured, methacrylate-based resin cement with amorphous silica filler. Radiopaque. Available in automix and double dispenser delivery systems. Conforms to ISO 4049. Biocompatible per ISO 10993. Storage at 2-8°C.

## Regulatory Identification

Zinc oxide-eugenol is a device composed of zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp. Dental cement other than zinc oxide-eugenol is a device composed of various materials other than zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.

## Predicate Devices

- BisCem ([K082449](/device/K082449.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 21, 2020

Vericom Co., Ltd. % Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA, Inc 1150 Roosevelt, STE 200 Irvine, California 92620

Re: K193260

Trade/Device Name: U-Cem Premium & MAZIC Cem Regulation Number: 21 CFR 872.3275 Regulation Name: Dental Cement Regulatory Class: Class II Product Code: EMA Dated: July 21, 2020 Received: July 24, 2020

Dear Priscilla Chung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic. Anesthesia. Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K193260

Device Name

U-Cem™ Premium & MAZIC® Cem

#### Indications for Use (Describe)

- 1. Luting metal crowns, bridge, inlay and including porcelain-fused-to-metal and composite-to-metal varieties
- 2. Luting resin crowns, bridges, inlays, onlays and veneers
- 3. Luting metal or non-metal/fiber posts
- 4. Luting orthodontic appliances
- 5. Luting porcelain inlays, onlays, crowns and veneers (includes alumina and zirconia)

| Type of Use (Select one or both, as applicable)                                                |                                                                                               |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: overline;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary (K193260)

This summary of 510(k) is being submitted in accordance with requirements of 21 CFR Part 807.92.

Date: July 16, 2020

# 1. 510K Applicant / Submitter:

Vericom Co., Ltd. 48.Toegyegongdan 1-Gil, Chuncheon-Si, Gangwon-Do, 200-944, Republic of Korea Tel: +82-31-441-2881 Fax: +82-31-441-2883

## 2. Submission Contact Person

LK Consulting Group USA, Inc. 1150 Roosevelt, STE 200, Irvine CA 92620 Priscilla Juhee Chung Phone: 714.202.5789 Fax: 714-409-3357 Email: juhee.c(@lkconsultinggroup.com

## 3. Device

- Proprietary Name: U-Cem™ Premium & MAZIC® Cem .
- Common Name: Luting Cement .
- Classification: Dental Cement, Class II (21 CFR 872.3275) .
- . Product Code: EMA

# 4. Predicate Device

BisCem (K082449) by Bisco, Inc.

# 5. Description:

The U-Cem™ Premium and the MAZIC® Cem are dual-cured self-adhesive resin cement. The U-Cem™ Premium and the MAZIC® Cem are the same product yet with different label design for different volume packages as shown below. The brand name, MAZIC® Cem, is specifically for 8.5g.

It is fluoride-releasing paste type, and has radiopaque. U-Cem™ Premium is available in both Automix and Double Dispenser types, and offers three different shades: Clear, Universal, and Opaque for esthetics.

{4}------------------------------------------------

It is used for cementation of ceramics, metals or composite inlays, onlays, crowns, bridges, posts and screws.

# 8. Indications for Use

1. Luting metal crowns, bridge, inlay and including porcelain-fused-to-metal and compositeto-metal varieties

- 2. Luting resin crowns, bridges, inlays, onlays and veneers
- 3. Luting metal or non-metal/fiber posts
- 4. Luting orthodontic appliances
- 5. Luting porcelain inlays, onlays, crowns and veneers (includes alumina and zirconia)

# 9. Substantial Equivalence Discussion:

The U-Cem™ Premium/ the MAZIC® Cem has the same intended use and the principle of operation, and also has similar physical properties demonstrating comparable performance specifications to the predicate device: BisCem (K082449) by Bisco, Inc.

The Indications for Use statement of the subject device is similar to that of the predicate device, but it is described in a slightly different way. However, fundamentally the predicate device's indications cover the subject device's claims.

One of the differences is the subject device has an additional delivery system which is the Double Dispenser type. This Double Dispenser has the same mixing ration as the auto mix type so this difference does not raise a question in substantial equivalence.

Another difference is that the subject device has three shades, whereas, the BisCem has two shades, and some raw materials used might be different. However, bench performance and biocompatibility testing performed demonstrates that these differences would not raise a new question in substantial equivalence. For the tests that were performed according to the standards, both devices satisfied the standards' requirements. For the testing that does not have an international standard, the test results of the subject device were substantially equivalent to the predicate device.

The Physical and Performance properties are also very similar except setting time. The subject device shows faster setting time which is 2 minutes and 12 seconds. Both devices meet the ISO 4049 requirement for the setting time, and it does not a raise an issue to support substantial equivalence.

Based on the information provided herein, it is concluded that the U-Cem™ Premium/ the MAZIC® Cem is substantially equivalent to the predicate device.

|              | Subject Device    | Predicate Device |
|--------------|-------------------|------------------|
| Device Name  | U-Cem™ Premium    | BisCem           |
| 510k #       | K193260           | K082449          |
| Manufacturer | VERICOM CO., LTD. | Bisco, Inc.      |

{5}------------------------------------------------

| Indications for use                     |                                                              | 1. Luting metal crowns, bridge, inlay and including porcelain-fused-to-metal and composite-to-metal varieties<br>2. Luting resin crowns, bridges, inlays, onlays and veneers<br>3. Luting metal or non-metal/fiber posts<br>4. Luting orthodontic appliances<br>5. Luting porcelain inlays, onlays, crowns and veneers (includes alumina and zirconia) | 1. Luting metal crowns, bridge, inlay and including porcelain-fused-to-metal and composite-to-metal varieties<br>2. Luting resin crowns, bridges, inlays, onlays and veneers<br>3. Luting metal or non-metal/fiber posts<br>4. Luting orthodontic appliances<br>5. Luting porcelain inlays, onlays, crowns and veneers (includes alumina and zirconia) |                                        |
|-----------------------------------------|--------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|
| Principle of operation                  |                                                              | Dual cured                                                                                                                                                                                                                                                                                                                                             | Dual cured                                                                                                                                                                                                                                                                                                                                             |                                        |
| Physical &<br>Performance<br>properties | Working Time<br>(at 60sec. after<br>completion of<br>mixing) | Physically homogeneous and forming the thin layer                                                                                                                                                                                                                                                                                                      | Physically homogeneous and forming the thin layer                                                                                                                                                                                                                                                                                                      |                                        |
|                                         | Setting time                                                 | 2 m 12s                                                                                                                                                                                                                                                                                                                                                | 9 m 08 s                                                                                                                                                                                                                                                                                                                                               |                                        |
|                                         | Compressive strength                                         | $236.35MPa$                                                                                                                                                                                                                                                                                                                                            | $189.15MPa$                                                                                                                                                                                                                                                                                                                                            |                                        |
|                                         | Flexural strength                                            | $135.87MPa$                                                                                                                                                                                                                                                                                                                                            | $101.93 MPa$                                                                                                                                                                                                                                                                                                                                           |                                        |
|                                         | Film thickness                                               | $11 μm$                                                                                                                                                                                                                                                                                                                                                | $23 μm$                                                                                                                                                                                                                                                                                                                                                |                                        |
|                                         | Shear<br>bond<br>strength                                    | Composite -Dentin                                                                                                                                                                                                                                                                                                                                      | $12.85MPa$                                                                                                                                                                                                                                                                                                                                             | $7.68MPa$                              |
|                                         |                                                              | Composite -Enamel                                                                                                                                                                                                                                                                                                                                      | $19.16MPa$                                                                                                                                                                                                                                                                                                                                             | $16.32MPa$                             |
|                                         |                                                              | Glass Ceramic -Litium disilicate                                                                                                                                                                                                                                                                                                                       | $6.21MPa$                                                                                                                                                                                                                                                                                                                                              | $5.49MPa$                              |
|                                         |                                                              | Ceramic Aluminum oxide                                                                                                                                                                                                                                                                                                                                 | $7.28MPa$                                                                                                                                                                                                                                                                                                                                              | $4.27MPa$                              |
|                                         |                                                              | Metal                                                                                                                                                                                                                                                                                                                                                  | $12.70MPa$                                                                                                                                                                                                                                                                                                                                             | $4.38MPa$                              |
|                                         |                                                              | Zirconia                                                                                                                                                                                                                                                                                                                                               | $9.05MPa$                                                                                                                                                                                                                                                                                                                                              | $5.12MPa$                              |
|                                         | Fluoride release                                             |                                                                                                                                                                                                                                                                                                                                                        | $2.42 ppm$                                                                                                                                                                                                                                                                                                                                             | $10.02 ppm$                            |
|                                         | Radio-opacity                                                |                                                                                                                                                                                                                                                                                                                                                        | 1.45                                                                                                                                                                                                                                                                                                                                                   | None                                   |
|                                         | Solubility                                                   |                                                                                                                                                                                                                                                                                                                                                        | $0.87 µg/mm²$                                                                                                                                                                                                                                                                                                                                          | $1.53µg/mm²$                           |
|                                         | Polymerization shrinkage                                     |                                                                                                                                                                                                                                                                                                                                                        | 4.38%                                                                                                                                                                                                                                                                                                                                                  | 4.18%                                  |
| Chemical<br>Composition                 | Filler                                                       |                                                                                                                                                                                                                                                                                                                                                        | Amorphous silica                                                                                                                                                                                                                                                                                                                                       | Amorphous silica                       |
|                                         | Resin composition                                            |                                                                                                                                                                                                                                                                                                                                                        | Methacrylate based                                                                                                                                                                                                                                                                                                                                     | Methacrylate based                     |
|                                         | Method of application<br>Ions released                       |                                                                                                                                                                                                                                                                                                                                                        | Bonding agent not required<br>Fluoride                                                                                                                                                                                                                                                                                                                 | Bonding agent not required<br>Fluoride |
| Standard Conformed                      |                                                              | ISO 4049                                                                                                                                                                                                                                                                                                                                               | ISO 4049                                                                                                                                                                                                                                                                                                                                               |                                        |
| Biocompatibility                        |                                                              | Yes                                                                                                                                                                                                                                                                                                                                                    | Yes                                                                                                                                                                                                                                                                                                                                                    |                                        |
| Use                                     |                                                              | Prescription / Hospital                                                                                                                                                                                                                                                                                                                                | Prescription / Hospital                                                                                                                                                                                                                                                                                                                                |                                        |
| Delivery system                         |                                                              | Dual syringe(Base:Catalyst=1:1), Automix and Double Dispenser type                                                                                                                                                                                                                                                                                     | Dual-syringe(Base:Catalyst=1:1), Automix type                                                                                                                                                                                                                                                                                                          |                                        |
| Storage condition                       |                                                              | 2~8°C<br>(at refrigerator temperature)                                                                                                                                                                                                                                                                                                                 | 2~8°C<br>(at refrigerator temperature)                                                                                                                                                                                                                                                                                                                 |                                        |

{6}------------------------------------------------

| Physical<br>properties | Shade | - Clear<br>- Universal<br>- Opaque | - Translucent<br>- Opaque |
|------------------------|-------|------------------------------------|---------------------------|
|------------------------|-------|------------------------------------|---------------------------|

#### 10. Performance Tests (Non-clinical)

- . Performance Test:
	- ISO 4049 Fourth edition 2009-10-01 Dentistry Polymer-based restorative materials

- ISO 9917-1 Second edition 2007-10-01 Dentistry - Water-based cements - Part 1: Powder/liquid acid-base cements, Annex D.

ISO /TS 11405 Third edition 2015-02-01 Dentistry - Testing of adhesion to tooth i structure

- Biocompatibility Tests ●
	- ISO 10993-1 Fourth edition 2009-10-15, biological evaluation of medical devices part 1: evaluation and testing within a risk management process [including: technical corrigendum 1 (2010)]. (Biocompatibility)
	- ISO 10993-3 Third edition 2014-10-1 Biological evaluation of medical devices Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity
	- ISO 10993-5 Third edition 2009-06-01 Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity
	- ISO 10993-10 Third Edition 2010-08-01 Biological evaluation of medical devices -Part 10: Tests for irritation and skin sensitization
	- ISO 10993-11 Second edition 2006-08-15 Biological evaluation of medical devices -Part 11: Tests for systemic toxicity
- Shelf Life Test Performance Tests in accordance with ISO 4049 and ISO 9917-1 .

The test results of non-clinical tests performed on the subject device supported that it is substantially equivalent to the predicate devices despite the differences.

## 11. Conclusions:

Based on the information provided in this premarket notification, Vericom Co., Ltd. concludes that the U-Cem™ Premium & MAZIC® Cem is substantially equivalent to the predicate device as described herein in.

---

**Source:** [https://fda.innolitics.com/device/K193260](https://fda.innolitics.com/device/K193260)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
