3M Attest Steam Chemical Integrator

K193254 · 3M Company · JOJ · Dec 23, 2019 · General Hospital

Device Facts

Record IDK193254
Device Name3M Attest Steam Chemical Integrator
Applicant3M Company
Product CodeJOJ · General Hospital
Decision DateDec 23, 2019
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 880.2800
Device ClassClass 2

Indications for Use

The 3M™ Attest™ Steam Chemical Integrators are designed to all critical parameters over a specified range of steam sterilization cycles. The integrating intended to be placed in each pack, pouch, container, tray or other containment device to function as an independent monitor of critical parameters for the following sterilization cycles: Gravity 250°F/121°C 30 minutes; Gravity 270°F/132°C 3, 4, 10, 15, 25 minute; Gravity 275°F/135°C 3, 10 minutes; Dynamic Air Removal 250°F/121°C 30 minutes; Dynamic Air Removal 270°F/132°C 4, 10 minutes; Dynamic Air Removal 273°F/134°C 3, 4 minutes; Dynamic Air Removal 275°F/135°C 3 minutes.

Device Story

Chemical indicator for steam sterilization monitoring; consists of paper wick and steam/temperature-sensitive chemical pellet in paper/film/foil laminate. Pellet melts and migrates as dark color along wick upon exposure to steam, time, and temperature. Migration visible through 'ACCEPT' or 'REJECT' window. Placed inside sterilization packs/trays by healthcare personnel. Provides visual confirmation of sterilization cycle parameters; aids in clinical decision-making regarding sterility of processed items. Benefits patient by ensuring medical devices have undergone required sterilization conditions.

Clinical Evidence

Bench testing only. Performance demonstrated via resistometer testing per ANSI/AAMI/ISO 11140-1:2014. Metrics included Stated Value (SV) verification (121°C-135°C), temperature coefficient (10-27°C), and correlation coefficient (≥0.9). Additional testing included simulated healthcare facility use, dry heat resistance, side-by-side comparison with biological indicators, and endpoint color stability (6 months). All tests passed.

Technological Characteristics

Chemical indicator; paper wick with proprietary steam/temperature-sensitive chemical pellet; paper/film/foil laminate housing. Operates via physical melting/migration principle. No energy source or software. Shelf-life: 1 year. Compliant with ANSI/AAMI/ISO 11140-1:2014.

Indications for Use

Indicated for use as an independent monitor of critical parameters (steam, time, temperature) in steam sterilization cycles for packs, pouches, containers, or trays. Applicable to gravity and dynamic air removal cycles ranging from 250°F to 275°F.

Regulatory Classification

Identification

Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. December 23, 2019 3M Company Mary Fretland Senior Regulatory Affairs Associate 3M Center, Building 275-5W-06 St. Paul, Minnesota 55144 Re: K193254 Trade/Device Name: 3M Attest Steam Chemical Integrator Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: JOJ Dated: November 25, 2019 Received: November 26, 2019 Dear Mary Fretland: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Sreekanth Gutala, PhD Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K193254 ### Device Name 3M™ Attest™ Steam Chemical Integrators ### Indications for Use (Describe) The 3M™ Attest™ Steam Chemical Integrators are designed to all critical parameters over a specified range of steam sterilization cycles. The integrating intended to be placed in each pack, pouch, container, tray or other containment device to function as an independent monitor of critical parameters for the following sterilization cycles: | Cycle Type | Temperature | Exposure Time | |---------------------|-------------|-------------------------| | Gravity | 250°F/121°C | 30 minutes | | Gravity | 270°F/132°C | 3, 4, 10, 15, 25 minute | | Gravity | 275°F/135°C | 3, 10 minutes | | Dynamic Air Removal | 250°F/121°C | 30 minutes | | Dynamic Air Removal | 270°F/132°C | 4, 10 minutes | | Dynamic Air Removal | 273°F/134°C | 3, 4 minutes | | Dynamic Air Removal | 275°F/135°C | 3 minutes | Minimum Stated Values for 3M™ Attest™ Steam Chemical Integrators as determined in a resistometer: 250°F/ 121°C: 16.5 minutes 270°F / 132°C: 2.0 minutes 273°F / 134°C: 1.4 minutes 275°F / 135°C: 1.2 minutes Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the 3M logo. The logo is red and consists of the number "3" and the letter "M" in a bold, sans-serif font. The logo is simple and recognizable. # 510(k) Summary for 3MTM Attest™ Steam Chemical Integrators ### Sponsor Information: 3M Health Care 3M Center, Bldg. 275-5W-06 St. Paul, MN 55144-1000 Contact: Mary Fretland Senior Regulatory Affairs Associate Phone Number: (651) 737-2296 Fax Number: (651) 737-5320 Email: mfretland@mmm.com Date of Summary: December 23, 2019 Submission Number: K193254 {4}------------------------------------------------ #### 1. Device Name and Classification: | Common or Usual Name: | Chemical Indicators | |------------------------|----------------------------------------------------| | Trade Name: | 3MTM AttestTM Steam Chemical Integrators | | Classification Name: | Physical/chemical sterilization process indicators | | Device Classification: | Class II, 21 CFR § 880.2800(b) | | Product Code: | JOJ | #### 2. Predicate Device: 3MTM Attest™ Steam Chemical Integrators (K191236) #### 3. Description of Device: 3M™ Attest™ Steam Chemical Integrators are chemical indicators consisting of a paper wick and a steam and temperature sensitive chemical pellet contained in a paper/film/foil laminate. The chemical pellet melts and migrates as a dark color along the paper wick. The migration is visible through a window marked ACCEPT or window marked REJECT; the extent of migration depends on steam, time, and temperature. {5}------------------------------------------------ #### 4. Indications for Use The 3M™ Attest™ Steam Chemical Integrators are designed to respond to all critical parameters over a specified range of steam sterilization cycles. The integrating indicator is intended to be placed in each pack, pouch, container, tray or other containment device to function as an independent monitor of critical parameters for the following sterilization cycles: | Cycle Type | Temperature | Exposure Time | |---------------------|-------------|---------------| | Gravity | 250°F/121°C | 30 minutes | | Gravity | 270°F/132°C | 3 minutes | | Gravity | 270°F/132°C | 4 minutes | | Gravity | 270°F/132°C | 10 minutes | | Gravity | 270°F/132°C | 15 minutes | | Gravity | 270°F/132°C | 25 minutes | | Gravity | 275°F/135°C | 3 minutes | | Gravity | 275°F/135°C | 10 minutes | | Dynamic Air Removal | 250°F/121°C | 30 minutes | | Dynamic Air Removal | 270°F/132°C | 4 minutes | | Dynamic Air Removal | 270°F/132°C | 10 minutes | | Dynamic Air Removal | 273°F/134°C | 3 minutes | | Dynamic Air Removal | 273°F/134°C | 4 minutes | | Dynamic Air Removal | 275°F/135°C | 3 minutes | Minimum Stated Values for 3M™ Attest™ Steam Chemical Integrators as determined in a resistometer: | 250°F/121°C | 270°F/132°C | 273°F/134°C | 275°F/135°C | |--------------|-------------|-------------|-------------| | 16.5 Minutes | 2.0 Minutes | 1.4 Minutes | 1.2 Minutes | #### ട. Nonclinical Comparison to the Predicate Device The design, fundamental technology and performance specifications for 3M™ Attest™ Steam Chemical Integrators are identical to the previously cleared device which is sold under the same tradename 3MTM Attest™ Steam Chemical Integrators (K191236). There has been no change to the device's performance specifications or fundamental scientific technology. The differences between the subject device and the predicate are limited to a change in shape and removal of the extender. The differences between 3MTM Attest™ Steam Chemical Integrator and the predicate do not raise any new questions of safety and effectiveness as demonstrated by the performance testing and biocompatibility assessment. {6}------------------------------------------------ #### Technical Characteristics 6. | Feature | Submission Device:<br>3MTM AttestTM Steam Chemical<br>Integrators (K193254) | Predicate Device:<br>3MTM AttestTM Steam Chemical<br>Integrators (K191236) | Comparison | |------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------| | Device Models | 1243A and 1243B | 1243RE and 1243RES | Similar | | Device Design | 3MTM AttestTM Steam Chemical<br>Integrators are chemical indicators<br>consisting of a paper wick and steam and<br>temperature sensitive chemical pellet<br>contained in a paper/film/foil laminate.<br>The chemical pellet melts and migrates as<br>a dark color along the paper wick. The<br>migration is visible through a window<br>marked ACCEPT or window marked<br>REJECT; the extent of migration depends<br>on steam, time, and temperature. | 3MTM AttestTM Steam Chemical<br>Integrators are chemical indicators<br>consisting of a paper wick and steam and<br>temperature sensitive chemical pellet<br>contained in a paper/film/foil laminate.<br>The chemical pellet melts and migrates as<br>a dark color along the paper wick. The<br>migration is visible through a window<br>marked ACCEPT or window marked<br>REJECT; the extent of migration depends<br>on steam, time, and temperature. | Identical | | Indicator Agent | Proprietary formulation. | Proprietary formulation. | Identical | | Sterilization<br>method and cycles | Steam sterilization processes 250°F to<br>275°F (121°C to 135°C) | Steam sterilization processes 250°F to<br>275°F (121°C to 135°C) | Identical | | Shelf-life | One (1) year | One (1) year | Identical | ### Technological Characteristics Comparison Between Subject and Predicate Device: # Indications for Use Comparison: | Indications for Use Comparison: | | | | | |---------------------------------|-------------|----------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|------------| | Cycle Type | Temperature | Submission Device (K193254):<br>3MTM AttestTM Steam Chemical<br>Integrators (1243A and 1243B)<br>Exposure Time | Predicate Device (K191236):<br>3MTM AttestTM Steam Chemical<br>Integrators (1243RE and 1243RES)<br>Exposure Time | Comparison | | Gravity | 250°F/121°C | 30 minutes | 30 minutes | Identical | | Gravity | 270°F/132°C | 3 minutes<br>4 minutes<br>10 minutes<br>15 minutes<br>25 minutes | 3 minutes<br>4 minutes<br>10 minutes<br>15 minutes<br>25 minutes | Identical | | Gravity | 275°F/135°C | 3 minutes<br>10 minutes | 3 minutes<br>10 minutes | Identical | | Dynamic Air<br>Removal | 250°F/121°C | 30 minutes | 30 minutes | Identical | | Dynamic Air<br>Removal | 270°F/132°C | 4 minutes<br>10 minutes | 4 minutes<br>10 minutes | Identical | | Dynamic Air<br>Removal | 273°F/134°C | 3 minutes<br>4 minutes | 3 minutes<br>4 minutes | Identical | | Dynamic Air<br>Removal | 275°F/135°C | 3 minutes | 3 minutes | Identical | {7}------------------------------------------------ | Feature | Submission Device (K193254):<br>3MTM AttestTM Steam Chemical<br>Integrators (1243A and 1243B) | Predicate Device (K191236):<br>3MTM AttestTM Steam Chemical Integrators<br>(1243RE and 1243RES) | Comparison | | | | | | | | | | | | | | | | | |----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|------------|--|--|--|--|--|--|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--| | | The minimum stated values for the 3MTM<br>AttestTM Steam Chemical Integrators as<br>determined using a resistometer are<br>provided in the table below. | The minimum stated values for the 3MTM<br>AttestTM Steam Chemical Integrators as<br>determined using a resistometer are<br>provided in the table below. | Identical | | | | | | | | | | | | | | | | | | Endpoint<br>Specifications<br>(Minimum Stated<br>Values) | Minimum Stated Values for 3MTM<br>AttestTM Steam Chemical Integrators<br>250°F/<br>121°C 270°F/<br>132°C 273°F/<br>134°C 275°F/<br>135°C 16.5<br>min 2.0<br>min 1.4<br>min 1.2<br>min | | | | | | | | | Minimum Stated Values for 3MTM<br>AttestTM Steam Chemical Integrators<br>250°F/<br>121°C 270°F/<br>132°C 273°F/<br>134°C 275°F/<br>135°C 16.5<br>min 2.0<br>min 1.4<br>min 1.2<br>min | | | | | | | | | | #### 7. Summary of Nonclinical Testing Nonclinical testing of the 3M™ Attest™ Steam Chemical Integrators was conducted in accordance with the FDA Guidance for Industry and FDA Staff: Premarket Notification [510(k)] Submissions for Chemical Indicators, issued December 19, 2003 as well as ANSI/AAM/ISO 11140-1:2014. The effectiveness of the 3MTM Attest™ Steam Chemical Integrator was demonstrated in the following tests: | Test Name | Purpose | Acceptance Criteria | Results | | |--------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|------| | Stated<br>Value (SV)<br>Testing | To identify the critical parameters required to<br>achieve a stated inactivation, by referring to a<br>stated test organism with stated D and z<br>values. The integrator must turn to "ACCEPT"<br>end point at the stated value time and must also<br>remain "REJECT" when exposed to conditions<br>of -1°C/-15% set point of the SV time. All<br>testing is completed in a saturated steam<br>resistometer. Integrator temperature coefficient<br>and correlation coefficient are calculated to<br>confirm alignment to biological indicator<br>performance. | SV at 121°C<br>SV at 132°C<br>SV at 134°C<br>SV at 135°C<br>Integrator<br>temperature<br>coefficient<br>Correlation<br>coefficient | ≥ 16.5 minutes<br>≥ 2.0 minutes<br>≥ 1.4 minutes<br>≥ 1.2 minutes<br>10 - 27°C<br>≥ 0.9 | PASS | | Health Care<br>Facility Simulated<br>Use Testing | Confirm integrators provide acceptable<br>performance in cleared customer use<br>sterilization cycles. | Device reaches “ACCEPT” endpoint<br>reaction when exposed to customer use<br>cycles.<br>Device does not reach endpoint<br>(“REJECT”) when exposed to failing<br>conditions in customer use cycles. | PASS | | {8}------------------------------------------------ | Test Name | Purpose | Acceptance Criteria | Results | |---------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|---------| | Dry Heat Testing | Verify device requires the presence of saturated steam to turn to reach endpoint. | Endpoint must not be met following dry heat exposure at 140°C for 30 min. | PASS | | Side-by-Side<br>Testing with<br>Biological<br>Indicator | Confirm integrators are parallel in performance to biological indicators (BI). | Chemical integrator parallels performance of BI and does not reach endpoint before BI is inactivated. | PASS | | Endpoint Color<br>Stability | Confirm endpoint color stability for samples exposed to passing and failing conditions in a steam resistometer. | Endpoint decision must remain unchanged after 6 months. | PASS | The results of performance testing on 3M™ Attest™ Steam Chemical Integrator demonstrate the device performs as intended in the claimed steam sterilization cycles. #### 8. Conclusion Based on the intended uses, technological characteristics and non-clinical performance data, the 3M™ Attest™ Steam Chemical Integrator is as safe, as effective and performs as well as or better than the legally marketed predicate, 3M™ Attest™ Steam Chemical Integrators cleared under K191236, Class II (CFR 880.2800, product code JOJ).
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...