K193247 · Tornier, Inc. · JDB · Aug 5, 2020 · Orthopedic
Device Facts
Record ID
K193247
Device Name
LATITUDE EV™ Total Elbow Arthroplasty
Applicant
Tornier, Inc.
Product Code
JDB · Orthopedic
Decision Date
Aug 5, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3160
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LATITUDE EV™ Total Elbow Arthroplasty system is intended for total elbow arthroplasty. Prosthetic replacement with this device may be indicated to relieve severe pain or significant disability following the effects of primary or secondary osteoarthritis and rheumatoid arthritis; correction of functional deformities; revision procedures where other treatments or devices have failed; treatment of fractures that are unmanageable using other techniques. The LATITUDE EV™ Total Elbow Arthroplasty system is intended for cemented use only.
Device Story
LATITUDE EV™ Total Elbow Arthroplasty system; semi-constrained or constrained elbow prosthesis; consists of humeral, ulnar, and optional radial implants. Implanted by orthopedic surgeons in clinical settings to replace elbow joint; reproduces natural flexion/extension axis and kinematics. Unlinked configuration provides semi-constrained function; ulnar cap linkage provides constrained function. All components intended for cemented use only. Device relieves pain, corrects deformities, and restores function in patients with arthritis or complex fractures.
Clinical Evidence
No clinical studies were performed. Substantial equivalence is supported by non-clinical performance testing, including fatigue testing, coating characterization (ASTM F1854-15, F1160-14, F1044-05, F1147-05, F1978-17), biocompatibility (ISO 10993-1:2018), packaging/shelf life (ISO 11607-1/2:2019, ASTM F1980-16), distribution testing, sterilization evaluation (ISO 11137-1/2:2006), and endotoxin testing (AAMI ST72:2011).
Technological Characteristics
Semi-constrained/constrained metal/polymer cemented prosthesis. Components: humeral, ulnar, and optional radial implants. Manufacturing: cast humeral and ulnar stems. Coating: plasma-sprayed. Standards: ASTM F1854-15, F1160-14, F1044-05, F1147-05, F1978-17; ISO 10993-1:2018; ISO 11607-1/2:2019; ISO 11137-1/2:2006. Sterilization: validated per ISO 11137. No software or electronic components.
Indications for Use
Indicated for patients requiring total elbow arthroplasty due to severe pain or disability from primary/secondary osteoarthritis or rheumatoid arthritis, functional deformities, failed prior treatments/devices, or unmanageable fractures.
Regulatory Classification
Identification
An elbow joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace an elbow joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that consist of a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a radial resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
COONRAD/MORREY TOTAL ELBOW, MODEL 8105 SERIES (K053189)
Submission Summary (Full Text)
{1}------------------------------------------------
August 5, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Tornier Inc. Renee Stoffel Principal Regulatory Affairs Specialist 10801 Nesbitt Ave South Bloomington, Minnesota 55347
## Re: K193247
Trade/Device Name: LATITUDE EV™ Total Elbow Arthroplasty system Regulation Number: 21 CFR 888.3160 Regulation Name: Elbow joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JDB, JDC Dated: July 8, 2020 Received: July 10, 2020
Dear Renee Stoffel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or
{2}------------------------------------------------
postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/trainingand-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance/contact-usdivision-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Michael Owens Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K193247
Device Name LATITUDE EV™ Total Elbow Arthroplasty
#### Indications for Use (Describe)
The LATITUDE EV™ Total Elbow Arthroplasty system is intended for total elbow arthroplasty. Prosthetic replacement with this device may be indicated to relieve severe pain or significant disability following the effects of primary or secondary osteoarthritis and rheumatoid arthritis; correction of functional deformities; revision procedures where other treatments or devices have failed; treatment of fractures that are unmanageable using other techniques.
The LATITUDE EV™ Total Elbow Arthroplasty system is intended for cemented use only.
Type of Use (Select one or both, as applicable)
| <span style="font-size:10pt"><b> <span style="font-family:Wingdings">X</span> Prescription Use (Part 21 CFR 801 Subpart D)</b></span> |
|---------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size:10pt"><b> <span style="font-family:Wingdings">□</span> Over-The-Counter Use (21 CFR 801 Subpart C)</b></span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "TORNIER" in a large, sans-serif, blue font. To the right of the word is a blue hexagon containing a stylized letter "T" inside. The letter "T" is also blue and has a unique design.
Date Prepared: July 8, 2020
## Administrative Information
| Name:<br>Address: | Tornier, Inc.<br>10801 Nesbitt Avenue South<br>Bloomington, MN 55437<br>United States of America |
|-------------------|--------------------------------------------------------------------------------------------------|
| Contact Person: | Renee Stoffel |
| Title: | Principal Regulatory Affairs Specialist |
| Phone: | 952-683-7471 |
| Fax: | 952-426-7601 |
## Device Information
| Name of Device: | LATITUDE EV™ Total Elbow Arthroplasty |
|-------------------|-------------------------------------------------------------------------------------|
| Common Name (s): | Elbow Prosthesis |
| Regulatory Class: | II |
| Regulation: | 21 CFR 888.3160, elbow joint metal/polymer semi-constrained cemented<br>prosthesis. |
| | 21 CFR 888.3150, elbow joint metal/polymer constrained cemented<br>prosthesis. |
| Product Codes: | JDB |
| | JDC |
## Predicate Device Information
| Primary Predicate: | K182461, LATITUDE EV™ Total Elbow Arthroplasty |
|-----------------------|-----------------------------------------------------------------------------------------------|
| Additional Predicate: | K171010, LATITUDE EV™ Total Elbow Arthroplasty |
| Reference Devices: | K070787, LATITUDE™ Elbow Prosthesis<br>K053189, COONRAD/MORREY TOTAL ELBOW, MODEL 8105 SERIES |
## Device Description
The LATITUDE EVI™ Total Elbow Arthroplasty system is a total elbow prosthesis consisting of a humeral, an ulnar, and an optional radial implant. The LATITUDE EV system is designed to facilitate the reproduction of the natural flexion/extension axis and kinematics of the prosthesis is a semiconstrained prosthesis when it is implanted in an unlinked configuration and becomes a constrained prosthesis when the ulnar cap is used to link the humeral and ulnar implant assemblies. All components are intended for cemented use only.
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "TORNIER" in a large, bold, blue font. To the right of the word is a blue hexagon containing a stylized letter "T". The letter "T" is also blue and is made up of two smaller hexagons.
## Indications for Use
The LATITUDE EV™ Total Elbow Arthroplasty system is intended for total elbow arthroplasty. Prosthetic replacement with this device may be indicated to relieve severe pain or significant disability following the effects of primary or secondary osteoarthritis and rheumatoid arthritis; correction of functional deformities; revision procedures where other treatments or devices have failed; treatment of fractures that are unmanageable using other techniques.
The LATITUDE EV™ Total Elbow Arthroplasty system is intended for cemented use only.
## Comparison to Predicate Device
The initial manufacturing process for the LATITUDE EV humeral and ulnar stems was changed from forging to casting for a subset of the stem sizes and lengths. Additional modifications for the subject components include sterile packaging materials, labeled shelf life, dimensional specifications, and material specifications. These manufacturing and design differences do not raise new issues of safety or effectiveness and are supported by performance testing and process validations.
## Non-clinical Performance Testing
Non-clinical testing was performed to demonstrate substantial equivalence to the predicate device.
- Humeral and ulnar stem fatigue testing ●
- Plasma-sprayed coating characterization and testing to ASTM F1854-15, ASTM F1160-14, . ASTM F1044-05, ASTM F1147-05, ASTM F1978-17
- Biocompatibility evaluation to ISO 10993-1:2018 ●
- . Packaging and shelf life evaluations to ISO 11607-1:2019, ISO 11607-2:2019, ASTM F1980-16
- Distribution testing to ISTA Procedure 3A (2011), ASTM D4169-16, ASTM F2096-11 .
- Sterilization evaluation to ISO 11137-1:2006 and ISO 11137-2:2006 .
- . Endotoxin testing to AAMI ST72:2011
# Clinical Testing
No clinical studies were performed.
## Conclusions
The LATITUDE EVIM Total Elbow Arthroplasty system does not raise new questions of safety or effectiveness. Differences in technological characteristics have been addressed with performance testing. The results of performance testing for the modified LATITUDE EV™ humeral and ulnar components support substantial equivalence to the current LATITUDE EV™ system (K182461).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.