K193190 · Becton, Dickinson and Company · QBP · Nov 9, 2020 · General Hospital
Device Facts
Record ID
K193190
Device Name
BD PureHub Disinfecting Cap
Applicant
Becton, Dickinson and Company
Product Code
QBP · General Hospital
Decision Date
Nov 9, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
BD PureHub™ Disinfecting Caps are intended to be used as a disinfecting cleaner for swabbable needle-free luer connectors prior to access and to act as a physical barrier between line accesses. BD PureHub™ Disinfecting Cap will disinfect the needle-free luer connector one (1) minute after application and act as a physical barrier for up to seven (7) days, if not removed.
Device Story
BD PureHub Disinfecting Cap is a sterile, single-use device for needle-free luer connectors. It consists of a high-density polyethylene housing and a polyester urethane sponge saturated with 70% isopropyl alcohol (IPA). The device is applied to the luer connector, where the IPA acts as an antimicrobial agent to disinfect the surface within one minute. It remains in place to serve as a physical barrier against contamination for up to seven days. Used in healthcare facilities or home settings by clinicians or patients. The device is supplied in bulk or on IV pole strips. It provides a passive disinfection method, reducing the risk of catheter-related bloodstream infections by maintaining a sterile barrier between line accesses.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including in vitro antimicrobial efficacy testing against S. aureus, S. epidermidis, E. coli, P. aeruginosa, C. albicans, C. glabrata, and A. baumannii, demonstrating >4-log reduction. Additional bench testing included particulate matter ingress (USP 788), IPA concentration, device retention, air leakage, weld retention, packaging integrity, and biocompatibility (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemocompatibility).
Technological Characteristics
Materials: High-density polyethylene (HDPE) housing, polyester urethane sponge. Disinfectant: 70% Isopropyl Alcohol (IPA). Form factor: Cap for needle-free luer connectors. Sterilization: Gamma irradiation (ISO 11137-2:2013, SAL 10^-6). Shelf life: 3 years. No software or electronic components.
Indications for Use
Indicated for use as a disinfecting cleaner for swabbable needle-free luer connectors prior to access and as a physical barrier between line accesses in patients requiring intravascular access.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the logo for the Department of Health & Human Services - USA. On the right is the logo for the U.S. Food & Drug Administration (FDA). The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.
Becton, Dickinson and Company Samhitha Mohan Regulatory Affairs Specialist 1 Becton Drive Franklin Lakes, New Jersey 07417
March 11, 2022
Re: K193190
Trade/Device Name: BD PureHub Disinfecting Cap Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: QBP
Dear Samhitha Mohan:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated November 9, 2020. Specifically, FDA is updating this SE Letter because FDA has better categorized your device technology under regulation 880.5440.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Payal Patel, OHT3: Office of GastroRenal, Ob-Gyn, General Hospital and Urology Devices, 240-402-6029, Payal.Patel(@fda.hhs.gov.
Sincerely,
FDA
Payal Patel Assistant Director for General Hospital Devices DHT3C: Division of Drug Delivery and General Hospital Devices and Human Factors OHT3: Office of GastroRenal, Ob-Gyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the logos of the Department of Health and Human Services (HHS) and the Food and Drug Administration (FDA). The HHS logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue font.
November 9, 2020
Becton, Dickinson and Company Samhitha Mohan Regulatory Affairs Specialist 1 Becton Drive Franklin Lakes, New Jersey 07417
Re: K193190
Trade/Device Name: BD PureHub™ Disinfecting Cap Regulatory Class: Unclassified Product Code: OBP Dated: October 8, 2020 Received: October 9, 2020
Dear Samhitha Mohan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{2}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Sapana Patel -S
for Payal Patel Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K193190
Device Name BD PureHub DisinfectingTM Cap
Indications for Use (Describe)
BD PureHub™ Disinfecting Caps are intended to be used as a disinfecting cleaner for swabbable needle-free luer connectors prior to access and to act as a physical barrier between line accesses.
BD PureHub™ Disinfecting Cap will disinfect the needle-free luer connector one (1) minute after application and act as a physical barrier for up to seven (7) days, if not removed.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the BD logo. The logo consists of two parts: a circular orange symbol on the left and the letters "BD" in blue on the right. The circular symbol features a stylized sun-like design with rays emanating from a central point. The letters "BD" are in a bold, sans-serif font.
## K193190 - 510(k) Summary (21 CFR §807.92)
## BD PureHub™ Disinfecting Cap
| Submitter<br>Information | Submitter Name: | Becton, Dickinson and Company |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| | Submitter Address: | 1 Becton Drive<br>Franklin Lakes<br>NJ 07417 |
| | Contact Person: | Samhitha Mohan<br>Staff Regulatory Affairs Specialist |
| | Email Address:<br>Phone Number:<br>Fax Number:<br>Date of Preparation: | Samhitha.Mohan@bd.com<br>(201) 847-5204<br>(201) 847-5307<br>November 3, 2020 |
| Subject Device | Trade Name:<br>Common Name:<br>Regulation Number:<br>Device Class:<br>Product Code:<br>Classification Panel: | BD PureHub™ Disinfecting Cap<br>Cap, Device Disinfectant<br>Unclassified<br>Unclassified<br>QBP<br>General Hospital |
| Predicate Device | Trade Name:<br>510(k) Reference:<br>Common Name:<br>Regulation Number:<br>Regulatory Class:<br>Product Code:<br>Classification Panel: | Curos™ Port Protector<br>K111992<br>Cap, Device Disinfectant<br>Unclassified<br>Unclassified<br>QBP<br>General Hospital |
| Device Description | BD PureHub™ Disinfecting Cap is a sterile, single use<br>disinfectant cap designed for needle-free Luer connectors. It<br>has high-density polyethylene housing and polyester urethane<br>sponge saturated with 70% Isopropyl Alcohol (IPA) solution.<br>PureHub™ disinfects needle-free Luer connectors one minute<br>after application and acts as a physical barrier for up to seven<br>days, if not removed. It is available in two packaging<br>configurations - Bulk Single Unit and IV Pole Strip. | |
| Indications for Use | BD PureHub™ Disinfecting Caps are intended to be used as a<br>disinfecting cleaner for swabbable needle-free luer connectors<br>prior to access and to act as a physical barrier between line<br>accesses. | |
{5}------------------------------------------------
BD PureHub™ Disinfecting Cap will disinfect the needle-free luer connector one (1) minute after application and act as a physical barrier for up to seven (7) days, if not removed.
## Technological Characteristics
The following table provides a comparison between the subject and predicate device -
| Attributes | Subject Device<br>(BD PureHub™<br>Disinfecting Cap) | Predicate Device<br>(Curos™ Port<br>Protector) | Comparison | |
|--------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | BD PureHub™<br>Disinfecting Caps are<br>intended to be used<br>as a disinfecting<br>cleaner for<br>swabbable needle-<br>free luer connectors<br>prior to access and<br>to act as a physical<br>barrier between line<br>accesses.<br><br>BD PureHub™<br>Disinfecting Cap will<br>disinfect the needle-<br>free luer connector<br>one (1) minute after<br>application and act<br>as a physical barrier<br>for up to seven (7)<br>days, if not<br>removed. | The Curos™ is intended<br>for use on swab-able<br>Luer access valves as a<br>disinfecting cleaner<br>prior to line access and<br>to act as a physical<br>barrier to<br>contamination between<br>line accesses. Curos™<br>will disinfect the valve<br>three (3) minutes after<br>application and act a s<br>a physical barrier to<br>contamination for up to<br>seven (7) days (168<br>hours) if not removed.<br><br>The effectiveness of<br>Curos™ Protectors were<br>tested <i>in vitro</i> against<br><i>Staphylococcus aureus</i> ,<br><i>Staphylococcus<br/>epidermis</i> , <i>Escherichia<br/>coli</i> and <i>Pseudomonas<br/>aeruginosa</i> , <i>candida<br/>glabrata</i> , <i>Canada<br/>albicans</i> and was found<br>to have >4-log<br>reduction. The Curos™<br>Port Protector may be<br>used in the home or<br>healthcare facility. | With the exception<br>of disinfection time<br>and the<br>microorganisms,<br>the intended use is<br>similar. <i>In vitro</i><br>antimicrobial<br>efficacy testing<br>was performed to<br>support PureHub's<br>disinfection time.<br><br>The micro-<br>oraganisms tested<br>are identical to<br>Curos™ Port<br>Protector with the<br>exception of the<br>addition of<br>antimicrobial<br>efficacy testing<br>conducted on<br><i>Acinetobacter<br/>baumannii</i> . The<br>subject device<br>does not list the<br>microorganisms<br>tested against in<br>the indications for<br>use. | |
| Condition of Use<br>(Environment) | Same as predicate | Home and health care<br>facility | Identical | |
| Cap Insertion<br>Site | Same as predicate | Needle-free Luer<br>connector | Identical | |
| | | | | |
| Device<br>Components | | Same as predicate | Ribbed external cap housing Sponge with 70% IPA solution Foil lid | Identical |
| User Population | | Same as predicate | General use | Identical |
| Operating<br>Principle | | Same as predicate | 70% IPA solution acts as an antimicrobial agent to disinfect needle-free Luer connector | Identical |
| Device<br>Materials | Cap<br>Housing | High Density<br>Polyethylene (HDPE) | Unknown | Material differences were assessed as per ISO 10993-1 |
| | Sponge | Polyester Urethane | Unknown | |
| | Disinfecta<br>-nt<br>Solution | Same as predicate | 70% IPA | |
| | Colorant | Teal green | Translucent green | |
| Packaging<br>Configurations | | Same as predicate | Bulk Single Unit and IV Pole Strip | Identical |
| Sterilization<br>Method | | Same as predicate | Gamma Irradiation | Identical |
| SAL | | Same as predicate | 10-6 | Identical |
| Antimicrobial<br>Efficacy | | Same as predicate | > 4-log reduction | Identical |
| Minimum<br>Disinfecting<br>Time | | 1 minute | 3 minutes | Similar; reduced disinfecting time was assessed through antimicrobial efficacy testing. |
| Target<br>microorganisms<br>for <i>in vitro</i><br>antimicrobial<br>efficacy testing | | Staphylococcus aureus Staphylococcus epidermidis Escherichia coli Pseudomonas aeruginosa Candida albicans Candida glabrata Acinetobacter baumannii | Staphylococcus aureus Staphylococcus epidermidis Escherichia coli Pseudomonas aeruginosa Candida albicans Candida glabrata | Identical except for Acinetobacter baumannii. <i>In vitro</i> antimicrobial efficacy testing was performed for all the microorganisms |
| Shelf Life | 3 years | 3 years | Identical | |
{6}------------------------------------------------
{7}------------------------------------------------
## Discussion:
The intended use of the subject and predicate device are similar in that both disinfect needle-free Luer connector and act as a physical barrier. However, the disinfection time of PureHub™ (1 minute) is shorter than Curos™ Port Protector (3 minutes) and PureHub™ tests an additional microorganism (Acinetobacter baumannii) compared to Curos™ Port Protector. In vitro antimicrobial efficacy testing was performed for all the above mentioned microorganisms to demonstrate that the shorter disinfection time does not raise any new or different questions of safety and effectiveness.
Additionally, since the predicate device materials were unknown, appropriate biocompatibility tests were performed as per ISO 10993-1 to ensure the safe use of PureHub™. The biocompatibility tests performed on the subject device are identified below.
| | | | The predicate device has not been subjected to design-related call. | | |
|--|--|--|---------------------------------------------------------------------|--|--|
| | | | | | |
| Non-Clinical<br>Testing | BD has performed the following performance tests in<br>accordance with 21 CFR §820.30 to demonstrate that the<br>PureHub™ Disinfecting Cap performs equivalent to the<br>predicate device. |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The following tests were performed on the subject device to an<br>internal specification or a Standard:<br>In vitro Antimicrobial Efficacy Staphylococcus aureus Staphylococcus epidermidis Escherichia coli Pseudomonas aeruginosa Candida albicans Candida glabrata Acinetobacter baumannii Particulate Matter Ingress USP 788 70% IPA Weight and % Isopropyl Alcohol/Water Solution Concentration Cap to Connector Interface Device Retention to Luer activated valve (LAV) Material Compatibility of subject device and LAV Air Leakage Weld Retention Packaging Integrity Tensile Strength on IV Pole Strip PureHub™ Connector Air Leak Testing Physical Barrier Testing |
| | As per ISO 10993-1:2009, the following biological tests were |
{8}------------------------------------------------
| | performed: |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | • Cytotoxicity<br>• Sensitization<br>• Irritation or Intracutaneous Activity<br>• Acute Systemic Toxicity<br>• Material Mediated Pyrogenicity<br>• LAL Endotoxin<br>• Hemocompatibility<br>• Subacute/Subchronic |
| | Additionally, the following tests were performed –<br>• Chemical Extractable Analysis<br>• 70% IPA Ingress Testing |
| | The BD PureHub™ Disinfecting Cap is sterilized by Gamma<br>radiation. The sterilization process was validated in accordance<br>with ISO 11137-2:2013 (VDmax25). |
| | The subject device met all the predetermined acceptance<br>criteria for the above listed performance tests and<br>demonstrated substantial equivalence to the predicate device. |
| Clinical Testing | Not applicable. |
| Summary of<br>Substantial<br>Equivalence | The differences between the predicate and the subject device<br>do not raise any new or different questions of safety or<br>effectiveness. The PureHub™ Disinfecting Cap is substantially<br>equivalent to Curos™ Port Protector cleared under K111992<br>with respect to the indications for use, target population,<br>treatment method and technological characteristics. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.