The ConforMiS Unicondylar Knee Replacement System (iUni) is intended for use in one compartment of the osteoarthritic knee to replace the damaged area of the articular surface with evidence of adequate healthy bone sufficient for support of the implanted components. Candidates for unicondylar knee replacement include those with: - Joint impairment due to osteoarthritis or traumatic arthritis of the knee - Previous femoral condyle or tibial plateau fracture, creating loss of function - Valgus or varus deformity of the knee - Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans. This implant is intended for cemented use only. The ConforMlS iDuo Bicompartmental Knee Repair System is intended for use in patients with severe knee joint pain and disability whose conditions cannot be solelyaddressed by the use of a prosthetic device that treats only a single knee compartment, such as unicondylar or patellofemoral prosthesis. The indications for use include restoring joint function and relief of pain due to: - Painful joint disease due to osteoarthritis - Traumatic arthritis of the knee - Post traumatic loss of joint function - Failed osteotomies, hemiarthroplasties and unicondylar implants The Duo Bicompartmental Knee Repair System may be utilized when the medial or lateral condyle and the patellofemoral areas have been affected by one or more of the above noted conditions. The iDuo implant is intended for cemented use only. The iTotal PS Knee Replacement System (KRS) is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis. The Indications for Use include: - Painful joint disease due to osteoarthritis, traumatoid arthritis, polyarthritis, polyarthritis or osteonecrosis of the knee. - Post traumatic loss of joint function. - Moderate varus, valgus or flexion deformity. - Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral implants. - Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans. This implant is intended for cemented use only. The iTotal Cruciate Retaining (CR) Knee Replacement System (KRS) is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis. The Indications for Use include: - Painful joint disease due to osteoarthritis, traumatic arthritis, theumatoid arthritis or osteonecrosis of the knee. - Post traumatic loss of joint function. - Moderate varus, valgus or flexion deformity in which the ligamentous structures can be returned to adequate function and stability. - Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral or bicompartmental implants. - Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans. This implant is intended for cemented use only.
Device Story
Patient-specific semi-constrained knee implants; designed using patient CT or MR imaging to match individual anatomy; femoral components made of cobalt chromium molybdenum (CoCrMo) alloy; tibial components include CoCrMo metal tray with UHMWPE inserts or all-polyethylene; patellar components made of UHMWPE; intended for cemented use only; used by orthopedic surgeons to treat severe knee pain/disability; restores joint function; provides patient-specific fit.
Clinical Evidence
Bench testing only. Sterilization validation performed to establish SAL of 1x10^-6, including VHP residual testing and product/packaging compatibility assessment.
Indicated for patients with knee joint pain/disability due to osteoarthritis, traumatic/rheumatoid arthritis, osteonecrosis, or post-traumatic loss of function. Applicable for unicondylar, bicompartmental, or total knee replacement (CR/PS) when single-compartment devices are insufficient. Includes revision procedures if anatomic landmarks are identifiable. Contraindicated for patients lacking adequate healthy bone for implant support.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Predicate Devices
iTotal Cruciate Retaining (CR) Knee Replacement System (K180906)
iUni Unicondylar Knee Replacement System (K132640, K133256)
iDuo Bicompartmental Knee Replacement System (K093513, K133256)
iTotal Posterior Stabilized (PS) Knee Replacement System (K161668)
Submission Summary (Full Text)
{1}------------------------------------------------
March 13, 2020
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Conformis, Inc. Nancy Giezen Manager Regulatory Affairs 600 Technology Park Drive Billerica. Massachusetts 01821
Re: K193105
Trade/Device Name: iUni Unicondylar Knee Replacement System, iDuo Bicompartmental Knee Repair System, iTotal CR Knee Replacement System, iTotal PS Knee Replacement System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH, OIY, OOG, NPJ, HSX Dated: January 8, 2020 Received: January 14, 2020
Dear Nancy Giezen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Dr.Ting Song, PhD, RAC Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193105
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
Device Name
iUni Unicondylar Knee Replacement System
Indications for Use (Describe)
The ConforMiS Unicondylar Knee Replacement System (iUni) is intended for use in one compartment of the osteoarthritic knee to replace the damaged area of the articular surface with evidence of adequate healthy bone sufficient for support of the implanted components.
Candidates for unicondylar knee replacement include those with:
- · Joint impairment due to osteoarthritis or traumatic
- arthritis of the knee
- · Previous femoral condyle or tibial plateau fracture,
- creating loss of function
- · Valgus or varus deformity of the knee
· Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans.
This implant is intended for cemented use only.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| <div> <span> <b> ☑ </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known)
Device Name
iDuo Bicompartmental Knee Repair System
#### Indications for Use (Describe)
The ConforMlS iDuo Bicompartmental Knee Repair System is intended for use in patients with severe knee joint pain and disability whose conditions cannot be solelyaddressed by the use of a prosthetic device that treats only a single knee compartment, such as unicondylar or patellofemoral prosthesis.
The indications for use include restoring joint function and relief of pain due to:
- · Painful joint disease due to osteoarthritis
- · Traumatic arthritis of the knee
- Post traumatic loss of joint function
- · Failed osteotomies, hemiarthroplasties and unicondylar implants
The Duo Bicompartmental Knee Repair System may be utilized when the medial or lateral condyle and the patellofemoral areas have been affected by one or more of the above noted conditions.
The iDuo implant is intended for cemented use only.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### Device Name
iTotal Posterior Stabilized {PS) Knee Replacement System
Indications for Use (Describe)
The iTotal PS Knee Replacement System (KRS) is intended for use as a total knee
replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis.
The Indications for Use include:
· Painful joint disease due to osteoarthritis, traumatoid arthritis, polyarthritis, polyarthritis or osteonecrosis of the knee.
· Post traumatic loss of joint function.
· Moderate varus, valgus or flexion deformity.
· Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral implants.
· Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans.
This implant is intended for cemented use only.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|
| <span style="text-decoration: underline;">Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="text-decoration: none;">Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
| <div> <span style="font-size: 20px;">☑</span> </div> | <div> <span>☐</span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### Device Name
iTotal Cruciate Retaining (CR) Knee Replacement System
#### Indications for Use (Describe)
The iTotal Cruciate Retaining (CR) Knee Replacement System (KRS) is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis.
The Indications for Use include:
- · Painful joint disease due to osteoarthritis, traumatic arthritis, theumatoid arthritis or osteonecrosis of the knee.
- · Post traumatic loss of joint function.
- · Moderate varus, valgus or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
- · Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral or bicompartmental implants.
- · Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans.
This implant is intended for cemented use only.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div> <span> <span style="font-size:16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
|-------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{7}------------------------------------------------
## 510(k) SUMMARY
| Submitter's Name and Address: | Conformis, Inc.<br>600 Technology Park Drive<br>Billerica, MA 01821 |
|---------------------------------------|-----------------------------------------------------------------------|
| Establishment Registration Number(s): | 3009844603 and 3004153240 |
| Date Summary was Prepared: | November 5, 2019 |
| Contact Person: | Nancy Giezen<br>Manager Regulatory Affairs<br>Telephone: 781-345-9058 |
## Trade/Device Name(s):
iUni Unicondylar Knee Replacement System iDuo Bicompartmental Knee Repair System iTotal Cruciate Retaining (CR) Knee Replacement System iTotal Posterior Stabilized (PS) Knee Replacement System
# Common Name:
Knee Replacement System
# Device Class:
Class II
# Regulation Numbers:
888.3560 888.3520
## Classification Names and Product Codes:
iUni
Prosthesis, knee, femorotibial, nonconstrained, cemented, metal/polymer Knee Arthroplasty Implantation System; OOG, HSX
iDuo
Prosthesis, Knee, Patellofemorotibial, Partial, Semi-Constrained, Cemented, Polymer/Metal/Polymer Knee Arthroplasty Implantation System; NPJ
{8}------------------------------------------------
iTotal CR and PS Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/PolymerKnee Arthroplasty Implantation System; JWH, OIY, OOG
# Legally Marketed Predicate Device (Primary Predicate):
iTotal Cruciate Retaining (CR) Knee Replacement System (K180906)
# (Ssecondary Predicates):
iUni Unicondylar Knee Replacement System (K132640, K133256) iDuo Bicompartmental Knee Replacement System (K093513, K133256) iTotal Posterior Stabilized (PS) Knee Replacement System (K161668)
# Device Description:
ConforMIS knee replacement systems are patient-specific semi-constrained knee implants which consist of a femoral, tibial, and/or patellar components. The products are intended for treatment of severe pain and/or disability of the knee damaged by osteoarthritis ortrauma.
Using patient imaging (either CT or MR scans), a patient-specific implant is designed that best meets the geometric and anatomic requirements of the specific patient. The femoral components of the devices are manufactured from cobalt chromium molybdenum (CoCrMo) alloy. The tibial component includes a metal tray manufactured from CoCrMo and polyethylene inserts manufactured from UHMWPE or an all polyethylene tibial component. The patellar components are manufactured from UHMWPE.
# Indications for Use:
iUni
The ConforMiS Unicondylar Knee Replacement System (iUni) is intended for use in one compartment of the osteoarthritic knee to replace the damaged area of the articular surface in patients with evidence of adequate healthy bone sufficient for support of the implanted components.
Candidates for unicondylar knee replacement include those with:
- · Joint impairment due to osteoarthritis or traumatic
- arthritis of the knee
- · Previous femoral condyle or tibial plateau fracture,
- creating loss of function
- · Valgus or varus deformity of the knee
- · Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans.
This implant is intended for cemented use only.
iDuo
The ConforMlS iDuo Bicompartmental Knee Repair System is intended for use in patients with
{9}------------------------------------------------
severe knee joint pain and disability whose conditions cannot be solelyaddressed by the use of a prosthetic device that treats only a single knee compartment, such as unicondylar or patellofemoral prosthesis. The indications for use include restoring joint function and relief of pain due to: • Painful joint disease due to osteoarthritis
· Traumatic arthritis of the knee
- Post traumatic loss of joint function
- · Failed osteotomies, hemiarthroplasties and unicondylar implants
The Duo Bicompartmental Knee Repair System may be utilized when the medial or lateral condyle and the patellofemoral areas have been affected by one or more of the above noted conditions.
The iDuo implant is intended for cemented use only.
# iTotal CR
The iTotal Cruciate Retaining (CR) Knee Replacement System (KRS) is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis. The Indications for Use include:
• Painful joint disease due to osteoarthritis, traumatic arthritis, rheumatoid arthritis or osteonecrosis of the knee.
· Post traumatic loss of joint function.
· Moderate varus, valgus or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
· Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral or bicompartmental implants.
• Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans. This implant is intended for cemented use only.
# iTotal PS
The iTotal PS Knee Replacement System (KRS) is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis. The Indications for Use include:
· Painful joint disease due to osteoarthritis, traumatic arthritis, rheumatoid arthritis, polyarthritis or osteonecrosis of the knee.
- Post traumatic loss of joint function.
- Moderate varus, valgus or flexion deformity.
{10}------------------------------------------------
· Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral or bicompartmental implants.
• Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans.
This implant is intended for cemented use only.
# Technological Characteristics:
The subject devices were compared to the predicates and the design features, materials, sizes and methods of sterilization (EO, VHP, steam) are identical. This 510(k) adds the option to sterilize device components using the Steris VHP LTS-V sterilizer.
# Non-Clinical Performance Evaluation:
The following non-clinical laboratory testing was performed, confirming that the device components are effectively sterilized using Vaporized Hydrogen Peroxide.
- · Sterilization Validation testing to establish a SAL of 1x10-6
- · VHP residual testing
- Product and packaging compatibility with VHP sterilization
# Conclusion:
Based on the testing conducted it is concluded that the subject devices are substantially equivalent to the predicate devices and can be sterilized to a SAL of 1x 10-6 ustilizing VHP sterilization.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.