← Product Code [ONF](/productcode/ONF) · K193072

# Lucent : IPL (K193072)

_Luvo Medical Technologies, Inc. · ONF · Mar 5, 2020 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K193072

## Device Facts

- **Applicant:** Luvo Medical Technologies, Inc.
- **Product Code:** [ONF](/productcode/ONF.md)
- **Decision Date:** Mar 5, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Lucent : IPL is Intended for use in aesthetic applications requiring selective photothermolysis (photocoagulation or coagulation) and hemostasis of soft tissue in the medical specialties of general and plastic surgery, and dermatology.

## Device Story

Lucent : IPL is a multi-wavelength, non-invasive, powered light-based system for dermatologic and general surgical procedures. It utilizes a console, electronics, software, cooling system, and a handpiece with six interchangeable filters (wavelength ranges 430-980nm). The device operates via selective photothermolysis, delivering light energy to target soft tissue for coagulation, hemostasis, or hair follicle destruction. It is intended for use by clinicians in medical settings. The provider selects the appropriate filter based on the target lesion or hair removal goal, adjusts pulse width and energy density, and applies the handpiece to the skin. The device output—controlled light pulses—induces thermal effects in targeted chromophores (melanin, hemoglobin), facilitating lesion clearance or hair reduction. This provides a non-invasive aesthetic treatment option for patients.

## Clinical Evidence

No clinical data was required for this submission. Substantial equivalence was established through bench testing and comparison of technical specifications to predicate devices.

## Technological Characteristics

Multi-wavelength non-laser light-based system. Wavelength range: 430-980nm. Pulse width: 3-35ms. Energy density: 6-40 J/cm2. Spot size: 15mm x 40mm. Includes system console, cooling system, and handpiece with six filters. Complies with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), and IEC 60601-2-57 (non-laser light source equipment).

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Alma Harmony XL Multi-Application Platform ([K072564](/device/K072564.md))
- Lumenis M22 System ([K142860](/device/K142860.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.

LUVO Medical Technologies, Inc. Mr. Gregory Berzak Regulatory Affairs Officer 125 Fleming Dr Cambridge. N1T 2B8 Ca

Re: K193072

Trade/Device Name: Lucent : IPL Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: ONF Dated: October 28, 2019 Received: November 4, 2019

Dear Gregory Berzak:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Mavadia-Shukla Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K193072

Device Name

Lucent : IPL

### Indications for Use (Describe)

The Lucent : IPL is Intended for use in aesthetic applications requiring selective photothermolysis (photocoagulation or coagulation) and hemostasis of soft tissue in the medical specialties of general and plastic surgery, and dermatology.

Indication for Use for the 430-980nm wavelength

· Benign pigmented epidermal and cutaneous lesions including dyschromia, hyperpigmentation, melasma, Ephelides (freckles), and striae.

· Benign cutaneous vascular lesions, including port wine stains, facial and truncal telangiectasias, angiomas and spider angiomas, and poikiloderma of Civatte.

· For use on Fitzpatrick skin types (I-V)

Indication for use for the 515-980nm wavelength

· Benign pigmented epidermal and cutaneous lesions including dyschromia, hyperpigmentation, melasma, Ephelides (freckles). and striae.

· Benign cutaneous vascular lesions, including port wine stains, facial and truncal telangiectasias, angiomas and spider angiomas, poikiloderma of Civatte and Lentigines.

· For use on Fitzpatrick skin types (I-V)

Indication for use for the 560-980nm wavelength

· Benign pigmented epidermal and cutaneous lesions including dyschromia, hyperpigmentation, melasma, Ephelides (freckles), and striae.

· Benign cutaneous vascular lesions, including port wine stains, facial and truncal telangiectasias, angiomas and spider angiomas, poikiloderma of Civatte and Lentigines.

· The removal of unwanted hair to effect stable long-term or permanent hair reduction

· For use on Fitzpatrick skin types (I-V)

Indication for use for the 585-980nm wavelength

· Benign pigmented epidermal and cutaneous lesions including dyschromia, hyperpigmentation, melasma, Ephelides (freckles), and striae.

· Benign cutaneous vascular lesions, including port wine stains, facial and truncal telangiectasias, angiomas and spider angiomas, and poikiloderma of Civatte.

· The removal of unwanted hair to effect stable long-term or permanent hair reduction

· For use on Fitzpatrick skin types (I-V)

{3}------------------------------------------------

Indication for use for the 640-980nm wavelength

· Benign pigmented epidermal and cutaneous lesions including dyschromia, hyperpigmentation, melasma, Ephelides (freckles), and striae.

- · Benign cutaneous vascular lesions, including port wine stains, facial and truncal telangiectasias, angiomas and spider angiomas, and poikiloderma of Civatte.
- · The removal of unwanted hair to effect stable long-term or permanent hair reduction
- · For use on Fitzpatrick skin types (I-V)

Indication for use for the 700-980nm wavelength

• The removal of unwanted hair and to effect stable long-term or permanent hair reduction. Permanent reduction in hair regrowth is defined as the long-term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime

· For use on Fitzpatrick skin types (I-V)

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

This 510(K) Summary of safety and effectiveness for the Lucent : IPL is submitted in accordance with the requirements of the SMDA 1990 and following guidance concerning the organization and content of a 510(K) summary.

| Applicant:                          | LUVO Medical Technologies, Inc.                                                                                                                                                                                                                                      |
|-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address:                            | LUVO Medical Technologies, Inc.<br>125 Fleming Dr<br>Cambridge, Ontario, Canada N1T 2B8                                                                                                                                                                              |
| Contact Person:                     | Mr. Gregory Berzak                                                                                                                                                                                                                                                   |
| Telephone:                          | 519-6203900- phone<br>gregoryb@clarionmedical.com                                                                                                                                                                                                                    |
| Preparation Date:                   | October 15, 2019                                                                                                                                                                                                                                                     |
| Device Trade Name:                  | Lucent : IPL                                                                                                                                                                                                                                                         |
| Common Name:                        | Powered light based non-laser surgical instrument with<br>thermal effect                                                                                                                                                                                             |
| Regulation Name:                    | Laser surgical instrument for use in general and<br>plastic surgery and dermatology                                                                                                                                                                                  |
| Regulation Number:                  | 21 CFR 878.4810 (Product Code:  ONF)                                                                                                                                                                                                                                 |
| Legally Marketed Predicate Devices: | Primary Predicate : Alma Harmony XL Multi-Application<br>Platform / Secondary Predicate: Lumenis M22 System                                                                                                                                                          |
| 510(K) number:                      | K072564/ K142860                                                                                                                                                                                                                                                     |
| Regulatory Class:                   | Class II Prescription Use                                                                                                                                                                                                                                            |
| Description of the Lucent : IPL:    | The LUVO Medical Technologies, Inc. Lucent : IPL isa<br>multi-wavelength non-invasive system for IPL skin<br>treatments, treatment of vascular and pigmented lesions,<br>and hair removal using multiple filters reflecting a<br>wavelength range of 430nm to 980nm. |
|                                     | The Lucent : IPL system consists of a system console,<br>electronics and software, cooling system, and handpiece<br>with six filters                                                                                                                                 |
| Intended use of Lucent : IPL:       | The Lucent : IPL is intended for use in dermatologicand<br>general surgical procedures. See the below Indications for<br>Use Comparison Table.                                                                                                                       |
| Performance Data:                   | The following performance data was provided in support<br>of<br>5-1                                                                                                                                                                                                  |

{5}------------------------------------------------

the substantial equivalence determination:

ES60601-1:2005/(R)2012 and A1:2012 Test for Medical Electrical equipment was performed for General Requirements for basic safety and essential performance;

IEC 60601-1-2 Edition 4.0 2014-02 Test for Medical Equipment for General Requirements for basic safety and essential performance: electromagnetic compatibility

IEC 60601-2-57 Edition 1.0 2011-01 Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use

Results of Clinical Study:

A human clinical study was not required as the device is substantially equivalent to the predicate devices.

Technical Specifications / Indications for Use Comparison:

|                    | 510(K) Submission<br>Lucent : IPL<br>(K193072)                                                                                                                                                                                                  | Predicate<br>K142860 - Lumenis<br>M22 System                                                                                                                                                                                                        |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic     |                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                     |
| Wavelength<br>(nm) | 430-980nm                                                                                                                                                                                                                                       | 400-1200nm                                                                                                                                                                                                                                          |
| Intended Use       | Intended for use in aesthetic<br>applications requiring selective<br>photothermolysis<br>(photocoagulation or coagulation)<br>and hemostasis of soft tissue in the<br>medical specialties of general<br>and plastic surgery, and<br>dermatology | Intended for use in aesthetic and cosmetic<br>applications requiring selective photothermolysis<br>(photocoagulation or coagulation) and hemostasis of<br>soft tissue in the medical specialties of general<br>and plastic surgery, and dermatology |

{6}------------------------------------------------

| Indications for<br>Use                    | Benign pigmented<br>epidermal and cutaneous<br>lesions including<br>dyschromia,<br>hyperpigmentation,<br>melasma, Ephelides<br>(freckles), and striae. Benign cutaneous vascular<br>lesions, including port wine<br>stains, hemangiomas, facial<br>and truncal telangiectasias,<br>angiomas and spider<br>angiomas, and<br>poikiloderma of Civatte. For use on Fitzpatrick skin<br>types (I-V) | Mild to Moderate inflammatory<br>(Acne Vulgaris) Benign epidermal<br>lesions, including dyschromia,<br>hyperpigmentation, melasma,<br>ephelides (freckles) and tattoos Cutaneous lesions, including warts,<br>scars and striae Benign cutaneous vascular lesions,<br>including port wine stains,<br>hemangiomas, facial, truncal and leg<br>telangiectasias, Erythema of rosacea, angiomas and<br>spider angiomas, poikiloderma of<br>Civatte, leg veins and venous<br>malformations Removal of unwanted hair from all<br>skin types, and to effect stable long<br>term, or permanent* hair reduction in<br>skin types I-V through selective targeting of<br>melanin in hair follicles |
|-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Pulse Width<br>(msec)                     | 3-35                                                                                                                                                                                                                                                                                                                                                                                           | 4-20                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Energy<br>Density<br>(Fluence)<br>(J/cm2) | 6-40                                                                                                                                                                                                                                                                                                                                                                                           | 10-35                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Spot Size<br>(mm)                         | 15mm x 40mm                                                                                                                                                                                                                                                                                                                                                                                    | 15mm x 35mm / 8mm x 15mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |

|                    | 510(K) Submission<br>Lucent : IPL<br>(K193072)                                                                                                                                                                                                  | Predicate<br>K072564 - Alma<br>VL515 IPL handpiece                                                                                                                                                                                                  |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic     |                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                     |
| Wavelength<br>(nm) | 515-980nm                                                                                                                                                                                                                                       | 515nm-950nm                                                                                                                                                                                                                                         |
| Intended Use       | Intended for use in aesthetic<br>applications requiring selective<br>photothermolysis<br>(photocoagulation or coagulation)<br>and hemostasis of soft tissue in the<br>medical specialties of general<br>and plastic surgery, and<br>dermatology | Intended for use in aesthetic and cosmetic<br>applications requiring selective photothermolysis<br>(photocoagulation or coagulation) and hemostasis of<br>soft tissue in the medical specialties of general<br>and plastic surgery, and dermatology |

{7}------------------------------------------------

| Indications for<br>Use                      | • Benign pigmented<br>epidermal and cutaneous<br>lesions including<br>dyschromia,<br>hyperpigmentation,<br>melasma, Ephelides<br>(freckles), and striae.<br><br>• Benign cutaneous vascular<br>lesions, including port wine<br>stains, hemangiomas, facial<br>and truncal telangiectasias,<br>angiomas and spider<br>angiomas, poikiloderma of<br>Civatte and Lentigines<br><br>• For use on Fitzpatrick skin<br>types (I-V) | • Moderate inflammatory acne (acne<br>vulgaris)<br><br>• Tattoos and benign pigmented epidermal<br>and cutaneous lesions including warts,<br>scars ,striae; dyschromia,<br>hyperpigmentation, melasma, epithelides<br>(freckles), lentigines, nevi, and café-au-lait<br>macules<br><br>• Benign cutaneous vascular lesions,<br>including port wine stains,hemangiomas,<br>facial, truncal and legtelangiectasias,<br>rosacea, erythema of rosacea, angiomas and<br>spider angiomas, poikiloderma of Civatte,<br>leg veins and venous malformations<br><br>• For use on skin types(I-V) |
|---------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Pulse Width<br>(msec)                       | 3-35                                                                                                                                                                                                                                                                                                                                                                                                                         | 10-15                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Energy<br>Density<br>(Fluence)<br>$(J/cm2)$ | 6-40                                                                                                                                                                                                                                                                                                                                                                                                                         | 10-30                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Spot Size<br>(mm)                           | 15mm x 40mm                                                                                                                                                                                                                                                                                                                                                                                                                  | 3cm2 and 6.4 cm2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |

|                | 510(K) Submission<br>Lucent : IPL<br>(K193072)                                                                                                                                                                                               | Predicate<br>K072564 - Alma IPL handpiece VL/PL, VP<br>and SSR                                                                                                                                                                                         |
|----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic |                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                        |
| Wavelength     | 560-980nm                                                                                                                                                                                                                                    | 540-950nm                                                                                                                                                                                                                                              |
| Intended Use   | Intended for use in aesthetic<br>applications requiring selective<br>photothermolysis (photocoagulation or<br>coagulation) and hemostasis of soft<br>tissue in the medical specialties of<br>general<br>and plastic surgery, and dermatology | Intended for use in aesthetic and cosmetic<br>applications requiring selective<br>photothermolysis (photocoagulation or<br>coagulation) and hemostasis of soft tissue in<br>the medical specialties of general and plastic<br>surgery, and dermatology |

{8}------------------------------------------------

| Indications for Use                    | Benign pigmented epidermal and<br>cutaneous lesions including<br>dyschromia, hyperpigmentation,<br>melasma, Ephelides (freckles),<br>and striae. Benign cutaneous vascular<br>lesions, including port wine<br>stains, hemangiomas, facial and<br>truncal telangiectasias, angiomas<br>and spider angiomas,<br>poikiloderma of Civatte and<br>Lentigines The removal of unwanted hair to<br>effect stable long-term or<br>permanent hair reduction For use on Fitzpatrick skin types<br>(I-V) | The treatment of moderate<br>inflammatory acne vulgaris. The treatment of benign pigmented<br>epidermal lesions including dyschromia,<br>hyperpigmentation, melasma, and<br>ephelides (freckles), lentigines, nevi,<br>melasma, and cafe-au-lait macules. The treatment of cutaneous<br>lesions including warts, scars and<br>striae. The treatment of benign cutaneous<br>vascular lesions including port wine<br>stains, hemangiomas, facial, truncal and<br>leg telangiectasias, rosacea, erythema or<br>rosacea, angiomas and spider angiomas,<br>poikiloderma of Civatte, leg veins and<br>venous malformations. The removal of unwanted hair to<br>effect stable long-term or permanent hair<br>reduction. Use on all skin types (Fitzpatrick I-VI). |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Pulse Width<br>(msec)                  | 3-35                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 10-15                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Energy Density<br>(Fluence)<br>(J/cm2) | 6-40                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 1-15 and 10-30                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Spot Size<br>(mm)                      | 15mm x 40mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 3 cm2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |

|                    | 510(K) Submission<br>Lucent : IPL<br>(K193072)                                                                                                                                                                                            | Predicate K072564 –<br>Alma<br>SR IPL handpiece                                                                                                                                                                                                        |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic     |                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                        |
| Wavelength<br>(nm) | 585-980nm                                                                                                                                                                                                                                 | 570-950nm                                                                                                                                                                                                                                              |
| Intended Use       | Intended for use in aesthetic applications<br>requiring selective photothermolysis<br>(photocoagulation or coagulation) and<br>hemostasis of soft tissue in the medical<br>specialties of general and plastic surgery,<br>and dermatology | Intended for use in aesthetic and cosmetic<br>applications requiring selective photothermolysis<br>(photocoagulation or coagulation) and hemostasis<br>of soft tissue in the medical specialties of general<br>and plastic<br>surgery, and dermatology |

{9}------------------------------------------------

| Indications for<br>Use                    | Benign pigmented epidermal and cutaneous lesions including dyschromia, hyperpigmentation, melasma, Ephelides (freckles), and striae. Benign cutaneous vascular lesions, including port wine stains, hemangiomas, facial and truncal telangiectasias, angiomas and spider angiomas, and poikiloderma of Civatte. The removal of unwanted hair to effect stable long-term or permanent hair reduction. For use on Fitzpatrick skin types (I-V) | The treatment of moderate inflammatory acnevulgaris. The treatment of benign pigmented epidermal lesions, including dyschromia, hyperpigmentation, melasma and ephelides (freckles). The treatment of face and body vascular and pigmentedlesions. The treatment of cutaneous lesions, including scars and striae. The treatment of benign cutaneous vascular lesions, including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, rosacea, melasma, angiomas and spider angiomas, poikiloderma of Civatte, leg veins and venous malformations. The removal of unwanted hair to effect stable long-term or permanent hair reduction. Use on all skin types (Fitzpatrick I-VI). |
|-------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Pulse Width<br>(msec)                     | 3-35                                                                                                                                                                                                                                                                                                                                                                                                                                         | 10-15                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Energy<br>Density<br>(Fluence)<br>(J/cm2) | 6-40                                                                                                                                                                                                                                                                                                                                                                                                                                         | 10-25                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Spot Size<br>(mm)                         | 15mm x 40mm                                                                                                                                                                                                                                                                                                                                                                                                                                  | 6.4 cm2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |

|                    | 510(K) Submission<br>Lucent : IPL<br>(K193072)                                                                                                                                                                                            | Predicate<br>K072564 - Alma<br>HR Module AFT<br>Handpiece                                                                                                                                                                                              |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic     |                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                        |
| Wavelength<br>(nm) | 640-980nm                                                                                                                                                                                                                                 | 650-950nm                                                                                                                                                                                                                                              |
| Intended Use       | Intended for use in aesthetic applications<br>requiring selective photothermolysis<br>(photocoagulation or coagulation) and<br>hemostasis of soft tissue in the medical<br>specialties of general and plastic surgery,<br>and dermatology | Intended for use in aesthetic and cosmetic<br>applications requiring selective photothermolysis<br>(photocoagulation or coagulation) and hemostasis<br>of soft tissue in the medical specialties of general<br>and plastic<br>surgery, and dermatology |

{10}------------------------------------------------

| Indications for<br>Use                    | Benign pigmented epidermal and cutaneous lesions including dyschromia, hyperpigmentation, melasma, Ephelides (freckles), and striae. Benign cutaneous vascular lesions, including port wine stains, hemangiomas, facial and truncal telangiectasias, angiomas and spider angiomas, and poikiloderma of Civatte. The removal of unwanted hair to effect stable long-term or permanent hair reduction. For use on Fitzpatrick skin types (I-V) | The treatment of moderate inflammatory acnevulgaris. The treatment of benign pigmented epidermal lesions, including dyschromia, hyperpigmentation, melasma and ephelides (freckles). The treatment of face and body vascular and pigmentedlesions. The treatment of cutaneous lesions, including scars and striae. The treatment of benign cutaneous vascular lesions, including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, rosacea, melasma, angiomas and spider angiomas, poikiloderma of Civatte, leg veins and venous malformations. The removal of unwanted hair to effect stable long-term or permanent hair reduction. Use on all skin types (Fitzpatrick I-VI) |
|-------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Pulse Width<br>(msec)                     | 3-35                                                                                                                                                                                                                                                                                                                                                                                                                                         | 30-50                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Energy<br>Density<br>(Fluence)<br>(J/cm2) | 6-40                                                                                                                                                                                                                                                                                                                                                                                                                                         | 5-25                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Spot Size<br>(mm)                         | 15mm x 40mm                                                                                                                                                                                                                                                                                                                                                                                                                                  | 6.4 cm²                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |

|                    | 510(K) Submission<br>Lucent : IPL<br>(K193072)                                                                                                                                                                                            | Predicate<br>K072564 - Alma SHR IPL                                                                                                                                                                                                                       |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic     |                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                           |
| Wavelength<br>(nm) | 700-980nm                                                                                                                                                                                                                                 | 780-950nm                                                                                                                                                                                                                                                 |
| Intended Use       | Intended for use in aesthetic applications<br>requiring selective photothermolysis<br>(photocoagulation or coagulation) and<br>hemostasis of soft tissue in the medical<br>specialties of general<br>and plastic surgery, and dermatology | Intended for use in aesthetic and cosmetic<br>applications requiring selective<br>photothermolysis (photocoagulation or<br>coagulation) and hemostasis of soft tissue in<br>the medical specialties of general and<br>plastic surgery, and<br>dermatology |

{11}------------------------------------------------

| Indications for Use                    |                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                            |         |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
|                                        | The removal of unwanted hair and to effect stable long-term or permanent hair reduction. Permanent reduction in hair regrowth is defined as the long- term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime Use on skin types (Fitzpatrick I-V) | The treatment of pseudofolliculitis barbae(PFB). The removal of unwanted hair and to effect stable long-term or permnanent hair reduction. Use on all skin types (Fitzpatrick I-VI), including tannedskin. |         |
| Pulse Width<br>(msec)                  | 3-35                                                                                                                                                                                                                                                                                                                              | ≤30msec/1,3 and 30 sec                                                                                                                                                                                     | ≤6msec  |
| Energy Density<br>(Fluence)<br>(J/cm2) | 6-40                                                                                                                                                                                                                                                                                                                              | 1-7 and 0.5-1.5                                                                                                                                                                                            | 1-7     |
| Spot Size<br>(mm)                      | 15mm x 40mm                                                                                                                                                                                                                                                                                                                       | 3 cm2                                                                                                                                                                                                      | 6.4 cm2 |

The Lucent : IPL's intended use, indications for use and technical specifications Conclusion: are substantially equivalent to the Harmony XL Multi-Application Platform and the M22 System.

---

**Source:** [https://fda.innolitics.com/device/K193072](https://fda.innolitics.com/device/K193072)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
