K193039 · Olympus Medical Systems Corp. · KNS · Dec 17, 2019 · Gastroenterology, Urology
Device Facts
Record ID
K193039
Device Name
Single Use Balloon Dilator V (with knife)
Applicant
Olympus Medical Systems Corp.
Product Code
KNS · Gastroenterology, Urology
Decision Date
Dec 17, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.4300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
These instruments have been designed to be used with an Olympus endoscope for papillotomy and for dilating the major papilla to retrieve biliary stones.
Device Story
Single Use Balloon Dilator V (with knife) is an endoscopic accessory used for papillotomy and biliary stone retrieval. Device consists of a balloon and a papillotomy knife. Operated by a physician during endoscopic procedures; device is inserted through a compatible Olympus endoscope. Principle of operation: high-frequency current applied to a cutting wire for tissue incision; inflation device attached to a balloon port injects fluid to dilate the papilla. V-marking on the insertion portion allows for visual confirmation of insertion length relative to the endoscope biopsy valve. Hook on the handle assists in manipulation. Device facilitates biliary stone retrieval by dilating the papilla and performing sphincterotomy, potentially reducing the need for more invasive surgical interventions.
Clinical Evidence
Bench testing only. Performance data included electrical safety (ANSI/AAMI ES 60601-1, IEC 60601-2-18), EMC (IEC 60601-1-2), sterilization/shelf-life (ASTM F1980-16), and biocompatibility (ISO 10993-1). Bench performance testing verified dimensions, endoscopic compatibility, knife operation, papillotomy resistance, tensile strength, balloon deflation time, balloon burst strength, balloon fatigue, and package integrity.
Technological Characteristics
4-lumen catheter with C-channel, integrated balloon, and papillotomy knife. Distal tip diameter 1.5mm; working length 1950mm. Balloon diameters: 8mm, 10mm, 12mm. Maximum pressure 3.5 bar. Pre-curved distal end. Materials evaluated per ISO 10993-1 for limited mucosal contact. Sterilized device. No software or electronic algorithm components.
Indications for Use
Indicated for use with an Olympus endoscope for papillotomy and dilation of the major papilla to retrieve biliary stones in patients requiring endoscopic biliary intervention.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
Predicate Devices
Single Use Balloon Dilator V (with knife) (K150142)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 17, 2019
Olympus Medical Systems Corp. Sheri Musgnung Manager, Regulatory Affairs Olympus Corporation of the Americas 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610
Re: K193039
> Trade/Device Name: Single Use Balloon Dilator V (with knife) Regulation Number: 21 CFR 876.4300 Regulation Name: Endoscopic Electrosurgical Unit and Accessories Regulatory Class: II Product Code: KNS, FGE Dated: October 30, 2019 Received: October 31, 2019
Dear Sheri Musgnung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Daniel G. Walter, Jr. Assistant Director THT3A3: Obesity and Hepatobiliary Devices DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
K193039
Device Name Single Use Balloon Dilator V (with Knife)
Indications for Use (Describe)
These instruments have been designed to be used with an Olympus endoscope for papillotomy and for dilating the major papilla to retrieve biliary stones.
Type of Use (*Select one or both, as applicable*)X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are slightly stylized, with a modern and clean font. Underneath the word "OLYMPUS", there is a horizontal yellow bar that spans the length of the word. The overall design is simple and corporate, likely representing the logo of the Olympus Corporation.
October 30, 2019
K193039 Page 1 of 5
# Section 5
# 510(k) Summary
#### 5.1 GENERAL INFORMATION
- 510(k) Submitter: OLYMPUS MEDICAL SYSTEMS CORP. 2951 Ishikawa-cho, Hachioji-shi, Tokyo, Japan 192-8507
- Sheri L. Musgnung Contact Person: Olympus Corporation of the America 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610, USA Phone: 484-896-3147 Fax: 484-896-7128 Email: sheri.musgnung@olympus.com
- Manufacturing site: Aomori Olympus Co., Ltd. 2-248-1 Okkonoki, Kuroishi-shi, Aomori 036-0357, Japan
#### 5.2 DEVICE IDENTIFICATION
| ■ Device Name | Single Use Balloon Dilator V (with knife) |
|---------------------|---------------------------------------------------------------------------------------|
| ■ Model Name | BD-VC431Q-1240-20<br>BD-VC431Q-1240-25<br>BD-VC431Q-1240-30 |
| ■ Common Name | Single Use Balloon Dilator (with knife) |
| ■ Regulation Number | 876.4300<br>876.5010 |
| ■ Regulation Name | Endoscopic electrosurgical unit and accessories.<br>Biliary catheter and accessories. |
| ■ Regulatory Class | II |
Section 5 510(k) Summary Page 1 of 5
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are evenly spaced and appear to be a sans-serif font. A thin, horizontal, yellow line is located underneath the word. The registered trademark symbol is located in the upper right corner of the image.
- Product Code KNS FGE
■ Classification Panel Gastroenterology/Urology
#### 5.3 PREDICATE DEVICE
#### Predicate device
| Device name | 510(k) Submitter | 510(k) No. |
|-------------------------------------------|-------------------------------|------------|
| Single Use Balloon Dilator V (with knife) | OLYMPUS MEDICAL SYSTEMS CORP. | K150142 |
#### 5.4 DEVICE DESCRIPTION
#### ■ General Description of the subject device
The subject device consists of the balloon and the papillotomy knife for endoscopic sphincterotomy.
The only difference from the predicate device which has been cleared under K150142 is the additional balloon diameter offerings
The intended use of the subject device is identical to the predicate device.
#### Principle of Operation
There are no differences in composition and basic principle compared with the predicate.
The subject device used in combination with compatible Olympus endoscopes, A-Cord (power cable), electrosurgical unit and compatible inflation device and the subject device is inserted endoscopically into the body cavity, and a tissue is cut by applying a high frequency current while the cutting wire is tightened.
In addition, an inflation device is attached to a balloon port, and the inflation fluid is injected by utilizing an inflation device to dilate the papilla.
The insertion portion of this product has a V-marking, and the relative insertion length of this product into the endoscope can be confirmed by the positional relationship between the V-marking and the biopsy valve of the endoscope.
The Hook provided on the handle is for attaching the subject device to an endoscope or an inflation device, which assists manipulation of the subject device.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. A thin, horizontal, yellow line is underneath the word. The registered trademark symbol is located to the right of the letter "S".
K193039 Page 3 of 5
#### 5.5 INDICATIONS FOR USE
These instruments have been designed to be used with an Olympus endoscope for papillotomy and for dilating the major papilla to retrieve biliary stones.
## 5.6 COMPARISON OF TECHNOLOGY CHARACTERISTICS WITH THE PREDICATE DEIVCE
The Single Use Balloon Dilator V (with knife) has the same technological characteristics and design as the predicate device except for the following new feature:
- । smaller balloon diameters
All other technological characteristics of both the subject and predicate device are identical.
Validation from non-clinical testing demonstrated that these technological features do not raise any new issues of safety or effectiveness of the subject device.
| Comparison Table | | |
|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| | Subject Device | Predicate Device |
| Item | BD-VC431Q-1240-20/25/30 | BD-VC431Q-1840-20/25/30 |
| Distal tip diameter (mm) | ø1.5 mm | ø1.5 mm |
| Tip length (mm) | 7 mm | 7 mm |
| Knife length (mm) | 20 mm / 25 mm / 30 mm | 20 mm / 25 mm / 30 mm |
| Maximum insertion portion<br>diameter (mm)<br>(before inflating the<br>balloon) | ø 3.65 mm | ø 3.65 mm |
| Working length (mm) | 1950 mm | 1950 mm |
| Balloon diameter (mm (Fr))<br>/ Pressure (bar (atm)) | ø 8 mm(24Fr) / 0.5bar(0.5atm)<br>ø 10mm(30Fr) / 2bar(2atm)<br>ø 12mm(36Fr) / 3.5bar(3.5atm) | ø 12 mm(36Fr) / 0.5bar(0.5atm)<br>ø 15mm(45Fr) / 2bar(2atm)<br>ø 18mm(54Fr) / 3.5bar(3.5atm) |
| Balloon length (mm) | 40mm | 40mm |
| Maximum pressure (bar<br>(atm)) | 3.5 bar (3.5 atm) | 3.5 bar (3.5 atm) |
| Shape of the distal end | Pre-curved | Pre-curved |
| Knife wire diameter (mm) | ø 0.2 mm | ø 0.2 mm |
| Coated portion | Provided | Provided |
| Lumen type | 4-Lumen with C-channel | 4-Lumen with C-channel |
Comparison Table
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, dark blue letters. The letters are closely spaced and appear to be a sans-serif font. A thin, horizontal gold bar is located beneath the word "OLYMPUS", adding a subtle visual element to the logo. The registered trademark symbol is located to the right of the letter "S".
K193039 Page 4 of 5
#### 5.7 PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
#### 1) Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the Single Use Balloon Dilator V (with knife). The system complies with the ANSVAAMI ES 60601-1:2005/(R)2012 and A1:2012 and IEC 60601-2-18:2009 standards for safety and the IEC 60601-1-2:2014 standards for EMC.
#### 2) Sterilization/Shelf-life testing
Sterilization/shelf-life testing for the Single Use Balloon Dilator V (with knife) was conducted in accordance with the FDA's Guidance for Industry and Food and Drug Administration Staff, "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile".
Accelerated aging test for the Single Use Balloon Dilator V (with knife) was conducted in accordance with ASTM F1980-16(2016), the standard guide for accelerated aging of sterile barrier systems for medical devices. The three years real-time aging test will be performed to demonstrate longer stability and support the results of the accelerated aging test.
#### 3) Biocompatibility testing
Biocompatibility testing for the Single Use Balloon Dilator V (with knife) were conducted in accordance with the FDA's Guidance for Industry and Food and Drug Administration Staff. Use of International Standard ISO 10993-1:2009. "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".
#### Contact classification:
The Single Use Balloon Dilator V (with knife) is considered a surface-contacting device in contact with mucosal membranes. The contact duration is limited exposure (i.e. contact is up to 24 hours).
#### 4) Performance testing - Bench
Bench testing for the Single Use Balloon Dilator V (with knife) as listed below was conducted to ensure that the subject device performs as intended and meet design specifications.
Dimensional verification Endoscopic compatibility
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are spaced closely together, creating a strong visual impact. Below the word, there is a horizontal yellow line that spans the width of the word. The overall design is simple and clean, with a focus on the brand name.
Traditional 510(k) Notification Single Use Balloon Dilator V (with knife) BD-VC431Q-1240
Knife operation Papillotomy resistance Tensile strength Balloon deflation time Balloon burst strength Balloon fatigue Package integrity
#### 5) Risk analysis
Risk analysis for the Single Use Balloon Dilator V (with knife) was conducted in accordance with established in-house acceptance criteria based on ISO 14971:2007. The design verification tests and their acceptance criteria were identified and performed as a result of this risk analysis assessment.
#### 5.8 CONCLUSIONS
Based on the indications for use, technological characteristics, performance testing and technological comparison to the predicate device, the Single Use Balloon Dilator V (with knife) raise no new issue of safety and effectiveness and are substantially equivalent to the predicate device in terms of safety, effectiveness, and performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.