← Product Code [GEI](/productcode/GEI) · K193019

# DANNIK Disposable Monopolar Laparoscopic Instrument (K193019)

_Dannik · GEI · Mar 4, 2020 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K193019

## Device Facts

- **Applicant:** Dannik
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** Mar 4, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The DANNIK Disposable Monopolar Laparoscopic Instrument have application in a variety of minimally invasive procedures to facilitate grasping, mobilization, dissection and transection of tissue.

## Device Story

Disposable monopolar laparoscopic instrument for minimally invasive surgery; used by surgeons to grasp, mobilize, dissect, and transect tissue. Device features handle and 360-degree rotation knob (ABS thermoplastic), 5mm diameter stainless steel shaft with high-density polyethylene insulation, and RF post for electrosurgical coagulation when connected to an external generator and ground electrode. Operated manually by surgeon in clinical/OR settings. Provides mechanical tissue manipulation and electrical coagulation; benefits include precise tissue handling and hemostasis during laparoscopic procedures.

## Clinical Evidence

No clinical data. Bench testing only, including performance and functionality evaluations: thermal effects on tissue, connection strength, clamping performance, cable connection, mechanical open/close performance, hardness, and corrosion resistance.

## Technological Characteristics

5mm diameter, 33cm length. Materials: ABS thermoplastic (handle/knob), medical-grade stainless steel (shaft), high-density polyethylene (insulation). Energy: Monopolar RF electrosurgery via external generator. Sterilization: Ethylene Oxide (ETO) per ISO 11135-1. Biocompatibility: ISO 10993 compliant. Manual operation.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Ethicon Endo-Surgery Endopath Endoscopic Instrument ([K984240](/device/K984240.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 4, 2020

DANNIK Olga Haberland Regulatory Compliance 941 W Morse Blvd., Suite 100 Winter Park, Florida 32789

Re: K193019

Trade/Device Name: DANNIK Disposable Monopolar Laparoscopic Instrument Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: February 18, 2020 Received: February 24, 2020

Dear Olga Haberland:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K193019

Device Name

DANNIK Disposable Monopolar Laparoscopic Instrument

Indications for Use (Describe)

The DANNIK Disposable Monopolar Laparoscopic Instrument have application in a variety of minimally invasive procedures to facilitate grasping, mobilization, dissection and transection of tissue.

| Type of Use (Select one or both, as applicable)                                                     |                                               |
|-----------------------------------------------------------------------------------------------------|-----------------------------------------------|
| <span style="font-family: Arial, sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |

| X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) /Summary

1. Contact Information DANNIK 941 West Morse Blvd. Suite #100 Winter Park, Florida 32789 Phone: (407) 745-1698 Olga Haberland, Regulatory Compliance January 30, 2020

- 2. Device Name
	- Trade Name DANNIK Disposable Monopolar Laparoscopic . Instrument
	- Common Name: Disposable Monopolar Laparoscopic Instrument .
	- Classification Name : Electrosurgical, Cutting & Coagulation Device . and Accessories
	- Classification: Class II General and Plastic Surgery Devices GEI, 21 . CFR878.4400

Performance Standards: Devices are manufactured according to Good Manufacturing Practices (G.M.P.), Association for Advancement of Medical Instrumentation (A.AM.I.) and American Society for Testing and Materials (A.S.T.M.) requirements and applicable Harmonized Standards ISO 13485.

- 3. Substantially Equivalent Device
	- Legally Marketed (unmodified Devices): .
	- Ethicon Endo-Surgery Endopath Endoscopic Instrument FDA 510K . (K984240)

005 1

{4}------------------------------------------------

# K193019

- 4. Device Description
The DANNIK Disposable Monopolar Laparoscopic Instrument consists of Handle which activates the instruments jaws and scissor blades, Rotator knob that provides 360 degrees rotation , both handle and rotation knob are made of ABS thermoplastic, and a Shaft made from Medical grade Stainless Steel and insulated with high-density polyethylene PE. 5mm in diameter with a working length of 33cm. RF Post for electrosurgery when attached to an approved electrosurgical generator to provide coagulation of tissue when used with appropriate ground electrode.

This device is packaged and sterilized for single use only. Do NOT re-use, reprocess or resterilize. Discard after use with care.

Biocompatibility Conforms to ISO 10993 Prescription Only Yes Sterilized by ETO ( Ethelyn Oxide )

- 5. Intended Use
The DANNIK Disposable Monopolar Laparoscopic Instrument have application in a variety of minimally invasive procedures to facilitate grasping, mobilization, dissection and transection of tissue.

- 6. Technological Characteristics of the Subject Device Compared to the Predicate Device
Comparison of Technological Characteristics: DANNIK Disposable Monopolar Laparoscopic Instrument is substantially equivalent to the predicate device.

There are no new technologies being added to this device from the predicate, in terms of finished device functions. The device has the same intended use and application as the predicate device.

{5}------------------------------------------------

| Device             | Dannik Disposable Monopolar<br>Laparoscopic Instrument                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Ethicon Endo-Surgery Endopath<br>Endoscopic Instrument FDA 510K<br>(K984240)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use       | The DANNIK Disposable<br>Monopolar Laparoscopic<br>Instrument have application in a<br>variety of minimally invasive<br>procedures to facilitate grasping,<br>mobilization, dissection and<br>transection of tissue.                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Product Picture    | Image: Dannik Disposable Monopolar Laparoscopic Instrument                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Image: Ethicon Endo-Surgery Endopath Endoscopic Instrument                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Device Description | The DANNIK Disposable Monopolar<br>Laparoscopic Instrument consists of<br>Handle which activates the<br>instruments jaws and scissor blades,<br>Rotator knob that provides 360<br>degrees rotation to shaft; both<br>handle and rotation knob are made<br>of ABS thermoplastic, and a Shaft<br>5mm in diameter with a working<br>length of 33cm made from Medical<br>grade Stainless Steel and insulated<br>with high-density polyethylene PE.<br>5mm in diameter with a working<br>length of 33cm. RF Post for<br>electrosurgery when attached to an<br>approved electrosurgical generator<br>to provide coagulation of tissue<br>when used with appropriate ground<br>electrode. | Same/Equivalent<br>"The Instruments have a rotating<br>insulated shaft with a diameter of<br>3mm, 5mm or 10mm. The Rotation<br>knob located on the handle rotates<br>the shaft 260 degrees in either<br>direction. The ring handles are<br>compressed and released to activate<br>the instrument jaws or scissor<br>blades. Each of the curved scissors<br>and dissectors has a monopolar<br>cautery connector that extends<br>from the top of the handle. The<br>connector is used for electrosurgery<br>when properly attached to a<br>standard cautery cable and proper<br>generator |
| Classification     | Class II General and Plastic Surgery<br>Devices GEI, 21 CFR878.4400                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Diameter           | 5mm x 33cm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same ( also offers additional diameters)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Materials          | Non-Patient Contact:<br>• ABS Thermoplastic-<br>Patient Contact:<br>• Stainless Steel<br>• PE (Polyethylene)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Sterilization      | Ethylene Oxide ( ETO)<br>I.S.O 11135-1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Prescription Only  | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Biocompatibility   | Conforms to ISO 10993                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |

{6}------------------------------------------------

## 7. Non-Clinical Tests

DANNIK Disposable Monopolar Laparoscopic Instrument has been evaluated by Design Engineers, through performance and functionality evaluations that include Thermal effects on Tissue Connection strength, clamping performance , cable connection, open and close performance, Hardness , corrosion resistance and outward.

- 8. Clinical Tests
No Clinical trials performed on the DANNIK Disposable Monopolar Laparoscopic Instrument

- 9. Conclusions
The subject device has equivalent indications for use as the predicate device. There are no new technologies being added to this device from the predicate, in terms of finished device functions. The device has the same intended use and application as the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K193019](https://fda.innolitics.com/device/K193019)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
