← Product Code [GEI](/productcode/GEI) · K193018

# Surgical Direct Hand-held Laparoscopic Instruments (K193018)

_Surgical Direct, Inc. · GEI · Dec 20, 2019 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K193018

## Device Facts

- **Applicant:** Surgical Direct, Inc.
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** Dec 20, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

Surgical Direct Hand-held Laparoscopic Instruments are intended for grasping, cutting, dissecting, retracting, clamping, biopsy, and/or cauterization of tissue. The devices are used in conjunction with a laparoscope during laparoscopic procedures. Instruments should be used only by personnel completely familiar with their operation. Using an instrument improperly for a task which it was not intended may result in a damaged or broken instrument.

## Device Story

Hand-held laparoscopic instruments consist of reusable handles, shafts, and various interchangeable tips (scissors, forceps, graspers, dissectors, needle holders). Device operates via mechanical linkage connecting handle to working tip; allows surgeon to manipulate tissue (cut, dissect, grasp, clamp, cauterize, retract) through a cannula/trocar during laparoscopic procedures. Used in operating room by physicians. Provides tactile feedback and mechanical control for tissue interaction. Benefits include minimally invasive access to surgical sites, reducing trauma compared to open surgery.

## Clinical Evidence

No clinical study was conducted. Evidence consists of bench testing, including electrical safety (IEC 60601-1, IEC 60601-2-2), EMC validation, biocompatibility (ISO 10993-1), cleaning/sterilization validation (AAMI TIR30), and mechanical performance testing (ex-vivo thermal tissue testing, scissors performance, expected life testing).

## Technological Characteristics

Reusable hand-held laparoscopic instruments; 5mm diameter; 20cm-50cm length; rigid shaft; pistol grip handle; monopolar electrosurgical capability. Materials evaluated per ISO 10993-1. Sterilization via steam pre-vacuum. Connectivity: None (mechanical).

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- ENDOPLUS Hand-held Laparoscopic Instruments ([K151567](/device/K151567.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

## December 20, 2019

Surgical Direct, Inc. % E. Smith Consultant Smith Associates 1468 Harwell Avenue Crofton, Maryland 21114

Re: K193018

Trade/Device Name: Surgical Direct Hand-held Laparoscopic Instruments Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: October 21, 2019 Received: October 29, 2019

Dear E. Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K193018

#### Device Name

Surgical Direct Hand-held Laparoscopic Instruments

#### Indications for Use (Describe)

Surgical Direct Hand-held Laparoscopic Instruments are intended for grasping, dissecting, retracting, clamping, biopsy, and/or cauterization of tissue. The devices are used in conjunction with a laparoscopic procedures. Instruments should be used only by personnel completely familiar with their operation. Using an instrument improperly for a task which it was not intended may result in a damaged or broken instrument.

| Type of Use (Select one or both, as applicable)                                                                   |                                                                                |
|-------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">✖</span> Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> | <div> <span>     Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## K193018 Page 1 of 11

## 510(k) Summary

## SPONSOR

| Company Name:   | Surgical Direct, Inc.                       |
|-----------------|---------------------------------------------|
| Address:        | 909 South Woodland Blvd<br>DeLand, FL 32720 |
| Telephone       | 386.738.1202 Ext. 222                       |
| Email:          | www.surgicaldirect.com                      |
| Contact Person: | David Murray<br>CEO/President               |

## Summary Preparation Date: December 19, 2019

### DEVICE NAME

| Trade Name:          | SURGICAL DIRECT Hand-held Laparoscopic Instruments   |
|----------------------|------------------------------------------------------|
| Common/Usual Name:   | Hand-held Surgical Instruments                       |
| Classification Name: | Electrosurgical, cutting & coagulation & accessories |
| Regulation Number:   | 21 CFR 878.4400                                      |
| Product Code:        | GEI                                                  |
| Device Class:        | Class II                                             |

## PREDICATE DEVICE;

| Manufacturer | Brand Name               | 510(k) Number |
|--------------|--------------------------|---------------|
| ENDOPLUS     | ENDOPLUS Hand-held       | K151567       |
|              | Laparoscopic Instruments |               |

{4}------------------------------------------------

## K193018 Page 2 of 11

### DEVICE DESCRIPTION

The SURGICAL DIRECT Hand-held Laparoscopic Instruments are commonly named hand-held laparoscopic instruments. Specifically, each instrument has a different tip or jaw to define the functional use, such as scissors, forceps, graspers, biopsy, dissectors or needle holders. The devices are designed to be used through a portal, the opening maintained by an introducer or cannula which allows for insertion and removal of the device without damage to surrounding soft tissue. The devices have a linkage mechanism connecting the working tips, or jaws to a handle. This allows the surgeon to cut, dissect, grasp, clamp, cauterize, palpate, or retract at the operating site in conjunction with a laparoscope.

### INDICATIONS FOR USE

Surgical Direct Hand-held Laparoscopic Instruments are intended for grasping, cutting, dissecting, retracting, clamping, biopsy, and/or cauterization of tissue. The devices are used in conjunction with a laparoscope during laparoscopic procedures. Instruments should be used only by personnel completely familiar with their operation. Using an instrument improperly for a task which it was not intended may result in a damaged or broken instrument.

|                               | SURGICAL DIRECT                                                                                                                                                                                                                                                                                                                                                                          | ENDOPLUS                                                                                                                                                                                                                                                                                                                                                                                       | Comment |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Indications for Use Statement | Surgical Direct Hand-held Laparoscopic Instruments are intended for grasping, cutting, dissecting, retracting, clamping, biopsy, and/or cauterization of tissue. The devices are used in conjunction with a laparoscope during laparoscopic procedures. Instruments should be used only by personnel completely familiar with their operation. Using an instrument improperly for a task | Endoplus Hand-held Laparoscopic instruments are intended for grasping, cutting, dissecting, retracting, clamping, biopsy, and/or cauterization of tissue. The devices are used in conjunction with a laparoscope during laparoscopic procedures. Instruments should be used only by personnel completely familiar with their operation. Using an instrument improperly for a task which it was | Same    |

## INDICATIONS FOR USE COMPARISON TABLE

{5}------------------------------------------------

## K193018 Page 3 of 11

| SURGICAL DIRECT                                                                  | ENDOPLUS                                                         | Comment |
|----------------------------------------------------------------------------------|------------------------------------------------------------------|---------|
| which it was not<br>intended may result in<br>a damaged or broken<br>instrument. | not intended may result<br>in a damaged or broken<br>instrument. |         |

### PREDICATE PRODUCT COMPARISON TABLE

| Parameters                   | SURGICAL DIRECT                                                      | ENDOPLUS                                                             | Comments |
|------------------------------|----------------------------------------------------------------------|----------------------------------------------------------------------|----------|
| 510(k) Number                |                                                                      | K151567                                                              | Same     |
| User Environment             | For use by physician<br>in the operating<br>room                     | For use by physician<br>in the operating<br>room                     | Same     |
| Monopolar                    | Yes                                                                  | Yes                                                                  | Same     |
| Length                       | 20cm to 50cm                                                         | 20cm to 50cm                                                         | Same     |
| Minimum Cannula<br>Size      | 5.5cm                                                                | 5.5cm                                                                | Same     |
| Handle Design                | Pistol Grip                                                          | Pistol Grip                                                          | Same     |
| Shaft Rotation               | Yes                                                                  | Yes                                                                  | Same     |
| Shaft Flexibility            | Rigid                                                                | Rigid                                                                | Same     |
| Shaft Diameter               | 5mm                                                                  | 5mm                                                                  | Same     |
| Articulating<br>Component    | Does not articulate                                                  | Does not articulate                                                  | Same     |
| Actuation<br>Mechanism       | Handle Operation<br>causes opening and<br>closing of end<br>effector | Handle Operation<br>causes opening and<br>closing of end<br>effector | Same     |
| Flush Port                   | Yes                                                                  | Yes                                                                  | Same     |
| Insertion through<br>trocar  | Yes                                                                  | Yes                                                                  | Same     |
| Form inserted into<br>trocar | Jaws closed                                                          | Jaws closed                                                          | Same     |
| Reusable/Disposable          | Reusable Handle,<br>Shaft and Insert                                 | Reusable Handle,<br>Shaft and Insert                                 | Same     |

## DISCUSSION OF TECHNOLOGICAL DIFFERENCES

Surgical Direct Hand-held Laparoscopic Instruments are same as the Endoplus Hand held Laparoscopic instruments in indications for use, use environment, user, monopolar, length, minimum cannula size, handle design, shaft rotations, shaft flexibility, shaft diameters,

{6}------------------------------------------------

## K193018 Page 4 of 11

articulating components, actuation mechanism, flush-port, insertion through trocar, form inserted into trocar, reusable handle, shaft and insert.

There are no differences between the SURGIAL DIRECT and the ENDOPLUS hand-held laparoscopic instruments.

## PERFORMANCE DATA

## Safety Testing:

- . IEC 60601-1: 2012 3rd edition - Medical electrical equipment—Part 1: General requirements for basic safety and essential performance
- . IEC 60601-1-2 Edition 3: 2007-03, medical electrical equipment - part 1-2: general requirements for basic safety and essential performance - collateral standard: electromagnetic compatibility - requirements and tests.
- . IEC 60601-2-2: 2009 Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories
- EMC Validation Test

## Cleaning and Sterilization:

The Reusable Hand-held Laparoscopic instruments are evaluated based on the following tests:

- . Manual cleaning validation study and Laparoscopic Instrument Rack TOC analysis(AAMI TIR30:2011)
- . Manual cleaning validation study and Laparoscopic Instrument Rack Protein analysis
- Steam Pre-vacuum Efficacy Validation Study ●
- Steam Sterilization Validation Study
- Shelf Life Protocol

## Biocompatibility:

In accordance with ISO 10993-1, Hand-held Laparoscopic instruments are categorized as externally communicating devices in contact with circulating blood for less than 24 hours. The following tests are recommended by ISO 10993-1:

- Cytotoxicity
- Sensitization ●
- Irritation
- Systemic toxicity ●
- . Intracutaneous
- Hemolysis

{7}------------------------------------------------

## K193018 Page 5 of 11

### Performance Testing:

- PLTW Comparative Ex-Vivo Thermal Tissue Testing Validation Study .
- Electrical Safety Validation Report (Electrical Safety Testing) ●
- Scissors Performance Testing
- . Expected Life Testing

### CLINICAL STUDY

No clinical study was conducted.

### CONCLUSION

Surgical Direct Hand-held Laparoscopic Instruments has an identical indication for use and identical technological characteristics as the predicate device. Biocompatibility testing, electrical testing and other performance tests were conducted to verify safe and effective use. Surgical Direct concludes that the Surgical Direct Hand-held Laparoscopic Instruments are substantially equivalent to the predicate device and introduces no new issues of safety and effectiveness.

| Surgical<br>Direct # | (mm) | Type        | Length | Descriptions                                              |
|----------------------|------|-------------|--------|-----------------------------------------------------------|
| SD8901               | 5    | insert      | Std    | Dolphin Nose Dissecting Forceps w/ cup                    |
| SD890145             | 5    | insert      | Bari   | Dolphin Nose Dissecting Forceps w/ cup                    |
| SD8902               | 5    | insert      | Std    | Curved Dissecting Forceps                                 |
| SD890245             | 5    | insert      | Bari   | Curved Dissecting Forceps                                 |
| SD8903               | 5    | insert      | Std    | Micro Dolphin Dissector w/ Cup                            |
| SD890345             | 5    | insert      | Bari   | Micro Dolphin Dissector w/ Cup                            |
| SD8904               | 5    | insert      | Std    | Short Curved Dissecting Forceps                           |
| SD890445             | 5    | insert      | Bari   | Short Curved Dissecting Forceps                           |
| SD8905               | 5    | insert      | Std    | Micro Dolphin Dissecting Forceps                          |
| SD890545             | 5    | insert      | Bari   | Micro Dolphin Dissecting Forceps                          |
| SD8906               | 5    | insert      | Std    | Fenestrated Grasping Forceps                              |
| SD890645             | 5    | insert      | Bari   | Fenestrated Grasping Forceps                              |
| SD8907               | 5    | insert      | Std    | Delicate Cone Dissector                                   |
| SD890745             | 5    | insert      | Bari   | Delicate Cone Dissector                                   |
| SD8908               | 5    | insert      | Std    | Alligator Forceps                                         |
| SD890845             | 5    | insert      | Bari   | Alligator Forceps                                         |
| SD8909               | 5    | insert      | Std    | Dolphin Nose Grasping Forceps                             |
| SD890945             | 5    | insert      | Bari   | Dolphin Nose Grasping Forceps                             |
| Surgical<br>Direct # | (mm) | Type        | Length | Descriptions                                              |
| SD8910               | 5    | insert      | Std    | Strong Fenestrated Grasping Forceps                       |
| SD891045             | 5    | insert      | Bari   | Strong Fenestrated Grasping Forceps                       |
| SD8911               | 5    | insert      | Std    | Maxi Fenestrated Grasping Forceps                         |
| SD891145             | 5    | insert      | Bari   | Maxi Fenestrated Grasping Forceps                         |
| SD8912               | 5    | insert      | Std    | Duckbill Dissector                                        |
| SD891245             | 5    | insert      | Bari   | Duckbill Dissector                                        |
| SD8913               | 5    | insert      | Std    | Curved Maryland Dissector                                 |
| SD891345             | 5    | insert      | Bari   | Curved Maryland Dissector                                 |
| SD8914               | 5    | insert      | Std    | Flat Dissector                                            |
| SD891445             | 5    | insert      | Bari   | Flat Dissector                                            |
| SD8915               | 5    | insert      | Std    | Straight Tapered Dissector                                |
| SD891545             | 5    | insert      | Bari   | Straight Tapered Dissector                                |
| SD8916               | 5    | insert      | Std    | Mini-Micro Dissecting Forceps                             |
| SD891645             | 5    | insert      | Bari   | Mini-Micro Dissecting Forceps                             |
| SD8917               | 5    | insert      | Std    | Mixter Dissector                                          |
| SD891745             | 5    | insert      | Bari   | Mixter Dissector                                          |
| SD8918               | 5    | insert      | Std    | Micro Dissecting Forceps                                  |
| SD891845             | 5    | insert      | Bari   | Micro Dissecting Forceps                                  |
| SD8919               | 5    | insert      | Std    | Mini Right-Angle Dissector                                |
| SD891945             | 5    | insert      | Bari   | Mini Right-Angle Dissector                                |
| SD8920               | 5    | insert      | Std    | Straight Mini Dissector                                   |
| SD892045             | 5    | insert      | Bari   | Straight Mini Dissector                                   |
| SD8921               | 5    | insert      | Std    | Pointed Dissector w/ cup                                  |
| SD892145             | 5    | insert      | Bari   | Pointed Dissector w/ cup                                  |
| SD8922               | 5    | insert      | Std    | Atraumatic Grasper S/A                                    |
| SD892245             | 5    | insert      | Bari   | Atraumatic Grasper S/A                                    |
| SD8923               | 5    | insert      | Std    | Atraumatic Grasper                                        |
| SD892345             | 5    | insert      | Bari   | Atraumatic Grasper                                        |
| SD8924               | 5    | insert      | Std    | Extracting Forceps Reverse Teeth                          |
| SD892445             | 5    | insert      | Bari   | Extracting Forceps Reverse Teeth                          |
| SD8925               | 5    | insert      | Std    | Traumatic Grasper 2x3 Teeth                               |
| SD892545             | 5    | insert      | Bari   | Traumatic Grasper 2x3 Teeth                               |
| SD8926               | 5    | insert      | Std    | Cobra Jaw Grasping Forceps                                |
| SD892645             | 5    | insert      | Bari   | Cobra Jaw Grasping Forceps                                |
| SD8927               | 5    | insert      | Std    | Claw Forceps 2x3 Teeth                                    |
| SD892745             | 5    | insert      | Bari   | Claw Forceps 2x3 Teeth                                    |
| Surgical<br>Direct # | (mm) | Type        | Length | Descriptions                                              |
| SD8928               | 5    | insert      | Std    | Glassman Forceps, 20mm                                    |
| SD892845             | 5    | insert      | Bari   | Glassman Forceps, 20mm                                    |
| SD8929               | 5    | insert      | Std    | Traumatic Grasper, 2x3 Teeth S/A                          |
| SD892945             | 5    | insert      | Bari   | Traumatic Grasper, 2x3 Teeth S/A                          |
| SD8930               | 5    | insert      | Std    | Retrograde Grasping Forceps                               |
| SD893045             | 5    | insert      | Bari   | Retrograde Grasping Forceps                               |
| SD8931               | 5    | insert      | Std    | Ripple Claw Forceps                                       |
| SD893145             | 5    | insert      | Bari   | Ripple Claw Forceps                                       |
| SD8932               | 5    | insert      | Std    | Grasping Forceps                                          |
| SD893245             | 5    | insert      | Bari   | Grasping Forceps                                          |
| SD8933               | 5    | insert      | Std    | Grasping Forceps w/ cup                                   |
| SD893345             | 5    | insert      | Bari   | Grasping Forceps w/ cup                                   |
| SD8934               | 5    | insert      | Std    | Long Grasping Forceps                                     |
| SD893445             | 5    | insert      | Bari   | Long Grasping Forceps                                     |
| SD8935               | 5    | insert      | Std    | Oviduct Atraumatic Forceps                                |
| SD893545             | 5    | insert      | Bari   | Oviduct Atraumatic Forceps                                |
| SD8936               | 5    | insert      | Std    | McKernan Tri Grasping Fenestrated Forceps                 |
| SD893645             | 5    | insert      | Bari   | McKernan Tri Grasping Fenestrated Forceps                 |
| SD8937               | 5    | insert      | Std    | Tri-Grasper                                               |
| SD893745             | 5    | insert      | Bari   | Tri-Grasper                                               |
| SD8938               | 5    | insert      | Std    | Atraumatic Tube Forceps                                   |
| SD893845             | 5    | insert      | Bari   | Atraumatic Tube Forceps                                   |
| SD8939               | 5    | insert      | Std    | Wave Jaw Grasping Forceps                                 |
| SD893945             | 5    | insert      | Bari   | Wave Jaw Grasping Forceps                                 |
| SD8940               | 5    | insert      | Std    | Allis forceps                                             |
| SD894045             | 5    | insert      | Bari   | Allis forceps                                             |
| SD8941               | 5    | insert      | Std    | Rodriguez Grasper                                         |
| SD894145             | 5    | insert      | Bari   | Rodriguez Grasper                                         |
| SD8942               | 5    | insert      | Std    | Angled Dissector                                          |
| SD894245             | 5    | insert      | Bari   | Angled Dissector                                          |
| SD8943               | 5    | insert      | Std    | Biopsy Forceps w/ Two Teeth                               |
| SD894345             | 5    | insert      | Bari   | Biopsy Forceps w/ Two Teeth                               |
| SD8944               | 5    | insert      | Std    | Forsythe Bump Grasper, Single Action                      |
| SD894445             | 5    | insert      | Bari   | Forsythe Bump Grasper, Single Action                      |
| SD8945               | 5    | insert      | Std    | Babcock Grasping Forceps                                  |
| SD894545             | 5    | insert      | Bari   | Babcock Grasping Forceps                                  |
| Surgical<br>Direct # | (mm) | Type        | Length | Descriptions                                              |
| SD8946               | 5    | insert      | Std    | Long Allis Forceps                                        |
| SD894645             | 5    | insert      | Bari   | Long Allis Forceps                                        |
| SD8947               | 5    | insert      | Std    | Endo Babcock Forceps                                      |
| SD894745             | 5    | insert      | Bari   | Endo Babcock Forceps                                      |
| SD8948               | 5    | insert      | Std    | Paddle Babcock                                            |
| SD894845             | 5    | insert      | Bari   | Paddle Babcock                                            |
| SD8949               | 5    | insert      | Std    | Debakey Tissue Forceps                                    |
| SD894945             | 5    | insert      | Bari   | Debakey Tissue Forceps                                    |
| SD8950               | 5    | insert      | Std    | Dorsey Fenestrated Grasper                                |
| SD895045             | 5    | insert      | Bari   | Dorsey Fenestrated Grasper                                |
| SD8951               | 5    | insert      | Std    | Glassman Forceps                                          |
| SD895145             | 5    | insert      | Bari   | Glassman Forceps                                          |
| SD8953               | 5    | insert      | Std    | Round Tip Allis Grasping Forceps                          |
| SD895345             | 5    | insert      | Bari   | Round Tip Allis Grasping Forceps                          |
| SD8956               | 5    | insert      | Std    | Medium Fenestrated Bowel Grasper                          |
| SD895645             | 5    | insert      | Bari   | Medium Fenestrated Bowel Grasper                          |
| SD8957               | 5    | insert      | Std    | Endo Allis Forceps                                        |
| SD895745             | 5    | insert      | Bari   | Endo Allis Forceps                                        |
| SD8958               | 5    | insert      | Std    | Pennington Forceps                                        |
| SD895845             | 5    | insert      | Bari   | Pennington Forceps                                        |
| SD8959               | 5    | insert      | Std    | Atraumatic Wave Grasper                                   |
| SD895945             | 5    | insert      | Bari   | Atraumatic Wave Grasper                                   |
| SD8960               | 5    | insert      | Std    | Kocher Grasping Forceps                                   |
| SD896045             | 5    | insert      | Bari   | Kocher Grasping Forceps                                   |
| SD8961               | 5    | insert      | Std    | Curved Metz Scissor                                       |
| SD896145             | 5    | insert      | Bari   | Curved Metz Scissor                                       |
| SD8962               | 5    | insert      | Std    | Straight Micro Scissor                                    |
| SD896245             | 5    | insert      | Bari   | Straight Micro Scissor                                    |
| SD8963               | 5    | insert      | Std    | Delicate Curved Metz Scissor                              |
| SD896345             | 5    | insert      | Bari   | Delicate Curved Metz Scissor                              |
| SD8964               | 5    | insert      | Std    | Curved Micro Scissor                                      |
| SD896445             | 5    | insert      | Bari   | Curved Micro Scissor                                      |
| SD8965               | 5    | insert      | Std    | Straight Metz Scissor                                     |
| SD896545             | 5    | insert      | Bari   | Straight Metz Scissor                                     |
| SD8966               | 5    | insert      | Std    | Hook Scissor                                              |
| SD896645             | 5    | insert      | Bari   | Hook Scissor                                              |
| Surgical<br>Direct # | (mm) | Type        | Length | Descriptions                                              |
| SD8967               | 5    | insert      | Std    | Delicate S/A Scissor                                      |
| SD896745             | 5    | insert      | Bari   | Delicate S/A Scissor                                      |
| SD8968               | 5    | insert      | Std    | Serrated Scissor                                          |
| SD896845             | 5    | insert      | Bari   | Serrated Scissor                                          |
| SD8969               | 5    | insert      | Std    | Biopsy Scoop Forceps                                      |
| SD896945             | 5    | insert      | Bari   | Biopsy Scoop Forceps                                      |
| SD8970               | 5    | insert      | Std    | Biopsy Forceps w/ One Tooth                               |
| SD897045             | 5    | insert      | Bari   | Biopsy Forceps w/ One Tooth                               |
| SD8971               | 5    | insert      | Std    | Dorsey Grasper, 40mm                                      |
| SD897145             | 5    | insert      | Bari   | Dorsey Grasper, 40mm                                      |
| SD8972               | 5    | insert      | Std    | Single Tooth Grasper                                      |
| SD897245             | 5    | insert      | Bari   | Single Tooth Grasper                                      |
| SD8973               | 5    | insert      | Std    | Glassman Forceps, 30mm                                    |
| SD897345             | 5    | insert      | Bari   | Glassman Forceps, 30mm                                    |
| SD8974               | 5    | insert      | Std    | Single Action Needle Holder                               |
| SD897445             | 5    | insert      | Bari   | Single Action Needle Holder                               |
| SD8975               | 5    | insert      | Std    | Suture Knot Pusher                                        |
| SD897545             | 5    | insert      | Bari   | Suture Knot Pusher                                        |
| SD8976               | 5    | insert      | Std    | Credi Curved Dissector                                    |
| SD897645             | 5    | insert      | Bari   | Credi Curved Dissector                                    |
| SD8977               | 5    | insert      | Std    | St. Joseph Dissecting Forceps                             |
| SD897745             | 5    | insert      | Bari   | St. Joseph Dissecting Forceps                             |
| SD8978               | 5    | insert      | Std    | Curved Debakey Tissue Forceps                             |
| SD897845             | 5    | insert      | Bari   | Curved Debakey Tissue Forceps                             |
| SD8979               | 5    | insert      | Std    | Petrovich Curved Dissector                                |
| SD897945             | 5    | insert      | Bari   | Petrovich Curved Dissector                                |
| SD8980               | 5    | insert      | Std    | Fenestrated Bowel Grasper, 40mm                           |
| SD898045             | 5    | insert      | Bari   | Fenestrated Bowel Grasper, 40mm                           |
| SD8981               | 5    | insert      | Std    | Byerley Grasper                                           |
| SD898145             | 5    | insert      | Bari   | Byerley Grasper                                           |
| SD8982               | 5    | insert      | Std    | Murray Grasper                                            |
| SD898245             | 5    | insert      | Bari   | Murray Grasper                                            |
| SD8983               | 5    | insert      | Std    | Fenestrated Bowel Grasper, 20mm                           |
| SD898345             | 5    | insert      | Bari   | Fenestrated Bowel Grasper, 20mm                           |
| SD8984               | 5    | insert      | Std    | Bump Grasper, Single Action                               |
| SD898445             | 5    | insert      | Bari   | Bump Grasper, Single Action                               |
| Surgical<br>Direct # | (mm) | Type        | Length | Descriptions                                              |
| SD8985               | 5    | insert      | Std    | Double Action Spoon Forceps, 5mm                          |
| SD898545             | 5    | insert      | Bari   | Double Action Spoon Forceps, 5mm                          |
| SD8986               | 5    | insert      | Std    | Biopsy Forceps w/ One Tooth (on bottom)                   |
| SD898645             | 5    | insert      | Bari   | Biopsy Forceps w/ One Tooth (on bottom)                   |
| SD8987               | 5    | insert      | Std    | 5mm Smoot Endo Cinch                                      |
| SD898745             | 5    | insert      | Bari   | 5mm Smoot Endo Cinch                                      |
| SD8988               | 5    | insert      | Std    | Andrew Babcock Forceps                                    |
| SD898845             | 5    | insert      | Bari   | Andrew Babcock Forceps                                    |
| SD8989               | 5    | insert      | Std    | Tenaculum, 5mm                                            |
| SD898945             | 5    | insert      | Bari   | Tenaculum, 5mm                                            |
| SD8990               | 5    | insert      | Std    | Maryland with Cross Serrations                            |
| SD899045             | 5    | insert      | Bari   | Maryland with Cross Serrations                            |
| SD8991               | 5    | insert      | Std    | Single Action Spoon Forceps                               |
| SD899145             | 5    | insert      | Bari   | Single Action Spoon Forceps                               |
| SD8992               | 5    | insert      | Std    | Harvey Grasper                                            |
| SD899245             | 5    | insert      | Bari   | Harvey Grasper                                            |
| SD8993               | 5    | insert      | Std    | 5mm Meah Grasper                                          |
| SD899345             | 5    | insert      | Bari   | 5mm Meah Grasper                                          |
| SD8994               | 5    | insert      | Std    | Strevy Grasper                                            |
| SD899445             | 5    | insert      | Bari   | Strevy Grasper                                            |
| SD8995               | 5    | insert      | Std    | Medium Curved Dissector                                   |
| SD899545             | 5    | insert      | Bari   | Medium Curved Dissector                                   |
| SD8996               | 5    | insert      | Std    | Heavy Curved Dissector                                    |
| SD899645             | 5    | insert      | Bari   | Heavy Curved Dissector                                    |
| SD8997               | 5    | insert      | Std    | Johan Forceps                                             |
| SD899745             | 5    | insert      | Bari   | Johan Forceps                                             |
| SD8998               | 5    | insert      | Std    | Delicate Maryland with Cross Serrations                   |
| SD899845             | 5    | insert      | Bari   | Delicate Maryland with Cross Serrations                   |
| SD8999               | 5    | insert      | Std    | Medium Dolphin Nose Dissecting Forceps w/ cup             |
| SD899945             | 5    | insert      | Bari   | Medium Dolphin Nose Dissecting Forceps w/ cup             |
| SD902                | 5    | handle/tube | Std    | Monopolar Handle with Rotation, Ratchet on Demand         |
| SD903                | 5    | handle/tube | Bari   | Monopolar Handle with Rotation, Ratchet on Demand         |
| SD904                | 5    | handle/tube | Std    | Take Apart Slide Lock with Rotation                       |
| SD905                | 5    | handle/tube | Bari   | Take Apart Slide Lock with Rotation                       |
| SD908                | 5    | handle/tube | Std    | Monopolar Handle with Non-Locking Rotation, NO<br>Ratchet |
| Surgical<br>Direct # | (mm) | Type        | Length | Descriptions                                              |
| SD909                | 5    | handle/tube | Bari   | Monopolar Handle with Non-Locking Rotation, NO<br>Ratchet |
| SD910                | 5    | handle/tube | Std    | Take Apart Slide Lock with Rotation                       |
| SD911                | 5    | handle/tube | Bari   | Take Apart Slide Lock with Rotation                       |
| SD9901               | 5    | insert      | Std    | Kowalski Hunter Bowel Grasper                             |
| SD990145             | 5    | insert      | Bari   | Kowalski Hunter Bowel Grasper                             |
| SD9902               | 5    | insert      | Std    | Devlin Micro-fenestrated Grasper                          |
| SD990245             | 5    | insert      | Bari   | Devlin Micro-fenestrated Grasper                          |
| SD9903               | 5    | insert      | Std    | Holmes Atraumatic Grasper                                 |
| SD990345             | 5    | insert      | Bari   | Holmes Atraumatic Grasper                                 |
| SD9904               | 5    | insert      | Std    | Graves Grasper                                            |
| SD990445             | 5    | insert      | Bari   | Graves Grasper                                            |
| SD9905               | 5    | insert      | Std    | 5mm Maryland with 90 Degree Tooth                         |
| SD990545             | 5    | insert      | Bari   | 5mm Maryland with 90 Degree Tooth                         |

#### Below is a complete listing of laparoscopic instruments covered by this 510(k) submission

{8}------------------------------------------------

# K193018
Page 6 of 11

{9}------------------------------------------------

# K193018
Page 7 of 11

{10}------------------------------------------------

# K193018
Page 8 of 11

{11}------------------------------------------------

# K193018
Page 9 of 11

{12}------------------------------------------------

# 

{13}------------------------------------------------

#

---

**Source:** [https://fda.innolitics.com/device/K193018](https://fda.innolitics.com/device/K193018)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
