Liebel-Flarsheim IBP Transfer Set

K193010 · Liebel-Flarsheim Company, LLC · PQH · Nov 18, 2020 · General Hospital

Device Facts

Record IDK193010
Device NameLiebel-Flarsheim IBP Transfer Set
ApplicantLiebel-Flarsheim Company, LLC
Product CodePQH · General Hospital
Decision DateNov 18, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5440
Device ClassClass 2

Indications for Use

The Liebel-Flarsheim™ IBP Transfer Set is intended for the transfer of fluids from bulk containers to empty sterile syringes on syringe-based contrast injection systems.

Device Story

The Liebel-Flarsheim™ IBP Transfer Set is a sterile, single-use tubing set used to connect Imaging Bulk Package (IBP) contrast media containers to empty sterile syringes on automatic contrast injection systems. The device consists of a 19-inch small-bore PVC tube with a vented IV spike at the proximal end for piercing the contrast container and a self-sealing Luer-activated connector (Smiths Medical Nu-Site®) at the distal end. It facilitates the transfer of contrast media from bulk containers to syringes for automated injection. The device is used in clinical settings where contrast-enhanced procedures are performed. By enabling the use of bulk contrast packages, it streamlines the filling process for automated injectors, potentially improving workflow efficiency. The device is discarded after the container is empty or after 12 hours of use.

Clinical Evidence

Bench testing only. Testing included biocompatibility (ISO 10993-1, -7, -18), sterility validation (ISO 11135), shelf-life verification, and performance testing per ISO 8536-4, ISO 594-2, and ISO 22413. Microbial ingress, chemical compatibility with Optiray® (Ioversol), and extractables/leachables testing were performed to verify functional requirements.

Technological Characteristics

19-inch small-bore (1/8" ID) PVC tubing; vented IV spike (ABS); self-sealing Luer-activated connector (polyurethane, silicone, acetal). Ethylene oxide (EtO) sterilized (SAL 10^-6). Complies with ISO 10993, ISO 8536-4, ISO 594-2, and ISO 22413. Single-use, non-DEHP PVC.

Indications for Use

Indicated for the transfer of Optiray® (Ioversol) Imaging Bulk Package contrast media to empty, single-use only, sterile syringes on syringe-based contrast injection systems for controlled, automatic intra-vascular administration of radiopaque contrast agents. Discard after container is empty or 12 hours after penetration, whichever occurs first. Contraindications determined by prescribing physician based on contrast media package inserts.

Regulatory Classification

Identification

An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.

Special Controls

*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. November 18, 2020 Liebel-Flarsheim Company LLC % Matthew Helmi Regulatory Affairs Associate II Guerbet LLC 821 Alexander Road Suite 204 Princeton, New Jersey 08540 Re: K193010 Trade/Device Name: Liebel-Flarsheim™ IBP Transfer Set Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular administration set Regulatory Class: Class II Product Code: PQH Dated: November 16, 2020 Received: November 17, 2020 Dear Matthew Helmi: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. for Payal Patel Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K193010 Device Name Liebel-Flarsheim™ IBP Transfer Set ### Indications for Use (Describe) The Liebel-Flarsheim™ IBP Transfer Set is indicated for the transfer of Optiray® (loversol) Imaging Bulk Package contrast media to empty, single-use only, sterile syringe based contrast injection systems indicated for the controlled, automatic intra-vascular administration of radiopaque contrast agents for their indicated procedures. The IBP Transfer Set is to be discarded after the container has been emptied or after 12 hours have elapsed since the container was penetrated, whichever occurs first. Contra-indications for the use of this device are determined by the prescribing physician at the time of used upon the contrast media package inserts. X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### Liebel-Flarsheim Company LLC 2111 East Galbraith Road Cincinnati, OH 45237 Tel. +1 513-948-2700 www.guerbet.com # Traditional Premarket Notification Submission - 510(k) Summary Liebel-Flarsheim™ IBP Transfer Set K193010 ### Submitter Liebel-Flarsheim Company LLC 2111 East Galbraith Road Cincinnati, OH 45237 Phone: (513)-948-4072 ## Primary Contact Person Matthew Helmi Guerbet LLC, Suite 204 821 Alexander Rd Princeton, NJ 08540 matthew.helmi@guerbet.com Phone: (609)-819-5213 ### Secondary Contact Ahmed Abdelal, PharmD Guerbet LLC, Suite 204 821 Alexander Rd Princeton, NJ 08540 Ahmed.Abdelal(@guerbet.com Phone: 609-480-7655 ## Date: November 16, 2020 ### Device Name and Classification | Trade Name: | Liebel-Flarsheim™ IBP Transfer Set | |---------------------|----------------------------------------------------| | Common Name: | Iodinated Contrast Media Transfer Tubing | | Set Classification: | Class II | | Regulation: | 21 CFR 880.5440, Intravascular Administration | | Set. Review Panel: | General Hospital | | Product Code: | PQH (Iodinated Contrast Media Transfer Tubing Set) | ### Predicate Device | Trade Name: | ulrich Transfer Set | |---------------------|----------------------------------------------------| | K Number: | K161723 | | Common Name: | Iodinated Contrast Media Transfer | | Set Classification: | Class II | | Regulation: | 21 CFR 880.5440, Intravascular Administration | | Set. Review Panel: | General Hospital | | Product Code: | PQH (Iodinated Contrast Media Transfer Tubing Set) | {4}------------------------------------------------ #### 5.1. Device Information ## Device Description and Intended Use The Liebel-Flarsheim™ IBP Transfer Set is a sterile tubing set connection between Optiray® (Ioversol) contrast media imaging bulk packages (hereinafter Optiray IBPs) and empty sterile singleuse syringes on syringe-based automatic contrast injection systems. The Liebel- Flarsheim™ IBP Transfer Set consists of a small bore (1/8" ID) clear polyvinyl chloride (PVC) tube with standard medical component end terminations. Bonded to the proximal end of the PVC tube is a vented IV spike that is used to pierce a rubber stopper belonging to an imaging bulk contrast media container. Distally, a self-sealing Luer type connector (Smiths Medical Nu- Site®) is bonded to the opposite end of the PVC tube. The Liebel-Flarsheim™ IBP Transfer Set, shown in the image below, has an overall length of 19-inches, a protective cap on the IV Spike, and both untethered and tethered protective caps on the Luer end to protect the Nu-Site® Luer Activated fitting between syringe fills. The transfer sets are individually packaged in blister packs with a perimeter heat sealed Tyvek lid. Sealed blister packs are packaged in an outer box containing 50 units (50 units/box) which are terminally sterilized with ethylene oxide. The Liebel-Flarsheim™ IBP Transfer Set is intended for the transfer of fluids from bulk containers to empty sterile syringes on syringe-based contrast injection systems. Image /page/4/Picture/4 description: The image shows a medical device, specifically an extension set with a needleless connector. The set includes a clear, flexible tube forming a loop, connected to a white connector at one end and a clear connector with a purple valve at the other. Two blue caps are also visible, one attached to the clear connector with a blue tether and the other detached, likely for covering the connector when not in use. #### Liebel-Flarsheim™ IBP Transfer Set Image 1: {5}------------------------------------------------ ### Indications for Use The Liebel-Flarsheim™ IBP Transfer Set is indicated for the transfer of Optiray® (loversol) Imaging Bulk Package contrast media to empty, single-use only, sterile syringes on syringe based contrast injection systems indicated for the controlled, automatic intra-vascular administration of radiopaque contrast agents for their indicated procedures. The IBP Transfer Set is to be discarded after the contrast media container has been emptied or after 12 hours have elapsed since the container was penetrated, whichever occurs first. Contra-indications for the use of this device are determined by the prescribing physician at the time of use based upon the contrast media package inserts. ### Risk Analysis Method The Liebel-Flarsheim™ IBP Transfer Set was assessed for risks to health associated with the use of intravascular administration sets and accessories. Risks including but not limited to device malfunctions, bio-incompatibility risks, infection risks, and improper use were assessed. This device risk analysis was conducted in accordance with ISO 14971:2012 Medical Devices – Application of Risk Management to Medical Devices. #### 5.2. Substantial Equivalence The Liebel-Flarsheim™ IBP Transfer Set is substantially equivalent to the ulrich Transfer Set (K161723) predicate device, currently legally marketed in the United States of America. The Liebel-Flarsheim™ IBP Transfer Set has the same intended use and substantially equivalent indications for use as the ulrich Transfer Set. The Liebel-Flarsheim™ IBP Transfer Set is indicated for the transfer of Optiray® (Ioversol) Imaging Bulk Package contrast media while the ulrich Transfer Set is indicated for the transfer of Omnipaque™ (Iohexol) Imaging Bulk Package contrast media. The Liebel-Flarsheim™ IBP Transfer Set uses equivalent overall design and operating principals to the ulrich Transfer Set. The materials of the components are comparable to the predicates. Similar tests and test methods performed in accordance with the same standards were used with the transfer set and the predicate device to validate the design. Safety questions have been evaluated for the transfer set through extensive design verification and validation testing. The above described differences do not adversely affect the safety or effectiveness of the transfer set as compared to the predicate device for the following reasons: - a. Biocompatibility: An evaluation of biocompatibility was performed in compliance with ISO 10993-1:2018. - b. Performance Testing (Bench): Bench testing, including but not limited to functional testing, chemical compatibility, and microbial ingress testing, was performed to verify the device for its intended use. The acceptance criteria were chosen to ensure that the device performs in an equivalent than the predicate device. {6}------------------------------------------------ #### 5.2.1 Summary Comparison of Subject and Predicate Device | Item | Subject Device<br>Liebel-Flarsheim™ IBP<br>Transfer Set<br>K193010 | Predicate Device<br>ulrich Transfer Set<br>K161723 | Comparison | |-----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | The Liebel-Flarsheim™ IBP<br>Transfer Set is intended for the<br>transfer of fluids from bulk<br>containers to empty sterile<br>syringes on syringe-based<br>contrast injection systems. | The transfer set is intended for<br>the transfer of fluids from bulk<br>containers to empty sterile<br>syringes on syringe based<br>contrast delivery systems<br>(injectors). | Identical to the<br>predicate<br>device. | | Indications for<br>Use | The Liebel-Flarsheim™ IBP<br>Transfer Set is indicated for the<br>transfer of Optiray® (Ioversol)<br>Imaging Bulk Package contrast<br>media to empty, single-use<br>only, sterile syringes on syringe<br>based contrast injection systems<br>indicated for the controlled,<br>automatic intra-vascular<br>administration of radiopaque<br>contrast agents for their<br>indicated procedures. The IBP<br>Transfer Set is to be discarded<br>after the contrast media<br>container has been emptied or<br>after 12 hours have elapsed<br>since the container was<br>penetrated, whichever occurs<br>first. | The ulrich Transfer Set is<br>indicated for the transfer of<br>Omnipaque™ (Iohexol)<br>contrast media as supplied in<br>an Imaging Bulk Package to<br>non pre-filled single-use only<br>sterile syringes on syringe-<br>based contrast injection<br>systems indicated for the<br>controlled, automatic<br>administration on the venous<br>side of contrast media for CT<br>procedures. The Transfer Set is<br>to be discarded after the<br>contrast media container has<br>been depleted or 8 hours have<br>elapsed since the container was<br>penetrated, whichever occurs<br>first. | Substantially<br>equivalent to the<br>predicate device. | | Item | Subject Device<br>Liebel-Flarsheim™ IBP<br>Transfer Set<br>K193010 | Predicate Device<br>ulrich Transfer Set<br>K161723 | Comparison | | Tubing Length | 19 inches | 20 inches | Substantially<br>equivalent to the<br>predicate device.<br>Does not impact<br>performance | | Single Use | Yes | Yes | Identical to the<br>predicate device. | | Biocompatible | Yes - ISO 10993-1 | Yes - ISO 10993-1 | Substantially<br>equivalent to the<br>predicate device. | | Sterile | Yes | Yes | Identical to the<br>predicate device. | | Sterility<br>Assurance<br>Level | $10^{-6}$ | $10^{-6}$ | Identical to the<br>predicate device. | | Sterilization<br>Method | Ethylene Oxide (EtO) | Ethylene Oxide (EtO) | Identical to the<br>predicate device. | | Shelf Life | Three (3) year shelf life. | One (1) year shelf life. | Substantially<br>equivalent to the<br>predicate device.<br>Shelf life testing<br>has been completed<br>and demonstrated<br>that the differences<br>do not raise new<br>questions of safety<br>and effectiveness. | | Item | Subject Device<br>Liebel-Flarsheim™ IBP<br>Transfer Set<br>K193010 | Predicate Device<br>ulrich Transfer Set<br>K161723 | Comparison | | Packaging | Individually packaged in a<br>Tyvek pouch. | Individually packaged in a<br>Tyvek pouch. | Substantially<br>equivalent to the<br>predicate device. | | Microbial<br>Ingress Testing | Yes | Yes | Substantially<br>equivalent to the<br>predicate device. | | Chemical<br>Compatibility<br>Testing | Yes - Optiray® (Ioversol)<br>Imaging Bulk Package | Yes - Omnipaque™<br>(Iohexol) Contrast Media | Substantially<br>equivalent to the<br>predicate device. | | Design-<br>Protective Cap<br>for Luer<br>Connection | Yes - Tethered Cap with<br>Smiths Medical Nu-Site®<br>Luer Actuated Connector | Yes - With Halkey-Roberts<br>Swabbable Valve | Substantially<br>equivalent to the<br>predicate device. | | Design- Spike | Protected IV Spike with Vent | Spike with Protected Cap<br>and Integrated Air Filter | Substantially<br>equivalent to the<br>predicate. | | Use Life (After<br>puncturing<br>IBP) | 12 Hours or disconnection<br>from IBP (whichever occurs<br>first) | 8 Hours or disconnection<br>from IBP (whichever occurs<br>first) | Substantially<br>equivalent to the<br>predicate. | | Material<br>Differences | Fluid Contacting Tubing<br>Non DEHP PVC<br>Spike<br>ABS<br>Luer<br>Polyurethane Hosing<br>Silicone Seal<br>Acetal Actuator | Unknown | Substantially<br>equivalent to the<br>predicate. | ### Comparison of Subject and Predicate Device Table 1: {7}------------------------------------------------ {8}------------------------------------------------ {9}------------------------------------------------ #### 5.2.2 Summary of substantial equivalence discussion In summary, the subject device's principle of operation, intended use, functional performance, and similar technological characteristics are substantially equivalent to those of the predicate device. Any differences in characteristics between the subject device and predicate device do not raise questions of safety or effectiveness. The subject device is substantially equivalent to the legally marketed predicate device. #### 5.3 Testing #### 5.3.1 Sterilization The Liebel-Flarsheim™ IBP Transfer Set is ethylene oxide (EtO) sterilized and was validated to a sterility assurance level (SAL) of 10-6 in accordance with the following standards: - ISO 11135:2014 Sterilization of health-care products -- Ethylene oxide -- Requirements for the . development, validation and routine control of a sterilization process for medical devices. - ISO 10993-7:2008 Biological evaluation of medical devices -- Part 7: Ethylene oxide sterilization . residuals EtO residuals are tested to ISO 10993-7:2008 limits. Validation results indicate that the Liebel-Flarsheim™ IBP Transfer Set complies with the standards. #### 5.3.2 Shelf Life The proposed labeled shelf life of the Liebel-Flarsheim™ IBP Transfer Set is three years. Verification and validation testing in accordance with the framework described in (but not in full conformity with) ISO 11607 has been conducted, validated the process, and verified that the primary package (sterile barrier) maintains its integrity throughout the course of the labeled three year shelf life as further described in Section 12 - Sterilization and Shelf Life. Maintenance of functional user requirements over the proposed three year shelf life was also verified as described in Section 16 - Performance Testing - Bench. #### 5.3.3 Biocompatibility The Liebel-Flarsheim™ IBP Transfer Set indirect patient contact materials were verified in accordance with the following standards: - . ISO 10993-1:2018 Biological evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process. - Cytotoxicity o - Hemocompatibility O - Irritation o - Sensitization o - Acute Systemic Toxicity o - Material Mediated Pyrogenicity Test o {10}------------------------------------------------ - ISO 10993-18:2005 Biological Evaluation of Medical Devices – Part 18: Chemical Characterization of Medical Device Materials Within a Risk Management Process - USP <788> Particulate Matter in Injections - USP <85> Bacterial Endotoxin Test ● Verification results indicate that the materials comply with the standards. #### 5.3.4 Performance - Bench The Liebel-Flarsheim™ IBP Transfer Set was tested for performance and verified in accordance with the following standards: - ISO 8536-4 - 2010, Infusion equipment for medical use - Part 4: Infusion sets for single use, gravity feed. - . ISO 594-2, Conical fittings with 6 % (Luer) taper for syringes, needles and certain other medical equipment -- Part 2: Lock fittings. - ISO 22413:2010, Transfer sets for pharmaceutical preparations -- Requirements and test methods. Additional testing included: - Microbial Ingress Testing ● - Chemical Compatibility through Approved Release Specifications for Optiray® (Ioversol) Imaging Bulk ● Package - Extractables and simulation testing for leachable compounds and particulates - Verification of functional performance according to product specifications #### 5.4 Conclusions The transfer set has the same intended use as the predicate device, differing only in the claimed compatible contrast media and access duration. The principle features of the device that were described, as well as the testing provided, show that the minor differences in device characteristics between the subject device and predicate devices do not raise any new questions of safety or effectiveness. Performance data, including Bench Testing and Biocompatibility, have established that the transfer set performs as intended. The subject device is substantially equivalent to the legally marketed predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...