Albumin BCP

K193001 · SENTINEL CH. SpA · CJW · Dec 19, 2019 · Clinical Chemistry

Device Facts

Record IDK193001
Device NameAlbumin BCP
ApplicantSENTINEL CH. SpA
Product CodeCJW · Clinical Chemistry
Decision DateDec 19, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1035
Device ClassClass 2

Intended Use

The Albumin BCP assay is an in vitro diagnostic test used for the determination of albumin in human serum or plasma. Albumin measurements are used in the diagnosis and treatment of numerous diseases primarily involving the liver or kidneys. The assay is intended for professional use only. For In Vitro Diagnostic use only.

Device Story

Albumin BCP is an in vitro diagnostic reagent for quantitative albumin measurement in human serum or plasma. The device utilizes a colorimetric bromcresol purple (BCP) dye-binding method; albumin reacts with BCP at pH 5.0–5.5 to form a colored complex. The intensity of this complex, measured at 600/700 nm, is proportional to albumin concentration. The assay is intended for professional use in clinical laboratories using the AU680 Automatic Analyzer. Healthcare providers use the resulting albumin concentration to assist in diagnosing and monitoring liver and kidney diseases. The device is supplied as a ready-to-use liquid reagent.

Clinical Evidence

Bench testing only. Performance evaluated on the AU680 analyzer. Studies included LoB (0.3 g/L), LoD (0.8 g/L), LoQ (1.5 g/L), and precision (total CV ≤ 2.2%). Linearity confirmed up to 70 g/L. Interference testing showed no significant bias for hemoglobin, unconjugated/conjugated bilirubin, or lipids up to 1200 mg/dL. Method comparison against the predicate (ADVIA 2400) using 128 samples showed high correlation (r=0.992, slope 0.94). Matrix comparison (serum vs. Li-Heparin/K-EDTA plasma) confirmed equivalence.

Technological Characteristics

Colorimetric dye-binding assay using bromcresol purple. Reagents: acetate buffer (6.8%), stabilizer (4.38%), BCP (0.028%), detergent (0.045%), sodium azide (<0.1%). Liquid format. Standardized to ERM-DA 470k/IFCC. Operates on AU680 Automatic Analyzer. Measuring range 6.0–70 g/L. Reagent storage 15-30 °C.

Indications for Use

Indicated for the quantitative determination of albumin in human serum or plasma for the diagnosis and treatment of liver or kidney diseases. For professional use only.

Regulatory Classification

Identification

An albumin test system is a device intended to measure the albumin concentration in serum and plasma. Albumin measurements are used in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. December 19, 2019 Sentinel CH. SpA Patricia Dupe Head of Quality System Via Robert Koch, 2 Milano, IT 20152 Re: K193001 Trade/Device Name: Albumin BCP Regulation Number: 21 CFR 862.1035 Regulation Name: Albumin test system Regulatory Class: Class II Product Code: CJW Dated: October 23, 2019 Received: October 28, 2019 Dear Patricia Dupe: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Marianela Perez-Torres, Ph.D. Acting Deputy Director Division of Chemistry and Toxicology Devices | OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality CDRH | Food and Drug Administration Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. 510(k) Number (if known) K193001 Device Name Albumin BCP Indications for Use (Describe) The Albumin BCP assay is an in vitro diagnostic test used for the determination of albumin in human serum or plasma. Albumin measurements are used in the diagnosis and treatment of numerous diseases primarily involving the liver or kidneys. The assay is intended for professional use only. For In Vitro Diagnostic use only. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### 510(k) SUMMARY (SUMMARY OF SAFETY AND EFFECTIVENESS) ട്. ## for the Albumin BCP (6x20mL), Albumin BCP (6x50 mL) This summary of the 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. #### 5.1. 510 (k) Number: K193001 #### 5.2. Applicant Name SENTINEL CH. S.p.A. Via Robert Koch, 2 Milano (MI) 20152, Italy +39 02 345 514 1 Primary contact person for all communications: Patricia Dupé Head of Quality System Phone: +39 02 34 551 496 Fax: +39 02 34 551 464 Email: patriciadupe@sentinel.it Date Summary Prepared: October 23th, 2019 Date Summary Revised: December 13th, 2019 {4}------------------------------------------------ #### 5.3. Device Name ## Albumin BCP Device Classification: Class II Classification Name: Bromcresol purple dye-binding, albumin Governing Regulation: 21CFR862.1035 Product Code: CJW Panel: Clinical Chemistry (75) #### 5.4. Predicate Device ADVIA Chemistry BCP assay (ALBP) (k132664) #### 5.5. Intended Use of the Device The Albumin BCP assay is an in vitro diagnostic test used for the determination of albumin in human serum or plasma. Albumin measurements are used in the diagnosis and treatment of numerous diseases primarily involving the liver or kidneys. The assay is intended for professional use only. For In Vitro Diagnostic use only. #### 5.6. Description of Device Albumin BCP reagent is ready to use liquid reagent that is supplied in two configurations: fill volume 20 mL in a 20 mL wedge or 50 mL in a 50 mL wedge, 6 wedges/kit. | Test per kit calculation (based on the minimum<br>reagent fill volume) | | |------------------------------------------------------------------------|------| | Albumin BCP (6x20 mL) | 2700 | | Albumin BCP (6x50 mL) | 6000 | {5}------------------------------------------------ | Reactive Ingredients | Concentration | |----------------------|---------------| | Acetate Buffer | 6.8 % | | Stabilizer | 4.38 % | | Bromcresol purple | 0.028 % | | Detergent | 0.045 % | | Sodium azide | <0.1 %. | #### 5.7. Principles of the Procedure Albumin at pH 5.0 – 5.5 reacts with bromocresol purple (BCP) forming a colored complex. The color intensity of this complex is proportional to the concentration of albumin present in the sample measurement as an endpoint reaction at 600/700 nm. Methodology: Colorimetric #### 5.8. Comparison of Technological Characteristics The Albumin BCP assay is intended for the quantitative measurement of albumin in human serum and plasma. For In Vitro Diagnostic use only. A comparison of the candidate device (Albumin BCP) and the predicate device [ADVIA Chemistry BCP assay (ALBP)] is presented in Table 5.1 {6}------------------------------------------------ | Predicate Assay (k132664) | | | |----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Characteristics | ADVIA Chemistry Albumin BCP assay (ALBP) | Candidate Assay<br>Albumin BCP | | Technical Characteristics | | | | Classification | Regulation name: Albumin BCP assay | Same | | Method Principle | Bromcresol purple (BCP) dye-binding method | Same | | Intended<br>Use/Indications<br>for Use | For in Vitro diagnostic use in the quantitative<br>measurement of albumin in human serum or plasma on<br>ADVIA Chemistry Systems.<br>Albumin measurements are used in the diagnosis and<br>treatment of numerous diseases primarily involving the<br>liver or kidneys | The Albumin BCP assay is an in vitro diagnostic test used<br>for the determination of albumin in human serum or plasma.<br>Albumin measurements are used in the diagnosis and<br>treatment of numerous diseases primarily involving the<br>liver or kidneys.<br>The assay is intended for professional use only.<br>For In Vitro Diagnostic use only. | | Analyte Measured | Albumin | Same | | Instrument to be<br>used | Advia 1650 Chemistry System | AU680 Automatic Analyzer | | Measurement | quantitative | Same | | Specimen type | Serum, plasma | Same | | Reference Value | 34 - 50 g/L | 35 - 52 g/L | | Reagents | Single reagent | Same | | Format | Liquid | Same | | Analytical<br>measuring<br>interval | 6.0 - 80 g/L | 6.0 - 70 g/L | | Standardization | ERM-DA 470k Reference Material | Same (ERM-DA 470k/IFCC) | | | Predicate Assay (k132664) | | | Characteristics | ADVIA Chemistry Albumin BCP assay (ALBP) | Candidate Assay<br>Albumin BCP | | Performance Characteristics | | | | Reagent storage<br>temperature | 2-8 °C | 15-30 °C | | Use of Calibrators | Yes | Same | | Use of Controls | Yes | Same | ## Table 5.1: Comparison of Albumin BCP to Predicate (k132664) ADVIA Chemistry Albumin BCP assay (ALBP) {7}------------------------------------------------ {8}------------------------------------------------ #### 5.9. Summary of Performance Testing All performance were established using the AU680 analyzer. #### 5.9.1. Limit of Blank (LoB) Study The Limit of Blank (LoB) Study was performed to determine the highest measurement result that is likely to be observed (with a stated probability) for a blank sample. The LoB, study was performed based on guidance from the Clinical and Laboratory Standards Institute (CLSI) document EP17-A2. Acceptance criteria: the Limit of Blank (LoB) should be ≤ 1 g/L A summary of the results is presented below: | Reagent | Calibrator | Limit of<br>Blank (LoB)<br>g/L | Acceptance<br>criteria<br>g/L | |-----------|------------|--------------------------------|-------------------------------| | Lot F0390 | | 0.2 | | | Lot F0391 | Lot E0179 | 0.3 | $ \leq 1 $ g/L | | Lot F0480 | | 0.3 | | The observed Limit of Blank supports the LoB claim of 0.3 g/L. #### Limit of Detection (LoD) Study 5.9.2. LoD is the lowest analyte concentration likely to be reliably distinguished from the LoB and at which detection is feasible. The Limit of Detection (LoD) was determined on the basis of CLSI EP17-A2 Guideline (Evaluation of detection capability for clinical laboratory measurement procedures). Acceptance criteria: the Limit of Detection (LoD) should be ≤ 3 g/L. A summary of the results is presented below: | Reagent | Calibrator | Limit of<br>Detection<br>(LoD)<br>g/L | Acceptance<br>criteria<br>g/L | |-----------|------------|---------------------------------------|-------------------------------| | Lot F0390 | | 0.6 | | | Lot F0391 | Lot E0179 | 0.8 | ≤ 3.0 g/L | | Lot F0480 | | 0.8 | | The results meet the acceptance criteria. The Limit of Detection is chosen to be the highest of the values obtained. The LoD observed supports the claim of 0.8 g/L. {9}------------------------------------------------ #### Limit of Quantitation (LoQ) Study 5.9.3. LoQ is the lowest amount of a measured in a material that can be quantitatively determined with stated accuracy (as total error or as independent for bias and precision), under stated experimental conditions. The results of the LoQ study are provided for completeness in conjunction with the LoB and LoD studies. The LoQ, study was performed based on guidance from the Clinical and Laboratory Standards Institute (CLSI) document EP17-A2. Acceptance criteria: the Limit of Quantitation (LoQ) should be ≤ 5 g/L A summary of the results is presented below: | Reagent | Calibrator | Limit of<br>Quantification<br>(LoQ)<br>g/L | Acceptance<br>criteria<br>g/L | |-----------|------------|--------------------------------------------|-------------------------------| | Lot F0390 | | 1.3 | | | Lot F0391 | Lot E0179 | 1.3 | ≤ 5.0 g/L | | Lot F0480 | | 1.3 | | The results meet the acceptance criteria. The LoQ observed supports the claim of 1.5 g/L. #### 5.9.4. Precision Study The Precision study verifies the agreement between indications or measured quantity values obtained by replicates measurements on the same or similar objects under specified conditions. One human serum sample pool and two human serum based controls were used. After calibration at Time zero (0), 2 replicates of each sample were performed on 2 different runs per day. The precision was determined on the basis of CLSI EP05-A3 Guideline (Evaluation of Precision of Quantitative Measurement Procedures). Acceptance Criteria: The Precision should be ≤ 2.5% across all tested concentrations {10}------------------------------------------------ | Reagent | Material | N | Concentration | Total | Acceptance | |--------------|----------|----|---------------|-------|------------| | | | | g/L | %CV | criteria | | Lot | Level 1 | 80 | 26.26 | 2.2% | | | F0390 | Level 2 | 80 | 40.53 | 2.0% | | | | Level 3 | 80 | 49.96 | 1.8% | | | Lot<br>F0391 | Level 1 | 80 | 26.50 | 2.1% | | | | Level 2 | 80 | 40.64 | 1.8% | | | | Level 3 | 80 | 50.30 | 1.9% | < 2.5% | | Lot<br>F0480 | Level 1 | 80 | 26.62 | 2.2% | | | | Level 2 | 80 | 40.67 | 2.1% | | | | Level 3 | 80 | 50.47 | 1.6% | | | Lot 90228 | Level 1 | 88 | 19.10 | 1.9% | | | | Level 2 | 88 | 40.18 | 1.2% | | | | Level 3 | 88 | 51.33 | 1.3% | | A summary of the results is presented below: All calculated CV's were well within the acceptance criteria of ≤ 2.5%. The highest %CV was 2.2%. The data show good precision across the concentration range from 18 to 55 g/L. #### 5.9.5 Intra Assay Precision Study The Intra Assay Precision Study was performed to verify the precision and the trueness of the method relative to the assigned values of materials with known concentrations. One human serum sample pool and two human serum based controls were used. After calibration, 20 replicates of each sample were run on 3 different runs (each run with a new calibration). The Intra Assay Precision was determined on the basis of CLSI EP15-A3 Guideline (User Verification of Precision and Estimation of Bias). Acceptance Criteria: The Intra Assay Precision should be ≤ 1.5 % across all tested concentrations {11}------------------------------------------------ | Reagent | Material | Concentration | Total | SD | Acceptance criteria | |-----------|----------|---------------|-------|-----|-----------------------------------| | | | g/L | %CV | g/L | % CV | | Lot F0390 | Level 1 | 21.4 | 0.75% | 0.2 | $ \u2264 $ 1.5% | | | Level 2 | 35.7 | 0.79% | 0.3 | | | | Level 3 | 50.2 | 0.40% | 0.2 | | | Lot F0391 | Level 1 | 21.5 | 0.74% | 0.2 | | | | Level 2 | 35.8 | 0.61% | 0.2 | | | | Level 3 | 50.3 | 0.49% | 0.2 | | | Lot F0480 | Level 1 | 21.5 | 0.95% | 0.2 | | | | Level 2 | 35.6 | 0.74% | 0.3 | | | | Level 3 | 50.3 | 0.44% | 0.2 | | A summary of the results is presented below: The results meet the acceptance criteria. All the samples gave %CV lower than 1.5% #### 5.9.6. Linearity (Measuring Range) The Study was performed to establish the Upper Limit of Measuring Range (MR) based upon the linearity of the Albumin BCP assay on the AU680 System. Linearity was evaluated using guideline CLSI EP6-A Guideline (Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach). Acceptance Criteria: the absolute bias should be between - 2 g/L and + 2 g/L or The relative bias should be between -6% and + 6% A summary of the results is presented below: | Reagent | Tested range (g/L) | Linear regression equations | |-----------|--------------------|------------------------------| | Lot F0390 | 4.17 to 78.30 | y = 0.00 + 1.000x, r = 0.999 | | Lot F0391 | 4.30 to 78.67 | y = 0.00 + 1.000x. r = 0.999 | | Lot F0480 | 4.20 to 78.03 | y = 0.00 + 1.000x, r = 0.999 | The results meet the acceptance criteria, demonstrating that the assay is linear up to 70 g/L. The upper limit of the MR is 70 g/L. The MR is claimed as 6.0 to 70 g/L. {12}------------------------------------------------ #### Endogenous Interferences Study 5.9.7 The Endogenous Interferences study was performed to evaluate the effect of some substances on the performances of the product. The interference was evaluated at the following Albumin concentrations: Low: ~ 35 g/L (±10%) High: ~ 50 g/L (±10%) For each concentration 2 aliquots of serum pool were prepared (Base and Test pool) Test pool was divided into 4 sub-aliquots, then spiked using the interfering substances under evaluation. The 4 Test Pool aliquots were diluted using the Base Pool to obtain additional dilution levels in the ratio: 100%, 90%, 80%, 70%, 60%, 30%, 20%, 10%, 0%. Potential interferences study was performed on the basis of CLSI document EP07-A2 -Interference Testing in Clinical Chemistry Acceptance Criteria: | Interfering substance | Tested<br>concentration up to | Acceptance criteria | |---------------------------|-------------------------------|---------------------| | Hemoglobin | 2000 mg/dL | | | Unconjugated bilirubin | 66 mg/dL | % bias: ±10% | | Conjugated bilirubin | 66 mg/dL | | | Lipids (as Triglycerides) | 2000 mg/dL | | A summary of the results is presented below: | Interfering<br>substance | Tested concentration<br>up to | Found concentration<br>without<br>Interferences | Acceptance<br>criteria | |------------------------------|-------------------------------|-------------------------------------------------|------------------------| | Hemoglobin | 2000 mg/dL | 2000 mg/dL | | | Unconjugated<br>bilirubin | 66 mg/dL | 66 mg/dL | | | Conjugated<br>bilirubin | 66 mg/dL | 66 mg/dL | % bias: ±10% | | Lipids (as<br>Triglycerides) | 2000 mg/dL | 1200 mg/dL | | The results meet the acceptance criteria. {13}------------------------------------------------ #### 5.9.8 Reagent stability The Stability study was performed to determine, both open on-board and calibration stability of the Albumin BCP assay on the AU680 System. The Stability study was determined following: European Standard EN 23640:2015 - Stability Testing of In Vitro Diagnostics Reagents Acceptance Criteria % Bias: within ± 10% vs initial measurement | Reagent | Material | %bias | % bias | Acceptance criteria | |--------------|----------|-------|--------|---------------------| | | | Min | Max | % bias vs Time 0 | | Lot<br>F0390 | Level 1 | 0.0% | 8.5% | | | | Level 2 | 1.2% | 7.9% | | | | Level 3 | 0.65% | 7.3% | | | Lot<br>F0391 | Level 1 | 0.0% | 9.8% | | | | Level 2 | 1.0% | 9.9% | | | | Level 3 | 0.1% | 9.6% | ±10% | | Lot<br>F0480 | Level 1 | 1.6% | 10.0% | | | | Level 2 | 0.6% | 9.8% | | | | Level 3 | 0.3% | 8.8% | | | Lot<br>90228 | Level 1 | -1.3% | 7.3% | | | | Level 2 | 0.0% | 5.9% | | | | Level 3 | 0.5% | 5.2% | | A summary of the results is presented below: The results meet the acceptance criteria. The on-board reagent stability claimed is 28 days on the AU680 analyzer. The calibration stability claimed is 14 days on the AU680 analyzer. #### 5.9.9 Method Comparison Method Comparison study was performed to demonstrate correlation to the predicate device (Siemens ADVIA® 2400). 128 serum samples with concentrations spanning the measuring interval 6.0 - 70 g/L were tested in duplicate on the AU680 System and on Siemens Advia 2400. 8 samples were altered samples, prepared for a better coverage of the measuring range. The method comparison study was performed based on guidance from the CLSI document EP09-A3. {14}------------------------------------------------ Acceptance Criteria: The method comparison study results were considered acceptable if the assay had a regression slope of 1.00 (± 0.10) and a correlation coefficient (r) of ≥ 0.975. A Passing-Bablok linear regression and a linear fit regression were performed, regressing the 1st replicate of the SCH Albumin BCP assay (y-axis) versus the mean of 2 replicates of each corresponding sample of the predicate (ADVIA® Chemistry Albumin BCP) assay (x-axis). A summary of the results is presented below: Passing & Bablok regression method: N = 128; y (g/L) = 0.94x + 1.01; r = 0.992 Linear fit regression method: N = 128; y (g/L) = 0.95x + 0.78; r = 0.992 SCH Albumin BCP assay on the AU680 demonstrated acceptable correlation to the predicate device (Siemens ADVIA 2400, ADVIA® Chemistry Albumin BCP) by meeting the evaluation criteria. ## 5.9.10 Matrix Comparison Study Matrix Comparison study was performed to demonstrate correlation between: - serum and Lithium-Heparin plasma samples A - ア serum and Potassium EDTA plasma samples 77 paired plasma/serum samples with albumin concentrations throughout the range of the assay were tested. 7 samples were altered samples, prepared for a better coverage of the measuring range. The matrix comparison study was performed based on guidance from the CLSI document EP09-A3. The matrix comparison study results were considered acceptable if the assay had a regression slope of 1.00 (± 0.10) and a correlation coefficient (r) of ≥ 0.975. A Passing-Bablok linear regression and a linear fit regression were performed, regressing the 1st replicate of the SCH Albumin BCP plasma, Li-Heparin (y-axis) versus the 1st replicate of each corresponding sample of the SCH Albumin BCP Serum (x-axis). A Passing-Bablok linear regression and a linear fit regression were performed, regressing the 1st replicate of the SCH Albumin BCP plasma, potassium EDTA (y-axis) versus the 1st replicate of each corresponding sample of the SCH Albumin BCP Serum (x-axis). {15}------------------------------------------------ A summary of the results, Serum vs Lithium-Heparin plasma samples is presented below: Passing & Bablok regression method: N = 77; y (g/L) = 1.01x - 0.34; r = 0.995 Linear fit regression method: N = 77; y (g/L) = 1.00x + 0.01; r = 0.995 A summary of the results, Serum vs Potassium EDTA plasma samples is presented below: Passing & Bablok regression method: N = 77; y (g/L) = 1.00x - 0.20; r = 0.996 Linear fit regression method: N = 77; y (g/L) = 0.995x − 0.05; r = 0.996 The results meet the acceptance criteria. #### 5.10. Conclusion Testing results indicate that the Candidate device (Albumin BCP) is safe and effective for the stated intended use and is substantially equivalent to the predicate device.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%