Ezra Plexo Software is a medical diagnostic application for viewing, manipulation, 3D- visualization, and comparison of MR medical images. The images can be viewed in a number of output formats including volume rendering. Ezra Plexo Software enables visualization of information that would otherwise have to be visually compared disjointedly. Ezra Plexo Software is designed to support the oncological workflow by helping the user to confirm the absence or presence of lesions, including evaluation, quantification, and documentation of any such lesions. Note: The clinician retains the ultimate responsibility for making the pertinent diagnosis based on their standard practices and visual comparison of the separate unregistered images. Ezra Plexo Software is a complement to these standard procedures.
Device Story
Web-based medical diagnostic application; enables 2D/3D visualization, manipulation, and comparison of MR images. Inputs: previously generated MR data files (no direct scanner interface). Processing: volumetric segmentation of regions of interest; manual and semi-automated segmentation tools; linear dimension and volume calculation. Outputs: on-screen analysis results, segmented images, and quantitative measurements. Used by radiologists in clinical settings to support oncological workflows; assists in confirming presence/absence of lesions. Clinician retains ultimate diagnostic responsibility; software serves as a complement to standard clinical practices.
Clinical Evidence
Non-clinical software validation performed on 150 patient exams. Consensus ground truth established by five U.S. board-certified expert radiologists. Testing confirmed device performance against design specifications and intended use; no clinical data provided.
Technological Characteristics
Web-based software (Chrome Desktop); DICOM compliant. Features include 2D/3D visualization, manual and semi-automated segmentation (MR prostate studies), and quantitative measurements. Developed per ISO 14971:2007, IEC 62304:2006/A1:2015, IEC 62366-1:2015, and NEMA PS 3.1-3.20 (2016).
Indications for Use
Indicated for use by clinicians to view, manipulate, 3D-visualize, and compare MR medical images to support oncological workflows, including the evaluation, quantification, and documentation of lesions.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a stylized human figure on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The text is in blue, with "FDA" in a square and the rest of the text in a smaller font.
July 10, 2020
Ezra AI Inc. % Mr. David Girard Quality Manager 79 Madison Ave., Ste 545 NEW YORK NY 10016
Re: K192969
Trade/Device Name: Ezra Plexo Software Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: QIH Dated: June 9, 2020 Received: June 11, 2020
Dear Mr. Girard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known)
K192969
Device Name Ezra Plexo Software
Indications for Use (Describe)
Ezra Plexo Software is a medical diagnostic application for viewing, manipulation, 3D- visualization, and comparison of MR medical images. The images can be viewed in a number of output formats including volume rendering.
Ezra Plexo Software enables visualization of information that would otherwise have to be visually compared disjointedly.
Ezra Plexo Software is designed to support the oncological workflow by helping the user to confirm the absence or presence of lesions, including evaluation, quantification, and documentation of any such lesions.
Note: The clinician retains the ultimate responsibility for making the pertinent diagnosis based on their standard practices and visual comparison of the separate unregistered images. Ezra Plexo Software is a complement to these standard procedures.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|-----------------------------------------------|
| X Prescription Use (Part 21 CER 801 Subnart D) | Over-The-Counter Ise (21 C.FR 801 Subnart C |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary 1
This 510(k) Summary is in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content of this 510(k) summary is provided in conformance with 21 CFR Part 807.92.
#### Submitter Information 1.1
| Submitter's Name | Ezra Al Inc. |
|-----------------------------|-------------------------------------------------------|
| Address | 79 Madison Avenue, Suite 545, New York, NY, 10016 USA |
| Telephone Number | 888-402-3972 |
| Contact Person | David Girard |
| Date of Summary Preparation | July 2, 2020 |
#### Subject Device Information 1.2
| Device Name | Ezra Plexo Software |
|----------------|-------------------------------------------------------------|
| Common Name | Plexo |
| Classification | II |
| Review Panel | Radiology |
| Product Code | QIH |
| Regulations | 892.2050 - Automated radiological image processing software |
#### 1.3 Predicate Device Information
Arterys Oncology DL Device Name 510(k) Number K173542
{4}------------------------------------------------
#### Device Description വ
Ezra Plexo is a medical image application for 2D and 3D visualization, comparison, and manipulation of medical images. Plexo can be accessed through a web browser. It provides radiologists the ability to view and manipulate volumetric data such as MRI. The product allows volumetric segmentation of regions of interest, while enabling users to edit such segmentations and to take quantitative measurements.
Plexo does not interface directly with the MR scanner or any other data collection equipment. Instead, it uploads data files previously generated by such equipment. Its functionality is independent of the acquisition equipment vendor. Its analysis results are available on screen and can be saved for review.
#### Indications for Use 3
Ezra Plexo Software is a medical diagnostic application for viewing, manipulation, and comparison of medical MR images can be viewed in a number of output formats including volume rendering.
Ezra Plexo Software enables visualization of information that would otherwise have to be visually compared disjointedly.
Ezra Plexo Software is designed to support the oncological workflow by helping the user confirm the absence or presence of lesions, including evaluation, quantification, and documentation of any such lesions.
Note: The clinician retains the ultimate responsibility for making the pertinent diagnosis based on their standard practices and visual comparison of the separate unregistered images. Ezra Plexo Software is a complement to these standard procedures.
{5}------------------------------------------------
### Summary of Technological 4 Characteristics Comparison
Table 1 shows that Plexo Software and the Predicate Device (K173542) are equivalent in technological characteristics.
| Feature | Subject Device:<br>Ezra Plexo Software | Predicate Device:<br>Arterys Oncology DL<br>(K173542) |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Platform | Chrome Desktop | Client-server, Chrome Desktop |
| DICOM Compliant | Yes | Yes |
| Type of Scans | MR | MR, CT |
| Image Navigation<br>Tools | Pan, zoom, rotate, slice scroll (view<br>multiple slices)<br>Adjust window level, minimize/maxi-<br>mize<br>Orientation labels, scroll sync, ruler | Pan, zoom, rotate, slice scroll (view<br>multiple slices)<br>Adjust window level, minimize/maxi-<br>mize<br>Orientation labels, cross-reference in-<br>dicator |
| Image Display Modes | Yes, static | Yes, static and cine |
| Text Annotation on<br>the Image | Yes | Yes |
| Layout | Can select a viewport layout<br>and add series to it | Can select a viewport layout<br>and add series to it |
| 2D Image Review | Yes | Yes |
| 3D Image Review | Yes, 3D review of segmented region | Yes, 3D review of the full image |
| Create MPR Images | No | Yes |
| Manual Segmentation | Yes | Yes |
| 2D Semi-Automated<br>Segmentation | Yes, available only for MR prostate<br>studies | Yes, available only for lung CT and<br>liver MR studies |
| Linear Dimension Cal-<br>culation | Yes | Yes |
| Volume Calculation | Yes | Yes |
Table 1: Subject and Predicate Device Comparison
{6}------------------------------------------------
### Performance Testing 5
The Ezra Plexo Software has been developed in a manner consistent with accepted standards for software development and evaluated in accordance with design specifications and applicable performance standards through software verification, validation, and usability testing.
The nonclinical software validation tests were performed on 150 patient exams utilizing consensus ground truth created by five U.S. board certified expert radiologists.
The test results demonstrated that the Ezra Plexo Software performs to its intended use, is deemed acceptable for clinical use, and does not introduce new questions of safety or efficacy. The testing was conducted in accordance to the software validation/verification plans and protocols. A full description of the software functionality, device hazard analysis, software requirements, verification, validation and usability study is provided in this submission.
### Conformance Standards б
There are no applicable FDA mandated performance standards for this device. However, voluntary standards have been utilized in the production of the software. The device was designed and developed in accordance to the following conformance standards:
- · ISO 14971:2007- Medical devices Application of risk management to medical devices
- · IEC 62304:2006/A1:2015 Medical device software Software life cycle processes
- · IEC 62366-1:2015 Medical Devices Application of usability engineering to medical devices
- · NEMA PS 3.1-3.20 (2016) Digital Imaging and Communications in Medicine (DICOM) set
### Substantial Equivalence Conclusion 7
The Ezra Plexo Software has the same intended use and similar technological characteristics as the predicate Arterys Oncology DL (K173542) device. As demonstrated in this submission, the subject and predicate device are identical in indications for use and intended use. The subject Ezra Plexo Software is substantially equivalent to the predicate Arterys Oncology DL (K173542) device, and the minor differences in the technological characteristics of the subject and predicate device do not raise any new or different questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.