K192966 · Laser Associated Sciences, Inc. · DPW · Jan 10, 2020 · Cardiovascular
Device Facts
Record ID
K192966
Device Name
FlowMet
Applicant
Laser Associated Sciences, Inc.
Product Code
DPW · Cardiovascular
Decision Date
Jan 10, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2100
Device Class
Class 2
Indications for Use
The FlowMet™ is a non-invasive probe that is affixed to the fingers or toes and intended to quantify tissue blood flow rate.
Device Story
FlowMet is a non-invasive peripheral blood flow monitor. It utilizes laser speckle imaging to quantify tissue blood flow rate in fingers or toes. The device consists of an infrared laser diode (785 nm) and a digital CMOS camera sensor. The laser light transilluminates the digit, and the camera captures changes in the laser speckle pattern caused by moving blood cells within the tissue. The device is affixed to the digit using medical tape. It is a single-use device. The system provides quantitative blood flow data to clinicians, assisting in the assessment of peripheral tissue perfusion. It is intended for prescription use.
Clinical Evidence
No clinical testing was performed. Substantial equivalence was established through bench testing comparing the subject device to the predicate using a tissue analog model with controlled volumetric flow rates ranging from 2-20 ml/min. Results showed high linearity (R2 > 0.99) and a coefficient of variation < 5% for both devices, with a correlation coefficient of R > 0.999 between the devices.
Technological Characteristics
Laser speckle imaging; 785 nm infrared laser diode (Class 1 per IEC 60825-1:2014); digital CMOS camera detector; transillumination geometry. Materials: ABS/PC housing, silicone, and acrylic adhesives (biocompatibility per ISO-10993). Single-use form factor affixed via medical tape. Non-sterile.
Indications for Use
Indicated for quantification of tissue blood flow rate in fingers or toes in patients requiring non-invasive peripheral blood flow monitoring.
Regulatory Classification
Identification
A cardiovascular blood flowmeter is a device that is connected to a flow transducer that energizes the transducer and processes and displays the blood flow signal.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 26, 2021
Laser Associated Sciences, Inc. Sean White President and CEO 5171 California Ave., Suite 150 Irvine, California 92617
Re: K192966 Trade/Device Name: FlowMet Regulation Number: 21 CFR 870.2100 Regulation Name: Cardiovascular blood flowmeter Regulatory Class: Class II Product Code: DPW
Dear Sean White:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated January 10, 2019. Specifically, FDA is updating this SE Letter to correct the SE letter date to January 10, 2020 as an administrative correction.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact LCDR Stephen Browning, OHT2: Office of Cardiovascular Devices, with phone number of 240-402-5241, and email address of Stephen.Browning@fda.hhs.gov.
Sincerely.
## Stephen C. Browning -S
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font below.
January 10, 2019
Laser Associated Sciences, Inc. Sean White President and CEO 5171 California Ave., Suite 150 Irvine, California 92617
Re: K192966
Trade/Device Name: FlowMet Regulation Number: 21 CFR 870.2100 Regulation Name: Cardiovascular Blood Flowmeter Regulatory Class: Class II Product Code: DPW Dated: December 10, 2019 Received: December 11, 2019
Dear Sean White:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{3}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
### Stephen C. Browning -S5
Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Indications for Use
510(k) Number (if known) K192966
Device Name FlowMet
Indications for Use (Describe)
The FlowMet is a non-invasive probe that is affixed to the fingers or toes and intended to quantify tissue blood flow rate.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510(k) Summary
#### Submitter's Name and Address
Laser Associated Sciences, Inc. 5171 California Ave. Suite 150 Irvine, CA 92617 Tel: (949) 662-8892 Contact Person for this submission: Sean White
#### Date of Summary
The summary was prepared 1 of September 2019 and revised on 6 of December 2019.
#### Device Information
| Trade name: | FlowMet™ |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------|
| Model No: | FlowMet™ |
| Type of product: | Finished product |
| Panel: | Cardiovascular |
| Common Name: | Peripheral Blood Flow Monitor |
| Classification Name: | Cardiovascular blood flow meter |
| Indications for Use: | The FlowMet™ is a non-invasive probe that is affixed to the fingers or toes and intended to quantify tissue blood flow rate. |
| Class: | II |
| Classification Regulation: | 870.2100 |
| Product Code: | DPW |
#### Predicate Device Information
| Trade name: | FlowMet-R™ |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| Model No: | FlowMet-R™ |
| Type of product: | Finished product |
| Panel: | Cardiovascular |
| Common Name: | Peripheral Blood Flow Monitor |
| Classification Name: | Cardiovascular blood flow meter |
| Indications for Use: | The FlowMet-R™ is a non-invasive probe that is affixed to<br>the fingers or toes and intended to quantify tissue blood<br>flow rate. |
{6}------------------------------------------------
| Class: | II |
|----------------------------|----------|
| Classification Regulation: | 870.2100 |
| Product Code: | DPW |
| 510(k) No. | K182494 |
#### Device Description
#### Intended Use / Indications for Use
The FlowMet™ is a non-invasive probe that is affixed to the fingers or toes and intended to quantify tissue blood flow rate.
#### Summary of technological characteristics of Device and Predicate Device
Both the FlowMet-R™ and the FlowMet™ use the same fundamental scientific technology for the assessment of peripheral tissue blood flow: Laser Speckle Imaging. Both devices also share the same indications for use. The predicate device, the FlowMet-R™, affixes to the digit via a spring-loaded clip-on mechanism; whereas the FlowMet™ device affixes to the digit using medical tape. Additionally, the FlowMet-R™ can be cleaned and re-used, whereas the FlowMet™ is a single use device.
#### Comparison to the predicate device K182494, FlowMet-R™
#### Laser safety
The FlowMet-R™ and FlowMet™ are classified as a Class I laser product according to IEC 60825-1:2014, and employ identical laser diodes as the energy source.
#### Measurement Site
The FlowMet-R™ and FlowMet™ use the same measurement site: finger or toe.
#### Sterility
The FlowMet-R™ and FlowMet™ are both supplied non-sterile.
#### Affixed sensor
The FlowMet™ probe is affixed to the digit (finger or toe) using a medical tape that is wrapped around the digit, whereas the FlowMet-R™ uses a clip-on probe designed to be affixed to the fingers or toes via pressure/friction.
#### Performance Data
FlowMet-R™ and FlowMet™ performance was verified under known flow rate conditions using the same performance test that established substantial equivalence between the FlowMet-R™ and its predicate. The FlowMet-R™ and FlowMet™ measured the same sample concurrently: a tissue analog containing fluid pumped through at controlled volumetric flow rates. The flow rates were varied from 2-20ml/min, which includes and exceeds the normal human physiological range. At each flow rate, data was collected concurrently from both devices. Both devices exhibited high linearity (FlowMet-R™ R2 > 0.99, FlowMet™ R2> 0.99) between data output and flow rate, and both devices exhibited a coefficient of variation between trials of less than 5%. Additionally, the correlation coefficient of measured flow rate between the devices was R>0.999.
{7}------------------------------------------------
#### Clinical Data
No clinical testing was performed.
A description of the technological characteristics of the FlowMet-R™ system and FlowMet™ system is provided below.
| Device | FlowMet-RTM,<br>K182494 | FlowMetTM,<br>Pending |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications Use | A non-invasive probe that is<br>affixed to the fingers or toes and<br>intended to quantify tissue blood<br>flow rate. | Same as predicate. |
| Intended Use | A non-invasive probe that is<br>affixed to the fingers or toes and<br>intended to quantify tissue blood<br>flow rate. | Same as predicate. |
| Fundamental<br>Scientific<br>Technology | Laser speckle imaging, wherein<br>changes in the contrast of a laser<br>speckle pattern are caused by<br>the movement of blood within<br>tissue, which is captured using a<br>camera sensor. | Same as predicate. |
| Type of Use | Reusable | Single-use |
| Light Source | Infra-red Laser Light,<br>785 nm,<br>Class 1 per IEC 60825-1:2014 | Same as predicate |
| Detector | Digital CMOS camera for laser<br>speckle imaging. | Same as predicate |
| Biocompatibility | Material intended to contact skin<br>(silicone rubber) tested for<br>biocompatibility per ISO-10993. | Materials intended to contact skin<br>(silicone and acrylic adhesives)<br>tested for biocompatibility per<br>ISO-10993. |
| Physical Structure | Light source (laser diode) and<br>detector (CMOS camera)<br>secured in clip-on ABS/PC<br>housing. Light source and<br>detector are oriented to allow<br>transillumination of digit. | Light source (laser diode) and<br>detector (CMOS camera)<br>secured in ABS/PC housings<br>which are affixed to the digit<br>using medical tape. Light source<br>and detector are oriented to allow<br>transillumination of digit. |
#### Summary
The FlowMet™ and the FlowMet-R™ both use the same fundamental scientific technology for the measurement of blood flow rate and both devices share the same indications for use. The primary change from the predicate to the FlowMet™ is the transition from a spring-loaded clip to medical tape for attachment of the digit. This change results in the FlowMet™ probe becoming single use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.