← Product Code [OYJ](/productcode/OYJ) · K192920

# Oragene®•Dx (K192920)

_DNA Genotek, Inc. · OYJ · Jan 14, 2020 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K192920

## Device Facts

- **Applicant:** DNA Genotek, Inc.
- **Product Code:** [OYJ](/productcode/OYJ.md)
- **Decision Date:** Jan 14, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1675
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

Oragene®●Dx is intended for use in the non-invasive collection of saliva samples for in vitro diagnostic testing of human DNA. Saliva may be collected by spitting directly into the Oragene®●Dx container or may be transferred into the Oragene®●Dx container using a sponge. Saliva samples may be collected by a healthcare professional or non-healthcare professional, such as a lay user. Saliva samples collected using Oragene®•Dx are stabilized and isolated for use in downstream diagnostic testing applications. Saliva samples collected using Oragene®•Dx can be transported and/or stored long term at ambient conditions.

## Device Story

Oragene®•Dx is a non-invasive saliva collection device for human DNA. It consists of a collection tube with a funnel lid containing a stabilizing liquid. Users expectorate saliva directly into the tube or use an optional sponge for assisted collection. Upon contact, the stabilizing liquid lyses cellular and nuclear membranes to release and stabilize nucleic acids. The device allows for transport and long-term storage at ambient conditions. It is used in clinical or home settings by healthcare professionals or lay users. The stabilized sample is subsequently used in downstream molecular diagnostic applications. The device benefits patients by providing a non-invasive alternative to blood collection for DNA-based testing.

## Clinical Evidence

No clinical diagnostic data. A lay-user study (n=206) evaluated collection and labeling comprehension for the OGD-610 model. 97.35% of samples met minimum DNA requirements for testing; 98.41% after re-extraction. Results demonstrate that lay users can follow instructions to obtain adequate samples in an OTC setting.

## Technological Characteristics

Non-sterile plastic collection tube with funnel lid and stabilizing liquid. Principle: chemical lysis and stabilization of nucleic acids. Dimensions/form factor: tube-based. Connectivity: N/A. Sterilization: N/A. Software: None.

## Regulatory Identification

A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.

## Predicate Devices

- Oragene®•Dx Collection Device, models OGD-500; OGD-575; OXD-525; OYD-500 ([K110701](/device/K110701.md))

## Reference Devices

- eSensor® Warfarin Sensitivity Saliva Test (k110786)
- 23andMe PGS ([DEN140044](/device/DEN140044.md))
- 23andMe PGS ([DEN160026](/device/DEN160026.md))
- Akonni TruDiagnosis® System (k183530)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA

U.S. FOOD &amp; DRUG

ADMINISTRATION

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

ASSAY ONLY

## I Background Information:

A 510(k) Number

K192920

B Applicant

DNA Genotek Inc.

C Proprietary and Established Names

Oragene®•Dx

D Regulatory Information

|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  OYJ | Class II | 21 CFR 862.1675 - Blood Specimen Collection Device | CH - Clinical Chemistry  |

## II Submission/Device Overview:

A Purpose for Submission:

Modification to an existing device

B Measurand:

Not applicable

C Type of Test:

Saliva Collection for DNA testing

## III Intended Use/Indications for Use:

A Intended Use(s):

See Indications for Use below.

Food and Drug Administration

10903 New Hampshire Avenue

Silver Spring, MD 20993-0002

www.fda.gov

{1}

K192920 - Page 2 of 5

## B Indication(s) for Use:

Oragene®•Dx is intended for use in the non-invasive collection of saliva samples for in vitro diagnostic testing of human DNA. Saliva may be collected by spitting directly into the Oragene®•Dx container or may be transferred into the Oragene®•Dx container using a sponge. Saliva samples may be collected by a healthcare professional or non-healthcare professional, such as a lay user. Saliva samples collected using Oragene®•Dx are stabilized and isolated for use in downstream diagnostic testing applications. Saliva samples collected using Oragene®•Dx can be transported and/or stored long term at ambient conditions.

## C Special Conditions for Use Statement(s):

OTC - Over The Counter

Device manufacturers must validate the use of Oragene®•Dx for use with their device.

## D Special Instrument Requirements:

None.

## IV Device/System Characteristics:

### A Device Description:

Oragene®•Dx (models OGD-500, OGD-510, OGD-575, OGD-600, OGD-610, and OGD-675) consists of a saliva collection tube with a funnel lid attached that contains a stabilizing liquid. Saliva is delivered directly by expectorating into the collection tube. A small cap is provided to close the tube for transport and storage. The difference in the six models is the amount and/or concentrations of the reagents in the tube, which vary because of the difference in the amount of saliva collected. The ratio of final sample to stabilizing liquid volume remains the same. The 600-series (OGD-600 and OGD-610) differs from the 500-series (OGD-510, and OGD-500) in labeling only (600 series labeling is in English only). The OGD-575 and OGD-675 models include a sponge for assisted collection of saliva.

Oragene-Dx can be used with both prescription use only and over-the-counter molecular diagnostic test systems.

### B Principle of Operation:

Oragene®•Dx collects and stabilizes human DNA from saliva; it can also be used for the transportation and long-term room temperature storage of a sample. Oragene®•Dx is a non-invasive alternative for collecting DNA for use with molecular diagnostic applications.

The collection device (all models) consist of a collection tube with a funnel lid attached that contains a stabilizing liquid. Saliva is delivered directly by spitting into the collection tube. After saliva is collected, the stabilizing liquid is mixed with the sample. A small cap is provided to close the tube for transport and storage (funnel with lid is removed and discarded). Upon contacting saliva cells, the stabilizing liquid lyses cellular and nuclear membranes to release and

{2}

stabilize nucleic acids. Samples can be immediately processed, transported or stored for future use.

## V Substantial Equivalence Information:

### A Predicate Device Name(s):

Oragene®•Dx Collection Device, models OGD-500; OGD-575; OXD-525; OYD-500

### B Predicate 510(k) Number(s):

K110701

### C Comparison with Predicate(s):

|  Similarities  |   |   |
| --- | --- | --- |
|  Device & Predicate Device(s): | K192920 | K110701  |
|  Intended Use | Intended for use in the non-invasive collection of saliva samples. | Same  |
|  Indications for use | For use in vitro diagnostic testing of human DNA | Same  |
|  Sample source | Human Saliva | Same  |
|  Collection Device Contents | Nucleic acid stabilization solution | Same  |
|  Differences  |   |   |
|  Device & Predicate Device(s): | K192920 | K110701  |
|  Tests intended for use with the saliva collection device | Downstream in vitro diagnostic tests of human germline DNA validated for use with the Oragene®•Dx | eSensor® Warfarin Sensitivity Saliva Test  |
|  Device model numbers | OGD-500, OGD-510, OGD-575, OGD0600, OGD-610, OGD-675 | OGD-500, OGD-575, OXD-525; OYD-500  |

## VI Standards/Guidance Documents Referenced:

- ISO 13485: Medical devices - Quality management systems — Requirements for regulatory purposes (2016)
- ISO 14971: Medical devices - Application of risk management to medical devices (2007)

## VII Performance Characteristics (if/when applicable):

### A Analytical Performance:

1. Precision/Reproducibility:

K192920 - Page 3 of 5

{3}

The reproducibility of the Oragene®•Dx collection device was established in k110701 and k141410.

2. Linearity:
Not applicable.

3. Analytical Specificity/Interference:
The effect of endogenous and exogenous interfering substances was established in k110701 and k141410.

4. Assay Reportable Range:
Not applicable.

5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Pre-collection shelf-life stability of the collection device, stability of samples post-saliva collection, and freeze-thaw stability of samples stored in the Oragene®•Dx collection device was established in k110701.

6. Detection Limit:
Sample volume tolerance studies were conducted in k110701.

7. Assay Cut-Off:
Not applicable.

B Comparison Studies:

1. Method Comparison with Predicate Device:
A method comparison study was previously performed in order to determine the accuracy of the genotype obtained on the eSensor® Warfarin Sensitivity Saliva Test (k110786) using saliva samples collected by the Oragene•Dx collection device as compared to bi-directional DNA sequencing in k110786.

2. Matrix Comparison:
A matrix comparison study was conducted in k110701.

C Clinical Studies:

1. Clinical Sensitivity:
Not applicable.

K192920 - Page 4 of 5

{4}

2. Clinical Specificity:

Not applicable.

3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):

Evaluation of lay user collection and labeling comprehension: A study to evaluate the lay user collection and labeling comprehension of the Oragene•Dx OGD-610 model was performed by over the counter users. The study consisted of 206 subjects 18 years or older and the demographics were representative of the US population. Participants independently collected a saliva sample at home using the instructions for use, answered the user comprehension questionnaire, and mailed both components according to the study instructions. Upon receipt at the testing laboratory, each study sample was assessed for compliance to collection instructions, sample volume, DNA concentration, and sequencing quality control metrics. User comprehension of test instructions, including comprehension of sample collection instructions was also assessed.

Of the samples evaluated, 97.35% contained the minimum amount of DNA required for testing after DNA extraction. Samples from two individuals were re-extracted from a second aliquot, for a total of 98.41% containing the minimum amount of DNA required for testing, demonstrating that customers were able to follow sample collection instructions in an over the counter setting to obtain adequate sample for testing.

A Flesch-Kincaid reading analysis was performed on the collection device labeling in k141410. Lay user collection and labeling comprehension for the Oragene•Dx OGD-500.001 device, when used with the over-the-counter 23andMe PGS Carrier Screening Test, was previously evaluated in k141410.

D Clinical Cut-Off:

Not applicable.

E Expected Values/Reference Range:

Not applicable.

VIII Proposed Labeling:

The labeling supports the finding of substantial equivalence for this device.

IX Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

K192920 - Page 5 of 5

---

**Source:** [https://fda.innolitics.com/device/K192920](https://fda.innolitics.com/device/K192920)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
