← Product Code [LLZ](/productcode/LLZ) · K192917

# ViewPoint 6 (K192917)

_Ge Medical Systems Ultrasound And · LLZ · Jan 6, 2020 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K192917

## Device Facts

- **Applicant:** Ge Medical Systems Ultrasound And
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Jan 6, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

ViewPoint 6 is intended to be used in medical practices and in clinical departments and serves the purposes of diagnostic interpretation of images, electronic documentation of examinations in the form of text and images, and generation of medical reports primarily for diagnostic ultrasound. ViewPoint 6 provides the user the ability to include images, drawings, and charts into medical reports. ViewPoint 6 is designed to accept, transfer, display, calculate, store, and process medical images and data, and enables the user to measure and annotate the images. The medical images, which 6 displays to the user, can be used for diagnostic purposes. ViewPoint 6 is intended for professional use only. ViewPoint 6 is not intended to be used as an automated diagnosis system. ViewPoint 6 is not intended to operate medical devices in surgery related procedures.

## Device Story

ViewPoint 6 is image archiving and reporting software for medical practices and clinical radiological departments. It accepts, transfers, displays, calculates, stores, and processes medical images and data. Users (clinicians) perform diagnostic interpretation, measurement, and annotation of images. The system provides exam-specific reporting forms with data entry fields, including images, drawings, and charts. It supports single workstation or client-server network configurations. The software integrates IOTA Classification ADNEX model and ACR TI-RADS risk-stratification tools to assist in diagnostic reporting. It is not an automated diagnosis system. The output is a professional medical report used by healthcare providers to support clinical decision-making and patient documentation.

## Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing, including module verification, system integration testing, design validation, and performance/safety testing. Compliance with standards: IEC 62366-1, IEC 62304, NEMA PS 3.1-3.20 (DICOM), ISO 14971, and IEC 82304-1.

## Technological Characteristics

Software-based image archiving and reporting system. Operates on Microsoft Windows hardware. Supports DICOM connectivity. Development follows IEC 62304 software lifecycle and ISO 14971 risk management. Features include IOTA ADNEX and ACR TI-RADS risk-stratification modules.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- ViewPoint 6 ([K173456](/device/K173456.md))

## Reference Devices

- Voluson E10 ([K192159](/device/K192159.md))
- Versana Balance ([K191792](/device/K191792.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized image of a bird, and the text on the right is in blue. The logo is simple and professional, and it is easily recognizable.

GE Medical Systems Ultrasound and Primary Care Diagnostics, LLC % Tracey Ortiz Regulatory Affairs Director 9900 W. Innovation Drive WAUWATOSA WI 53226

January 6, 2020

### Re: K192917

Trade/Device Name: ViewPoint 6 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: November 14, 2019 Received: November 15, 2019

Dear Tracey Ortiz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

K192917

510(k) Number (if known)

Device Name ViewPoint 6

Indications for Use (Describe)

ViewPoint 6 is intended to be used in medical practices and in clinical departments and serves the purposes of diagnostic interpretation of images, electronic documentation of examinations in the form of text and images, and generation of medical reports primarily for diagnostic ultrasound.

ViewPoint 6 provides the user the ability to include images, drawings, and charts into medical reports. ViewPoint 6 is designed to accept, transfer, display, calculate, store, and process medical images and data, and enables the user to measure and annotate the images. The medical images, which 6 displays to the user, can be used for diagnostic purposes.

ViewPoint 6 is intended for professional use only. ViewPoint 6 is not intended to be used as an automated diagnosis system.

ViewPoint 6 is not intended to operate medical devices in surgery related procedures.

Type of Use (Select one or both, as applicable)

|X Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in white inside. The letters are stylized and connected, and there are three white swirls around the circle. The logo is simple and recognizable, and it is associated with a well-known company.

# K192917

# GE Healthcare

510(k) Premarket Notification Submission

## 510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided: Date: October 11, 2019

| Submitter:                                                                  | GE Medical Systems Ultrasound and Primary Care<br>Diagnostics<br>9900 Innovation Drive<br>Wauwatosa, WI 53226                                                                                                                              |
|-----------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Primary Contact Person:                                                     | Tracey Ortiz<br>Regulatory Affairs Director<br>GE Healthcare<br>T:(262)676-6120                                                                                                                                                            |
| Secondary Contact<br>Person:                                                | Wei Liwen<br>Regulatory Affairs Leader<br>GE Healthcare                                                                                                                                                                                    |
| Trade Name:<br>Common/Usual Name:<br>Classification Names:<br>Product Code: | ViewPoint 6<br>PACS-Picture archiving and communications system<br>Class II<br>LLZ, Picture archiving and communications system,<br>21 CFR 892.2050, 90-LLZ                                                                                |
| Primary Predicate Device:<br>Product Code:                                  | ViewPoint 6 (K173456)<br>LLZ, Picture archiving and communications system,<br>21 CFR 892.2050, 90-LLZ                                                                                                                                      |
| Reference Device(s):<br>Product Code:                                       | Voluson E10 (K192159)<br>Versana Balance (K191792)<br>IYN (primary) & IYO (secondary)<br>IYN, Ultrasonic Pulsed Doppler Imaging System,<br>21CFR 892.1550, 90-IYN<br>IYO, Ultrasonic Pulsed Echo Imaging System, 21CFR<br>892.1560, 90-IYO |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo is a blue circle with the letters "GE" in a stylized font in the center. There are three water droplet shapes surrounding the circle. The logo is simple and recognizable, and it is associated with a well-known company.

510(k) Premarket Notification Submission

ViewPoint 6 is an image archiving and reporting Device Description: software for medical practices and clinical radiological departments. It is used for diagnostic interpretation of images and other data. It provides different calculations and tools to allow for the assessment of the images and data.

> ViewPoint 6 is for professional use only and enables quick diagnostic reporting with standardized terminology. It has an intuitive graphical user interface (GUI) and is based on Microsoft Windows® with defined hardware requirements for the user to install on their computer.

> ViewPoint 6 provides exam type specific reporting forms for various medical care areas. Forms are composed of different sections with data entry fields. The documentation can include measurements, exam findings, images, and graphs. All data is saved in the ViewPoint 6 database and can be compiled to a professional report. Images and image sequences can be reviewed in the ViewPoint 6 display area based on user preference.

> ViewPoint 6 supports both a single workstation and a client - server setup. The number of user licenses determines how many workstations in the network have concurrent access to the database. Access can be limited to read - only functionality.

ViewPoint 6 is intended to be used in medical Intended Use/Indication for Use: practices and in clinical departments and serves the purposes of diagnostic interpretation of images, electronic documentation of examinations in the form of text and images, and generation of medical reports primarily for diagnostic ultrasound.

ViewPoint 6 provides the user the ability to include images, drawings, and charts into medical reports. ViewPoint 6 is designed to accept, transfer, display, calculate, store, and process medical images and data, and enables the user to measure and annotate the images. The medical images, which ViewPoint 6 displays to the user, can be used for diagnostic purposes.

ViewPoint 6 is intended for professional use only. ViewPoint 6 is not intended to be used as an

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" in a stylized script, enclosed within a circle. There are three water droplet shapes surrounding the circle. The logo is colored in a bright blue hue.

# GE Healthcare

510(k) Premarket Notification Submission

automated diagnosis system. ViewPoint 6 is not intended to operate medical devices in surgery related procedures.

Technology: The ViewPoint 6 employs the same fundamental scientific technology as its predicate devices.

Determination of Comparison to Predicate Devices

Substantial Equivalence: The proposed ViewPoint 6 system is substantially equivalent to the ViewPoint 6 (K173456) with regards intended use, capabilities, technological to characteristics, safety and effectiveness.

The proposed ViewPoint 6 is adding

- IOTA Classification ADNEX Model which was ● cleared with Voluson E10 (K192159).
- . ACR TI-RADS Risk-stratification which was cleared in Versana Balance (K191792).
- Thyroid Nodule Overview Table comparison feature
- User defined section and fields

## Summary of Non-Clinical Tests:

The ViewPoint 6 and its applications comply with voluntary standards:

- IEC 62366 1:2015 Medical devices-Application . of usability engineering to medical devices
- IEC 62304:2006, Medical device software-● Software life cycle process
- NEMA PS 3.1-3.20 (2016), Digital Imaging and ● Communications in Medicine (DICOM) Set. (Radiology)
- . ISO 14971:2012 Medical Devices-Application of risk management to medical devices
- . IEC 82304-1:2016, Health software-General requirements for product safety

The following quality assurance measures are applied to the development of the system:

- . Risk Analysis
- Requirements Reviews
- . Design Reviews

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo, which consists of the letters 'GE' intertwined within a blue circle. Three stylized water droplets or splashes surround the circle, adding a dynamic element to the design. The logo is simple, recognizable, and represents the brand identity of General Electric.

# GE Healthcare

510(k) Premarket Notification Submission

- Testing on unit level (Module verification) ●
- Integration testing (System verification) ●
- Final Acceptance Testing (Design Validation ● and Service Validation)
- . Performance testing (Verification)
- . Safety testing (Verification)

## Summary of Clinical Tests:

The subject of this premarket submission, ViewPoint 6, did not require clinical studies to support substantial equivalence.

GE Healthcare considers the ViewPoint 6 to be as safe, Conclusion: as effective, and performance is substantially equivalent to the predicate devices.

---

**Source:** [https://fda.innolitics.com/device/K192917](https://fda.innolitics.com/device/K192917)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
