K192900 · Nonin Medical, Inc. · DQA · Jun 29, 2020 · Cardiovascular
Device Facts
Record ID
K192900
Device Name
H500 Multi-Sensing Oximetry System
Applicant
Nonin Medical, Inc.
Product Code
DQA · Cardiovascular
Decision Date
Jun 29, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Nonin Medical CO-Pilot™ Model H500 Multi-Sensing Oximetry System is intended for noninvasive measuring of functional oxygen saturation of arterial hemoglobin (%SpO2), pulse rate, carboxyhemoglobin saturation (%COHb), methemoglobin saturation (%MetHb), and cerebral or somatic hemoglobin oxygen saturation (%rSO2). This device is not meant for sole use in clinical decision making; it must be used in conjunction with additional methods of assessing clinical signs and symptoms. · For %SpO2 and pulse rate, the H500 System is intended for spot-checking and/or measuring during clinician assessment of adult, pediatric, infant, and neonate patients who are well or poorly perfused, during both motion and non-motion conditions in professional healthcare facilities, mobile, and EMS settings. · For %rSO2, the H500 System is intended for spot-checking and/or measuring during clinician assessment of adult, pediatric, infant, and neonate patients in professional healthcare facilities, mobile, and EMS settings. · For %COHb and %MetHb, the H500 System is intended for spot-checking, multiple spot-checks to observe change, and/ or measuring during clinician assessment of adult and pediatric patients in professional healthcare facilities, mobile, and EMS settings.
Device Story
Handheld, wireless multi-sensing oximetry system; consists of display, signal processor, and compatible sensors. Inputs: light-based signals from sensors applied to patient. Transformation: signal processor acquires data, translates into %SpO2, pulse rate, %COHb, %MetHb, and %rSO2. Output: physiological parameters displayed on device. Used in professional healthcare, mobile, and EMS settings by clinicians. Output used as adjunct to clinical assessment; aids in patient evaluation. Benefits: provides noninvasive, portable monitoring of hemoglobin saturation and pulse rate.
Clinical Evidence
Clinical accuracy testing conducted on healthy adult subjects (18+) for %COHb, %MetHb, %SpO2 (including in presence of COHb/MetHb), and %rSO2 (adults). Pediatric rSO2 testing conducted on patients (4 days to 10 years) in cardiac catheterization labs. Accuracy calculated as root-mean-squared (Arms) per ISO 80601-2-61. Results demonstrate performance equivalent to predicate devices.
Technological Characteristics
Handheld, wireless device; uses light-based sensing (optics). Compatible with Nonin-branded sensors. Connectivity via Bluetooth Low-Energy (BLE). Ingress protection IP33. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, and ISO 80601-2-61. Software-based signal processing.
Indications for Use
Indicated for noninvasive measurement of %SpO2, pulse rate, %COHb, %MetHb, and %rSO2 in adult, pediatric, infant, and neonate patients. Used for spot-checking or measurement during clinician assessment in professional healthcare, mobile, and EMS settings. Not for sole use in clinical decision-making.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Nonin's SenSmart X-100 Universal Oximetry System (K132402)
Nonin's X-100C/Multi-Sensing Oximetry Systems (K160231)
Nonin's X-100C/Multi-Sensing Oximetry Systems (K172625)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Nonin Medical, Inc. Brent Geiger Vice President Quality, Regulatory,Clinical Affairs 13700 1st Avenue North Plymouth, Minnesota 55441
Re: K192900
Trade/Device Name: Nonin Medical CO-Pilot Model H500 Multi-Sensing Oximetry System Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA, MUD Dated: May 21, 2020 Received: May 26, 2020
Dear Brent Geiger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192900
#### Device Name
Nonin Medical CO-Pilot™ Model H500 Multi-Sensing Oximetry System
Indications for Use (Describe)
The Nonin Medical CO-Pilot™ Model H500 Multi-Sensing Oximetry System is intended for noninvasive measuring of functional oxygen saturation of arterial hemoglobin (%SpO2), pulse rate, carboxyhemoglobin saturation (%COHb), methemoglobin saturation (%MetHb), and cerebral or somatic hemoglobin oxygen saturation (%rSO2). This device is not meant for sole use in clinical decision making; it must be used in conjunction with additional methods of assessing clinical signs and symptoms.
· For %SpO2 and pulse rate, the H500 System is intended for spot-checking and/or measuring during clinician assessment of adult, pediatric, infant, and neonate patients who are well or poorly perfused, during both motion and non-motion conditions in professional healthcare facilities, mobile, and EMS settings.
· For %rSO2, the H500 System is intended for spot-checking and/or measuring during clinician assessment of adult, pediatric, infant, and neonate patients in professional healthcare facilities, mobile, and EMS settings.
· For %COHb and %MetHb, the H500 System is intended for spot-checking, multiple spot-checks to observe change, and/ or measuring during clinician assessment of adult and pediatric patients in professional healthcare facilities, mobile, and EMS settings.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for NONIN. The logo consists of two overlapping orange squares on the left side of the image. To the right of the squares is the word "NONIN" in blue, with each letter being capitalized.
### K192900 510(k) Summary – H500 Multi-Sensing Oximetry System
| Submitter:<br>Phone:<br>Fax: | Nonin Medical, Inc.<br>13700 1st Ave. North<br>Plymouth, MN 55441-5443<br>763 553 9968<br>763 553 7807 |
|------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|
| Contact Person: | Brent Geiger, MS, RAC<br>Vice President Quality, Regulatory and Clinical Affairs |
| Date Prepared:<br>Trade Names: | January 31st, 2020<br>CO-Pilot™ Model H500 Multi-Sensing Oximetry System |
| Common Name:<br>Classification Name:<br>Regulation Number:<br>Product Code, Panel: | Oximeter<br>Oximeter<br>Class II, 21 CFR 870.2700 (Oximeter)<br>DQA, Anesthesiology<br>MUD, Cardiovascular |
#### Predicate Devices
Primary Predicate Device - Nonin's SenSmart X-100 Universal Oximetry System (refer to clearance K132402) is the primary predicate and is a modular system that is indicated for use in simultaneously measuring, displaying, monitoring, and recording up to six (6) channels of functional oxygen saturation of arterial hemoqlobin (%SpO2) and pulse rate or cerebral or somatic hemoqlobin oxygen saturation (%rSO2) of blood underneath the sensor. Patient populations include adult, pediatric, infant, and neonate through the use of SenSmart compatible sensors. The SenSmart system is intended for use in hospitals, long-term care, medical facilities, sleep laboratories and subacute environments. The X-100 SenSmart system may be used for spot-checking and continuous monitoring with patient alarms. The SenSmart pulse oximetry (%SpO2) functionality is suitable for use in both motion and non-motion conditions, including patients who are well or poorly perfused.
Reference Predicate Devices - Nonin's X-100C/Multi-Sensing Oximetry Systems (refer to clearances K160231 and K172625) are additional reference predicate devices for this submission and are indicated for noninvasive spot-checking and/or continuous monitoring of carboxyhemoglobin (%COHb), methemoglobin (%MetHb), functional oxygen saturation of arterial hemoglobin (%SpO2) and pulse rate of adult and pediatric patients. The systems are indicated for use by trained personnel in clinical and nonclinical settings, including Emergency Medical Service (EMS), hospitals, medical facilities, mobile environments, and home healthcare environments.
#### Indications for Use
The Nonin Medical CO-Pilot™ Model H500 Multi-Sensing Oximetry System is intended for noninyasive measuring of functional oxygen saturation of arterial hemoglobin (%SpO2), pulse rate, carboxyhemodlobin saturation (%COHb), methemodlobin saturation (%MetHb), and cerebral or somatic hemoglobin oxygen saturation (%rSO2). This device is not meant for sole use in clinical decision making; it must be used in conjunction with additional methods of assessing clinical signs and symptoms.
- For %SpO2 and pulse rate, the H500 System is intended for spot-checking and/or measuring . during clinician assessment of adult, pediatric, infant, and neonate patients who are well or poorly perfused, during both motion and non-motion conditions in professional healthcare facilities, mobile, and EMS settings.
- . For %rSO2, the H500 System is intended for spot-checking and/or measuring dinician assessment of adult, pediatric, infant, and neonate patients in professional healthcare facilities, mobile, and EMS settings.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image contains the logo for Nonin. The logo consists of two overlapping orange squares on the left, followed by the word "NONIN" in blue. The letters are bold and sans-serif.
- For %COHb and %MetHb, the H500 System is intended for spot-checking, multiple spot-checks to observe change, and/or measuring during clinician assessment of adult and pediatric patients in professional healthcare facilities, mobile, and EMS settings.
### Device Description
The Nonin Medical CO-Pilot™ Model H500 Multi-Sensing Oximetry System is a small handheld wireless device intended to measure functional oxygen saturation of arterial hemoglobin (%SpO2), pulse rate, carboxyhemoglobin saturation (%COHb), methemoglobin saturation (%MetHb), and cerebral or somatic hemoglobin oxygen saturation (%rSO2) of adult, pediatric and neonate patients. It is intended for professional use only, in healthcare facilities, mobile and EMS environments. The system is not provided sterile and is not a reprocessed single-use device. The H500 System consists of three components which are the display, the signal processor and associated sensors. It is intended to be used with specific parts, accessories and compatible sensors which are outlined in Table 1 below.
| Parts and Accessories | Compatible Sensors | | |
|--------------------------------------------------------------------|--------------------|---------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Sensor Model | 510(k) | Description |
| • H500-PS – Power Supply,<br>25W and cable | 8100A Series | K160865 | Reusable, Finger Clip Pulse Oximeter Sensor |
| • D-HH – H500 System<br>Display | | | 8100AA: Measures SpO2 and pulse rate of adult<br>and pediatric patients (> 30 kg / 66 lb) who are<br>well or poorly perfused, during both motion and<br>non-motion conditions. |
| • SP-BLE – Signal<br>Processor for use with the<br>H500 System | | | 8100AP: Measures SpO2 and pulse rate of<br>pediatric patients (8 – 60 kg / 18 – 132 lb) who<br>are well or poorly perfused, during both motion<br>and non-motion conditions |
| • X-100SL – Signal<br>Processor Sensor Lock, 2-<br>pack | | | |
| • H500CC – Carrying case | | | |
| • INT-100 Intermediate<br>Cable for use with 8204CA<br>only. | 8100S Series | K132402 | Reusable, Soft Pulse Oximeter Sensor |
| • INT-200 Intermediate<br>Cable for use with SpO2-<br>only sensors | | | 8100SL: Measures SpO2 and pulse rate of adult<br>and pediatric patients who are well or poorly<br>perfused, during both motion and non-motion<br>conditions, with digit height (thickness) of 12.5 –<br>25.5 mm (0.5 – 1.0 in.). |
| | | | 8100SM: Measures SpO2 and pulse rate of<br>adult and pediatric patients who are well or<br>poorly perfused, during both motion and non-<br>motion conditions, with digit height (thickness) of<br>10 – 19 mm (0.4 – 0.75 in.). |
| | | | 8100SS: Measures SpO2 and pulse rate of adult<br>and pediatric patients who are well or poorly<br>perfused, during both motion and non-motion<br>conditions, with digit height (thickness) of 7.5 –<br>12.5 mm (0.3 – 0.5 in.). |
| | 6100C Series | K160865 | Single-Patient Use, Disposable Pulse Oximeter<br>Sensors |
| | | | 6100CA: Measures SpO2 and pulse rate of adult<br>patients (> 30 kg / 66 lb) who are well or poorly<br>perfused, during both motion and non-motion<br>conditions. |
| | | | 6100CP: Measures SpO2 and pulse rate of<br>pediatric patients (> 10 kg / 22 lb) who are well |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains the logo for Nonin. The logo consists of two overlapping orange squares on the left side of the image. To the right of the squares is the company name "NONIN" in blue, with each letter being capitalized.
| | | or poorly perfused, during both motion and non-motion conditions. |
|-----------|---------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | 6100CI: Measures SpO2 and pulse rate of infant patients (> 2 kg / 4 lb) who are well or poorly perfused, during both motion and non-motion conditions. |
| | | 6100CN: Measures SpO2 and pulse rate of neonate patients (< 2 kg / 4 lb) and adult patients (> 30 kg / 66 lb) who are well or poorly perfused, during both motion and non-motion conditions. |
| 8100Q2 | K160865 | Reusable Ear Clip Pulse Oximeter Sensor |
| | | Measures SpO2 and pulse rate of adult and pediatric patients (>40 kg / 88 lb) who are well or poorly perfused, during non-motion conditions. |
| 8004CA | K132402 | Single-Patient Use, Non-Sterile, Disposable Regional Oximetry Sensor |
| | | Measures rSO2 of adult and pediatric patients weighing ≥ 88 pounds (40 kilograms). |
| 8004CB | K132402 | Single-Patient Use, Non-Sterile, Disposable Regional Oximetry Sensor |
| | | Measures rSO2 of neonate, infant, and pediatric patients weighing ≤ 88 pounds (40 kilograms). |
| 8004CB-NA | K132402 | Non-Adhesive, Single-Patient Use, Non-Sterile, Disposable Regional Oximetry Sensor |
| | | Measures rSO2 of neonate, infant, and pediatric patients weighing ≤ 88 pounds (40 kilograms). |
| 8204CA | K151305 | Single-Patient Use, Non-Sterile, Disposable Regional Oximetry Sensor |
| | | Measures rSO2 of adult and pediatric patients weighing ≥ 88 pounds (40 kilograms). |
| 8330AA | K172625 | Multi-Sensing Reusable Finger Clip Sensor |
| | | Measures SpO2, COHb, MetHb and pulse rate of adult and pediatric patients (> 66 lbs/30 kg). |
Table 1: H500 Parts, Accessories and Compatible Sensors
The H500 system is compatible and intended to be used only with Nonin-branded oximeter sensors. Such sensors are manufactured to meet the accuracy specifications for Nonin oximeters and using other manufacturers' sensors can result in improper oximeter performance. Sensor application sites are as specified in the respective instructions for use.
The main function of the H500 System software is to acquire signal data for oximetry measurements and translate that data into pulse rate, %SpO2, %COHb, %MetHb, and/or %rSO2 data for the display device. The H500 system utilizes software separated into three distinct modules: display software, signal processor software and Bluetooth Low-Energy (BLE) radio software modules control and monitor interdependent system operations, generate errors, monitor and report on patient measurements, and provide the operator with a user interface. The software communicates oximetry data using the standard wireless BLE technology and determines when to turn the device on and off based on the user input to the display device. The H500 is not an alarm system and is not intended to be used for monitoring with patient alarms. Required software documentation deliverables and applicable software
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Nonin. The logo consists of two overlapping orange squares on the left, followed by the word "NONIN" in blue. The letters are in a sans-serif font and are slightly rounded.
testing have been completed in accordance with FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.
### System Performance Testing
The H500 Multi-Sensing Oximetry System and its associated sensors are supported by safety, electromagnetic compatibility, device performance, and clinical testing to ensure appropriate functionality and to demonstrate substantial equivalence to the predicate devices were tested with the H500 system. A summary of such testing is as follows:
### Wireless Coexistence Testing
Wireless Coexistence testing has been conducted following the ANSI C63.27:2017 procedure for cochannel and adjacent channel interference including overstress per Annex F.6. The Tier 1 (most rigorous) band-specific guidance per Annex A, Sub-clause A.2.2 for Bluetooth Low Energy was also followed.
Functional and Safety Testing - The results of the testing demonstrate equivalency with the predicate devices and compliance to recognized standards. Table 2 summarizes test results for the proposed devices, which met the relevant requirements of the applicable recognized standards.
| Test | Reference | Result |
|-------------------------------------------|--------------------------------------------------------------------------------------------------------------|--------|
| Electrical Safety | IEC 60601-1 | Pass |
| Temperature and Humidity | IEC 60601-1<br>EN 1789 | Pass |
| Atmospheric Pressure (Altitude) | IEC 60601-1 | Pass |
| Electromagnetic Immunity and<br>Emissions | IEC 60601-1-2 | Pass |
| Performance | ISO 80601-2-61<br>IEC 60601-1<br>IEC 60601-1-6<br>IEC 60601-1-12<br>IEC 62304<br>ANSI/AAMI EC13<br>ISO 14155 | Pass |
| Ingress Protection | ISO 80601-2-61 | Pass |
| Diaphoretic related ingress | Internal performance<br>characterization | Pass |
| Mechanical Durability | IEC 60601-1<br>ISO 80601-2-61<br>ISTA 2A<br>ASTM D-4169 | Pass |
| Biocompatibility | ISO 10993-1 | Pass |
Table 2 - Summary of Functional and Safety Testing
# Clinical Accuracy Testing
COHb
COHb accuracy testing was conducted at an independent research laboratory on healthy, male and female, non-smoking, light to dark-skinned subjects that were 18 years of age and older. The measured carboxyhemoglobin value (%COHb) of the sensors was compared to simultaneous arterial blood samples as assessed by CO-oximetry. The accuracy of the sensors in comparison to the COoximeter samples measured over the COHb range of 0-15% with 95 - 100% SaO2. Accuracy data was calculated using the root-mean-squared (Arms value) for all subjects, per ISO 80601-2-61. Medical Electrical Equipment-Particular requirements for basic safety and essential performance of pulse oximeter equipment.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image contains the logo for NONIN. The logo consists of two overlapping orange squares on the left side of the image. To the right of the squares is the company name, NONIN, in blue font.
#### MetHb
MetHb accuracy testing was conducted at an independent research laboratory on healthy, male and female, non-smoking, light to dark-skinned subjects that were 18 years of age and older. The measured methemoqlobin value (%MetHb) of the sensors was compared to simultaneous arterial blood samples as assessed by CO-oximetry. The accuracy of the sensors in comparison to the COoximeter samples measured over the MetHb range of 0 – 15% with 95 – 100% SaO2. Accuracy data was calculated using the root-mean-squared (Arms value) for all subjects, per ISO 80601-2-61. Medical Electrical Equipments for basic safety and essential performance of pulse oximeter equipment.
# SpO2
During no-motion conditions at an independent research laboratory, SpO2 accuracy testing was conducted during induced hypoxia studies on healthy, male and female, non-smoking, light- to darkskinned subjects that were 18 years of age and older. The measured arterial hemoglobin saturation value (SpO2) of the sensors was compared to arterial hemoglobin oxygen (SaO2) value, determined from blood samples with a laboratory CO-oximeter. The accuracy of the sensors in comparison to the CO-oximeter samples measured over the SpO2 range of 70 – 100%. Accuracy data was calculated using the rootmean-squared (Arms value) for all subjects, per ISO 80601-2-61, Medical Electrical Equipment-Particular requirements for basic safety and essential performance of pulse oximeter equipment.
#### Sp02 in Presence of COHb and MetHb
During no-motion conditions at an independent research laboratory, SpO2 accuracy testing in the presence of COHb and MetHb was conducted during induced hypoxia studies on healthy, male and female, non-smoking, light- to dark-skinned subjects that were 18 years of age and older. The measured arterial hemoglobin saturation value (SpO2) of the sensors was compared to arterial hemoglobin oxygen (SaO2) value, determined from blood samples with a laboratory CO-oximeter. The accuracy of the sensors in comparison to the CO-oximeter samples measured over the SpO2 range of 80 – 100%, range 0 – 15% COHb and SpO2 range of 80 – 100%, range 0 – 15% MetHb. Accuracy data was calculated using the root-mean-squared (Arms value) for all subjects, per ISO 80601-2-61, Medical Electrical Equipment-Particular requirements for basic safety and essential performance of pulse oximeter equipment.
#### rSO2
rSO2 accuracy testing using 8004CA/8204CA sensors was conducted during induced hypoxia studies on healthy, non-smoking, light- to dark-skinned subjects that were 18 years of age and older. The measured regional hemoglobin saturation value (rSO2) of the sensors was compared to arterial/venous hemoglobin oxygen (SavO2) value, determined from venous and arterial blood samples. The model used for blood in the brain was 70% venous and 30% arterial, which is applicable under normocapnic conditions. The venous blood was drawn from the right jugular bulb. The accuracy of the sensors in comparison to the blood gas analyzer samples measured over the rSO2 range of 45 - 100%. Accuracy data was calculated using the root-mean-squared (Arms value) for all subjects, per ISO 80601-2-61. Medical Electrical Equipment—Particular requirements for basic safety and essential performance of pulse oximeter equipment.
rSO2 accuracy testing using 8004CB/8004CB-NA sensors was conducted in cardiac catheterization laboratories on sick, male and female, pediatric patients ranging in age from 4 days to 10 years with lightto dark-skin. The measured regional hemoglobin saturation value (rSO2) of the sensors is compared to arterial/venous hemoglobin oxygen (SavO2) value, determined from venous and arterial blood samples. The model used for blood in the brain was 70% venous and 30% arterial. The venous blood was drawn from the right jugular bulb. The accuracy of the sensors in comparison to the blood gas analyzer samples measured over the rSO2 range of 45 – 95%. Accuracy data was calculated using the root-mean-squared (Arms value) for all subjects, per ISO 80601-2-61, Medical Electrical Equipment—Particular requirements for basic safety and essential performance of pulse oximeter equipment.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for NONIN. The logo consists of two overlapping orange squares that are rotated 45 degrees. To the right of the squares is the word "NONIN" in blue, with each letter being capitalized.
# Testing Conclusion
The subiect H500 Multi-Sensing Oximetry System meets all testing acceptance criteria. Based on test results, and comparison to the legally marketed predicates, the H500 Multi-Sensing Oximetry System performance is substantially equivalent to the predicate devices for its intended use.
### Summary of Substantial Equivalence
The subject and predicate devices are all designed, developed and manufactured by Nonin Medical, Inc. All share the same intended use in that thev are intended to be used to noninvasively measure physiological parameters including %SpO2, pulse rate, %COHb, %MetHb, and %rSO2 and similar indications for use including use environment and intended patient populations. The systems all share the same measurement technology, principles of operation, safety specifications, sensor application sites, critical sensor optics technology and level of ingress protection. There are no changes in the optics performance specifications, optics wavelengths, and signal processing algorithms between the subject and predicates. The systems all use the same compatible sensor accessories and share the same specified environmental use conditions. The accuracy specifications are the same among the subject and predicates for their respective parameters. The subject H500 system uses the same SenSmart oximetry signal processing as the X-100C systems. As such, there is no accuracy difference between the predicate devices and the CO-Pilot H500 systems. The subject and predicate devices primarily differ in that the H500 System is small, handheld and wireless while the predicate systems are larger, wired and tabletop designs. Please refer to Table 3 for predicate device comparison table.
| Table 3: Predicate Device Comparison Table | | | |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Important<br>Parameters | Subject Device<br>CO-Pilot TM Model H500 Multi-<br>Sensing Oximetry System | Primary Predicate Device<br>X-100 System (K132402) | Reference Predicate Devices<br>X-100C/Multi-Sensing<br>Oximetry System (K160231<br>and K172625) |
| Intended Use | Same<br>All systems are intended to be used to noninvasively measure physiological parameters including<br>functional oxygen saturation of arterial hemoglobin (%SpO2), pulse rate, carboxyhemoglobin<br>saturation (%COHb), methemoglobin saturation (%MetHb), and cerebral or somatic hemoglobin<br>oxygen saturation (%rSO2) | | |
| Product Code | DQA/MUD | DQA/MUD | DQA |
| FDA Regulation | | Same<br>870.2700 | |
| Indications for<br>Use | The Nonin Medical CO-Pilot™ Model<br>H500 Multi-Sensing Oximetry System is<br>intended for noninvasive measuring of<br>functional oxygen saturation of arterial<br>hemoglobin (%SpO2), pulse rate,<br>carboxyhemoglobin saturation (%COHb)<br>methemoglobin saturation (%MetHb),<br>and cerebral or somatic hemoglobin<br>oxygen saturation (%rSO2). This device<br>is not meant for sole use in clinical<br>decision making; it must be used in<br>conjunction with additional methods of<br>assessing clinical signs and symptoms.<br>• For %SpO2 and pulse rate, the H500<br>System is intended for spot-checking<br>and/or measuring during clinician<br>assessment of adult, pediatric, infant,<br>and neonate patients who are well or<br>poorly perfused, during both motion and<br>non-motion conditions in professional<br>healthcare facilities, mobile, and EMS<br>settings.<br>• For %rSO2, the H500 System is<br>intended for spot-checking and/or<br>measuring during clinician assessment of<br>adult, pediatric, infant, and neonate<br>patients in professional healthcare<br>facilities, mobile, and EMS settings.<br>• For %COHb and %MetHb, the H500<br>System is intended for spot-checking,<br>multiple spot-checks to observe change,<br>and/or measuring during clinician<br>assessment of adult and pediatric<br>patients in professional healthcare<br>facilities, mobile, and EMS settings. | The X-100 system is a modular system<br>and is indicated for use in<br>simultaneously measuring, displaying,<br>monitoring, and recording up to six (6)<br>channels of functional oxygen saturation<br>of arterial hemoglobin (%SpO2) and<br>pulse rate or cerebral or somatic<br>hemoglobin oxygen saturation (%rSO2)<br>of blood underneath the sensor. Patient<br>populations include adult, pediatric,<br>infant, and neonate through the use of<br>SenSmart compatible sensors.<br>The SenSmart system is intended for<br>use in hospitals, long-term care, medical<br>facilities, sleep laboratories, sub-acute<br>environments, and Emergency Medical<br>Services (EMS), including patient<br>transport. The X-100 SenSmart system<br>may be used for spot-checking and<br>continuous monitoring with patient<br>alarms. The SenSmart pulse oxlmetry<br>(SpO2) functionality Is suitable for use<br>In both motion and non-motion<br>conditions, Including patients who are<br>well or poorly perfused. | The X-100C/Multi-Sensing Oximetry<br>system is indicated for noninvasive<br>measuring of functional oxygen<br>saturation of arterial hemoglobin<br>(%SpO2), carboxyhemoglobin<br>saturation (%COHb), methemoglobin<br>saturation (%MetHb), and pulse rate of<br>adult and pediatric patients (>66 lbs / 30<br>kg) using the Model 8300AA/8330AA<br>sensor. The measurements may be<br>multiple spot-checks to observe change<br>and/or continuous monitoring. The<br>systems are indicated for use by trained<br>personnel in clinical and non-clinical<br>settings, including Emergency Medical<br>Service (EMS), hospitals, medical<br>facilities, and mobile environments. This<br>device is not meant for sole use in<br>clinical decision making; it must be used<br>in conjunction with additional methods<br>of assessing clinical signs and<br>symptoms. |
| Use<br>Environment | Professional healthcare facilities,<br>mobile, and EMS settings | Hospitals, long-term care,<br>medical facilities, sleep<br>laboratories, sub-acute<br>environments, and EMS<br>settings | Hospitals, medical facilities,<br>mobile environments and EMS<br>settings |
| Compatible<br>Sensor<br>Accessories | Same<br>Compatible and intended to be used only with Nonin-branded oximeter sensors. Such sensors are<br>manufactured to meet the accuracy specifications for Nonin oximeters and using other<br>manufacturers' sensors can result in improper oximeter performance. | | |
| Compatible<br>Sensor<br>Application<br>Sites | Same<br>As specified in the respective sensor instructions for use | | |
| Handheld or<br>Tabletop? | Handheld | Tabletop | Tabletop |
| Wireless? | Yes | No | No |
| Includes Patient<br>Alarms? | No | Yes | Yes |
| Principles of<br>Operation | Same<br>Non-invasive oximetry using light-based measurement sharing the same measurement technology<br>including optics specifications and signal processing algorithms. | | |
| Range of SpO2 | Same<br>0-100%…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.