The SentEP system is intended for use as a medical imaging system that allows the review, analysis, communication, and media interchange of multi-dimensional digital images. It is also intended for intraprocedural use. The SentEP system is designed as an additional visualization modality to assist the clinician. The SentEP system is indicated for use in electrophysiology (EP) procedures to assist the clinician in visualization of the heart electroanatomic data.
Device Story
SentEP system is a mixed-reality medical imaging system for electrophysiology (EP) procedures; imports multi-dimensional digital images from an Electroanatomic Mapping System (EAMS). System components include a Head Mounted Display (HMD) (Microsoft HoloLens with proprietary software), a Data Manager PC/Spectator Display, and network cabling. The EP wears the HMD to view a stereoscopic 3D holographic anatomical model in real-time; supporting staff view the same model on a 2D spectator display. The device provides the EP with direct visualization of EAMS data without controlling the EAMS directly. Used in sterile EP labs; non-patient contacting; non-sterile. Benefits include improved visualization of heart anatomy during mapping and ablation, potentially enhancing navigation accuracy and clinician workflow efficiency.
Clinical Evidence
Prospective, acute, single-center study of 16 pediatric patients (ages 7.3-19.4). Evaluated navigational capability and usability compared to standard EAMS. Endpoints included navigation to 5 locations within 60s, point location accuracy, and physician exit surveys. No adverse events reported. Results demonstrated safety and effectiveness profile similar to EAMS.
Technological Characteristics
Non-patient contacting, non-sterile. Components: Microsoft HoloLens HMD, Data Manager PC, spectator display. Connectivity: Networked to EAMS. Software: Proprietary rendering and real-time data integration. Standards: IEC 62304 (software lifecycle), IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC).
Indications for Use
Indicated for use in electrophysiology (EP) procedures to assist clinicians in visualizing heart electroanatomic data. Intended for use in pediatric and adult patient populations undergoing cardiac catheter mapping and ablation procedures.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the full name of the agency on the right. The symbol on the left is a stylized representation of a human figure, while the FDA acronym is in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in a sans-serif font.
September 18, 2020
SentiAR, Inc Berk Tas Ceo and President 212 Kingshighway Blvd., Suite 115 Mailbox 28 St. Louis, Missouri 63108
Re: K192890
Trade/Device Name: SentEP system Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving And Communications System Regulatory Class: Class II Product Code: LLZ Dated: August 17, 2020 Received: August 18, 2020
Dear Berk Tas:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mark Fellman Assistant Director DHT2A: Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K192890
Device Name SentEP system
#### Indications for Use (Describe)
The SentEP system is intended for use as a medical imaging system that allows the review, and media interchange of multi-dimensional digital images. It is also intended for intraprocedural use. The SentEP system is designed as an additional visualization modality to assist the clinician. The SentEP system is indicated for use in electrophysiology (EP) procedures to assist the clinician in visualization of the heart electroanatomic data.
Type of Use (*Select one or both, as applicable*)
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# SENTIAF
## 510(K) SUMMARY
#### 1 SUBMITTER INFORMATION
| Date Prepared: | August 7, 2020 |
|---------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| Submitter's Name<br>and Address | SentiAR, Inc.<br>212 N. Kingshighway Blvd., Suite 115<br>Mailbox 28<br>St. Louis, MO 63108<br>Telephone: (314) 499-0552 |
| Contact Person: | Berk Tas, Chief Executive Officer |
#### 2 DEVICE INFORMATION
| Name of Medical<br>Device | Proprietary Name: SentEP system<br>Common Name: Imaging software system |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>Classification | Classification Name: System, Image Processing Radiological<br>Regulation Name: Picture archiving and communication system<br>Regulation Number: 21 CFR 892.2050<br>Regulatory Class: Class II<br>Product Code: LLZ<br>Classification Panel: Radiology |
| Device Description | The SentEP system is an imaging system intended to be used as an<br>adjunct product to assist the clinician in visualization of the heart<br>anatomy during cardiac catheter mapping and ablation procedures.<br>Images are imported from an EAMS (electroanatomic mapping<br>system) and displayed through the SentEP system. As a result, users<br>can view both the 2D EAMS display monitor and the 3D SentEP<br>system display at the same time. |
#### 3 PREDICATE/REFERENCE DEVICE INFORMATION
| Predicate<br>Device | Ceevra Reveal 2.0 cleared via K173274 on July 10, 2018 |
|---------------------|------------------------------------------------------------------|
| Reference<br>Device | Abbott EnSite Velocity System cleared via K130594 on May 3, 2013 |
#### 4 DEVICE DESCRIPTION
The SentEP system is a medical imaging system which allows for the review, analysis, communication, and media interchange of multi-dimensional digital images. Multi-dimensional images are imported from an Electroanatomic Mapping System (EAMS) which are then displayed by the SentEP system software to assist physicians in the visualization of heart electroanatomic data. The SentEP system Head Mounted Display (HMD) provides a view through, stereoscopic 3D view of the data, allowing the user to see the environment as well as the specific holographic image. The SentEP system software additionally provides a 2D view of the
{5}------------------------------------------------
data on the Data Manager PC / Spectator Display for others not using the HMD. Components of the SentEP system are detailed in the table below.
| Component | Function |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| HMD | Used to display the anatomical hologram with real-time data. Used to provide the Operator with the ability to interact with data and images. The system is compatible with up to 5 HMDs in a single shared session |
| Data Manager PC/<br>Spectator Display | Used to coordinate transmission of anatomical and real-time data between EAMS and HMDs and to provide a spectator display. |
| Enclosure | Used to provide network connectivity to EAMS, Data Manager PC/Spectator Display, and HMD |
| Cables | Used to provide power and network connectivity to components |
| Proprietary Software | Used to render the model and provide real-time data within the user interface |
An Electrophysiologist (EP) is the intended user of the SentEP system. The EP performs procedures in a cardiac electrophysiology laboratory, which is a sterile operating room located in a hospital or clinic. The SentEP system provides the EP with a mixed reality user interface design displayed through the HMD (a commercially available Microsoft HoloLens installed with proprietary software and provided as a component of the SentEP system). Supporting staff members can view the anatomical model on an additional CC display from the same perspective as the controlling EP, while the controlling EP manipulates the display on the HMD. The SentEP system is a non-patient contacting device and is provided non-sterile.
Manipulation of the current EAMS display requires relaying commands from the EP to a nurse or technician within the EP lab or between the EP lab and control room outside of the EP lab. The SentEP system provides the EP direct visualization and control of the EAMS output. The SentEP system does not provide control of the EAMS directly.
In summary, the SentEP system does not deliver therapy, nor does it intervene with therapy. It is intended to provide users with information to assist the clinician in visualization of the heart electroanatomic data.
### 5 INDICATIONS FOR USE
The SentEP system is intended for use as a medical imaging system that allows the review, analysis, communication, and media interchange of multi-dimensional digital images. It is also intended for intraprocedural use. The SentEP system is designed as an additional visualization modality to assist the clinician. The SentEP system is indicated for use in electrophysiology (EP) procedures to assist the clinician in visualization of the heart electroanatomic data.
#### 6 COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The SentEP system has similar intended use and indications for use, principles of operation, and technological characteristics as the legally marketed predicate device.
- . Both devices are intended as medical imaging systems that allow the review, analysis, communication, and media interchange of multi-dimensional digital images.
{6}------------------------------------------------
- . Both devices support viewing 3D digital images on a head mounted display.
- Both devices are intended for intraprocedural use by healthcare professionals.
- Both devices are provided non-sterile and are not intended to be sterilized. ●
- Both devices are biocompatible for their intended use. .
Differences between the SentEP system and the predicate device have been evaluated for impact to safety and performance. Primary differences between the SentEP system and the predicate device are identified below.
- . The SentEP system is intended only for intraprocedural use. The predicate device is intended for both preprocedural and intraprocedural use. The SentEP system intended use is a subset of the predicate device intended use, which does not raise any different questions of safety and effectiveness.
- The source data for the SentEP system is derived from the reference device. The . predicate device uses a DICOM protocol to generate the 3D dimensional images. While the proposed and predicate devices use different communication protocols, both the SentEP system and the predicate device offer proprietary software for the same purpose of displaying anatomical image data in 3D format, and therefore do not raise any different questions of safety and effectiveness.
- The SentEP system processes streaming live data (dynamic) collected in real-time. The . predicate device can process either historical static images or historical dynamic recordings. Both devices process dynamic images that provide the ability to view the data for the intended purpose of each of the devices, and therefore the difference does not raise any different questions of safety and effectiveness.
Although the SentEP system and the predicate device technological characteristics differ in the preprocedural use, communication source, and image data format, performance testing has demonstrated that the SentEP system is as safe and effective as the predicate and that the differences do not raise any different questions concerning safety and effectiveness. The implemented design controls, risk management activities, labeling, and performed verification and validation tests demonstrate the safety and performance of the SentEP system. Based on the comparison information provided above. it is concluded that the SentEP system is substantially equivalent to the predicate device.
#### 7 PERFORMANCE DATA
Verification and validation testing data summarized below were provided in support of the substantial equivalence determination.
#### Non-Clinical Performance Testing:
Non-clinical performance testing included an evaluation of biocompatibility, mechanical testing, electromagnetic compatibility and electrical safety, wireless capability, and human factors testing.
The safety and performance of the SentEP system have been evaluated and verified according to the specified standards. Verification and validation testing have demonstrated compliance to the software specifications and performance standards. The SentEP system demonstrates conformance specifically to:
- IEC 62304:2006/AC: 2008- Medical device software Software life cycle processes ●
- IEC 60601-1:2005 Ed.3+C1; C2 Medical Electrical Equipment Part 1: General . requirements for basic safety and essential performance
{7}------------------------------------------------
# SENTIAR
- . IEC 60601-1-2 ed 4.0 (2014-02) Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
#### Clinical Study:
A clinical study was completed to assess SentEP system usability and performance during EP procedures. The clinical study was a prospective, acute, single center study of 16 pediatric EP patients representing the intended patient population (age 7.3-19.4, 6 males-10 females, 3 ethnicities represented). The study was performed in the United States.
The purpose of this study was to evaluate both the navigational capability of the SentEP system when compared to the EAMS system and the usability of the SentEP system by assessing the user comfort of the device. EP procedures were performed in accordance with normal standard of care for the condition being treated. For a portion of the procedure, the operator had the electroanatomic imaging information displayed on the SentEP system HMD in addition to the traditional EAMS system display while navigating to intended targets.
The equivalency between the SentEP system and the EAMS system was evaluated by comparing: 1) the ability to navigate to 5 locations within allocated time (60s per location), 2) point location accuracy and 3) Physician Exit Survey responses. No adverse events were reported during this study. The comparison of the performance data and safety data collected support the conclusion that the SentEP system was found to have a safety and effectiveness profile that is similar to the legally marketed EAMS system.
#### 8 SAFETY AND PERFORMANCE CONCLUSION
Non-clinical and clinical performance testing demonstrated that the SentEP system is as safe and effective as the legally marketed predicate and reference devices. Based on evidence submitted in this premarket notification, including similarities in the indications for use, technological characteristics, and performance between the SentEP system and the predicate and reference devices, the SentEP system raises no new questions of safety or effectiveness and is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.