Acclarix AX8 Diagnostic Ultrasound System / Acclarix LX9 Series Diagnostic Ultrasound System
K192879 · Edan Instruments, Inc. · IYN · May 21, 2020 · Radiology
Device Facts
Record ID
K192879
Device Name
Acclarix AX8 Diagnostic Ultrasound System / Acclarix LX9 Series Diagnostic Ultrasound System
Applicant
Edan Instruments, Inc.
Product Code
IYN · Radiology
Decision Date
May 21, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Acclarix LX9 Series/ Acclarix AX8 Diagnostic Ultrasound System is intended for use by a qualified physician or allied health professional for ultrasound evaluation in hospitals and clinics. Clinical applications include Abdominal, Gynecology, Obstetric, Cardiac, Small parts, Urology, Musculoskeletal, Peripheral vascular, Intra-operative, Pediatric, Neonatal and Adult Cephalic.
Device Story
Software-controlled diagnostic ultrasound system; consists of main unit and various transducers. Acquires and displays ultrasound data in multiple imaging modes (B, M, Doppler, Harmonic, Elastography, Contrast). Used in hospitals/clinics by physicians or allied health professionals. System processes acoustic signals to generate real-time images for clinical evaluation. Includes ECG module for waveform display; Auto OB and Auto NT measurement functions for automated quantification. Output displayed on system monitor; assists clinicians in diagnostic decision-making across various clinical applications. Benefits include non-invasive visualization of internal structures and physiological assessment.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by non-clinical bench testing, including electrical safety (ANSI AAMI ES60601-1), electromagnetic compatibility (IEC 60601-1-2), ultrasonic safety (IEC 60601-2-37, IEC 62359), and biocompatibility (ISO 10993 series).
Technological Characteristics
Software-controlled ultrasound system. Modes: B, M, Doppler, Harmonic, Elastography, Contrast. Transducers: C7-2XQ, E10-3BQ, E10-3HQ, C5-1Q, L10-4Q, L12-5Q, L17-7HQ, L17-7SQ, P5-1Q. Biocompatibility per ISO 10993. ECG module (external for AX8, built-in for LX9). Connectivity: ECG waveform display. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, IEC 62359.
Indications for Use
Indicated for ultrasound evaluation in hospitals and clinics by qualified physicians or allied health professionals. Clinical applications: Abdominal, Gynecology, Obstetric, Cardiac, Small parts, Urology, Musculoskeletal, Peripheral vascular, Intra-operative, Pediatric, Neonatal, and Adult Cephalic.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Edan Instruments, Acclarix AX8 Diagnostic Ultrasound System (K180862)
Reference Devices
Shenzhen Mindray, DC-80 Diagnostic Ultrasound System (K173471)
Submission Summary (Full Text)
{1}------------------------------------------------
May 21, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION".
Edan Instruments, Inc. % Melody Huang Regulatory Engineer #15 Jinhui Road, Jinsha Community Kengzi Sub-District, Pingshan District Shenzhen, Guangdong 518122 P.R. CHINA
Re: K192879
Trade/Device Name: Acclarix AX8 Diagnostic Ultrasound System / Acclarix LX9 Series Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: April 17, 2020 Received: April 20, 2020
Dear Melody Huang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192879
Device Name
Acclarix AX8 Diagnostic Ultrasound System / Acclarix LX9 Series Diagnostic Ultrasound System
#### Indications for Use (Describe)
The Acclarix LX9 Series/ Acclarix AX8 Diagnostic Ultrasound System is intended for use by a qualified physician or allied health professional for ultrasound evaluation in hospitals and clinics. Clinical applications include Abdominal, Gynecology, Obstetric, Cardiac, Small parts, Urology, Musculoskeletal, Peripheral vascular, Intra-operative, Pediatric, Neonatal and Adult Cephalic.
The Modes of Operation for Acclarix LX9 series include B mode, Doppler mode, Harmonic Imaging, Elastography Imaging, Contrast imaging and their combination modes.
The Modes of Operation for Acclarix AX8 include B mode, M mode, Harmonic Imaging, Elastography Imaging and their combination modes.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><span><span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</span></div> | <div style="display:flex; align-items:center;"><span><span style="font-size: 20px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</span></div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
| K192879 | Prepared in accordance with the requirements of 21 CFR Part 807.92 |
|-----------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <b>1. Submitter:</b> | Edan Instruments, Inc.<br>#15 Jinhui Road, Jinsha Community, Kengzi Sub-District, Pingshan District,<br>Shenzhen, 518122 P.R.China.<br>Tel: +86-755-2685 8736 Fax: +86-755-2689 8330 |
| Contact person: | Melody Huang |
| Preparing date: | Sep. 25, 2019 |
| <b>2. Device name and<br/>classification:</b> | Device Name: Acclarix AX8 Diagnostic Ultrasound System / Acclarix LX9 Series<br>Diagnostic Ultrasound System<br>Model: Acclarix AX8, Acclarix LX9, Acclarix LX9 Exp, Acclarix LX9 Super,<br>Acclarix LX85, Acclarix LX88<br>Classification Name/ Product code:<br>21 CFR 892.1550 System, Imaging, Pulsed Doppler, Ultrasonic / IYN<br>21 CFR 892.1560, Ultrasonic, Pulsed echo, Imaging / IYO<br>21 CFR 892.1570, Transducer, Ultrasonic, Diagnostic / ITX<br>Regulatory Class: Class II |
| <b>3.Premarket Notification<br/>Class III Certification<br/>and Summary</b> | Not applicable, the subject device is Class II. |
| <b>4. Predicate Device(s):</b> | 1) Edan Instruments, Acclarix AX8 Diagnostic Ultrasound System, cleared under<br>K180862 (Primary)<br>2) Shenzhen Mindray, DC-80 Diagnostic Ultrasound System, cleared under K173471<br>(Reference) |
| <b>5. Reason for<br/>Submission</b> | By submission of the Traditional 510(k), Edan Instruments is requesting<br>clearance for an updated version of the Acclarix AX8 Diagnostic Ultrasound Systems<br>and new models of the Acclarix LX9 series Diagnostic Ultrasound Systems. |
| <b>6.Pre-Submission,<br/>IDE</b> | Not applicable, there is no pre-submission. |
| <b>7. Device Description:</b> | Acclarix AX8/ Acclarix LX9 Series is a software controlled Diagnostic Ultrasound<br>System, which consists of a main unit along with associated transducers. It is<br>intended for use by a qualified physician or allied health professional for ultrasound<br>evaluations. This system is a Track 3 device to acquire and display ultrasound data in |
{5}------------------------------------------------
various imaging modes.
8. Indication for Use The Acclarix LX9 Series/ Acclarix AX8 Diagnostic Ultrasound System is intended for use by a qualified physician or allied health professional for ultrasound evaluation in hospitals and clinics. Clinical applications include Abdominal, Gynecology, Obstetric, Cardiac, Small parts, Urology, Musculoskeletal. Peripheral vascular, Intra-operative, Pediatric, Neonatal and Adult Cephalic.
> The Modes of Operation for Acclarix LX9 series include B mode, M mode, Doppler mode, Harmonic Imaging, Elastography Imaging, Contrast imaging and their combination modes. The Modes of Operation for Acclarix AX8 include B mode, M mode, Doppler mode, Harmonic Imaging, Elastography Imaging and their combination modes.
## 9. Predicate Device Comparison
The subject devices Acclarix AX8 and Acclarix LX9 series have technological characteristics, is comparable in key safety and effectiveness features, and has the same intended uses and basic operating modes as the primary predicated device Acclarix AX8 (K180862), except the following main differences.
The main changes of the subject devices Acclarix AX8 and Acclarix LX9 series compared with the primary predicated device Acclarix AX8 (K180862) are as the followings:
- 1) Addition of new transducers: C7-2XQ, E10-3BQ, E10-3HQ, C5-1Q, which have similar specifications, same intended use and same modes of operation as the transducers of Acclarix AX8 (K180602).
- 2) Addition of Color M mode, which is substantially equivalent to that of the reference predicate device DC-80 (K173471).
- 3) Addition of Strain Elastography mode on Linear transducers, including L10-4Q, L12-5Q, L17-7HQ and L17-7SQ, which is substantially equivalent to that of the reference predicate device DC-80 (K173471).
- 4) Addition of ECG module to support ECG waveform display on the ultrasound image, which is substantially equivalent to that of the reference predicate device DC-80 (K173471). The ECG function is supported on P5-1Q transducer. The ECG function on Acclarix AX8 is provided by an external ECG module, while the ECC function on Acclarix LX9 series is provided by a built-in ECG module.
- 5) Addition of Auto OB and Auto NT measurement function, which is substantially equivalent to the reference predicate device DC-80 (K173471).
{6}------------------------------------------------
The additional changes of the subject device Acclarix LX9 series compared with primary predicated device Acclarix AX8 (K180862) are as the followings:
- 1) Acclarix LX9 series additionally supports a transducer C6-2MQ for 3D/4D imaging is not a new feature for EDAN device which was cleared in transducer C5-2MQ of Acclarix AX8 (K180602).
- 2) Acclarix LX9 series does not support C5-2XQ, L10-40, P5-1XO and C5-2MQ transducers, which were cleared in Acclarix AX8 (K180862).
- 3) Acclarix LX9 series additionally supports Contrast Imaging mode for Liver on C5-1Q transducers, and the Contrast Imaging mode is substantially equivalent to the reference predicate device DC-80 (K173471).
The subject and predicate devices have similar design features and performance specifications. The technological differences between the subject and predicate devices do not raise different questions of safety or effectiveness.
## 10. Performance Data:
## Clinical test:
Clinical testing is not required.
#### Non-clinical test:
The Acclarix AX8/ Acclarix LX9 series Diagnostic Ultrasound Systems comply with:
(1) ANSI AAMI ES60601-1 Electrical Safety
(2) IEC 60601-1-2 Electromagnetic Compatibility
(3) IEC 60601-2-37 Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
(4) Guidance for Industry and FDA Staff—Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers issued on June 27, 2019
(5) IEC 62359 Ultrasonics - Field characterization - Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields
The following biocompatibility standards are conducted on the subject device:
(1) ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 10993-12
The tests were selected to show substantial equivalence between the subject devices and the predicate devices.
The non-clinical performance testing showed that the subject devices are as safe and as effective as the predicate devices.
### 11. Conclusion
Verification and validation testing has been conducted on the Acclarix LX9 Series Diagnostic Ultrasound Systems. This premarket notification submission demonstrates that Acclarix LX9 Series
{7}------------------------------------------------
Diagnostic Ultrasound Systems are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.