Retrospective multi reader multi case study: 240 mammographic screening images
14 (radiologists)
Breast cancer detection
Deep learning modules
—
AUC = 0.883 (mammogram level, combined)
—
—
Standalone performance testing: mammograms of women acquired with Hologic and GE devices
—
Indications for Use
MammoScreen™ is intended for use as a concurrent reading aid for interpreting physicians, to help identify findings on screening FFDM acquired with compatible mammography systems and assess their level of suspicion. Output of the device includes marks placed on findings on the mammogram and level of suspicion scores. The findings could be soft tissue lesions or calcifications. The level of suspicion score is expressed at the finding level, for each breast and overall for the mammogram. Patient management decisions should not be made solely on the basis of analysis by MammoScreen™.
Device Story
Software-only device; processes screening FFDM images from compatible mammography systems. Uses deep learning algorithms trained on biopsy-proven databases to detect and characterize suspicious calcifications and soft tissue lesions. Operates as a DICOM Web-compliant node; receives images from a DICOM Web Server. Outputs include CAD marks on findings and a 1-10 suspicion score at finding, breast, and exam levels. Used by radiologists in clinical settings concurrently with mammogram interpretation. Radiologists review marks and scores to assist in identifying potential cancer; findings with scores >=5 are highlighted by default. Does not replace clinical judgment; patient management decisions require physician review. Benefits include improved diagnostic performance and cancer detection sensitivity.
Clinical Evidence
Retrospective multi-reader multi-case study (n=14 radiologists, 240 cases). Primary endpoint: AUC improvement. Results: Assisted AUC (0.798) significantly higher than unaided (0.769, p=0.035). Standalone performance (AUC 0.79) non-inferior to unaided radiologists. Reading time impact was negligible. Breast and lesion-level performance improvements were statistically significant.
Technological Characteristics
Software-only; DICOM Web-compliant. Deep learning-based image analysis for lesion detection. Connectivity: Networked via DICOM. Standards: IEC 62304 (software lifecycle), IEC 62366-1 (usability). Moderate level of concern.
Indications for Use
Indicated for women undergoing screening full-field digital mammography (FFDM). Used as a concurrent reading aid by physicians to identify and assess suspicion levels of soft tissue lesions or calcifications.
Regulatory Classification
Identification
A radiological computer-assisted detection and diagnostic software is an image processing device intended to aid in the detection, localization, and characterization of fracture, lesions, or other disease-specific findings on acquired medical images (e.g., radiography, magnetic resonance, computed tomography). The device detects, identifies, and characterizes findings based on features or information extracted from images, and provides information about the presence, location, and characteristics of the findings to the user. The analysis is intended to inform the primary diagnostic and patient management decisions that are made by the clinical user. The device is not intended as a replacement for a complete clinician's review or their clinical judgment that takes into account other relevant information from the image or patient history.
Special Controls
A radiological computer assisted detection and diagnosis software must comply with the following special controls: Design verification and validation must include: 1. i. A detailed description of the image analysis algorithm, including but not limited to a description of the algorithm inputs and outputs, each major component or block, how the algorithm and output affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide improved assisted-read detection and diagnostic performance as intended in the indicated user population(s), and to characterize the standalone device performance for labeling. Performance testing includes standalone test(s), side-by-side comparison(s), and/or a reader study, as applicable. iii. Results from standalone performance testing used to characterize the independent performance of the device separate from aided user performance. The performance assessment must be based on appropriate diagnostic accuracy measures (e.g., receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Devices with localization output must include localization accuracy testing as a component of standalone testing. The test dataset must be representative of the typical patient population with enrichment made only to ensure that the test dataset contain a sufficient number of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, concomitant disease, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment. iv. Results from performance testing that demonstrate that the device provides improved assisted-read detection and/or diagnostic performance as intended in the indicated user population(s) when used in accordance with the instructions for use. The reader population must be comprised of the intended user population in terms of but not limited to clinical training, certification, and years of experience. The performance assessment must be based on appropriate diagnostic accuracy measures (e.g., receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Test datasets must meet the requirements described in 1(iii) above. v. Appropriate software documentation, including device hazard analysis, software requirements specification document, software design specification document, traceability analysis, system level test protocol, pass/fail criteria, testing results, and cybersecurity measures. 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the device instructions for use, including the intended reading protocol and how the user should interpret the device output. iii. A detailed description of the intended user, and any user training materials as programs that addresses appropriate reading protocols for the device to ensure that the end user is fully aware of how to interpret and apply the device output. iv. A detailed description of the device inputs and outputs. v. A detailed description of compatible imaging hardware and imaging protocols. vi. Warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality or for certain subpopulations), as applicable. vii. A detailed summary of the performance testing, including: test methods, dataset characteristics, results, and a summary of sub-analyses on case distributions stratified by relevant confounders, such as anatomical characteristics, patient demographics and medical history, user experience, and imaging equipment.
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the image analysis algorithm, including a description of the algorithm inputs and outputs, each major component or block, how the algorithm and output affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide improved assisted-read detection and diagnostic performance as intended in the indicated user population(s), and to characterize the standalone device performance for labeling. Performance testing includes standalone test(s), side-by-side comparison(s), and/or a reader study, as applicable.
(iii) Results from standalone performance testing used to characterize the independent performance of the device separate from aided user performance. The performance assessment must be based on appropriate diagnostic accuracy measures (
*e.g.,* receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Devices with localization output must include localization accuracy testing as a component of standalone testing. The test dataset must be representative of the typical patient population with enrichment made only to ensure that the test dataset contains a sufficient number of cases from important cohorts (*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, concomitant disease, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment.(iv) Results from performance testing that demonstrate that the device provides improved assisted-read detection and/or diagnostic performance as intended in the indicated user population(s) when used in accordance with the instructions for use. The reader population must be comprised of the intended user population in terms of clinical training, certification, and years of experience. The performance assessment must be based on appropriate diagnostic accuracy measures (
*e.g.,* receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Test datasets must meet the requirements described in paragraph (b)(1)(iii) of this section.(v) Appropriate software documentation, including device hazard analysis, software requirements specification document, software design specification document, traceability analysis, system level test protocol, pass/fail criteria, testing results, and cybersecurity measures.
(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use.
(ii) A detailed description of the device instructions for use, including the intended reading protocol and how the user should interpret the device output.
(iii) A detailed description of the intended user, and any user training materials or programs that address appropriate reading protocols for the device, to ensure that the end user is fully aware of how to interpret and apply the device output.
(iv) A detailed description of the device inputs and outputs.
(v) A detailed description of compatible imaging hardware and imaging protocols.
(vi) Warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality or for certain subpopulations), as applicable.(vii) A detailed summary of the performance testing, including test methods, dataset characteristics, results, and a summary of sub-analyses on case distributions stratified by relevant confounders, such as anatomical characteristics, patient demographics and medical history, user experience, and imaging equipment.
{1}------------------------------------------------
March 25, 2020
Image /page/0/Picture/1 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Therapixel % Ms. Cindy Domecus Principal Domecus Consulting Services LLC 1171 Barroihet Drive HILLSBOROUGH CA 94010
Re: K192854
Trade/Device Name: MammoScreen Regulation Number: 21 CFR 892.2090 Regulation Name: Radiological computer assisted detection and diagnosis software Regulatory Class: Class II Product Code: QDQ Dated: February 10, 2020 Received: February 11, 2020
Dear Ms. Domecus:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR
{2}------------------------------------------------
803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192854
Device Name MammoScreen
#### Indications for Use (Describe)
MammoScreen™ is intended for use as a concurrent reading aid for interpreting physicians, to help identify findings on screening FFDM acquired with compatible mammography systems and assess their level of suspicion. Output of the device includes marks placed on findings on the mammogram and level of suspicion scores. The findings could be soft tissue lesions or calcifications. The level of suspicion score is expressed at the finding level, for each breast and overall for the mammogram. Patient management decisions should not be made solely on the basis of analysis by MammoScreen™.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary K192854
This 510(k) summary of safety and effectiveness information is prepared in accordance with the requirements of 21 CFR § 807.92.
## Applicant Information:
510(k) Owner: Therapixel Village By CA - Le Theseus, Rue Claude Daunesse, 06560 Valbonne France Phone: +33 9 72 55 20 39
#### Submission Correspondent:
Cindy Domecus, R.A.C. (US & EU) Regulatory Consultant to Therapixel Phone: 650.343.4813 Fax: 650.343.7822 Email: domecusconsulting@comcast.net
Additional Contact:
Quentin de Snoeck RA/QA/CA Manager at Therapixel Phone: +33 9 72 55 20 39 Email: qdesnoeck(@therapixel.com
Date Summary Prepared: March 18th, 2020
{5}------------------------------------------------
## Device Information:
| Trade Name: | MammoScreen |
|-----------------------------|--------------------------------------------------------------------------------------------------|
| Common Name: | Computer-Assisted Detection Device |
| Device Classification Name: | Radiological Computer Assisted Detection/Diagnosis Software For<br>Lesions Suspicious For Cancer |
| Regulation Number: | 892.2090 |
| Regulation Class: | Class II |
| Product Code: | QDQ |
| Submission type | Traditional 510(k) |
## Predicate Device:
The predicate device is Transpara™, cleared under K181704.
#### Device Description:
MammoScreen is a software-only device for aiding interpreting physicians in identifying focal findings suspicious for breast cancer in screening FFDM (full-field digital mammography) acquired with compatible mammography systems. The product consists of a processing server and a web interface. The software applies algorithms for recognition of suspicious calcifications and soft tissue lesions. These algorithms have been trained on large databases of biopsy proven examples of breast cancer, benign lesions and normal tissue.
MammoScreen automatically processes FFDM and the output of the device can be used by radiologists concurrently with the reading of mammograms. The user interface of MammoScreen has several functions:
a) Activation of computer aided detection (CAD) marks to highlight locations, known as findings, where the device detected calcifications or soft tissue lesions suspicious for cancer.
b) Association of findings with a score, known as the MammoScreen Score, which characterizes findings on a 1-10 scale, with increasing level of suspicion. Only the most suspicious findings (with a MammoScreen score equal or greater than 5) are initially marked to limit the number of findings to review. The user shall also review findings with score of 4 or lower.
c) Indication, with matching markers, when findings corresponding to the same findings are detected in multiple views of the FFDM.
MammoScreen is configured as a DICOM Web compliant node in a network and receives its input images from another DICOM node, called "the DICOM Web Server". The MammoScreen output will be displayed on the screen of a personal computer compliant with requirements specified in the User Manual.
The image analysis unit includes machine learning components trained to detect positive findings (calcifications and soft tissue lesions).
{6}------------------------------------------------
## Indication for Use:
MammoScreen™ is intended for use as a concurrent reading aid for interpreting physicians, to help identify findings on screening FFDM acquired with compatible mammography systems and assess their level of suspicion. Output of the device includes marks placed on findings on the mammogram and level of suspicion scores. The findings could be soft tissue lesions or calcifications. The level of suspicion score is expressed at the finding level, for each breast and overall for the mammogram. Patient management decisions should not be made solely on the basis of analysis by MammoScreen™.
#### Intended patient population
The device is intended to be used in the population of women undergoing screening FFDM.
## Warnings and precautions
Patient management decisions should not be made solely on the basis of analysis by MammoScreen.
| | Subject Device<br>MammoScreen | Predicate Device<br>Transpara™<br>K181704 | Substantially<br>Equivalent? |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|------------------------------|
| Classification<br>Regulation | 21 CFR 892.2090<br>Radiological Computer Assisted<br>Detection And Diagnosis Software | SAME | Yes, identical. |
| Medical<br>Device<br>Classification | Class II | SAME | Yes, identical. |
| Product Code | QDQ | SAME | Yes, identical. |
| Level of<br>Concern | Moderate | SAME | Yes, identical. |
| Intended Use | A concurrent reading aid for<br>physicians interpreting screening<br>FFDM acquired with compatible<br>mammography systems, to identify<br>findings and assess their level of<br>suspicion. | SAME | Yes, identical. |
| Target patient<br>population | Women undergoing FFDM screening mammography | SAME | Yes, identical. |
## Predicate and Subject Device Comparison:
{7}------------------------------------------------
| | Subject Device | Predicate Device | Substantially<br>Equivalent? |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|
| | MammoScreen | Transpara™ | |
| | | K181704 | |
| Target user<br>population | Physicians interpreting FFDM<br>screening mammograms | SAME | Yes, identical |
| Design | Software-only device | SAME | Yes, identical |
| Indication for<br>Use | MammoScreen™ is intended for<br>use as a concurrent reading aid for<br>interpreting physicians, to help<br>identify findings on screening<br>FFDM acquired with compatible<br>mammography systems and assess<br>their level of suspicion. Output of<br>the device includes marks placed on<br>findings on the mammogram and<br>level of suspicion scores. The<br>findings could be soft tissue lesions<br>or calcifications. The level of<br>suspicion score is expressed at the<br>finding level, for each breast and<br>overall for the mammogram. Patient<br>management decisions should not<br>be made solely on the basis of<br>analysis by MammoScreen™. | The ScreenPoint Transpara™<br>system is intended for use as a<br>concurrent reading aid for<br>physicians interpreting<br>screening mammograms, to<br>identify regions suspicious for<br>breast cancer and assess their<br>likelihood of malignancy.<br>Output of the device includes<br>marks placed on suspicious<br>soft tissue lesions and<br>suspicious calcifications;<br>region-based scores, displayed<br>upon the physician's query,<br>indicating the likelihood that<br>cancer is present in specific<br>regions; and an overall score<br>indicating the likelihood that<br>cancer is present on the<br>mammogram.<br>Patient management decisions<br>should not be made solely on<br>the basis of analysis by<br>Transpara™. | Yes, identical |
{8}------------------------------------------------
| | Subject Device | Predicate Device | Substantially<br>Equivalent? |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | MammoScreen | Transpara™<br>K181704 | |
| Score | Finding level:<br>10-point scale score indicating the<br>level of suspicion of malignancy<br>(from low suspicion to high<br>suspicion).<br><br>Breast level:<br>The same 10-point scale score as<br>finding level. The score of a breast<br>is equal to the maximum score of<br>the findings detected in this breast. | Finding level:<br>Continuous score 1-100<br>indicating the level of<br>suspicion of malignancy (from<br>low suspicion to high<br>suspicion).<br><br>Breast level:<br>None | Both scores are<br>substantially<br>equivalent.<br><br>Both scores increase<br>with the level of<br>suspicion. The<br>minimum (resp. the<br>maximum) of the both<br>scores describes the<br>same status.<br><br>At the Exam level,<br>both scores have a 10-<br>point scale. |
| | Exam level:<br>Exam-level of suspicion resulting<br>directly from the maximum score of<br>both breasts (1-to-1 mapping<br>between the score and the exam-<br>level of suspicion). | Exam level:<br>10-point scale score indicative<br>of higher frequency of cancer<br>positive | |
| Finding<br>discovery | Findings are by-default displayed<br>when score is equal or higher to 5.<br><br>Upon user request for findings of<br>score equal or less to 4. | Upon user request by clicking<br>in a position of the image also<br>detected by Transpara™. | Both are the<br>demonstration of the<br>same intention:<br>reducing the number<br>of findings the user<br>has to review. In this<br>sense, both are<br>equivalent. |
{9}------------------------------------------------
| | Subject Device | Predicate Device | Substantially |
|---------------------|------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| | MammoScreen | Transpara™ | Equivalent? |
| | | K181704 | |
| Performances | Reader study: | Reader study: | |
| | - 240 cases<br>- 14 radiologists | - 240 cases<br>- 14 radiologists | Despite a slightly<br>higher performance of<br>MammoScreen<br>compared to |
| | Reading time (two sessions mean): | Reading time: | Transpara™, gains |
| | - 57,67 seconds (unaided<br>session)<br>- 64, 13 seconds (with<br>MammoScreen) | - 146 seconds (unaided<br>session)<br>- 149 seconds (with<br>Transpara™) | are still comparable<br>and do not raise new<br>questions regarding<br>safety and<br>effectiveness of the<br>device. |
| | AUC: | AUC: | |
| | - radiologists AUC<br>(unaided) = 0,769<br>- standalone AUC = 0,786 | - radiologists AUC<br>(unaided) = 0,866<br>- standalone AUC =<br>0,887 | |
| Features | Distinguishes two types of<br>suspicious findings (calcifications<br>and soft tissue lesions). | Distinguishes two types of<br>suspicious findings<br>(calcifications and soft tissue<br>lesions). | Despite some<br>differences between<br>the predicate device<br>and MammoScreen, |
| | The CAD output provided by the<br>server includes the location and the<br>outline of findings. | The CAD output provided by<br>the server includes the location<br>and the outline of findings. | features are still<br>comparable and do<br>not raise new<br>questions regarding<br>safety and |
| | MammoScreen processing server is<br>a standalone system WITH a user<br>interface. | Transpara™ processing server<br>is a standalone system<br>WITHOUT a user interface. | effectiveness of the<br>device. |
| Imaging<br>Modality | FFDM | SAME | Yes, identical |
{10}------------------------------------------------
| | Subject Device<br>MammoScreen | Predicate Device<br>Transpara™ | Substantially<br>Equivalent? |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|------------------------------|
| | | K181704 | |
| Fundamental<br>scientific<br>technology | In MammoScreen, a range of<br>medical image processing and<br>machine learning techniques are<br>implemented. The system includes<br>'deep learning' modules for<br>recognition of suspicious<br>calcifications and soft tissue<br>lesions. These modules are trained<br>with very large databases of biopsy-<br>proven examples of breast cancer<br>and normal tissue. | SAME | Yes, identical |
Table 7.1 Comparison between subject and predicate device
## Discussion:
The indication for use of MammoScreen is very similar to that of the predicate device. Both devices are intended to be used by clinicians interpreting digital mammogram images, to help them with localizing and characterizing suspicious findings. The devices are both intended to be used concurrently with the reading of images and are not intended as a replacement for the review of a clinician or their clinical judgement. Both devices target the same findings, the same patient and user populations and use the same imaging modality. Also, both devices have a Moderate Level of Concern.
Both devices give a score. The scoring system is not the same, but is based on similar principles: providing users with a level of suspicion for malignancy from low (score 1) to high (score 10 or 100). TransparaTM exhibits some minor differences though:
- at the finding level, Transpara™ uses a 1-100 score of Transpara™ while the MammoScreen score has a 10-point scale. Indeed, Therapixel believes, based on questionnaire submitted to 15 radiologists, that interpretability of less granular scale is easier for users (how shall a difference between a score of 41 and 43 be interpreted for instance?)
- at the breast level. Transpara™ does not provide a score. Therapixel considers that this level is required for better interpretability at exam level, scores are equivalent. However, using Transpara™, the user cannot determine with precision to which of the breast (if not both), and where in the breast(s) this overall score applies. On the contrary, Therapixel's score (and the indication of the exam-level of suspicion that directly results from it) is directly connected to one of the breasts (or both) and indicated as such, and exactly where in the breast (finding) thanks to the scoring consistency (a breast inherits from the maximum score of its findings, and the exam inherits from the maximum score of both breasts). Doing so makes the algorithm decision more explicit, and easier to interpret from a user point of view.
{11}------------------------------------------------
Both scores increase with the level of suspicion. The minimum (resp. the maximum) of the both scores describes the same status. At the Exam level, both scores have a 10-point scale.
Both scores are substantially equivalent, the minor differences do not raise further or different questions of performance or safety.
In conclusion, these differences do not raise different questions of safety and effectiveness of the device when used as labeled. The overall design of MammoScreen is very similar to that of the predicate device. Both devices have the same intended use, very similar indications for use, and minor differences between MammoScreen and the predicate device do not raise different questions of safety and effectiveness which are equivalent.
## Non-clinical Testing
MammoScreen is a software-only device. The level of concern for the device is determined as Moderate Level of Concern.
Tests have been performed in compliance with the following recognized consensus standards:
- IEC 62304 Edition 1.1 2015-06 Medical device software Software life-cycle processes ●
- IEC 62366-1 Edition 1.0 2015-02 Medical devices Application of usability engineering ● to medical devices.
MammoScreen has successfully completed integration and verification testing and beta validation. In addition, potential hazards have been evaluated and mitigated, and have acceptable levels.
{12}------------------------------------------------
## Standalone Performance Testing
Standalone performance testing of MammoScreen assesses the performance of MammoScreenalgorithms in the absence of a clinician and includes mammograms of women acquired with devices from two manufacturers: Hologic® and GE®.
The following graph shows the ratio of positive cases over all cases processed by both GE® and Hologic® devices that fall into each score:
Image /page/11/Figure/3 description: The image is a bar graph titled "Repartition by MS score (combined)". The x-axis is labeled "MammoScreen score" and ranges from 1 to 10. The y-axis is labeled "Repartition of positive cases" and ranges from 0.0 to 1.0. The bar graph shows an increase in the repartition of positive cases as the MammoScreen score increases.
Figure 7.2.-Repartition of positive cases per MammoScreen scores for Hologic® and GE® combined
## Mammogram level
Standalone performance of the MammoScreen-AI algorithm in characterizing positive and negative US FFDM acquired on Hologic® devices, and performance comparison with FFDM acquired on GE® devices, are given in the table below:
| | Hologic® | GE® | Combined | FOMGE-FOMHOL |
|------------|----------------------|----------------------|----------------------|-----------------------|
| ROC<br>AUC | 0.868 (0.851, 0.885) | 0.887 (0.875, 0.898) | 0.883 (0.873, 0.892) | 0.018 (-0.002, 0.039) |
{13}------------------------------------------------
| Sensitivity | 0.844 (0.815, 0.872) | 0.849 (0.827, 0.871) | 0.847 (0.829, 0.864) | 0.005 (-0.032, 0.042)1 |
|-------------|----------------------|----------------------|----------------------|------------------------|
| Specificity | 0.705 (0.689, 0.722) | 0.738 (0.728, 0.747) | 0.729 (0.721, 0.737) | 0.032 (0.014, 0.051) |
Image /page/12/Figure/1 description: The image is a Receiver Operating Characteristic (ROC) curve, which is a graphical plot that illustrates the diagnostic ability of a binary classifier system as its discrimination threshold is varied. The plot shows the true positive rate (y-axis) versus the false positive rate (x-axis). There are three ROC curves plotted: AUC Hologic = 0.868, AUC GE = 0.887, and AUC Combined = 0.882, along with data points labeled MS6.
Figure 7.3: ROC curves of MammoScreen-AI algorithm at mammogram level on US Hologic® FFDM (blue), GE® FFDM (orange), and on both types combined (green). Sensitivity and specificity at the chosen standalone regime (MS6) are shown with a blue dot for each case.
#### Breast level
Standalone performance of the MammoScreen-AI algorithm in characterizing positive and negative breasts on US FFDM acquired on acquired on Hologic® devices, and performance comparison with FFDM acquired on GE® devices, are given in the table below:
| | Hologic® | GE® | Combined | FOMGE-FOMHOL |
|-------------|----------------------|----------------------|----------------------|-----------------------|
| ROC<br>AUC | 0.901 (0.886, 0.915) | 0.916 (0.906, 0.926) | 0.911 (0.902, 0.919) | 0.015 (-0.002, 0.033) |
| Sensitivity | 0.813 (0.781, 0.843) | 0.830 (0.808, 0.853) | 0.823 (0.805, 0.841) | 0.017 (-0.023, 0.056) |
| Specificity | 0.840 (0.831, 0.848) | 0.846 (0.840, 0.851) | 0.844 (0.839, 0.849) | 0.006 (-0.004, 0.016) |
<sup>4</sup> Due to the lower bound of the 95% Cl being close to the non-inferiority margin (-0.03), 10,000 bootstrap replicates were used instead of 2,000.
{14}------------------------------------------------
Image /page/13/Figure/0 description: The image shows a receiver operating characteristic (ROC) curve. The ROC curve plots the true positive rate against the false positive rate. There are three ROC curves plotted on the graph, representing different models: AUC Hologic = 0.901, AUC GE = 0.916, and AUC Combined = 0.911. The ROC curves are all relatively close to each other, indicating that the models have similar performance.
Figure 7.4: ROC curves of ManmoScreen-AI algorithm at breast level on US Hologic® FFDM (blue), GE® FFDM (orange), and on both types combined (green). Sensitivity at the chosen standalone regime (MSG) are shown with a blue dot for each case.
## Finding level
Standalone performance of the MammoScreen-AI algorithm in detecting and characterizing positive findings (soft tissue lesions and calcifications) on FFDM acquired on Hologic® devices, and comparison with FFDM acquired on GE® devices, are given in the tables below.
Two characteristics are considered: The Free-Response ROC (FROC) and the Localized ROC (LROC).
FROC curve shows the sensitivity of finding detection and characterization as a function of the average number of false marks per image. It illustrates how performant an algorithm is at detecting and characterizing positive findings while keeping the number of false marks as low as possible. The drawback of FROC curves is that the x-axis has virtually no limit and that the curve area is not bound to unit square. Therefore, no simple unique measure (such as the AUC) may be derived from FROC curves to compare between several of such curves. However, it is still valuable information as it accounts for the average number of false marks per image.
LROC curves are generally used when only one positive finding per case shall be localized. It shows the sensitivity of correctly marking a positive case (a case is considered as correctly marked
{15}------------------------------------------------
if the algorithm placed a mark at a reasonable distance from the ground truth location – here in a 15mm radius from the ground truth) as a function of the false positive rate (a negative case is considered a false positive if the algorithm has placed one or more marks on it). LROC curve areas are bound to unit square, allowing to derive single measures to describe them such as AUC.
For MammoScreen, LROC curves at the breast level are used. This allows to measure how performant the algorithm is at marking the correct finding within a breast (in either of the 2 views of the breast) as a function of the false positive breast rate (i.e., rate of placing false marks in either of the 2 views of a negative breast).
In addition to FROC and LROC analysis, the performance of four detection Operating Points (OP) is reported. Those correspond to the three OP made corresponding to visualization levels made available to users via the MammoScreen user interface, using the Filtering slider (refer to section 5.3.1.1 Filtering of findings) and corresponding to MammoScreen scores 1, 3 and 5 respectively.
| MammoScreen Score | MS1 | MS3 | MS5 |
|---------------------------------|----------------------------------|-------------------------------------|---------------------------------|
| Filtering<br>slider<br>position | Left (Lowest<br>Suspicion shown) | Middle (Lowest<br>Suspicion hidden) | Right (Low Suspicion<br>hidden) |
| Soft tissue lesions | | | | | |
|---------------------|-----------------------|----------------------|----------------------|----------------------|------------------------|
| | | Hologic® | GE® | Combined | FOMGE-FOMHOL |
| LROC | AUC (primary) | 0.837 (0.811, 0.861) | 0.900 (0.884, 0.916) | 0.877 (0.862, 0.890) | 0.064 (0.034, 0.095) |
| | Sensitivity @ MS1 | 0.942 (0.921, 0.963) | 0.976 (0.964, 0.987) | 0.962 (0.951, 0.973) | 0.034 (0.010, 0.060) |
| | Sensitivity @ MS3 | 0.926 (0.902, 0.949) | 0.966 (0.951, 0.978) | 0.950 (0.937, 0.963) | 0.039 (0.012, 0.067) |
| | Sensitivity @ MS5 | 0.774 (0.735, 0.811) | 0.852 (0.824, 0.879) | 0.822 (0.799, 0.843) | 0.080 (0.032, 0.128) |
| | Specificity @ MS1 | 0.109 (0.102, 0.116) | 0.166 (0.161, 0.172) | 0.150 (0.146, 0.155) | 0.058 (0.048, 0.067) |
| | Specificity @ MS3 | 0.233 (0.222, 0.242) | 0.222 (0.216, 0.228) | 0.225 (0.219, 0.230) | -0.011 (-0.022, 0.002) |
| | Specificity @ MS5 | 0.780 (0.771, 0.791) | 0.801 (0.795, 0.807) | 0.795 (0.790, 0.800) | 0.021 (0.010, 0.032) |
| FROC | Sensitivity @ MS1 | 0.895 (0.872, 0.915) | 0.952 (0.940, 0.963) | 0.930 (0.918, 0.940) | 0.057 (0.033, 0.081) |
| | Sensitivity @ MS3 | 0.878 (0.854, 0.901) | 0.943 (0.930, 0.955) | 0.918 (0.906, 0.929) | 0.064 (0.038, 0.090) |
| | Sensitivity @ MS5 | 0.750 (0.718, 0.782) | 0.835 (0.814, 0.855) | 0.802 (0.784, 0.819) | 0.084 (0.047, 0.121) |
| | Avg false marks @ MS1 | 1.819 (1.815, 1.824) | 1.268 (1.266, 1.270) | 1.424 (1.422, 1.426) | NA |
| | Avg false marks @ MS3 | 1.182 (1.171, 1.193) | 1.026 (1.021, 1.031) | 1.070 (1.066, 1.075) | NA |
| | Avg false marks @ MS5 | 0.294 (0.286, 0.302) | 0.218 (0.214, 0.222) | 0.239 (0.235, 0.243) | NA |
{16}------------------------------------------------
| | | Calcifications | | | |
|------|-----------------------|----------------------|----------------------|----------------------|-------------------------|
| | | Hologic® | GE® | Combined | FOMGE-FOMHOL |
| LROC | AUC (primary) | 0.974 (0.959, 0.985) | 0.930 (0.912, 0.948) | 0.942 (0.928, 0.954) | -0.044 (-0.067, -0.021) |
| | Sensitivity @ MS1 | 0.994 (0.979, 1.000) | 0.971 (0.953, 0.987) | 0.978 (0.964, 0.989) | -0.023 (-0.043, -0.001) |
| | Sensitivity @ MS3 | 0.994 (0.979, 1.000) | 0.971 (0.953, 0.987) | 0.978 (0.964, 0.989) | -0.023 (-0.043, -0.001) |
| | Sensitivity @ MS5 | 0.962 (0.930, 0.988) | 0.804 (0.764, 0.844) | 0.851 (0.820, 0.879) | -0.158 (-0.209, -0.108) |
| | Specificity @ MS1 | 0.549 (0.537, 0.561) | 0.645 (0.638, 0.652) | 0.619 (0.613, 0.625) | 0.096 (0.083, 0.110) |
| | Specificity @ MS3 | 0.584 (0.573, 0.597) | 0.658 (0.652, 0.665) | 0.638 (0.632, 0.644) | 0.074 (0.061, 0.088) |
| | Specificity @ MS5 | 0.869 (0.861, 0.878) | 0.909 (0.905, 0.913) | 0.898 (0.894, 0.901) | 0.040 (0.031, 0.048) |
| FROC | Sensitivity @ MS1 | 0.994 (0.984, 1.000) | 0.941 (0.923, 0.956) | 0.957 (0.944, 0.969) | -0.052 (-0.072, -0.033) |
| | Sensitivity @ MS3 | 0.994 (0.984, 1.000) | 0.938 (0.920, 0.955) | 0.955 (0.941, 0.968) | -0.055 (-0.075, -0.036) |
| | Sensitivity @ MS5 | 0.942 (0.916, 0.968) | 0.779 (0.747, 0.809) | 0.828 (0.805, 0.851) | -0.163 (-0.202, -0.124) |
| | Avg false marks @ MS1 | 0.870 (0.867, 0.873) | 0.531 (0.529, 0.532) | 0.627 (0.625, 0.628) | NA |
| | Avg false marks @ MS3 | 0.753 (0.748, 0.759) | 0.479 (0.477, 0.482) | 0.557 (0.554, 0.559) | NA |
| | Avg false marks @ MS5 | 0.238 (0.232, 0.245) | 0.122 (0.119, 0.125) | 0.155 (0.152, 0.158) | NA |
Image /page/15/Figure/1 description: The image contains two ROC curves, one titled "Breast Localized ROC (LROC)" and the other titled "Image Free-Response ROC (FROC)". Each plot shows three curves representing different methods: Hologic, GE, and Combined, with corresponding AUC values of 0.837, 0.900, and 0.877 respectively. The plots also include data points for MS1, MS3, MS5, and MS6, represented by red, green, blue, and yellow markers, respectively, on each curve.
Figure 7.5: LROC (left) and FROC (right) curves on US soft tissue lesions only.
{17}------------------------------------------------
Image /page/16/Figure/0 description: The image contains two ROC curves, one titled "Breast Localized ROC (LROC)" and the other titled "Image Free-Response ROC (FROC)". Each ROC curve plots the true positive rate against the false positive rate for different models. The LROC plot shows the AUC for Hologic is 0.974, GE is 0.930, and Combined is 0.941. Both plots also show data points for MS1, MS3, MS5, and MS6.
Figure 7.6: LROC (left) and FROC (right) curves on US calcifications only.
{18}------------------------------------------------
# Per-MammoScreen score analysis
#### Mammogram level
Image /page/17/Figure/2 description: This image contains two bar charts comparing Hollogic and GE mammogram scores. The left chart shows the proportion of screening mammograms per score, while the right chart shows the probability of positive mammograms per score. The MammoScreen score ranges from 1 to 10 on the y-axis of both charts. The proportion of screening mammograms per score is highest for Hollogic at a score of 4, with 33.744%, while GE is highest at a score of 4, with 37.931%.
Image /page/17/Figure/3 description: The image contains text describing figure 7.8. The figure shows the repartition of screening FFDM per MammoScreen score at the mammogram level. The left side of the figure shows the repartition of screening FFDM per score for Hologic and GE manufacturers. The right side of the figure shows the probability of discovering a positive FFDM per score.
{19}------------------------------------------------
## Breast level
Image /page/18/Figure/1 description: This image contains two horizontal bar charts comparing the performance of Hologic and GE mammography systems. The left chart shows the proportion of screening mammograms per score, while the right chart displays the probability of positive mammograms per score, with scores ranging from 1 to 10. The proportion of screening mammograms per score is highest for both systems at score 4, with Hologic at 30.729% and GE at 33.821%, while the probability of positive mammograms per score is 100% for both systems at score 10.
Image /page/18/Figure/2 description: The image shows the title of a figure, "Figure 7.9: Repartition of screening FFDM per MammoScreenTM score at breast level." There are two descriptions below the title. The first description is "Left: Repartition of screening FFDM per score for Hologic® and GE® manufacturers." The second description is "Right: Probability of discovering a positive FFDM per score."
{20}------------------------------------------------
Table 7.2: NPV, specificity, NPV, sensitivity at mannoscreen™ scores on a similated screening distribution (9% confidence interval in brackets).
# Hologic®
| FOM | MS1 | MS2 | MS3 | MS4 | MS5 | MS6 | MS7 | MS8 | MS9 | MS10 |
|-------------|-----------------------------|-----------------------------|-----------------------------|-----------------------------|-----------------------------|-----------------------------|-----------------------------|-----------------------------|-----------------------------|-----------------------------|
| NPV | 1.0000<br>(1.0000 - 1.0000) | 1.0000<br>(1.0000 - 1.0000) | 0.9991<br>(0.9987 - 0.9995) | 0.9984<br>(0.9977 - 0.9991) | 0.9979<br>(0.9970 - 0.9987) | 0.9956<br>(0.9946 - 0.9964) | 0.9939<br>(0.9927 - 0.9948) | 0.9928<br>(0.9916 - 0.9938) | 0.9922<br>(0.9911 - 0.9932) | 0.9922<br>(0.9911 - 0.9932) |
| Specificity | 0.0162<br>(0.0129 - 0.0196) | 0.0460<br>(0.0390 - 0.0533) | 0.2175<br>(0.2051 - 0.2309) | 0.5632<br>(0.5424 - 0.5805) | 0.7185<br>(0.7015 - 0.7332) | 0.9690<br>(0.9627 - 0.9739) | 0.9942<br>(0.9922 - 0.9968) | 0.9987<br>(0.9968 - 0.9998) | 1.0000<br>(1.0000 - 1.0000) | 1.0000<br>(1.0000 - 1.0000) |
| PPV | 0.0078<br>(0.0068 - 0.0089) | 0.0079<br>(0.0069 - 0.0091) | 0.0082<br>(0.0071 - 0.0094) | 0.0097<br>(0.0084 - 0.0111) | 0.0158<br>(0.0134 - 0.0180) | 0.0221<br>(0.0185 - 0.0250) | 0.1049<br>(0.0863 - 0.1267) | 0.2443<br>(0.1804 - 0.3734) | 0.3788<br>(0.1662 - 0.7885) | 1.0000<br>(1.0000 - 1.0000) |
| Sensitivity | 1.0000<br>(1.0000 - 1.0000) | 1.0000<br>(1.0000 - 1.0000) | 1.0000<br>(1.0000 - 1.0000) | 0.9764<br>(0.9617 - 0.9862) | 0.8882<br>(0.8397 - 0.9302) | 0.8073<br>(0.7331 - 0.8737) | 0.4584<br>(0.4040 - 0.5177) | 0.2294<br>(0.1956 - 0.2693) | 0.0844<br>(0.0643 - 0.1036) | 0.0000<br>(0.0000 - 0.0000) |
# GE®
| FOM | MS1 | MS2 | MS3 | MS4 | MS5 | MS6 | MS7 | MS8 | MS9 | MS10 |
|-------------|-------------------|-------------------|-------------------|-------------------|-------------------|-------------------|-------------------|-------------------|-------------------|-------------------|
| NPV | 1.0000 | 1.0000 | 0.9996 | 0.9992 | 0.9988 | 0.9973 | 0.9965 | 0.9955 | 0.9953 | 0.9952 |
| | (1.0000 - 1.0000) | (1.0000 - 1.0000) | (0.9994 - 0.9997) | (0.9987 - 0.9994) | (0.9982 - 0.9991) | (0.9969 - 0.9978) | (0.9960 - 0.9969) | (0.9950 - 0.9960) | (0.9947 - 0.9958) | (0.9947 - 0.9958) |
| Specificity | 0.0409 | 0.0610 | 0.4238 | 0.6339 | 0.7741 | 0.9808 | 0.9973 | 1.0000 | 1.0000 | 1.0000 |
| | (0.0372 - 0.0448) | (0.0558 - 0.0655) | (0.4108 - 0.4338) | (0.6232 - 0.6442) | (0.7659 - 0.7822) | (0.9786 - 0.9838) | (0.9962 - 0.9982) | (1.0000 - 1.0000) | (1.0000 - 1.0000) | (1.0000 - 1.0000) |
| PPV | 0.0048 | 0.0050 | 0.0051 | 0.0079 | 0.0115 | 0.0166 | 0.1015 | 0.3154 | 1.0000 | 1.0000 |
| | (0.0042 - 0.0053) | (0.0044 - 0.0055) | (0.0045 - 0.0056) | (0.0070 - 0.0088) | (0.0103 - 0.0127) | (0.0148 - 0.0184) | (0.0861 - 0.1235) | (0.2385 - 0.4166) | (1.0000 - 1.0000) | (1.0000 - 1.0000) |
| Sensitivity | 1.0000 | 1.0000 | 1.0000 | 0.9626 | 0.8936 | 0.7992 | 0.4513 | 0.2579 | 0.0516 | 0.0078 |
| | (1.0000 - 1.0000) | (1.0000 - 1.0000) | (1.0000 - 1.0000) | (0.9529 - 0.9731) | (0.8288 - 0.9233) | (0.7325 - 0.8514) | (0.4054 - 0.5004) | (0.2236 - 0.2983) | (0.0398 - 0.0635) | (0.0035 - 0.0130) |
{21}------------------------------------------------
## Clinical Testing
This investigation was a retrospective multi reader multi case study meant to compare cancer detection performance of radiologists reading with the aid of MammoScreen to the reading results of the same cohort of radiologists without any decision support.
The main objective of this investigation was to determine whether the radiologist performance when using MammoScreen is superior to unaided radiologist performance for interpretation of 2D Full Field Digital Screening Mammograms.
To do this, 240 mammographic screening images acquired at a US center have been collected. For each exam, the cancer status has been verified by either biopsy results (for all cancer positive cases and some of the negative cases) or an adequate follow-up (for negative cases only) and used as gold standard. The images have been read by 14 radiologists with and without the aid of MammoScreen, the interpretation of the standalone MammoScreen has been recorded as well. Finally, the algorithm's interpretation, the radiologist interpretation using the help of the algorithm and the unaided radiologist interpretation of every mammogram have been compared to the "gold standard". Those pairwise comparisons, have been used to numerically compute the primary and secondary endpoints (AUC under the ROC curve, Sensitivity and Specificity) to compare radiologist performance when using the system and without using it and standalone algorithm performance.
The performance characteristics of radiologists taking part to the clinical investigation was improved when using MammoScreen support, with the average AUC going from 0.77 to 0.8 (difference = 0.028; P = 0.035) (Figure 7.10.1). The AUC was higher with the aid of MammoScreen for 11 of the 14 radiologists (Figure 7.10.2).
Image /page/20/Figure/5 description: The image is a plot of the receiver operating characteristic (ROC) curve. The x-axis is the false positive rate, and the y-axis is the true positive rate. There are two ROC curves plotted on the graph, one for the unassisted case and one for the assisted case. The area under the curve (AUC) for the unassisted case is 0.769, and the AUC for the assisted case is 0.798.
Figure 7.10.1– Average ROC curves of all readers when unassisted (blue) and assisted (orange) with MammoScreen.
{22}------------------------------------------------
Image /page/21/Figure/0 description: The image contains two receiver operating characteristic (ROC) curves, one labeled "unassisted" and the other "assisted". Each ROC curve plots the true positive rate against the false positive rate for multiple readers and a multi-specialty (MS) reader. The area under the curve (AUC) values for each reader are displayed in the legend, ranging from 0.694 to 0.845 for the unassisted ROC and 0.745 to 0.847 for the assisted ROC, with the AUC for the MS reader being 0.790 in both cases.
Figure 7.10.2- Left: ROC curves of all readers when unassisted. Right: ROC curves of all readers when assisted with MammoScreen.
Performances were also measured at breast and lesion level; the overall performance improvement was found to be statistically significant at both breast (in terms of AUC) and lesion (in terms of pAUC) level confirming the trend of the analysis at mammogram level (Figure 7.10.3).
Image /page/21/Figure/3 description: The image contains two receiver operating characteristic (ROC) curves, labeled 'a' and 'b'. Each plot shows two curves, one for 'AUC unassisted' and another for 'AUC assisted'. In plot 'a', the AUC for unassisted is 0.794 and for assisted is 0.827, while in plot 'b', the AUC for unassisted is 0.592 and for assisted is 0.644. Both plots also include a dashed line representing the line of no discrimination.
Figure 7.10.3 – Average ROC curves of all readers when unassisted (blue) and assisted (orange) with MammoScreen at breast level (a) and lesion level (b).
{23}------------------------------------------------
On average, reading time per case increased when using MammoScreen in both reading sessions. During the first session the average reading time was 60.82 seconds (95% confidence interval: 59.25 seconds, 62.39 seconds) for the unaided reading condition and 68.65 seconds (95% confidence interval: 66.92 seconds, 70.39 seconds) when using MammoScreen. For the second reading session the average time was lower with respect to the first session, 54.52 seconds (95% confidence interval: 52.97 seconds, 56.07 seconds) for the unaided reading condition and 59.61 seconds (95% confidence interval: 58.05 seconds, 61.17 seconds) when using MammoScreen. Differences in reading time with and without the use of MammoScreen changed as a function of the MammoScreen score (P < 0.05): for score equal or lower than 4 during the first reading session radiologists increased their reading time by 1% when using MammoScreen, while during the second reading session the use of MammoScreen made them decrease their reading time by 2%. For score higher than 4 in both reading sessions the use of MammoScreen increased the average reading time of about 14%, the maximum increase in reading time did not exceed 15s.
The standalone analysis of MammoScreen has shown that the algorithm exhibits performances comparable (non-inferior) to those of an experienced radiologist as confirmed by the absence of statistical effect (p>0.05) and by the lower confidence interval of the difference of AUC being equal or superior to the effect size (-0.03). Indeed, the performance of the standalone MammoScreen (AUC = 0.79) was found to be non-inferior to the average performance of unaided radiologists (AUC = 0.77) (Figure 7.10.4).
Image /page/22/Figure/2 description: This image is a receiver operating characteristic (ROC) curve. The plot shows the true positive rate on the y-axis and the false positive rate on the x-axis. There are three ROC curves plotted: one for an unassisted model with an AUC of 0.768, one for an assisted model with an AUC of 0.797, and one for a standalone model with an AUC of 0.790.
Figure 7.10.4- Comparison of unassisted radiologists, assisted radiologists, and standalone MammoScreen ROC curve
{24}------------------------------------------------
## Conclusions
Non-clinical and clinical performance tests demonstrate that MammoScreen is safe and effective.
Results of the primary analysis of the clinical test demonstrate that use of MammoScreen improves detection of breast cancer in mammograms. Descriptively, improvement was observed to depend neeligibly on lesion type and the total reading time was not observed to increase with the use of MammoScreen. In addition, the sensitivity of the readers tended to increase with the use of MammoScreen without decreasing specificity.
Radiologists improved their diagnostic performance in the detection of breast cancer with 2D FFDM (Full-Field Digital Mammography) by using MammoScreen. The overall conclusion of this clinical investigation is that the MammoScreen improves the diagnostic performance of radiologists in the detection of breast cancer without slowing down their average reading time. Finally, in standalone testing, MammoScreen breast cancer detection performance was observed to approach the average performance of the clinical study radiologists when reading mammograms unaided.
Based on the Intended Use. Indications for Use, product technical information, performance evaluation, and standards compliance provided in this premarket notification, MammoScreen has been shown to be substantially equivalent to the cited predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.