K192807 · Medical Components, Inc. (dba MedComp) · MPB · Apr 23, 2020 · Gastroenterology, Urology
Device Facts
Record ID
K192807
Device Name
Duo-Flow Side x Side Double Lumen Catheter
Applicant
Medical Components, Inc. (dba MedComp)
Product Code
MPB · Gastroenterology, Urology
Decision Date
Apr 23, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5540
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Duo-Flow Side x Side Double Lumen Catheter is intended for short-term central venous access for hemodialysis, apheresis, and infusion.
Device Story
Duo-Flow Side x Side Double Lumen Catheter is a non-implanted blood access device for short-term central venous access. It features two lumens: a proximal lumen for arterial outflow and a distal lumen for venous return. Available in 9F, 11F, and 12F sizes with straight or curved extensions. Used in clinical settings, typically the operating room or bedside, by healthcare professionals. The device facilitates hemodialysis, apheresis, and infusion procedures. It is sterilized via ethylene oxide (EO). The catheter provides a pathway for blood flow between the patient and external hemodialysis or apheresis equipment, enabling extracorporeal blood treatment.
Clinical Evidence
Bench testing only. Performance testing included air leak, liquid leak, peak tensile force, gravity flow, transit/shelf-life testing, and luer taper gauging per ISO 10555-1, ISO 11607-1/2, ISTA 3A, and ISO 594-1/2. Biocompatibility testing conducted per ISO 10993 series, including cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, subacute toxicity, genotoxicity, implantation, and hemocompatibility (ASTM F576).
Technological Characteristics
Double-lumen catheter; 9F, 11F, 12F sizes; straight or curved extensions; proximal/distal lumen configuration. Materials evaluated for biocompatibility per ISO 10993-1. Sterilization: Ethylene Oxide (EO). Connectivity: None (mechanical device).
Indications for Use
Indicated for adult patients requiring short-term (<30 days) central venous access for hemodialysis, apheresis, and infusion via subclavian, jugular, or femoral sites.
Regulatory Classification
Identification
A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device. (2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle). (3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).
Special Controls
*Classification.* (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided.
(ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated.
(B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided.
(C) Priming volumes must be established.
(D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use.
(E) Air leakage testing and liquid leakage testing must be conducted.
(F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage.
(G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern.
(H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established.
(iii) Performance data must demonstrate the sterility of the device.
(iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing.
(v) Labeling of implanted blood access devices for hemodialysis must include the following:
(A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated.
(B) Labeling must specify the forward and reverse recirculation rates.
(C) Labeling must provide the arterial and venous priming volumes.
(D) Labeling must specify an expiration date.
(E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device.
(F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient.
(G) Labeling must include a patient implant card.
(H) The labeling must contain comprehensive instructions for the following:
(
*1* ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;(
*2* ) Proper care and maintenance of the device and device exit site;(
*3* ) Removal of the device;(
*4* ) Anticoagulation;(
*5* ) Management of obstruction and thrombus formation; and(
*6* ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following:
(A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port.
(B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following:
(A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied.
(B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties.
(C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens.
(D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness.
(viii) The following must be included for A-V shunt cannulae (with vessel tips):
(A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply.
(B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(2) Class II (performance standards) for the nonimplanted blood access device.
(3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section.
(4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services-USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a symbol of the agency's mission to protect and promote public health through the regulation and supervision of food safety, tobacco, pharmaceuticals, medical devices, and other related products.
April 23, 2020
Medical Components Inc. (dba MedComp) Courtney Nix Regulatory Affairs Director, North America and Europe 1499 Delp Drive Harleysville, PA 19438
Re: K192807
> Trade/Device Name: Duo-Flow Side x Side Double Lumen Catheter Regulation Number: 21 CFR 876.5540 Regulation Name: Blood access device and accessories Regulatory Class: II Product Code: MPB Dated: March 13, 2020 Received: March 16, 2020
Dear Courtney Nix:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Carolyn Y. Neuland, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192807
Device Name
Duo-Flow Side x Side Double Lumen Catheter
Indications for Use (Describe)
The Duo-Flow Side x Side Double Lumen Catheter is intended for short-term central venous access for hemodialysis, apheresis, and infusion.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/15 description: The image shows the logo for medCOMP. The logo features a red geometric shape that resembles a stylized house or building. The word "med" is in red, while "COMP" is in black. The logo is simple and modern, with a focus on the company name and a memorable visual element.
Harleysville, PA 19438
Tel: 215-256-4201
Fax. 215-256-1787
www.medcompnet.com
Medcomp®: Duo-Flow® Side x Side Double Lumen Catheter
| Section 6 | | 510(k) SUMMARY | Traditional 510K |
|--------------------------|-------------------------|----------------------------------------------------------------------------------------|------------------|
| A. Submitter Information | | | |
| | Submitter Name: | Medical Components Inc.<br>(dba Medcomp®)<br>1499 Delp Drive<br>Harleysville, PA 19438 | |
| | Address: | 2518902 | |
| | Registration Number: | Courtney Nix<br>Regulatory Affairs Director,<br>North America and EU | |
| | Contact Person: | 09/27/2019 | |
| | Date of Preparation: | | |
| B. Subject Device | | | |
| | Trade Name: | Duo-Flow® Side x Side Double Lumen Catheter | |
| | Device: | Catheter, Hemodialysis, Non-Implanted<br>Blood access device and accessories | |
| | Regulation Description: | MPB | |
| | Product Code: | 876.5540 | |
| | Regulation Number: | II | |
| | Class: | Gastroenterology/Urology | |
| | Review Panel: | | |
| C. Predicate Device | | | |
| | Predicate Trade Name: | Mahurkar™ Acute Dual Lumen Catheter | |
| | 510(k) Number: | K955002 | |
| | 510(k) Holder: | Quinton, Inc. | |
| | Device: | Catheter, Hemodialysis, Non-Implanted<br>Blood access device and accessories | |
| | Regulation Description: | MPB | |
| | Product Code: | 876.5540 | |
| | Regulation Number: | II | |
| | Class: | Gastroenterology/Urology | |
| | Review Panel: | | |
| D. Device Description: | | | |
- Red Adapter = proximal lumen O
- O Blue Adapter = distal lumen
The proximal lumen provides "arterial" outflow from the patient; the distal lumen provides "venous" return.
Medcomp®: Duo-Flow® Side x Side Double Lumen Catheter Section 6: 510(k) Site Summary
6-1
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for medCOMP. The logo consists of a red geometric shape that resembles a stylized letter "C" or an incomplete square with rounded corners. Below the geometric shape, the word "medCOMP" is written in a stylized font, with "med" in red and "COMP" in black.
Harleysville, PA 19438
Tel: 215-256-4201
Fax. 215-256-1787
www.medcompnet.com
Medcomp®: Duo-Flow® Side x Side Double Lumen Catheter
The catheter comes in a variety of sizes and is offered with curved or straight extensions.
#### E. Indications for Use:
The Duo-Flow® Side x Side Double Lumen Catheter is intended for short-term central venous access for hemodialysis, apheresis, and infusion.
#### F. Comparison to Predicate Device(s):
Table 6.1: 510(k) Summary Design Comparison Matrix
| Attribute | Subject Device<br>Duo-Flow® Side x Side Double Lumen Catheter | Predicate Device<br>Mahurkar™ Acute Dual Lumen Catheter (K955002) | | |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|---------------------------------------------------------------------------------------------------------------|
| Prescription<br>Indications<br>for Use | Prescription Use<br>The Duo-Flow® Side x Side Double Lumen Catheter is intended for short-term central venous access for hemodialysis, apheresis, and infusion | Prescription Use<br>The catheter is intended for short-term central venous access for hemodialysis, apheresis, and infusion. | | |
| Definition<br>(From<br>Product<br>Code) | Short-term (< 30 days) central venous access for hemodialysis and apheresis. | Short-term (< 30 days) central venous access for hemodialysis and apheresis. | | |
| Location of<br>Use | The operating room is the preferred location for insertion; however, bedside insertion is acceptable if sterile technique is followed. | The operating room is the preferred location for insertion; however, bedside insertion is acceptable if sterile technique is followed. | | |
| French Size | 9F, 11F, and 12F | 8F, 10F, 11.5F and 13.5F | | |
| Catheter<br>Configuration<br>Lengths | Straight and Curved Extensions | Straight, Curved Extensions, and Pre-Curved | | |
| | French<br>Size<br>9<br>11<br>12 | Length(s)<br>(cm) 7.5, 10, 12,<br>15, 20<br>10, 12, 15,<br>20, 24<br>13, 15, 20,<br>24 | French<br>Size 8F 10 11.5 13.5 | Length(s) (cm) 9cm (Straight<br>only), 12cm,<br>15cm 12, 15, 19.5 13.5, 16, 19.5,<br>24 13.5, 16, 19.5,<br>24 |
| Duration of<br>Use | Short-term (less than 30 days) | Per K-DOQI guidelines: (1)<br>Remove the catheter when no | | |
Medcomp®: Duo-Flow® Side x Side Double Lumen Catheter Section 6: 510(k) Site Summary
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the MedComp logo. The logo consists of a red geometric shape that resembles a stylized house or building. Below the geometric shape, the word "medCOMP" is written in a stylized font, with "med" in red and "COMP" in black.
- 1499 Delp Drive
Harleysville, PA 19438
Tel: 215-256-4201
Fax. 215-256-1787
18
www.medcompnet.com
Medcomp®: Duo-Flow® Side x Side Double Lumen Catheter
| | | longer needed, (2) Remove<br>femoral catheters within 3-4 days,<br>and (3) Replace subclavian and<br>jugular catheters every 21 days. |
|-------------------------|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Sterilization<br>Method | EO | EO |
| Number of<br>Lumens | 2 | 2 |
| Patient<br>Population | Adult | Adult |
| Insertion Site | Subclavian, jugular or femoral | Subclavian, jugular or femoral |
| Kit Type | Convenience | Legally Marketed |
#### G. Bench/Performance Data/ Non-Clinical Testing:
Table 6.2: Applicable Standards and Performance Testing
| Standard | Standard Title | Revision/Date | Performance<br>Testing |
|-------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|---------------------------------------------------------------|
| ISO 10555-1 | Intravascular catheters --<br>sterile and single-use<br>intravascular catheters --<br>part 1: general<br>requirements. | Second Edition<br>2013-06-15 | Air Leak, Liquid Leak,<br>Peak Tensile Force,<br>Gravity Flow |
| ISO 11607-1 | Packaging for terminally<br>sterilized medical devices<br>- part<br>1: requirements for<br>materials, sterile barrier<br>systems and packaging<br>systems<br>[including: amendment 1<br>(2014)]. | First Edition<br>2006-04-15 | Transit and Shelf Life<br>testing |
| ISO 11607-2 | Packaging for terminally<br>sterilized medical devices<br>- part<br>2: validation requirements<br>for forming, sealing and<br>assembly processes<br>[including: amendment 1<br>(2014)]. | First Edition<br>2006-04-15 | Transit and Shelf Life<br>testing |
| ISTA 3A | Packaged products for<br>parcel delivery system<br>shipment 70 KG (150 lbs)<br>or less | 2008 | Transit Testing |
| ISO 594-1 | Conical fittings with a 6% | First edition | Gauging |
Medcomp®: Duo-Flow® Side x Side Double Lumen Catheter Section 6: 510(k) Site Summary
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the MedComp logo. The logo features a stylized red geometric shape above the word "medCOMP". The "med" portion of the word is in red, while the "COMP" portion is in black.
Harleysville, PA 19438
Tel: 215-256-4201
Fax. 215-256-1787
www.medcompnet.com
Medcomp®: Duo-Flow® Side x Side Double Lumen Catheter
| | (Luer) taper for syringes,<br>needles and certain other<br>medical equipment - Part<br>1: General requirements | 1986-06-15 | |
|-----------|-------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 594-2 | Conical fittings with a 6%<br>(luer) taper for syringes,<br>needles and<br>certain other medical<br>equipment - part 2: lock<br>fittings. | Second Edition<br>1998-09-01 | Liquid Leakage<br>Air Leakage<br>Separation Force<br>Unscrewing Torque<br>Ease of Assembly<br>Resistance to<br>Overriding<br>Stress Cracking |
#### H. Biocompatibility
The biocompatibility evaluation for the Duo-Flow® Side x Side Double Lumen Catheter was conducted in accordance with the FDA quidance document: Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing Within a Risk Management Process" and International Standard ISO 10993-1 "Biocompatibility Evaluation of Medical Devices - Part 1 Evaluation and Testing Within a Risk Management Process", as recognized by the FDA. The Duo-Flow® Side x Side Double Lumen Catheter met the biocompatibility requirements for externally communicating medical device intended to primarily be in contact with circulating blood for a prolonged contact duration (>24 hours to less than 30 days). The biological endpoints that were met as are follows:
#### CYTOTOXICITY
- o ISO 10993-5:2009, Biological Evaluation of Medical Devices Part 5: Tests for In Vitro Cytotoxicity
#### SENSITIZATION
- o ISO 10993-10:2010, Biological Evaluation of Medical Devices Part 10: Tests for Irritation and Skin Sensitization
#### IRRITATION OR INTRACUTANEOUS REACTIVITY
- · ISO 10993-10:2010, Biological Evaluation of Medical Devices Part 10: Tests for Irritation and Skin Sensitization
#### SYSTEMIC TOXICITY
- o ISO 10993-11:2006, Biological Evaluation of Medical Devices Part 11: Tests for Systemic Toxicity
#### PYROGENICITY
- ISO 10993-11:2006, Biological Evaluation of Medical Devices Part 11: 0 Tests for Systemic Toxicity, Annex F
#### SUBACUTE TOXICITY
Medcomp®: Duo-Flow® Side x Side Double Lumen Catheter Section 6: 510(k) Site Summary
6-4
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image contains the logo for medCOMP. The logo consists of a red geometric shape that resembles a stylized letter "M" or an abstract building. Below the geometric shape, the word "med" is written in red, lowercase letters, while "COMP" is written in black, uppercase letters. The overall design is clean and modern.
Harleysville, PA 19438
Tel: 215-256-4201
Fax. 215-256-1787
www.medcompnet.com
### Medcomp®: Duo-Flow® Side x Side Double Lumen Catheter
- O ISO 10993-11:2006, Biological Evaluation of Medical Devices Part 11: Tests for Systemic Toxicity
### GENOTOXICITY
- o ISO 10993-3:2014, Biological Evaluation of Medical Devices Part 3: Tests for Genotoxicity, Carcinogenicity, and Reproductive Toxicity
### IMPLANTATION
- · ISO 10993-6:2016, Biological Evaluation of Medical Devices Part 6: Test for Local Effects After Implantation
#### HEMOCOMPATIBILITY
- 0 ASTM F576:2017, Standard Practice for Assessment of Hemolytic Properties of Materials
- · ISO 10993-4:2002/A1:2006, Biological Evaluation of Medical Devices Part 4: Selection of Tests for Interactions with Blood
- ISO 10993-4:2017, Biological Evaluation of Medical Devices Part 4: O Selection of Tests for Interactions with Blood
### OTHER
- ISO 10993-18, 2005, Biological Evaluation of Medical Devices, Part 18: O Chemical Characterization of Materials
#### ـــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــ Summary of Substantial Equivalence
Based on the indications for use, design, safety and performance testing. the proposed device, Duo-Flow® Side x Side Double Lumen Catheter, raises no new questions of effectiveness compared to the predicate device and is substantially equivalent to the predicate device, Mahurkar™ Acute Dual Lumen Catheter (K955002).
Medcomp®: Duo-Flow® Side x Side Double Lumen Catheter Section 6: 510(k) Site Summary
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.