← Product Code [FDF](/productcode/FDF) · K192793

# Evis Exera III Colonovideoscope Olympus PCF-H190TL, Evis Exera III Colonovideoscope Olympus PCF-H190TI, Evis Exera III Colonovideoscope Olympus PCF-HQ190L, Evis Exera III Colonovideoscope Olympus PCF-HQ190I (K192793)

_Olympus Medical Systems Corp. · FDF · Jul 17, 2020 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K192793

## Device Facts

- **Applicant:** Olympus Medical Systems Corp.
- **Product Code:** [FDF](/productcode/FDF.md)
- **Decision Date:** Jul 17, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

This instrument is intended to be used with an Olympus video system center, light source, documentation equipment, monitor. EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. The EVIS EXERA III COLONOVIDEOSCOPE PCF-H190TL/I is indicated for use within the lower digestive tract (including the anus, rectum, sigmoid colon, colon, and ileocecal valve). This instrument is intended to be used with an Olympus video system center, endoscope position detecting unit, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. The EVIS EXERA III COLONOVIDEOSCOPE PCF-HO190L/I is indicated for use within the lower digestive tract (including the anus, rectum, sigmoid colon, colon, and ileocecal valve).

## Device Story

Colonovideoscope for lower digestive tract visualization; used with Olympus video system center, light source, monitor, and EndoTherapy accessories. Inputs: optical images of internal anatomy; outputs: real-time video feed for clinician viewing. Operates via insertion into lower GI tract; clinician controls distal tip angulation and stiffness. Features include White Light Imaging (WLI) and Narrow Band Imaging (NBI). NBI serves as adjunctive tool for real-time histological prediction of diminutive polyps using NICE classification. Used in clinical settings by physicians to perform diagnostic and therapeutic endoscopic procedures. Benefits include enhanced visualization and real-time assessment of polyp histology, aiding clinical decision-making during colonoscopy.

## Clinical Evidence

No prospective clinical trials performed. Evidence consists of a meta-analysis of independent and Olympus-sponsored clinical literature regarding NBI performance for differentiating diminutive colorectal polyps. Meta-analysis evaluated predictive accuracy of endoscopists using NICE classification criteria to distinguish neoplastic vs. non-neoplastic lesions.

## Technological Characteristics

Flexible endoscope with control, insertion, and connector sections. Imaging modes: WLI and NBI. Complies with ANSI/AAMI ES 60601-1, IEC 60601-2-18, and IEC 60601-1-2. Biocompatibility per ISO 10993-1. Features variable stiffness, forward water jet, and CCD imaging sensor.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- EVIS EXERA III VIDEO SYSTEM COLONOVIDEOSCOPE PCF-PH190L/I ([K131780](/device/K131780.md))
- EVIS EXERA III VIDEO SYSTEM COLONOVIDEOSCOPE CF-HQ190L/I ([K131780](/device/K131780.md))

## Reference Devices

- EVIS EXERA III VIDEO SYSTEM COLONOVIDEOSCOPE PCF-H190L/I ([K131780](/device/K131780.md))
- PENTAX VIDEO COLONOSCOPES EC-3490TLi ([K131855](/device/K131855.md))
- EVIS EXERA III VIDEO SYSTEM COLONOVIDEOSCOPE CF-H190L/I ([K131780](/device/K131780.md))
- EVIS EXERA III video system center CV-190 ([K131780](/device/K131780.md))
- EVIS EXERA III xenon light source CLV-190 ([K131780](/device/K131780.md))
- High definition LCD monitor OEV262H ([K102379](/device/K102379.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 17, 2020

Olympus Medical Systems Corp. % Lisa M. Boyle Regulatory Affairs Specialist II Olympus Corporation of the Americas 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610

Re: K192793

> Trade/Device Name: Evis Exera III Colonovideoscope Olympus PCF-H190TL Evis Exera III Colonovideoscope Olympus PCF-H190TI Evis Exera III Colonovideoscope Olympus PCF-HQ190L Evis Exera III Colonovideoscope Olympus PCF-HQ190I Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FDF, NWB Dated: September 26, 2019 Received: September 30, 2019

Dear Lisa M. Boyle:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for Shanil P. Haugen, PhD Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

#### K192793

Device Name

EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-H190TL/I EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-HQ190L/I

Indications for Use (Describe)

#### EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-H190TL/I

This instrument is intended to be used with an Olympus video system center, light source, documentation equipment, monitor. EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery.

The EVIS EXERA III COLONOVIDEOSCOPE PCF-H190TL/I is indicated for use within the lower digestive tract (including the anus, rectum, sigmoid colon, colon, and ileocecal valve).

#### EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-HQ190L/I

This instrument is intended to be used with an Olympus video system center, endoscope position detecting unit, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery.

The EVIS EXERA III COLONOVIDEOSCOPE PCF-HO190L/I is indicated for use within the lower digestive tract (including the anus, rectum, sigmoid colon, colon, and ileocecal valve).

Type of Use (Select one or both, as applicable)

| <span style="font-size: 10px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 10px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The font is sans-serif and the letters are closely spaced together. The word appears to be a logo or brand name. There is a registered trademark symbol to the right of the word.

Date Prepared: May 29, 2020

# 510(k) Summary

## 1. GENERAL INFORMATION

- 510(k) Submitter: OLYMPUS MEDICAL SYSTEMS CORP. 2951 Ishikawa-cho, Hachioji-shi, Tokyo, Japan 192-8507
- Contact Person: Lisa M. Boyle Olympus Corporation of the Americas 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610, USA Phone: 484-896-3676 Fax: 484-896-7128 Email: lisa.boyle@olympus.com
- Manufacturing site: Aizu Olympus Co., Ltd., 500 Muranishi, Niidera, Monden-machi, Aizuwakamatsu-shi, Fukushima 965-8520, Japan

## 2. DEVICE IDENTIFICATION

- Device Name EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-H190TL/I EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-HQ190L/I
- Model Name PCF-H190TL, PCF-H190TI PCF-HQ190L, PCF-H190I
- Common Name COLONOVIDEOSCOPE
- Regulation Number 21 CFR 876.1500
- Regulation Name Endoscope and accessories
- Regulatory Class II

510(k) Summary Page 1 of 8

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. There is a registered trademark symbol to the right of the word.

K192793

- Product Code FDF, NWB
- Gastroenterology/urology ■ Classification Panel

## 3. PREDICATE DEVICE

#### ■ Predicate device

#### PCF-H190TL/I

| Device name                                                  | 510(k) Submitter                 | 510(k) No. |
|--------------------------------------------------------------|----------------------------------|------------|
| EVIS EXERA III VIDEO SYSTEM<br>COLONOVIDEOSCOPE PCF-PH190L/I | OLYMPUS MEDICAL<br>SYSTEMS CORP. | K131780    |

#### PCF-HQ190L/I

| Device name                                                 | 510(k) Submitter                 | 510(k) No. |
|-------------------------------------------------------------|----------------------------------|------------|
| EVIS EXERA III VIDEO SYSTEM<br>COLONOVIDEOSCOPE CF-HQ190L/I | OLYMPUS MEDICAL<br>SYSTEMS CORP. | K131780    |

## ■ Reference device

#### PCF-H190TL/I

| Device name                                                 | 510(k) Submitter                 | 510(k) No. |
|-------------------------------------------------------------|----------------------------------|------------|
| EVIS EXERA III VIDEO SYSTEM<br>COLONOVIDEOSCOPE PCF-H190L/I | OLYMPUS MEDICAL<br>SYSTEMS CORP. | K131780    |
| PENTAX VIDEO COLONOSCOPES<br>EC-3490TLi                     | PENTAX MEDICAL<br>COMPANY        | K131855    |

#### PCF-HQ190L/I

| Device name                                                | 510(k) Submitter                 | 510(k) No. |
|------------------------------------------------------------|----------------------------------|------------|
| EVIS EXERA III VIDEO SYSTEM<br>COLONOVIDEOSCOPE CF-H190L/I | OLYMPUS MEDICAL<br>SYSTEMS CORP. | K131780    |
| PENTAX VIDEO COLONOSCOPES<br>EC-3490TLi                    | PENTAX MEDICAL<br>COMPANY        | K131855    |

## 4. DEVICE DESCRIPTION

## 1) General Description of the subject device

The EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-H190TL/I and

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are spaced closely together, and the word is centered in the image. There is a registered trademark symbol to the right of the letter "S". A thin yellow line is underneath the word.

PCF-HQ190L/I are intended to be used with an Olympus video system center, light source, endoscope position detecting unit (for PCF-HQ190L/I only), documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. The PCF-H190TL/I and PCF-HQ190L/I are indicated for use within the lower digestive tract (including the anus, rectum, sigmoid colon, colon, and ileocecal valve).

The following items are components and accessories to be marketed with the EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-H190TL/I and PCF-HQ190L/I.

- Single use combination cleaning brush (BW-412T)
- Injection tube (MH-946)
- Channel plug (MH-944)
- AW channel cleaning adapter (MH-948)
- Suction cleaning adapter (MH-856)
- Auxiliary water tube (MAJ-855)
- ETO cap (MB-156)

Available imaging modes are listed below:

| Available imaging modes |
|-------------------------|
| WLI                     |
| NBI                     |

# 2) Principle of Operation and Mechanism of Action

The EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-H190TL/I and PCF-HQ190L/I consist of three parts: the control section, the insertion section, and the connector section. The basic principle including user interface and operation for the procedure of the PCF-H190TL/I and PCF-HQ190L/I are identical to that of the predicate devices.

| Model        | Device Name                     | 510(k) No.                 |
|--------------|---------------------------------|----------------------------|
| PCF-H190TL/I | EVIS EXERA III COLONOVIDEOSCOPE | Part of this<br>submission |

## ■ Components list of PCF-H190TL/I

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The font is sans-serif and the letters are closely spaced together. The word is centered and takes up most of the frame. There is a registered trademark symbol to the right of the S.

#### Traditional 510(k) Notification COLONOVIDEOSCOPE PCF-H190TL/I COLONOVIDEOSCOPE PCF-HQ190L/I

| MH-856  | Suction cleaning adapter              | K131780                          |
|---------|---------------------------------------|----------------------------------|
| MH-948  | AW channel cleaning adapter           | K131780                          |
| MB-156  | ETO cap                               | K131780                          |
| MH-944  | Channel plug                          | K131780                          |
| MH-946  | Injection tube                        | K131780                          |
| MAJ-855 | Auxiliary water tube                  | K131780                          |
| BW-412T | Single use combination cleaning brush | 510(k) exempt<br>(876.1500, MNL) |

## ■ Components list of PCF-HQ190L/I

| Model        | Device Name                           | 510(k) No.                       |
|--------------|---------------------------------------|----------------------------------|
| PCF-HQ190L/I | EVIS EXERA III COLONOVIDEOSCOPE       | Part of this<br>submission       |
| MH-856       | Suction cleaning adapter              | K131780                          |
| MH-948       | AW channel cleaning adapter           | K131780                          |
| MB-156       | ETO cap                               | K131780                          |
| MH-944       | Channel plug                          | K131780                          |
| MH-946       | Injection tube                        | K131780                          |
| MAJ-855      | Auxiliary water tube                  | K131780                          |
| BW-412T      | Single use combination cleaning brush | 510(k) exempt<br>(876.1500, MNL) |

#### Components list in the complete system for PCF-H190TL/I and PCF-HQ190L/I

| Model   | Device Name                        | 510(k) No. |
|---------|------------------------------------|------------|
| CV-190  | EVIS EXERA III video system center | K131780    |
| CLV-190 | EVIS EXERA III xenon light source  | K131780    |
| OEV262H | High definition LCD monitor        | K102379    |

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are slightly blurred, giving the image a soft focus. There is a thin, horizontal, gradient line below the word, transitioning from yellow to white.

## 5. INDICATIONS FOR USE

## ■ EVIS EXERA III COLONOVIDEOSCOPE PCF-H190TL/I

This instrument is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, Endotherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. The EVIS EXERA III COLONOVIDEOSCOPE PCF-H190TL/I is indicated for use within the lower digestive tract (including the anus, rectum, sigmoid colon, colon, and ileocecal valve).

## ■ EVIS EXERA III COLONOVIDEOSCOPE PCF-HQ190L/I

This instrument is intended to be used with an Olympus video system center, endoscope position detecting unit, light source, documentation equipment, monitor, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery.

The EVIS EXERA III COLONOVIDEOSCOPE PCF-HQ190L/I is indicated for use within the lower digestive tract (including the anus, rectum, sigmoid colon, colon, and ileocecal valve).

# 6. COMPARISON OF TECHNOLOGY CHARACTERISTICS WITH THE PREDICATE DEIVCE

The EVIS EXERA III VIDEO SYSTEM COLONOVIDEOSCOPE PCF-H190TL/I has the same technological characteristics and design as the predicate devices except for the following new features:

- Bending section UP Angulation
- Forward water jet
- Insertion section variable stiffness
- Outer Diameter of Distal End
- Outer Diameter of Insertion Tube
- Inner structure of Air/Water nozzle

The EVIS EXERA III VIDEO SYSTEM COLONOVIDEOSCOPE PCF-HQ190L/I has the same technological characteristics and design as the predicate device except for the following new features:

- Outer Diameter of Distal End
- Outer Diameter of Insertion Tube
- Inner Diameter of Instrument Channel
- Inner structure of Air/Water nozzle
- Downsizing of CCD

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. There is a small registered trademark symbol to the right of the word.

K192793 Page 6 of 8

All other technological characteristics of both the subject and predicate devices are identical. Validation from non-clinical testing demonstrated that these technological features do not raise any new issues of safety or effectiveness of the subject device.

## 7. PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination.

#### 1) Reprocessing validation testing

Reprocessing instruction and reprocessing method validation testing for the EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-H190TL/I and PCF-HQ190L/I were conducted and documentation was provided as recommended by Guidance for Industry and Food and Drug Administration Staff, "Reprocessing Medical Devices in Health Care Setting: Validation Methods and Labeling".

#### 2) Biocompatibility testing

Biocompatibility testing for the EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-H190TL/I and PCF-HQ190L/I were conducted in accordance with the FDA's Guidance for Industry and Food and Drug Administration Staff, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The biocompatibility testing included the following tests:

- Cytotoxicity Study Using the Colony Assay
- Intracutaneous Study in Rabbits
- Guinea Pig Maximization Sensitization Test

## 3) Software verification and validation testing

Software verification and validation testing for the EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-H190TL/I and PCF-HQ190L/I were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices".

## 4) Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-H190TL/I and PCF-HQ190L/I. The system complies with the ANSVAAMI ES 60601-1:2005/(R)2012 and A1:2012 and IEC 60601-2-18:2009 standards for safety and the IEC 60601-1-2:2014 standards for EMC.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The font is sans-serif and the letters are closely spaced together. There is a registered trademark symbol to the right of the letter "S".

## 5) Performance testing - Bench

Bench testing for the EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-H190TL/I and PCF-HQ190L/I as listed below was conducted to ensure that the subject device performs as intended and meet design specifications.

- Thermal Safety Test
- Composite Durability Test
- Photobiological Safety Test
- Retroflexed Withdrawal and Detection of Hidden Polyps (for PCF-H190TL/L only)

#### 6) Performance testing - Animal

No animal study was performed to demonstrate substantial equivalence.

#### 7) Performance testing - Clinical

A meta-analysis of the clinical performance of NBI for differentiation of diminutive colorectal polyps has been provided in Section 20.1. The clinical literature represents independent studies performed by clinicians with the exception of two Olympus-sponsored studies. The objective of the meta-analysis was to determine the predictive accuracy of endoscopists utilizing NBI and the NBI International Colorectal Endoscopic (NICE) classification or all available classification criteria in general to differentiate diminutive colorectal polyps as neoplastic or non-neoplastic lesions.

The evidence being presented in this premarket notification supports our labeling change of the subject device for the use of Narrow Band Imaging as an adjunctive tool to assist the endoscopist in making predictions of the underlying histology of diminutive polyps. This application does not seek to show superiority or non-inferiority of NBI to the gold standard histopathology.

The results show that pooled data from NBI-experienced and inexperienced endoscopists making high confidence predictions of diminutive polyp histology in real-time provide reasonable certainty of neoplastic vs. non-neoplastic identity. The NICE classification, simplified criteria for distinguishing neoplastic from non-neoplastic polyps, helps the endoscopist achieve similar performance.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are evenly spaced and appear to be a sans-serif font. A thin, horizontal, yellow line is located underneath the word. 

#### 8. CONCLUSIONS

Based on the indications for use, technological characteristics, performance testing and technological comparison to the predicate devices, the EVIS EXERA III COLONOVIDEOSCOPE OLYMPUS PCF-H190TL/I and PCF-HQ190L/I raise no new issue of safety and effectiveness and are substantially equivalent to the predicate devices in terms of safety, effectiveness and performance.

---

**Source:** [https://fda.innolitics.com/device/K192793](https://fda.innolitics.com/device/K192793)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
