K192791 · Edan Instruments, Inc. · IYO · Nov 27, 2019 · Radiology
Device Facts
Record ID
K192791
Device Name
Diagnostic Ultrasound System
Applicant
Edan Instruments, Inc.
Product Code
IYO · Radiology
Decision Date
Nov 27, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1560
Device Class
Class 2
Indications for Use
The Acclarix AX3 series Diagnostic Ultrasound System is intended for use by a qualified physician or allied health professional for ultrasound evaluations in hospitals and clinical applications include: Abdominal, Gynecology, Obstetric, Cardiac, Small parts, Urology, Musculoskeletal, Peripheral vascular, Adult Cephalic.
Device Story
Acclarix AX3 Series is a software-controlled diagnostic ultrasound system consisting of a main unit and various transducers. It acquires and displays ultrasound data in multiple imaging modes. Used in hospitals and clinics by physicians or allied health professionals to perform diagnostic evaluations. The system processes acoustic signals received via transducers to generate real-time images for clinical assessment. Healthcare providers view these images on the system display to aid in diagnosis and clinical decision-making. The device benefits patients by providing non-invasive diagnostic imaging for various anatomical regions and clinical conditions.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Software-controlled diagnostic ultrasound system. Complies with ANSI/AAMI ES60601-1, IEC 60601-1-2, and IEC 60601-2-37. Biocompatibility testing per ISO 10993-1, ISO 10993-5, and ISO 10993-10. Acoustic output meets FDA guidelines. System supports various transducers including L17-7HO and L17-70.
Indications for Use
Indicated for ultrasound evaluations in hospitals and clinics for abdominal, gynecological, obstetric, cardiac, small parts, urological, musculoskeletal, peripheral vascular, and adult cephalic applications. Intended for use by qualified physicians or allied health professionals.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Edan Instruments, Acclarix AX4 Diagnostic Ultrasound System (K171900)
Reference Devices
Edan Instruments, Acclarix U50 Diagnostic Ultrasound System (K173003)
Edan Instruments, Acclarix AX8 Diagnostic Ultrasound System (K180862)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 27, 2019
Edan Instruments, Inc. % Stella Guo Regulatory Engineer #15 Jinhui Road, Jinsha Community, Kengzi Sub-District Pingshan District Shenzhen, Guangdong 518122 CHINA
Re: K192791
Trade/Device Name: Acclarix AX3 Series Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: Class II Product Code: IYO. ITX. IYN Dated: September 25, 2019 Received: September 30, 2019
Dear Stella Guo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
#### K192791
#### Device Name
Acclarix AX3 Series Diagnostic Ultrasound System (including Acclarix AX3 Exp/Acclarix AX3 Super/Acclarix AX25/Acclarix AX28/Acclarix AX2/Acclarix AX2 Exp/Acclarix AX2 Super/Acclarix AX15/Acclarix AX18)
#### Indications for Use (Describe)
The Acclarix AX3 series Diagnostic Ultrasound System is intended for use by a qualified physician or allied health professional for ultrasound evaluations in hospitals and clinical applications include:
- □ Abdominal
- □ Gynecology
- Obstetric
- Cardiac
- □ Small parts
- □ Urology
- □ Musculoskeletal
- □ Peripheral vascular
- □ Adult Cephalic
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{4}------------------------------------------------
| | 510(k) Summary | K192791 |
|-----------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| | Prepared in accordance with the requirements of 21 CFR Part 807.92 | |
| <b>1. Submitter:</b> | Edan Instruments, Inc.<br>#15 Jinhui Road, Jinsha Community, Kengzi Sub-District, Pingshan District,<br>Shenzhen, 518122 P.R.China.<br>Tel: +86-755-2685 8736 | Fax: +86-755-2689 8330 |
| <b>Contact person:</b> | Stella Guo | |
| <b>Preparing date:</b> | September 25, 2019 | |
| <b>2. Device name and<br/>classification:</b> | Device Name: Diagnostic Ultrasound System | |
| | Model: Acclarix AX3, Acclarix AX3 Exp, Acclarix AX3 Super, Acclarix | |
| | AX25, Acclarix AX28, Acclarix AX2, Acclarix AX2 Exp, Acclarix AX2 Super<br>Acclarix AX15, Acclarix AX18 | |
| | Classification Name/ Product code: | |
| | 21 CFR 892.1550 System, Imaging, Pulsed Doppler, Ultrasonic / IYN | |
| | 21 CFR 892.1560, Ultrasonic, Pulsed echo, Imaging / IYO | |
| | 21 CFR 892.1570, Transducer, Ultrasonic, Diagnostic / ITX | |
| | Regulatory Class: Class II | |
| <b>3.Premarket Notification<br/>Class III Certification<br/>and Summary</b> | Not applicable, the subject device is Class II. | |
| <b>4. Predicate Device(s):</b> | 1) Edan Instruments, Acclarix AX4 Diagnostic Ultrasound System, cleared<br>under K171900 (Primary) | |
| | 2) Edan Instruments, Acclarix U50 Diagnostic Ultrasound System, cleared<br>under K173003 (Reference) | |
| | 3) Edan Instruments, Acclarix AX8 Diagnostic Ultrasound System, cleared<br>under K180862 (Reference) | |
| <b>5. Reason for<br/>Submission</b> | By submission of the Traditional 510(k), Edan Instruments is requesting<br>clearance for new devices Acclarix AX3 Series Diagnostic Ultrasound System | |
| <b>6.Pre-Submission</b> | Not applicable, there is no pre-submission. | |
IDE
{5}------------------------------------------------
- 7. Device Description: Acclarix AX3 Series is a software controlled Diagnostic Ultrasound System, which consists of a main unit along with associated transducers. It is intended for use by a qualified physician or allied health professional for ultrasound evaluations. This system is a Track 3 device to acquire and display ultrasound data in various imaging modes.
- 8. Indication for Use The Acclarix AX3 series Diagnostic Ultrasound System is intended for use by a qualified physician or allied health professional for ultrasound evaluations in hospitals and clinics. General clinical applications include:
- Abdominal ●
- . Gynecology
- Obstetric ●
- Cardiac ●
- Small Parts ●
- Urology
- . Musculoskeletal
- Peripheral Vascular ●
- Adult Cephalic ●
### 9. Predicate Device Comparison
The subject device Acclarix AX3 Series is same as the primary predicated device Acclarix AX4 (K171900) in items such as:
- 1) Intended Use/ Indications for Use;
- 2) Mode of Operations;
- 3) Transducer Types
- 4) Acoustic Output, which are below the limits of FDA;
The differences of subject device Acclarix AX3 Series with the predicated devices are described as below:
- 1) The subject device Acclarix AX3 Series is not intended to do Intra-operative. Pediatric and Neonatal clinical ultrasound evaluation, which are already cleared in the primary predicated device Acclarix AX4 (K171900).
{6}------------------------------------------------
- 2) The subject device Acclarix AX3 Series does not support 3D/4D, which are already cleared in the primary predicated device Acclarix AX4 (K171900).
- 3) The subject device does not support four transducers that are already cleared in the primary predicated device Acclarix AX4 (K171900)
- 4) The subject device additionally support two new transducers L17-7HO and L17-70. L17-7HO is already cleared in the reference predicate device Acclarix AX8 (K180862). And L17-70 is modified from the transducer L15-7b cleared by the reference predicate device U50 (K173003) by only changing the shape of the transducer connector.
The subject and predicate devices have similar design features and performance specifications. The technological differences between the subject and predicate devices do not raise different questions of safety or effectiveness.
# 10. Performance Data:
# Clinical test:
Clinical testing is not required.
## Non-clinical test:
The Acclarix AX3 Series Diagnostic Ultrasound System comply with:
(1) ANSI/AAMI ES60601-1 Electrical Safety
(2) IEC 60601-1-2 Electromagnetic Compatibility
(3) IEC 60601-2-37 Requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
(4) Acoustic output testing as per the guideline "Marketing Clearance of Diagnostic Ultrasound Systems and
Transducers" dated June 27, 2019.
The following biocompatibility standards are conducted on the subject device:
(5) ISO 10993-1, ISO 10993-5 and ISO 10993-10
The tests were selected to show substantial equivalence between the subject device and the predicate.
The non-clinical performance testing showed that the subject devices are as safe and as effective as the predicate devices.
# 11. Conclusion
Verification and validation testing has been conducted on the Acclarix AX3 Series Diagnostic Ultrasound System. This premarket notification submission demonstrates that Acclarix AX3 Series Diagnostic Ultrasound System are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.