C4 Fiducial Marker

K192771 · C4 Imaging, LLC · NEU · Oct 21, 2020 · General, Plastic Surgery

Device Facts

Record IDK192771
Device NameC4 Fiducial Marker
ApplicantC4 Imaging, LLC
Product CodeNEU · General, Plastic Surgery
Decision DateOct 21, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4300
Device ClassClass 2

Indications for Use

The C4 Fiducial Marker is indicated to be implanted into the body in situations where the location of specific anatomy, normal or diseased, needs to be marked for a future medical procedure. The device can be visualized using MRI, CT or x-ray, and provides a reference from which other procedures can be guided.

Device Story

C4 Fiducial Marker is an implantable radiographic marker used to identify specific anatomy for future medical procedures. Device consists of a sealed PEEK polymer tube containing a solution of cobalt chloride and N-Acetylcysteine, with zirconium oxide sealed within the shell. It is implanted into the body by a physician. The marker is visualized using MRI, CT, or x-ray imaging modalities. By providing a stable reference point, it assists clinicians in guiding subsequent medical procedures, potentially improving accuracy and patient outcomes. It is a passive device; no software or active processing is involved.

Clinical Evidence

Bench testing only. Performance testing included structural and functional integrity (negative/positive pressure, axial/lateral load), MRI, CT, and x-ray imaging in tissue-equivalent phantoms, biocompatibility assessment per ISO 10993, and endotoxin testing.

Technological Characteristics

Materials: PEEK polymer shell, cobalt chloride (up to 1%), N-Acetylcysteine (up to 2%), zirconium oxide. Dimensions: 5.5-10.0 mm length, 1.0 mm (+/- 0.2 mm) diameter. Passive implantable marker. Visualization: MRI, CT, x-ray. Biocompatibility: ISO 10993 compliant.

Indications for Use

Indicated for patients requiring anatomical marking of normal or diseased tissue for future medical procedures; implantable for visualization via MRI, CT, or x-ray to guide subsequent interventions.

Regulatory Classification

Identification

An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that. October 21, 2020 C4 Imaging LLC Andrew Bright President and CEO 196 West Ashland Street Doylestown, Pennsylvania 18901 Re: K192771 Trade/Device Name: C4 Fiducial Marker Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable Clip Regulatory Class: Class II Product Code: NEU Dated: September 26, 2019 Received: September 30, 2019 Dear Mr. Bright: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for {1}------------------------------------------------ devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Cindy Chowdhury, Ph.D., M.B.A. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K192771 Device Name C4 Fiducial Marker Indications for Use (Describe) The C4 Fiducial Marker is indicated to be implanted into the body in situations where the location of specific anatomy, normal or diseased, needs to be marked for a future medical procedure. The device can be visualized using MRI, CT or x-ray, and provides a reference from which other procedures can be guided. | Type of Use (Select one or both, as applicable) | | |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--| | <div> <span> <svg height="12" width="12"> <polygon points="0,0 12,0 12,12 0,12" style="fill:none;stroke:black;stroke-width:1"></polygon> <polyline points="0,0 12,12" style="fill:none;stroke:black;stroke-width:1"></polyline> <polyline points="12,0 0,12" style="fill:none;stroke:black;stroke-width:1"></polyline> </svg> </span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | | | <div> <span> <svg height="12" width="12"> <polygon points="0,0 12,0 12,12 0,12" style="fill:none;stroke:black;stroke-width:1"></polygon> </svg> </span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> | | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary [as required by section 807.92(c)] This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92. ## General Information | Submitted by: | C4 Imaging, LLC<br>196 West Ashland Street<br>Suite 109<br>Doylestown, PA 18901 USA<br>Phone: +1 (609) 933-5895<br>Email: abright@c4imaging.com<br><br>Contact Person: Andrew Bright | |----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Date Prepared: | October 20, 2019 | # Device Name | Trade Name: | C4 Fiducial Marker | |-----------------|--------------------------| | Common Name(s): | Anatomical tissue marker | # Classification | Regulation: | 21 CFR §878.4300 | |----------------------|-----------------------------------| | Class: | Class II | | Product Code: | NEU | | Classification name: | Marker, Radiographic, Implantable | {4}------------------------------------------------ ## Predicate Devices | Mixed media marker | K102506 | Cortex Manufacturing | |-------------------------|---------|--------------------------------| | BiomarC Fiducial Marker | K063193 | Carbon Medical<br>Technologies | #### Device Description The C4 Fiducial Marker consists of a sealed polyether ether ketone (PEEK) polymer tube containing a solution of up to 1% cobalt chloride and up to 2% N-Acetylcysteine, The ratio of cobalt chloride to N-Acetylcysteine is 1:2. Zirconium oxide is sealed within the polymer shell of the device. The length of the polymer capsule is between 5.5 mm and 10.0 mm, and the diameter is 1.0 mm (+/- 0.2 mm). #### Indication The C4 Fiducial Marker is indicated to be implanted into the body in situations where the location of specific anatomy, normal or diseased, needs to be marked for a future medical procedure. The device can be visualized using MRI, CT or x-ray, and provides a reference from which other procedures can be guided. ## Comparison to Predicate Devices The indication and technological characteristics of the C4 Fiducial Marker are substantially equivalent to the Mixed Media Marker (Cortex Manufacturing -K102506) and the BiomarC Fiducial Marker (Carbon Medical Technologies -K063193). Like the Mixed Media Marker, the C4 Fiducial Marker is comprised of a sealed polyether ether ketone (PEEK) capsule that encloses an internal marker for MRI. CT and X-ray imaging. In both cases the PEEK capsule is the only patient contacting material. Like the BiomarC Fiducial, the C4 Fiducial Marker has a sealed biocompatible shell that encloses a zirconium oxide marker. The C4 Fiducial Marker was also compared to a reference device. the C4 Imaging MRI Marker NS (K171487), which has similar technological characteristics. The C4 Fiducial Marker and the MRI Marker NS both comprise a PEEK shell enclosing a cobalt chloride N-acetyl cysteine solution. A risk evaluation, based on a Failure Mode Effect Analysis (FEMA), determined that technological differences were not considered to have an impact on the safety or effectiveness of the C4 Fiducial Marker. Adoption evaluation of biocompatibility, MR compatibility, sterilization, structural integrity, packaging and shelf life data during the risk assessment confirmed the C4 Fiducial Marker is substantially equivalent to the predicate devices and reference device. {5}------------------------------------------------ # Performance Testing The C4 Fiducial Marker was subject to structural and functional integrity testing involving negative and positive pressure, as well as axial and lateral load testing. These integrity tests were followed by MRI, CT and x-ray imaging demonstrating functional performance. Separate MRI, CT and x-ray imaging tests were performed in tissue equivalent phantoms. A biocompatibility (ISO 10993) assessment and endotoxin tests were performed on the device. ### Conclusion The C4 Fiducial Marker is substantially equivalent to the Mixed Media Marker (K102506) and to the BiomarC Fiducial Marker (K063193). There are no differences that would affect the substantial equivalence of the proposed device
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...