CALCIUM-CRESOLPHTHALEIN: Reagent for the measurement of calcium concentration in human serum, plasma or urine. The obtained values are useful as an aid in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms). This reagent is for use in the BioSystems BA analyzers. Only for in vitro use in the clinical laboratory. GLUCOSE: Reagent for the measurement of glucose concentration in human serum and plasma. The obtained values are useful as an aid in the diagnosis and treatment of the diabetes mellitus. This reagent is for use in the BioSystems BA analyzers. Only for in vitro use in the clinical laboratory. UREA/BUN - UV: Reagent for the measurement of urea concentration in human serum, plasma or urine. The obtained values are useful as an aid in the diagnosis of certain renal and metabolic diseases. This reagent is for use in the BioSystems BA analyzers. Only for in vitro use in the clinical laboratory.
Device Story
Quantitative photometric assays for calcium, glucose, and urea/BUN; reagents provided ready-to-use for BioSystems BA400 analyzers. Calcium measured via o-cresolphthalein complexone reaction; glucose via glucose oxidase/peroxidase reaction; urea via urease/glutamate dehydrogenase reaction. Analyzers perform spectrophotometric measurement of colored complexes. Used in clinical laboratories by trained personnel. Output provides analyte concentration values; assists clinicians in diagnosing/monitoring metabolic, renal, and endocrine conditions. No wireless or external data transmission capabilities.
Clinical Evidence
Bench testing only. Precision studies (n=80 per level) showed CVs ranging from 0.2% to 5.0%. Linearity (r² ≥ 0.999) and detection limits (LoB, LoD, LoQ) were established per CLSI guidelines. Method comparison against predicate devices (n=77-153 samples) showed high correlation (R ≥ 0.992). Interference testing confirmed performance within 10% threshold for common endogenous and exogenous substances.
Technological Characteristics
Photometric assays; reagents include enzymes (glucose oxidase, peroxidase, urease, glutamate dehydrogenase) and chemical complexes (o-cresolphthalein). Dimensions/form factor: ready-to-use liquid reagents. Energy source: spectrophotometric light source in BA400 analyzer. Connectivity: standalone, no wireless/network transmission. Sterilization: not applicable (reagents).
Indications for Use
Indicated for measurement of calcium, glucose, and urea/BUN in human serum, plasma, or urine to aid in diagnosis/treatment of parathyroid, bone, renal, metabolic diseases, tetany, and diabetes mellitus. For professional in vitro clinical laboratory use only.
Regulatory Classification
Identification
A calcium test system is a device intended to measure the total calcium level in serum. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).
Submission Summary (Full Text)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K192730
B Applicant
BioSystems S.A.
C Proprietary and Established Names
CALCIUM-CRESOLPHTHALEIN, GLUCOSE, UREA/BUN-UV
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| CHW | Class II | 21 CFR 862.1145 - Calcium Test System | CH - Clinical Chemistry |
| CGA | Class II | 21 CFR 862.1345 - Glucose test system | CH - Clinical Chemistry |
| CDQ | Class II | 21 CFR 862.1770 - Urea nitrogen test system | CH - Clinical Chemistry |
## II Submission/Device Overview:
A Purpose for Submission:
New device
B Measurand:
Calcium, Glucose and Urea/BUN (blood urea nitrogen)
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
C Type of Test:
Calcium, Glucose and Urea/BUN are quantitative photometric assays.
III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
CALCIUM-CRESOLPHTHALEIN: Reagent for the measurement of calcium concentration in human serum, plasma or urine. The obtained values are useful as an aid in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms). This reagent is for use in the BioSystems BA analyzers. Only for in vitro use in the clinical laboratory.
GLUCOSE: Reagent for the measurement of glucose concentration in human serum and plasma. The obtained values are useful as an aid in the diagnosis and treatment of the diabetes mellitus. This reagent is for use in the BioSystems BA analyzers. Only for in vitro use in the clinical laboratory.
UREA/BUN - UV: Reagent for the measurement of urea concentration in human serum, plasma or urine. The obtained values are useful as an aid in the diagnosis of certain renal and metabolic diseases. This reagent is for use in the BioSystems BA analyzers. Only for in vitro use in the clinical laboratory.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
D Special Instrument Requirements:
BioSystems BA400
IV Device/System Characteristics:
A Device Description:
Reagents are provided ready to use.
CALCIUM-CRESOLPHTHALEIN: Reagent for the measurement of calcium concentration in human serum, plasma or urine. The CALCIUM-CRESOLPHTHALEIN assay consists of the following reagents:
- Ethanolamine 900 mmol/L.
K192730 - Page 2 of 11
{2}
- Reagent: O-Cresolphthalein Complexone 0.3 mmol/L, 8 hydroxyquinoline 28 mmol/L, hydrochloric acid 100 mmol/L.
The GLUCOSE assay consists of the following reagents:
Phosphate 100 mmol/L, phenol 5 mmol/L, glucose oxidase > 10 U/mL, peroxidase >1 U/mL, 4-aminoantipyrine 0.4 mmol/L, pH 7.5
The UREA/BUN - UV assay consists of the following reagents:
- Tris 100 mmol/L, 2-oxoglutarate 5.6 mmol/L, urease > 140 U/mL, glutamate dehydrogenase > 140 U/mL, ethyleneglicol 220 g/L, sodium azide 0.95, pH 8.0.
- NADH 1.5 mmol/L, sodium azide 9.5 g/L.
## B Principle of Operation:
The CALCIUM-CRESOLPHTHALEIN reagent is a photometric assay in which Calcium in the sample reacts with o-cresolphthalein complexone (o-CPC) forming a colored complex that can be measured by spectrophotometry.
The GLUCOSE reagent is a photometric assay in which glucose is oxidized by glucose-oxidase to gluconate and hydrogen peroxide. The hydrogen peroxide is then oxidized by peroxidase in the presence of 4-aminoantipyrine and phenol to form a colored complex that can be measured by spectrophotometry.
The UREA-BUN/UV reagent is a photometric assay in which BUN-Urea is hydrolyzed by urease to produce ammonia and carbon dioxide. The ammonia reacts with 2-oxoglutarate in the presence of NADH to yield glutamate and NAD+.
| Modes of Operation | Yes | No |
| --- | --- | --- |
| Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device? | ☐ | ☑ |
| Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission? | ☐ | ☑ |
| Software | | |
| FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types. | ☐ | ☑ |
## V Substantial Equivalence Information:
A Predicate Device Name(s):
CALCIUM GEN. 2
GLUCOSE HK GEN. 3
Urea/BUN
B Predicate 510(k) Number(s):
K100853
K192730 - Page 3 of 11
{3}
Comparison with Predicate(s):
| Device & Predicate Device(s): | K192730 | K100853 |
| --- | --- | --- |
| Device Trade Name | CALCIUM-CRESOLPHTHALEIN | CALCIUM GEN. 2 |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | Reagent for the measurement of calcium concentration | Same |
| Measurand | Calcium | Same |
| Sample type | Serum, Lithium Heparin plasma or urine. | Same |
| General Device Characteristic Differences | | |
| Method | O-CRESOLPHTHALEIN COMPLEXONE | NM-BAPTA |
| Measuring range | Serum/plasma: 0.80 - 20 mg/dL
Urine: 3.97-30 mg/dL | Serum/plasma: 0.80 - 20.1 mg/dL
Urine: 0-140 mg/dL |
| Device & Predicate Device(s): | K192730 | K100853 |
| --- | --- | --- |
| Device Trade Name | GLUCOSE | GLUCOSE HK GEN. 3 |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | Reagent for the measurement of glucose concentration | Same |
| Measurand | Glucose | Same |
| General Device Characteristic Differences | | |
| Method | Oxidase | Hexokinase |
| Sample type | Serum or Lithium Heparin plasma | Serum, plasma, urine or cerebrospinal fluid |
K192730 - Page 4 of 11
{4}
| Device & Predicate Device(s): | K192730 | K100853 |
| --- | --- | --- |
| Measuring range | Serum/plasma: 2.0 - 500 mg/dL | Serum/plasma: 2.0 - 750 mg/dL |
| Device & Predicate Device(s): | K192730 | K100853 |
| --- | --- | --- |
| Device Trade Name | UREA-BUN/UV | Urea/BUN |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | Reagent for the measurement of urea concentration | Same |
| Measurand | Urea | Same |
| Sample type | Serum, Lithium Heparin plasma or urine. | Same |
| Method | Urease/Glutamate Dehydrogenase | Same |
| General Device Characteristic Differences | | |
| Measuring range | Serum/plasma: 10 mg/dL–240 mg/dL
Urine: 635-12000 mg/dL | Serum/plasma: 3.0-240 mg/dL
Urine: 6-12000 mg/dL |
VI Standards/Guidance Documents Referenced:
CLSI - EP05-A3: Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline-Third Edition.
CLSI - EP06-A: Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline.
CLSI - EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline-Second Edition.
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
Internal precision studies were performed according to the CLSI EP5-A3 guideline. For CALCIUM-CRESOLPHTHALEIN, GLUCOSE and UREA/BUN - UV, three levels of commercial biochemistry control serum (human serum matrix) Level I, II and III or Urine,
K192730 - Page 5 of 11
{5}
Level I, II and III were tested on two runs per day, two measurements per run, for twenty days with a total of 80 observations using the BioSystems BA400 analyzer. The results of the precision studies are summarized in the tables below:
CALCIUM-CRESOLPHTHALEIN:
| Matrix | Samples | n | Mean (mg/dL) | Repeatability | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | SD | CV% | SD | CV% |
| Serum | Level 1 | 80 | 5.10 | 0.1 | 1.6 | 0.1 | 2.9 |
| | Level 2 | 80 | 10.0 | 0.2 | 2.2 | 0.3 | 2.7 |
| | Level 3 | 80 | 14.0 | 0.1 | 1.0 | 0.3 | 1.9 |
| Urine | Level 1 | 80 | 7.95 | 0.2 | 1.9 | 0.3 | 3.3 |
| | Level 2 | 80 | 15.5 | 0.2 | 1.2 | 0.4 | 2.3 |
| | Level 3 | 80 | 26.6 | 0.3 | 1.0 | 0.9 | 3.5 |
GLUCOSE:
| Matrix | Samples | n | Mean (mg/dL) | Repeatability | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | SD | CV% | SD | CV% |
| Serum | Level 1 | 80 | 48.6 | 0.3 | 0.7 | 0.4 | 0.9 |
| | Level 2 | 80 | 211 | 0.9 | 0.4 | 2.2 | 1.0 |
| | Level 3 | 80 | 389 | 0.9 | 0.2 | 2.4 | 0.6 |
UREA/BUN - UV:
| Matrix | Samples | n | Mean (mg/dL) | Repeatability | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | SD | CV% | SD | CV% |
| Serum | Level 1 | 80 | 26.8 | 1.0 | 3.6 | 1.3 | 5.0 |
| | Level 2 | 80 | 137 | 1.5 | 1.1 | 2.3 | 1.7 |
| | Level 3 | 80 | 172 | 2.2 | 1.3 | 3.4 | 2.0 |
| Urine | Level 1 | 80 | 1003 | 39.5 | 3.9 | 45.5 | 4.5 |
| | Level 2 | 80 | 1918 | 44.6 | 2.3 | 59.5 | 3.1 |
| | Level 3 | 80 | 7293 | 87.3 | 1.2 | 180 | 2.5 |
2. Linearity:
Linearity studies were conducted according to the CLSI EP06-A guideline. Eleven levels of dilutions were prepared by mixing different proportions of samples with high and low concentration of each analyte. Samples were analyzed on one BA400 instrument in duplicate. The results of the regression analysis are shown below:
| Analyte | Matrix | Slope | Intercept | r2 | Range Conc. Tested (mg/dL) | Claimed range (mg/dL) |
| --- | --- | --- | --- | --- | --- | --- |
| CALCIUM-CRESOLPH-THALEIN | Serum | 1.004 | -0.42 | 0.999 | 0.0 – 20.0 | 0.80 - 20.0 |
| | Urine | 1.010 | -0.119 | 1.000 | 0.0 – 30.0 | 3.97 – 30.0 |
K192730 - Page 6 of 11
{6}
| Analyte | Matrix | Slope | Intercept | r² | Range Conc. Tested (mg/dL) | Claimed range (mg/dL) |
| --- | --- | --- | --- | --- | --- | --- |
| GLUCOSE | Serum | 1.002 | 2.61 | 1.000 | 0.0 - 500 | 2.0 - 500 |
| UREA/BUN - UV | Serum | 1.004 | -0.42 | 0.999 | 0.0 - 240 | 10.0 - 240 |
| | Urine | 0.996 | 0.19 | 0.999 | 0.0 - 12000 | 635 - 12000 |
The linearity studies support the sponsor’s claimed measuring ranges as described in the table above.
## 3. Analytical Specificity/Interference:
Interference studies were performed for CALCIUM-CRESOLPHTHALEIN, GLUCOSE and UREA/BUN – UV. Specimens consisted of human serum and urine representing low and high concentrations of the analyte, with and without the substance tested for interference. Serum samples for the CALCIUM-CRESOLPHTHALEIN assay were tested at concentrations of approximately 9.7 and 13.8 mg/dL, and urine samples were tested at concentrations of 5.7 and 14.9 mg/dL. Serum samples for the GLUCOSE assay were tested at concentrations of approximately 87 and 219 mg/dL. Serum samples for the UREA/BUN - UV assay were tested at concentrations of approximately 21.9 and 117 mg/dL, and urine samples were tested at concentrations of 866 and 1638 mg/dL. Samples were spiked with increasing concentrations of the potential interferents and analyzed in duplicate using the BioSystems BA400 analyzer. Interference was calculated as: (b-a)/a, where a and b are the concentration of measurand found in samples without (a) and with (b) interferent. The sponsor stated that interference was considered to be non-significant if the difference between the samples with and without interferent was within 10%.
The results of the interference studies are summarized in the table below.
| Analyte | Interferent | Highest concentration tested that did not show significant interference |
| --- | --- | --- |
| CALCIUM-CRESOLPHTHALEIN | Bilirubin | 30 mg/dL |
| | Hemoglobin | 500 mg/dL |
| | Triglycerides | 1300 mg/dL |
| GLUCOSE | Bilirubin | 12 mg/dL |
| | Hemoglobin | 604 mg/dL |
| | Triglycerides | 222 mg/dL |
| UREA/BUN-UV | Bilirubin | 30 mg/dL |
| | Hemoglobin | 500 mg/dL |
| | Triglycerides | 1625 mg/dL |
The following limitation is stated in the GLUCOSE device labeling:
This device has significant positive interference with triglycerides at levels above 222 mg/dL. Do not use lipemic samples with this assay.
K192730 - Page 7 of 11
{7}
Exogenous interferences:
| Analyte | Matrix | Interferent | Highest concentration tested that did not show significant interference |
| --- | --- | --- | --- |
| CALCIUM-CRESOLPHTHALEIN | Serum | Acetaminophen | 20 mg/dL |
| | | Ascorbic Acid | 30 mg/dL |
| | | Acetylsalicylic Acid | 70 mg/dL |
| | Urine | Ascorbic Acid | 30 mg/dL |
| GLUCOSE | Serum | Acetaminophen | 20 mg/dL |
| | | Uric Acid | 20 mg/dL |
| | | Ascorbic Acid | 6 mg/dL |
| UREA/BUN-UV | Serum | Acetaminophen | 20 mg/dL |
| | | Ascorbic Acid | 30 mg/dL |
| | | Acetylsalicylic Acid | 70 mg/dL |
| | Urine | Ascorbic Acid | 30 mg/dL |
4. Assay Reportable Range:
| Analyte | Matrix | Claimed range (mg/dL) |
| --- | --- | --- |
| CALCIUM-CRESOLPHTHALEIN | Serum | 0.80 - 20.0 |
| | Urine | 3.97 – 30.0 |
| GLUCOSE | Serum | 2.0 - 500 |
| UREA/BUN-UV | Serum | 10.0 - 240 |
| | Urine | 635 - 12000 |
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Traceability:
The sponsor provided the following information on the traceability of the assays:
| Assay | Traceability |
| --- | --- |
| CALCIUM-CRESOLPHTHALEIN | SRM 956 (NIST) Standard Reference Materials. |
| GLUCOSE | SRM 965 (NIST) Standard Reference Materials. |
| UREA/BUN-UV | SRM 909 (NIST) Standard Reference Materials |
6. Detection Limit:
Detection capabilities studies for each analyte were evaluated based upon the CLSI EP17-A2 guideline.
Limit of blank (LoB) studies were performed by testing 5 blank samples in replicates of 4 over 3 days, on two lots for a total of 120 observations. LoB was defined as the highest result that can reasonably be expected from a blank sample for a given error probability with $\alpha = 0.05$.
K192730 - Page 8 of 11
{8}
Limit of detection (LoD) studies were performed by testing 5 pools of human serum or urine with analyte concentrations close to expected detection limit for each analyte. Samples were tested in replicates of 4 over 3 days, on two lots for a total of 120 observations. The LoD was estimated based on a probability of error of 5% (β=0.05).
Limit of quantitation (LoQ) studies were performed using 4 to 5 pools of human serum or urine with analytes concentrations closed to the expected LoQ of the corresponding assay. Samples were tested in replicates of 4 over 3 days, on two lots for a total of 120 observations. The sponsor defines LoQ as the lowest concentration of a measurand with a CV ≤10%.
The results of the detection limit studies are presented in the table below:
| Assays | Matrix | LoB | LoD | LoQ | Claimed Range |
| --- | --- | --- | --- | --- | --- |
| CALCIUM-CRESOLPHTHALEIN | Serum | 0.22 | 0.46 | 0.80 | 0.80 - 20.0 mg/dL |
| | Urine | 0.19 | 0.47 | 3.97 | 3.97 - 20.0 mg/dL |
| GLUCOSE | Serum | 0.39 | 1.09 | 2.00 | 2.0 - 500 mg/dL |
| Urea/BUN-UV | Serum | 1.67 | 3.78 | 10.0 | 10.0-240 mg/dL |
| | Urine | 169 | 356 | 635 | 635-12000 mg/dL |
7. Assay Cut-Off:
Not applicable.
B Comparison Studies:
1. Method Comparison with Predicate Device:
Method comparison studies were conducted by testing native human serum, Lithium Heparin plasma and urine samples (as applicable) on CALCIUM-CRESOLPHTHALEIN, GLUCOSE and UREA/BUN-UV candidate assays on the BA400 analyzer and the predicate devices. To supplement testing across the reportable range of the assays, some samples were altered (spiked or diluted) and used to cover the full measuring range of each analyte.
The results of the regression analysis are summarized below:
| Analyte | Specimen Type | N | Slope | Intercept | R | Test range mg/dL | Claimed Measuring Range mg/dL |
| --- | --- | --- | --- | --- | --- | --- | --- |
| CALCIUM-CRESOLPH-THALEIN | Serum | 153 | 1.056 | -0.57 | 0.992 | 0.88 -17.9 | 0.80 - 20.0 |
| | Li Heparin Plasma | 146 | 1.033 | -0.336 | 0.996 | 1.20 - 19.8 | |
| | Urine | 77 | 0.97 | 0.0034 | 0.995 | 4.32 - 29.4 | 3.97 - 30.0 |
| GLUCOSE | Serum | 146 | 1.004 | 2.10 | 0.999 | 8.43 - 485 | 2.0 - 500 |
K192730 - Page 9 of 11
{9}
| Analyte | Specimen Type | N | Slope | Intercept | R | Test range mg/dL | Claimed Measuring Range mg/dL |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | Li Heparin Plasma | 127 | 0.997 | 0.28 | 0.996 | 5.58 - 432 | |
| UREA/BUN - UV | Serum | 111 | 0.99 | 2.34 | 0.997 | 10.8 - 232 | 10.0 - 240 |
| | Li Heparin Plasma | 112 | 1.00 | 3.02 | 0.998 | 12.4 -238 | |
| | Urine | 97 | 1.057 | -22.0 | 0.998 | 650-11462 | 635-12000 |
2. Matrix Comparison:
Not applicable. Different sample types were assessed in other analytical studies described above.
C Clinical Studies:
1. Clinical Sensitivity:
Not applicable.
2. Clinical Specificity:
Not applicable.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable.
D Clinical Cut-Off:
Not applicable.
E Expected Values/Reference Range:
The following reference ranges are cited from literature:
CALCIUM-CRESOLPHTHALEIN$^{(1)}$
8.6-10.0 mg/dL (2.15-2.50 mmol/L)
Urine: 100-300 mg/24-h (2.5-7.5 mmol/24-h)
GLUCOSE
Serum and plasma (Adult): 74 -100 mg/dL (4.10-5.60 mmol/L)$^{(2)}$
According to American Diabetes Association)$^{(3)}$, elevation of fasting plasma glucose values over
126 mg/dL (7.0 mmol/L) on more than one occasion is diagnostic of diabetes mellitus.
K192730 - Page 10 of 11
{10}
UREA/BUN-UV$^{(2)}$
Serum and plasma: 14.4 – 48 mg/dL urea, 6 - 20 mg/dL BUN (5.14 – 17.14 mmol/L urea).
Urine: 26-43 g/24-h urea, 12-20 g/24 h BUN (428-714 mmol/24-h urea)
$^{(1)}$Tietz Textbook of Clinical Chemistry and Molecular Diagnostics, 3rd ed. Burtis C.A., Ashwood E.R., Bruns D.E., (WB Saunders Co, 1987).
$^{(2)}$Tietz Textbook of Clinical Chemistry and Molecular Diagnostics, 5th ed. Burtis C.A., Ashwood E.R., Bruns D.E., (WB Saunders Co, 2012).
$^{(3)}$American Diabetes Association. Classification and Diagnosis of Diabetes: Standards of Medical care in Diabetes. Diabetes Care 2020 Jan; 43(Supplement 1): S14-S31.
## VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
## IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K192730 - Page 11 of 11
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.