The Padlock Clip™ is indicated for use in flexible Endoscopy and for the compression of tissue in the gastrointestinal tract, for hemostasis or for treating lesions of the wall of the gastrointestinal organs. The Padlock Clip™ is indicated for clip placement within the gastrointestinal (GI) tract for the purpose of: Endoscopic marking of lesions Hemostasis for: Mucosal/Submucosal defects Bleeding Ulcers Arteries <2mm Polyps <1.5cm in diameter Diverticula in the Colon Closure of GI tract luminal perforations <20mm that can be treated conservatively
Device Story
The Padlock Clip is a single-use, radiopaque, coin-sized ligation clip composed of superelastic shape memory alloy (Nitinol). It is mounted on the distal tip of a flexible endoscope via a delivery system featuring a central 'tissue chamber'. The device is operated by a physician during endoscopic procedures; an independent hand control deploys the clip. Tissue is pulled into the chamber to facilitate approximation of larger volumes than standard clips. The device is compatible with endoscopes having distal tip diameters of 9.5mm to 14mm. It provides mechanical tissue compression to achieve hemostasis or close perforations, aiding in lesion management and potentially reducing the need for more invasive surgical interventions.
Clinical Evidence
Bench testing only. Verification activities included biocompatibility testing (cytotoxicity, sensitization, irritation, systemic toxicity, intramuscular implantation, genotoxicity, pyrogenicity) per ISO 10993 standards and corrosion susceptibility testing per ASTM F2129. Functional testing confirmed the device successfully deploys and secures defects.
Technological Characteristics
Materials: Nitinol (superelastic shape memory alloy), Polypropylene housing, Type 304/302/303/18-2 stainless steel, black fluorocarbon securing strap. Dimensions: Compatible with 9.5mm-14mm endoscope distal tips; clip opening 6-24mm. Principle: Mechanical tissue compression via shape memory alloy clip. Connectivity: None. Sterilization: Not specified. Standards: ISO 10993 (biocompatibility), ASTM F2129 (corrosion).
Indications for Use
Indicated for patients requiring endoscopic tissue compression in the GI tract for hemostasis (mucosal/submucosal defects, bleeding ulcers, arteries <2mm, polyps <1.5cm, colonic diverticula), lesion marking, or closure of GI luminal perforations <20mm treatable conservatively.
Regulatory Classification
Identification
A hemorrhoidal ligator is a device used to cut off the blood flow to hemorrhoidal tissue by means of a ligature or band placed around the hemorrhoid.
Special Controls
*Classification.* Class II (special controls). Except for a hemostatic metal clip intended for use in the gastrointestinal tract, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in large, bold, blue letters, and the word "ADMINISTRATION" in smaller, blue letters below.
October 24, 2019
STERIS Corporation Tony Piotrkowski Regulatory Affairs Director 5960 Heisley Road Mentor, OH 44060
Re: K192722
> Trade/Device Name: Padlock Clip Pro-Select™ defect closure system, Padlock Pro-Select™ defect closure device Regulation Number: 21 CFR 876.4400 Regulation Name: Hemorrhoidal Ligator Regulatory Class: II Product Code: PKL Dated: September 26, 2019 Received: September 27, 2019
Dear Tony Piotrkowski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS)
regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For Martha W. Betz, Ph.D. Acting Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
### 510(k) Number (if known) K192722
### Device Name
Padlock Clip defect closure system Padlock Clip Pro-Select defect closure system
### Indications for Use (Describe)
The Padlock Clip™ is indicated for use in flexible Endoscopy and for the compression of tissue in the gastrointestinal tract, for hemostasis or for treating lesions of the gastrointestinal organs.
The Padlock Clip™ is indicated for clip placement within the gastrointestinal (GI) tract for the purpose of:
- · Endoscopic marking of lesions
- · Hemostasis for:
- o Mucosal/Submucosal defects
- o Bleeding Ulcers
- o Arteries <2mm
- o Polyps <1.5cm in diameter
- o Diverticula in the Colon
· Closure of GI tract luminal perforations <20mm that can be treated conservatively
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| | |
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word STERIS in large, bold letters, with the registered trademark symbol next to it. Below the word STERIS is a graphic of several horizontal wavy lines stacked on top of each other. At the top of the image are the words "K192722" and "Page 1 of 5".
# 510(k) Summary For Padlock Clip Pro-Select™ defect closure system and Padlock Clip™ defect closure system
STERIS Corporation 5960 Heisley Road Mentor, OH 44060
Contact: Anthony Piotrkowski Regulatory Affairs Director Tel: 440-392-7458 Fax: 440-357-9198 Email: tony_piotrkowski@steris.com
Summary Date: September 26, 2019
STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600
{5}------------------------------------------------
## STERIS Special 510(k) PREMARKET NOTIFICATION Modification to Padlock Clip and Padlock Clip Pro-Select defect closure systems
#### 1. Device Name
| Trade Name: | Padlock Clip defect closure system and<br>Padlock Clip Pro-Select defect closure system |
|------------------------|-----------------------------------------------------------------------------------------|
| Device Classification: | Class II |
| Common/usual Name: | Ligator clip |
| Classification Name: | Hemorrhoidal ligator |
| Classification Number: | 21 CFR 876.4400 |
| Product Code: | PKL |
#### 2. Predicate Device
| K180689 | Padlock Clip defect closure device and Padlock Clip Pro-Select defect closure device |
|---------|--------------------------------------------------------------------------------------|
|---------|--------------------------------------------------------------------------------------|
Originally cleared as:
#### 3. Description of Device
The Padlock Clip™ ligation clip consists of a preloaded, radiopaque, single use, coin sized ligation clip made of superelastic shape memory alloy for tissue approximation with opening sizes of 6 to 24mm on a flexible delivery system.
#### 4. Intended Use
The Padlock Clip™ is indicated for use in flexible Endoscopy and for the compression of tissue in the gastrointestinal tract, for hemostasis or for treating lesions of the wall of the gastrointestinal organs.
The Padlock Clip™ is indicated for clip placement within the gastrointestinal (GI) tract for the purpose of:
- · Endoscopic marking of lesions
- · Hemostasis for:
- o Mucosal/Submucosal defects
- o Bleeding Ulcers
- o Arteries <2mm
- o Polyps <1.5cm in diameter
- o Diverticula in the Colon
- · Closure of GI tract luminal perforations <20mm that can be treated conservatively
{6}------------------------------------------------
#### 5. Description of Technological Similarities and Differences
The Padlock Clip™ is identical to the predicate device except for the specific modification described in this submission. The differences between the proposed and predicate devices are limited to the addition of an alternate supplier for the clip component of the device. It is the same raw material from the same source with only the supplier of the final, processed component changing and this proposed change raises no new concerns of safety and effectiveness when compared to the predicate device as evidenced through the testing summarized in this submission.
## Device Comparison Table
| | Proposed<br>Padlock Clip | K180689<br>Padlock Clip | Comparison |
|----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|
| Feature | | | |
| Indications<br>for Use | The Padlock ClipTM is<br>indicated for use in flexible<br>Endoscopy and for the<br>compression of tissue in the<br>gastrointestinal tract, for<br>hemostasis or for treating<br>lesions of the wall of the<br>gastrointestinal organs.<br><br>The Padlock ClipTM is<br>indicated for clip placement<br>within the gastrointestinal<br>(GI) tract for the purpose of:<br>Endoscopic marking of<br>lesions Hemostasis for:<br>Mucosal/Submucosal<br>defects Bleeding Ulcers Arteries <2mm Polys <1.5cm in<br>diameter Diverticula in the Colon Closure of GI tract<br>luminal perforations<br><20mm that can be<br>treated conservatively | The Padlock ClipTM is<br>indicated for use in flexible<br>Endoscopy and for the<br>compression of tissue in the<br>gastrointestinal tract, for<br>hemostasis or for treating<br>lesions of the wall of the<br>gastrointestinal organs.<br><br>The Padlock ClipTM is<br>indicated for clip placement<br>within the gastrointestinal<br>(GI) tract for the purpose of:<br>Endoscopic marking of<br>lesions Hemostasis for:<br>Mucosal/Submucosal<br>defects Bleeding Ulcers Arteries <2mm Polys <1.5cm in<br>diameter Diverticula in the Colon Closure of GI tract<br>luminal perforations<br><20mm that can be treated<br>conservatively | Identical |
| Feature | Proposed<br>Padlock Clip | K180689<br>Padlock Clip | Comparison |
| Operating<br>Principles/<br>Technology | The Padlock Clip™ defect<br>closure device is mounted and<br>secured at the distal tip on the<br>outside surface of a flexible<br>endoscope. The Padlock<br>Clip™ is deployed using an<br>independent hand control.<br>The linking cable to the hand<br>control is not located within<br>the endoscopic accessory<br>channel. The Padlock Clip™<br>delivery system includes a<br>central "tissue chamber" that<br>resides on the distal tip of the<br>endoscope. Clinically<br>efficacious tissue<br>manipulation techniques may<br>be used to pull target tissue<br>into this "tissue chamber" to<br>approximate a larger volume<br>of tissue than would<br>otherwise be approximated by<br>deploying the clip alone. The<br>Padlock Clip™ delivery<br>system is compatible with<br>flexible endoscopes with<br>distal tip outer diameters<br>ranging from 9.5mm to<br>14mm. | The Padlock Clip™ defect<br>closure device is mounted and<br>secured at the distal tip on the<br>outside surface of a flexible<br>endoscope. The Padlock<br>Clip™ is deployed using an<br>independent hand control.<br>The linking cable to the hand<br>control is not located within<br>the endoscopic accessory<br>channel. The Padlock Clip™<br>delivery system includes a<br>central "tissue chamber" that<br>resides on the distal tip of the<br>endoscope. Clinically<br>efficacious tissue<br>manipulation techniques may<br>be used to pull target tissue<br>into this "tissue chamber" to<br>approximate a larger volume<br>of tissue than would<br>otherwise be approximated by<br>deploying the clip alone. The<br>Padlock Clip™ delivery<br>system is compatible with<br>flexible endoscopes with<br>distal tip outer diameters<br>ranging from 9.5mm to<br>14mm. | Identical |
| Clip | Radiopaque, single use, coin<br>sized clip made of<br>superelastic shape memory<br>alloy (Nitinol®) | Radiopaque, single use, coin<br>sized clip made of<br>superelastic shape memory<br>alloy (Nitinol®) | Identical -<br>new source<br>for<br>processing<br>to its final<br>form |
| Housing | Polypropylene | Polypropylene | Identical |
| Stainless<br>Steel | Type 304<br>Type 302<br>Type 303<br>Type 18-2 | Type 304<br>Type 302<br>Type 303<br>Type 18-2 | Identical |
| Securing<br>strap | Black fluorocarbon | Black fluorocarbon | Identical |
### Table 5-1. Padlock Device Comparison Table
{7}------------------------------------------------
## STERIS Special 510(k) PREMARKET NOTIFICATION Modification to Padlock Clip and Padlock Clip Pro-Select defect closure systems
Table 5-2 summarizes the verification activity that was performed with its respective acceptance criteria to ensure that this modification does not affect the safety or effectiveness of the Padlock Clip defect closure system. All testing to
{8}------------------------------------------------
## STERIS Special 510(k) PREMARKET NOTIFICATION Modification to Padlock Clip and Padlock Clip Pro-Select defect closure systems
support this change was done following well established methods such as recognized consensus (i.e. ISO 10993 parts 3, 5, 6 10 and 11, ASTM F2129) standards, FDA guidance or using the same test methodology used to verify the predicate device design as noted in K180689 or K120814.
| Test | Acceptance<br>Criteria | Result |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|--------|
| Cytotoxicity | Per ISO 10993-5 | Pass |
| Sensitization | Per ISO 10993-10 | Pass |
| Irritation | Per ISO 10993-10 | Pass |
| 28-Day Systemic Toxicity | Per ISO 10993-11 | Pass |
| 13-Week Intramuscular Implantation | Per ISO 10993-6 | Pass |
| 1-Week Intramuscular Implantation | Per ISO 10993-6 | Pass |
| Genotoxicity -Ames Assay and Mouse Lymphoma | Per ISO 10993-3 | Pass |
| Pyrogenicity | Per ISO 10993-11 | Pass |
| Corrosion Susceptibility per ASTM F2129 and<br>Technical Considerations for Non-Clinical<br>Assessment of Medical Devices containing Nitinol<br>Draft Guidance for Industry and Food and Drug<br>Administration Staff, April 19, 2019. | The breakdown<br>potential of the test<br>clip must be greater<br>than or equal to the<br>predicate clip | Pass |
| Functional Testing | Deploys<br>Secures defect | Pass |
| | | | Table 5-2. Summary of verification activities. | |
|--|--|--|------------------------------------------------|--|
|--|--|--|------------------------------------------------|--|
#### 6. Conclusion
Based on the intended use, technological characteristics and non-clinical performance data, the subject device is as safe, as effective and performs as well as the legally marketed predicate device (K180689), Class II (21 CFR 876.4400), product code PKL.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.