K192711 · Dannik · FZP · Dec 20, 2019 · General, Plastic Surgery
Device Facts
Record ID
K192711
Device Name
Dannik Titanium Ligation Clip
Applicant
Dannik
Product Code
FZP · General, Plastic Surgery
Decision Date
Dec 20, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The DANNIK Titanium Ligation Clips is designed for the intended use of ligating blood vessels .The Clip has been specially designed to insure occlusion of vessels and prevent any slippage once applied. The clip has applications in many types of surgical procedures where hemostasis is required or radiographic marking is necessary. Choose the size of clip to fit the procedure making certain the tissue to be occluded fits completely within the clip. The DANNIK Titanium Ligation, clip can be left in vivo without sequela as it is biologically inert.
Device Story
DANNIK Titanium Ligation Clip is a passive, implantable surgical device used for vessel ligation and hemostasis. The device consists of titanium clips supplied in cartridges. It is intended for use by surgeons in various surgical procedures. The clip is applied to tissue or vessels to ensure occlusion and prevent slippage. Being biologically inert, it can be left in vivo. The device is provided sterile for single use. It functions as a mechanical clamp; no electronic or software components are involved.
Clinical Evidence
No clinical data. Bench testing only. Performance evaluation included appearance, dimensions, surface roughness, hardness, toughness, clamping (retention) performance, and corrosion resistance. All tests passed.
Technological Characteristics
Material: Grade I Titanium (ASTM F-67 95, ISO 5832-2-93). Form factor: Implantable clips in four sizes (small, medium, large, medium-large). Sterilization: Ethylene Oxide (ISO 11135-1). Biocompatibility: ISO 10993 compliant. Single-use, non-electronic, passive mechanical device.
Indications for Use
Indicated for ligation of blood vessels and other anatomic structures in surgical procedures requiring hemostasis or radiographic marking. For use by surgeons selecting appropriate clip size based on clinical judgment.
Regulatory Classification
Identification
An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.
Predicate Devices
Vitalitec International Inc Implantable Titanium Hemostatic Clip (K981645)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 20, 2019
Dannik Olga Haberland Regulatory Compliance 941 W Morse Blvd. Suite 100 Winter Park, Florida 32789
Re: K192711
Trade/Device Name: Dannik Titanium Ligation Clip Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable Clip Regulatory Class: Class II Product Code: FZP Dated: September 19, 2019 Received: September 27, 2019
Dear Olga Haberland:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Cindy Chowdhury, Ph.D., M.B.A. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K192711
Device Name DANNIK Titanium Ligation Clip
#### Indications for Use (Describe)
The DANNIK Titanium Ligation Clips is designed for the intended use of ligating blood vessels .The Clip has been specially designed to insure occlusion of vessels and prevent any slippage once applied. The clip has applications in many types of surgical procedures where hemostasis is required or radiographic marking is necessary.
Choose the size of clip to fit the procedure making certain the tissue to be occluded fits completely within the clip. The DANNIK Titanium Ligation, clip can be left in vivo without sequela as it is biologically inert.
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|----------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) /Summary
- 1. Contact Information DANNIK 941 West Morse Blvd. Suite #100 Winter Park, Florida 32789 Phone: (407) 745-1698 Olga Haberland, Regulatory Compliance December 16, 2019
- 2. Device Name
- Trade Name DANNIK Titanium Ligation Clip ●
- Common Name: Titanium Ligation Clip ●
- Classification Name Clip Implantable ●
- Classification: Class II General and Plastic Surgery Devices #79 FZP, ● 21 CFR 878.4300
Performance Standards: Devices are manufactured according to Good Manufacturing Practices (G.M.P.), Association for Advancement of Medical Instrumentation (A.AM.I.) and American Society for Testing and Materials (A.S.T.M.) requirements and applicable Harmonized Standards ISO 13485.
Material Composition: A.S.T.M F-67 95, Grade I Titanium I.S.O. #5832-2-93, Grade I Titanium
- 3. Substantially Equivalent Device
- Legally Marketed (unmodified Devices): .
- o Primary Predicate: Vitalitec International Inc Implantable Titanium Hemostatic Clip FDA 510K (K981645)
- o Additional Predicate: Weck Metal Ligating Clips (K132658)
{5}------------------------------------------------
## 4. Device Description
The DANNIK Titanium Ligation Clips are exclusively made of titanium and supplied in four different sizes: small, medium-large, and large. Packaged in cartridge of six clip units.
The titanium used meets all requirements for A.S.T.M. specifications F-67 95 "Unalloyed Titanium for Surgical Implant Applications", Grade I and ISO 5832-2-93 and 10993 "Implants for Surgery - Metallic Materials - Part 2: Unalloyed Titanium".
This device is packaged and sterilized for single use only. Do NOT re-use, reprocess or resterilize. Discard after use with care.
Biocompatibility Conforms to ISO 10993 Prescription Only Yes Sterilized by ETO (Ethylene Oxide)
### 5. Intended Use
The DANNIK Titanium Ligation Clips is designed for the intended use of ligating blood vessels. The Clip has been specially designed to insure occlusion of vessels and prevent any slippage once applied. The clip has applications in many types of surgical procedures where hemostasis is required or radiographic marking is necessary.
Choose the size of clip to fit the procedure making certain the tissue to be occluded fits completely within the clip. The DANNIK Titanium Ligation, clip can be left in vivo without sequela as it is biologically inert
- 6. Technological Characteristics of the Subject Device Compared to the Predicate Device
Comparison of Technological Characteristics: The titanium clip material and disposable holder material substantially equivalent to the predicate device. In function, the clips are the substantially equivalent to the predicate device as well.
Safety and Efficacy Information: Titanium is well recognized as being safe and effective for long term implantation. Millions of titanium clips are applied yearly and since inception in the 1960's attest to the wide
{6}------------------------------------------------
acceptance of this method of hemostatic and ligation control.
There are no new technologies being added to this device from the predicate, in terms of finished device functions. The device has the same intended use and application as the predicate device.
| Device | DANNIK Titanium<br>Ligation Clip | Vitalitec K981645 | Weck Metal Ligating<br>Clips K132658 |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use | The DANNIK Titanium<br>Ligation Clips is<br>designed for the<br>intended use of<br>ligating blood vessels<br>.The Clip has been<br>specially designed to<br>insure occlusion of<br>vessels and prevent<br>any slippage once<br>applied. The clip has<br>applications in many<br>types of surgical<br>procedures where<br>hemostasis is<br>required or<br>radiographic marking<br>is necessary.<br>Choose the size of<br>clip to fit the<br>procedure making<br>certain the tissue to<br>be occluded fits<br>completely within<br>the clip. The DANNIK<br>Titanium Ligation,<br>clip can be left in vivo<br>without sequela as it<br>is biologically inert. | The VITALITEC<br>INTERNATIONAL,<br>INC,. Hemostatic Clip<br>is designed for the<br>intended use of<br>ligating blood<br>vessels. The clip has<br>been specially<br>designed to insure<br>occclusion of vessels<br>and prevent any<br>slippage once<br>applied. The clip has<br>applications in many<br>types of surgical<br>procedures where<br>hemostasis is<br>required or<br>radiographic marking<br>is necessary.<br>Choose the size of<br>clip to fit the<br>procedure making<br>certain the tissue to<br>be occluded fits<br>completely within<br>the clip. The<br>VITALITEC<br>INTERNATIONAL,<br>INC., clip can be left<br>in vivo without<br>sequela as it is<br>biologically inert. | Weck@ Ligating Clips<br>are intended for use<br>in procedures<br>involving vessels or<br>anatomic structures<br>for which the<br>surgeon determines<br>ligating clips are the<br>best choice. Surgeons<br>should select the<br>size, type and<br>material of the clip<br>based upon their<br>experience,<br>judgment and needs. |
| Device Description | Titanium Ligation<br>Clips made<br>exclusively made of<br>titanium | Same | Weck® Metal Ligating<br>Clips are<br>manufactured from<br>medical grade<br>titanium, tantalum or<br>stainless steel alloys |
| Clip Material | Titanium<br>The titanium used<br>meets all<br>requirements for<br>A.S.T.M.<br>specifications F-67 95<br>"Unalloyed Titanium<br>for Surgical Implant<br>Applications", Grade I<br>and ISO 5832-2-93<br>and 10993 "Implants<br>for Surgery - Metallic<br>Materials - Part 2:<br>Unalloyed Titanium". | Same | Medical Grade<br>Titanium |
| Sterilization | Ethylene Oxide (<br>ETO)<br>I.S.O 11135-1 | Unknown | Unknown |
| Prescription Only | Yes | Same | Same |
| Biocompatibility | Conforms to ISO<br>10993 | Unknown | Conforms to ISO<br>10993 |
| Enviromental<br>Conditions | | | Weck® Metal Ligating<br>Clips are "MR<br>Conditional" up to<br>and including 3-Tesla<br>MR environments. |
{7}------------------------------------------------
K192711
## Page 4 of 5
# 7. Non-Clinical Tests
The DANNIK Titanium Ligation Clips has been evaluated by our Design Engineer and through performance studies and bench testing, as attached in ANNEX F. Testing encompassed appearance, dimension, surface roughness, hardness toughness , clamping( retention) performance and corrosion resistance.
{8}------------------------------------------------
All the test results were "PASS". The performance of Titanium Clip meets the technical standard requirements of Dannik as compared to the predicate.
- 8. Clinical Tests No Clinical trials performed on the DANNIK Titanium Ligation Clips .
### 9. Conclusions
The subject device has equivalent indications for use as the predicate device. The technological characteristics, non-clinical testing and performance and bench testing of the DANNIK Titanium Ligation Clips show that the device is substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.