InMode System with the Morpheus8 (Fractora) Applicators

K192695 · Inmode MD , Ltd. · GEI · Dec 27, 2019 · General, Plastic Surgery

Device Facts

Record IDK192695
Device NameInMode System with the Morpheus8 (Fractora) Applicators
ApplicantInmode MD , Ltd.
Product CodeGEI · General, Plastic Surgery
Decision DateDec 27, 2019
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The InMode System with the Morpheus8 (Fractora) Applicators is intended for use in Dermatological and General Surgical procedures for Electrocoagulation and Homeostasis. At higher energy levels greater than 62 ml/pin, use of the Morpheus8 (Fractora) Applicators is limited to Skin Types I-V.

Device Story

InMode System is a computerized, RF-based electrosurgical device for dermatological electrocoagulation and hemostasis. System comprises an AC/DC power supply, RF generator, controller, and LCD touch screen. Operators (clinicians) use a foot switch to activate energy delivery through disposable, single-use fractional RF electrode pin tips (12, 24, 40, or T-pin configurations). Device delivers RF energy to skin via multi-electrode pin arrays. System monitors RF parameters to enhance safety and minimize side effects. Output is used by clinicians to perform surgical procedures; clinical benefit is achieved through controlled tissue coagulation. Device is intended for clinical use.

Clinical Evidence

No clinical data. Bench testing only. Performance testing compared RF output (pulse duration, energy per pin, total energy, peak-to-peak voltage) of new tip heads against predicate device across various resistance loads, demonstrating equivalent performance.

Technological Characteristics

Electrosurgical RF device; 1 MHz output frequency; 65W max power. Components: AC/DC power supply, RF generator, controller, LCD interface. Applicators: fractional multi-electrode pin arrays (12, 24, 40, T-pin). Materials: biocompatible, identical to predicate. Standards: AAMI/ANSI 60601-1, IEC 60601-1-2, IEC 60601-2-2. Sterilization: Gamma radiation (ISO 11137-1).

Indications for Use

Indicated for dermatological and general surgical procedures requiring electrocoagulation and hemostasis. Use at energy levels >62 mJ/pin is restricted to patients with Skin Types I-IV.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. December 27, 2019 Inmode MD Ltd. % Amit Goren Regulatory Manager A. Stein- Regulatory Affairs Consulting Ltd. 20 Hata'as Str., Suite 102 Kfar Saba, 4442520 Il Re: K192695 Trade/Device Name: InMode System with the Morpheus8 (Fractora) Applicators Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: November 25, 2019 Received: November 27, 2019 Dear Amit Goren: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. For Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ### Indications for Use 510(k) Number (if known) Device Name InMode System with the Morpheus8 (Fractora) Applicators Indications for Use (Describe) The InMode System with the Morpheus8 (Fractora) Applicators is intended for use in Dermatological and General Surgical procedures for Electrocoagulation and Homeostasis. At higher energy levels greater than 62 ml/pin, use of the Morpheus8 (Fractora) Applicators is limited to Skin Types I-V. Type of Use (Select one or both, as applicable) | > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(K) SUMMARY # INMODE RF SYSTEM # 510(k) Number K192695 ## Applicant Name: | Company Name: | InMode Ltd | |-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------| | Address: | Tabor Building, Shaar Yokneam<br>POB 44, Yokneam 20692<br>Israel<br>Tel: +972-4-9097470<br>Fax: +972-4-9097471<br>E-mail: amit@asteinrac.com | | Contact Person: | Amit Goren | | Official Correspondent: | Amit Goren | | Company Name: | A. Stein - Regulatory Affairs Consulting Ltd. | | Address: | 20 Hata'as Str., Suite 102<br>Kfar Saba 4442520 Israel<br>Tel: +972-9-7670002<br>Fax: +972-9-7668534<br>E-mail: amit@asteinrac.com | | Date Prepared: | November 25, 2019 | | Trade Name: | InMode System with the Morpheus8 (Fractora) Applicators | | Classification Name: | CFR Classification section 878.4400; (Product code GEI) | | Classification: | Class II Medical Device | ## Predicate Device: The subject device is substantially equivalent to the following predicate device: | Device | Manufacturer | 510(k) No. | |--------------------------------------------------------|----------------|------------| | InMode System with the<br>Fractora3D/3D-90 Applicators | InMode MD Ltd. | K180189 | {4}------------------------------------------------ ## Device Description: The InMode System with the Morpheus8 (Fractora) Applicators is a computerized, programmed, RF technology-based device intended for dermatological applications which requires skin electrocoagulation and hemostasis. The device platform is basically constituted on the same system platform as FDA cleared for the proposed predicate device; The InMode System with the Fractora3D/3D-90 Applicators (K180189). The device applicators employ fractional RF multi-electrode technology for procedures requiring electrocoagulation and hemostasis. The Morphues8 (Fractora) Applicators are designed to deliver radiofrequency energy to the skin in a fractional manner, via an array of multielectrode pins. The Device provides enhanced safety while minimizing possible side effects by monitoring RF parameters. The InMode System with the Morpheus8 (Fractora) Applicators consists of an AC/DC power supply unit, RF generator, controller and user interface including LCD touch screen. The Morpheus8 (Fractora) Applicators are connected to the console via a cable and a foot switch activates the energy delivery to the applicator. The subject device applicator comprises a disposable, single use, fractional RF electrode pins tip head. The InMode System with the Morpheus8 (Fractora) Applicators comprises the following applicator tip heads: - Morpheus8 24 Pin Applicator (Fractora3D) tip (FDA Cleared in K180189) - Morpheus8 40 Pin treatment tip (New tip) . - . Morpheus8 12 Pin treatment tip (New tip) - . Morpheus8 T Pin treatment tip (New tip) ### Device Specifications: | Main Line Frequency (nominal) | 50 - 60 Hz | |----------------------------------------|----------------------| | Input Voltage (nominal) | 100 - 240 VAC | | Electrosurgical Unit dimensions (inch) | 18.2"W x18.2"D x40"H | | Platform weight (lb.) | 70.5 | | Applicator weight (lb.) | 1.26 | | RF Max Output Power (Watt) | 65 | | RF Output Frequency (MHz) | $1\pm 2%$ | {5}------------------------------------------------ ## Intended Use/Indication for Use: The InMode System with the Morpheus8 (Fractora) Applicators is intended for use in Dermatologic and General Surgical procedures for Electrocoagulation and Hemostasis. At higher energy levels greater than 62 mJ/pin, use of the Morrpheus8 (Fractora) Applicator is limited to Skin Types I-IV. ## Performance Standards: The InMode System with the Morpheus8 (Fractora) Applicators complies with the following FDA recognized consensus standards: - AAMI/ANSI 60601-1 (2012), Medical Electrical Equipment Part 1: General . Requirements For Basic Safety And Essential Performance (IEC 60601-1:2005, Mod). - IEC 60601-1-2 (Edition 3.0, 2007), Medical Electrical Equipment Part 1-2: . General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests. - IEC 60601-2-2 Edition 5.0 2009-02, Medical Electrical Equipment Part 2-2: . Particular Requirements For The Basic Safety And Essential Performance Of High Frequency Surgical Equipment And High Frequency Surgical Accessories. ## Non-Clinical (Bench) Performance Data: The InMode System with the Morpheus8 (Fractora) Applicators was tested for its applicator tip RF output performance specifications. The device utilizing the Morphues8 12, 24, 40 pin and T tip heads preset to different RF energy levels, was tested under different resistance loads with the following parameters measured and compared to the results obtained in similar testing conducted on the predicate device: pulse duration, energy per pin, total energy, and pick to pick voltage. The results demonstrate equivalent measurements obtained for the subject and predicate devices. ## Animal Performance Data / Histology Data: Not Applicable ## Clinical Performance Data: Not Applicable {6}------------------------------------------------ ## Biocompatibility All of the modified device materials which come in direct contact with the patient skin are biocompatible and identical to the materials used in the predicate device manufacturing. ## Substantial Equivalence: A comparison table is provided below comparing the intended use and basic technological characteristics of the InMode System with the Morpheus8 (Fractora) Applicators to the intended use and basic technological characteristics of the predicate device. | Technological<br>Characteristic | InMode System with the<br>Morpheus8 (Fractora)<br>Applicators<br>InMode Ltd.<br>(K192695) | InMode System with the<br>Fractora3D/3D-90<br>Applicators<br>InMode MD Ltd.<br>(K180189) | |---------------------------------|------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------| | Dimensions | 46cm W x 46cm D x<br>100cm H<br>[18.2" W x 18.2" D x<br>40" H] | 46cm W x 46cm D x<br>100cm H<br>[18.2" W x 18.2" D x<br>40" H] | | Weight | | | | Platform weight | 32 Kg (70.5 lbs.) | 32 Kg (70.5 lbs.) | | Applicator weight | 0.4 Kg (1.26 lbs.)<br>Tip weight - 0.02 Kg | 0.4 Kg (1.26 lbs.)<br>Tip weight - 0.02 Kg | | Applicator Tip Heads: | Morpheus8 (Fractora) 12 tip head Morpheus8 (Fractora) 24 tip head Morpheus8 (Fractora) 40 tip head Morpheus8 (Fractora) T tip head | Morpheus8 (Fractora3D)<br>24 tip head | | Number of pins | 12-40 pins | 24 pins | | Maximal Treatment<br>depth | 4.0mm<br>T tip head - 0.5mm<br>(Fixed) | 4.0mm | | Cable Dimensions: | 250 cm | 250 cm | | Performance | Frequency: 1 MHz | Frequency: 1 MHz | {7}------------------------------------------------ | Technological<br>Characteristic | InMode System with the<br>Morpheus8 (Fractora)<br>Applicators<br>InMode Ltd.<br>(K192695) | InMode System with the<br>Fractora3D/3D-90<br>Applicators<br>InMode MD Ltd.<br>(K180189) | |---------------------------------|--------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------| | | Maximal RF output<br>power: 65W | Maximal RF output<br>power: 65W | | Standards Met | IEC 60601-1<br>IEC 60601-1-2<br>ANSI AAMI 60601-2-2<br>for safety of high<br>frequency surgical<br>equipment | IEC 60601-1<br>IEC 60601-1-2<br>ANSI AAMI 60601-2-2<br>for safety of high<br>frequency surgical<br>equipment | | Reprocessing &<br>Sterilization | For single use.<br>All applicator tip heads<br>are gamma sterilized. | For single use.<br>Non-sterilized. To be<br>sterilized prior to use. | ## Comparison Discussion: The indications for use and technological characteristics of the modified device are substantially equivalent to the indications for use and technological characteristics of the predicate device. The design and components of the modified device, including the console (with power supply, RF generator, controller and display panel) and the Applicator handle (with cable and connector to console) are identical to the design and components found in the predicate device. The modified device comprises the predicate device Morpheus8 (Fractora) 24 pin tip head and additional three compatible tip heads; the Morpheus8 (Fractora) 12, 40 pin and T tip heads. The performance specifications (including RF frequency, pulse duration, RF current distribution and RF energy per pin) of the modified device are comparable to those of the predicate device. The safety features and compliance with safety standards in the modified device are identical to the safety features and compliance with safety standards found in the predicate device. Patient contact materials are also identical. Any minor differences in the technological characteristics do not raise new safety or effectiveness concerns. Furthermore, the modified device underwent bench performance testing, including software validation testing, electrical and mechanical safety testing to IEC 60601-1 and electromagnetic compatibility testing according to IEC 60601-1-2 and IEC 60601-2-2. Sterilization methods applied on the device applicator tip heads were validated in accordance with ISO 11137-1. {8}------------------------------------------------ The performance tests demonstrated that the minor differences in the device design and specifications meet the system requirements and do not raise new safety or effectiveness concerns. Consequently, it can be concluded that the InMode System with the Morpheus8 (Fractora) Applicators is substantially equivalent to the predicate InMode System with the Fractora3D/3D-90 Applicators, FDA cleared in 510(k) K180189, and therefore, may be legally marketed in the USA.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...